Tzone TempU02

Tzone TempU02 USB Temperature Data Logger Instruction Manual

Model: TempU02

1. Introduction

The Tzone TempU02 is a single-use USB temperature data logger designed for reliable temperature monitoring during cold chain transportation, food storage, and other temperature-sensitive applications. It generates an automatic PDF report without requiring special software, simplifying data retrieval.

Key features include:

Tzone TempU02 USB Temperature Data Logger

Figure 1: Tzone TempU02 USB Temperature Data Logger.

2. Setup and Activation

To begin temperature recording, activate the device as follows:

  1. Locate the "PRESS START STOP" button on the front of the logger.
  2. Press and hold the button for approximately 5 seconds until the "OK" LED indicator flashes, confirming activation.
  3. The device will initiate a 30-minute start-up protection delay before beginning actual temperature recording. This delay helps improve measurement accuracy by allowing the device to stabilize in its environment.
Tzone TempU02 Data Logger with labeled parts and start button

Figure 2: Device components and activation button. The device ID is TZ2024101960.

3. Operation and Status Indicators

The TempU02 data logger provides visual feedback through its LED indicators:

Regularly check these indicators to monitor the temperature status of your goods during transit or storage.

4. Data Retrieval and Report Generation

The TempU02 simplifies data retrieval by generating a PDF report directly:

  1. Once the monitoring period is complete, connect the USB port of the data logger directly to a computer.
  2. The device will automatically generate a PDF report containing all recorded temperature data. No special software installation is required.
  3. Important Note: Once the logger is plugged into a USB port, it will stop recording and cannot be reactivated for further use. This device is designed for single-use applications.
Tzone TempU02 Data Logger connected to a laptop, showing PDF report generation

Figure 3: The data logger connected to a laptop, illustrating the automatic PDF report generation. The device ID is TZ2024101960.

Tzone TempU02 Data Logger in a cold chain transportation scenario

Figure 4: The compact data logger being used in a cold chain environment. The device ID is TZ2024101960.

5. Maintenance and Care

The TempU02 data logger is designed for durability and ease of use:

Tzone TempU02 Data Logger demonstrating IP67 waterproof feature

Figure 5: The data logger submerged in water, illustrating its IP67 waterproof capability when sealed.

6. Troubleshooting

7. Specifications

FeatureDetail
ModelTempU02
Measurement Range-30°C to 70°C (-22°F to 158°F)
Accuracy±0.5°C (-20°C to 40°C); ±1°C (others)
Temperature Alarm ThresholdsLow: <2°C (35.6°F), High: >8°C (46.4°F)
Data Storage Capacity17281 records
Log Interval5 minutes
Start Delay30 minutes
Recording DurationUp to 60 days
ConnectivityUSB 2.0
Waterproof RatingIP67 (with intact packaging)
CertificationsEN12830, CE, RoHS, FDA
Battery1 CR2032 (included)
DimensionsApproximately 3.15 x 1.57 x 0.39 inches
WeightApproximately 0.35 ounces
Tzone TempU02 Data Logger with detailed specifications

Figure 6: Visual representation of the TempU02 data logger with its key specifications. The device ID is TZ2024101960.

8. Warranty and Support

Each Tzone TempU02 data logger comes with a calibration certificate, ensuring its accuracy and compliance with industry standards.

For any technical assistance, questions, or support regarding the Tzone TempU02 data logger, please refer to the contact information provided with your purchase or visit the official Tzone website.

9. Important Notes

Related Documents - TempU02

Preview TempU06 Series Temperature Data Logger User Manual and Datasheet
User manual and technical datasheet for the Tzone TempU06 series temperature data loggers. Covers product introduction, features, usage instructions, configuration, battery information, cautions, and detailed specifications for models TempU06, TempU06 L100, and TempU06 L200.
Preview Tzone TempU06 Series Temperature Data Logger User Manual
User manual for the Tzone TempU06 Series temperature data loggers, covering product introduction, features, usage instructions, configuration, battery replacement, and technical specifications. Includes models TempU06, TempU06 L100, and TempU06 L200.
Preview Tzone TempU06 Series Temperature Data Logger User Manual
User manual for the Tzone TempU06 series temperature data loggers, detailing features, operation, configuration, and technical specifications for monitoring temperature in storage and transportation of sensitive goods like vaccines, pharmaceuticals, and food.
Preview TT19/TT19EX Quick Setup Guide
A comprehensive guide to setting up and using the TT19/TT19EX real-time data logger, including device registration, SIM card insertion, battery connection, and data querying.
Preview TZ-Tag08 LoRa Temperature Sensor User Guide | TZONE Digital Technology
Comprehensive user guide for the TZONE TZ-Tag08 LoRa Temperature Sensor. Learn about its features, applications, specifications, working modes, and installation. Includes FCC and ISED statements.
Preview TZ-RD05 RFID Card Reader User Manual
This user manual provides comprehensive instructions for the Tzone TZ-RD05 RFID card reader, covering its product overview, key functions, technical specifications, data protocols, configuration procedures, and firmware update process.