Triplett GSM100
Contents
Triplett GSM100 Combustible Gas Leak Detector
User Instruction Manual
1. Introduction
The Triplett GSM100 is a highly sensitive and reliable combustible gas leak detector designed to identify the presence of various combustible gases. Its flexible gooseneck probe allows for easy access to hard-to-reach areas, making it an essential tool for detecting potential gas leaks in residential, commercial, and industrial settings.
This manual provides detailed instructions for the safe and effective operation, maintenance, and troubleshooting of your GSM100 detector. Please read it thoroughly before use.

Image 1.1: The Triplett GSM100 Combustible Gas Leak Detector, showing its main body and flexible gooseneck probe.
2. Safety Information
Always observe the following safety precautions to prevent injury and damage to the instrument:
- Do not attempt to detect gas leaks in areas with extremely high concentrations of gas, as this may pose an explosion hazard.
- Ensure proper ventilation when investigating gas leaks.
- Do not expose the sensor to water or other liquids.
- Do not use the device if it appears damaged.
- Only use the provided Li-Po battery and AC charger.
- Keep the device away from children.
3. Product Features
The Triplett GSM100 offers a range of features for effective gas detection:
- Identifies combustible gases including Methane, Hydrogen, Carbon Monoxide, Propane, Ethylene, Ethane, Hexane, Benzene, Iso-Butane, Ethanol, Acetaldehyde, Formaldehyde, Toluene, P-Xylene, Ammonia, and Hydrogen Sulfide.
- TRI Mode Detection: Provides visual (LEDs), audible (85dB tick), and vibration alerts.
- Five red LEDs display threshold levels of measurement.
- Two sensitivity settings: Low and High.
- 16" (406mm) adjustable gooseneck probe with replaceable heated sensor.
- Auto zero calibration in fresh air.
- Hold button to freeze measurements.
- Auto power off with disable function to conserve battery life.
4. Components and Controls
Familiarize yourself with the main components of the GSM100:
- Flexible Gooseneck Probe: Allows for precise positioning of the sensor.
- Heated Sensor: Located at the tip of the gooseneck, detects gas.
- LED Display: Five red LEDs indicate gas concentration levels.
- Sensitivity Buttons (H/L): Toggles between High and Low sensitivity.
- Auto Zero Button: Initiates fresh air calibration.
- Hold Button: Freezes the current reading on the display.
- Power Button: Turns the device on/off.
- Speaker: Emits audible alerts.
- Vibration Motor: Provides tactile alerts.

Image 4.1: The GSM100 being used to detect gas, demonstrating the flexible gooseneck probe in action.
5. Setup
5.1. Battery Installation and Charging
The GSM100 comes with a rechargeable Li-Po battery. Before first use, ensure the battery is fully charged.
- Locate the battery compartment on the device.
- Insert the Li-Po battery, ensuring correct polarity.
- Connect the provided AC charger to the charging port on the device.
- Plug the charger into a standard wall outlet. The charging indicator will illuminate.
- Allow the battery to charge completely before disconnecting.
5.2. Initial Power On and Calibration
The device performs an auto-zero calibration upon startup.
- Ensure you are in a fresh air environment, free from any gas contaminants.
- Press and hold the Power button until the device turns on.
- The device will initiate a warm-up and auto-zero process. Wait for this process to complete (typically a few seconds) before proceeding.
6. Operation
6.1. Gas Detection Procedure
- Turn on the device in a fresh air environment as described in Section 5.2.
- Select the desired sensitivity using the H/L button. 'H' indicates High sensitivity for detecting small leaks, 'L' for larger concentrations.
- Slowly move the tip of the flexible gooseneck probe towards the suspected leak area.
- Observe the LED display and listen for audible alerts. The number of illuminated LEDs and the frequency of the audible tick/vibration will increase as the gas concentration rises.
- Pinpoint the leak by identifying the area where the alerts are strongest.
6.2. Using the Hold Function
To freeze the current measurement on the LED display:
- While detecting gas, press the Hold button. The current LED indication will remain lit.
- Press the Hold button again to release the measurement and return to live detection.
6.3. Auto Power Off (APO)
The GSM100 features an Auto Power Off function to conserve battery life. If no activity is detected for a certain period, the device will automatically shut down.
- To disable APO, consult the full user manual (PDF) for specific button sequences, if available.
7. Maintenance
7.1. Cleaning
To maintain the device's performance and appearance:
- Wipe the exterior of the device with a soft, damp cloth. Do not use abrasive cleaners or solvents.
- Keep the sensor tip clean and free from dust or debris.
7.2. Sensor Replacement
The heated sensor at the tip of the gooseneck is replaceable. If the sensor's performance degrades over time, it may need to be replaced. Refer to the official Triplett documentation or contact customer support for instructions on obtaining and replacing the sensor.

Image 7.1: The Triplett GSM100 detector shown alongside its protective carrying case, emphasizing portability and care.
8. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| Device does not turn on. | Low or dead battery. | Charge the battery using the provided AC charger. |
| Inaccurate readings or no detection. | Sensor contaminated or worn out. Not calibrated in fresh air. | Perform auto-zero calibration in fresh air. Clean the sensor tip. Consider sensor replacement if issues persist. |
| Audible/vibration alerts are weak or absent. | Low battery. Speaker/vibration motor malfunction. | Charge the battery. If problem persists, contact customer support. |
9. Specifications
| Feature | Specification |
|---|---|
| Model Number | GSM100 |
| Product Dimensions | 11.4 x 5.5 x 2.4 inches |
| Item Weight | 1.06 Pounds (480 Grams) |
| Power Source | Battery Powered (1 Lithium Polymer battery, included) |
| Alarm Type | Audible, Visual (LEDs), Vibration |
| Operating Humidity | Up to 95% |
| Gooseneck Length | 16 inches (406mm) |
| Manufacturer | Triplett Test Equipment and Tools |
| Country of Origin | China |
10. What's in the Box
Your Triplett GSM100 package includes:
- Triplett GSM100 Combustible Gas Leak Detector
- Rechargeable Li-Po Battery (pre-installed or separate)
- AC Charger
11. Warranty and Support
For warranty information, technical support, or to purchase replacement parts (such as the heated sensor), please visit the official Triplett website or contact their customer service department. Retain your proof of purchase for warranty claims.
You can also refer to the official User Manual (PDF) for more detailed information, available at: Triplett GSM100 User Manual
Ask a question about this manual