1. Introduction
The MUXWELL DH-801 is a versatile programmable DC electronic load designed for testing power supplies, batteries, and other electronic components. It supports multiple operating modes including Constant Current (CC), Constant Voltage (CV), Constant Resistance (CR), and Constant Power (CP). This manual provides essential information for the safe and efficient operation of your device.
2. Safety Information
- Always connect the device to a properly grounded AC power source.
- Do not operate the device in wet or damp conditions.
- Ensure proper ventilation to prevent overheating.
- Do not exceed the maximum input voltage (150V) or current (20A) ratings.
- Disconnect power before making or changing connections.
- Refer to qualified personnel for any servicing or repairs.
3. Product Overview
3.1 Front and Rear Panels
The following image illustrates the key components and controls on the front and rear panels of the MUXWELL DH-801 electronic load.

Image: Front and rear panel layout of the MUXWELL DH-801 Programmable DC Electronic Load, showing display, control buttons, input terminals, power input, and cooling vents.
Front Panel Controls:
- Display Area: Shows settings, measured values (voltage, current, power), and other operational information.
- MODE Key: Used to select the operating mode (CC, CV, CR, CP).
- INTEGRATOR Key: Starts or stops the integrator calculation for Watt-hours (WH) and Amp-hours (AH).
- SHORT Key: Enables or disables the short-circuit test mode.
- LOAD Key: Turns the electronic load function on or off.
- Load Input Terminal +: Connects to the positive pole of the device under test.
- ROTARY/PUSH Knob: Pushing enters the parameter setting menu or switches the numeric segment. Rotating increases or decreases values in setting mode, or switches measurement display items in main menu.
- Load Input Terminal -: Connects to the negative pole of the device under test.
Rear Panel Components:
- RS-232 Connector: Reserved, no function for this model.
- AC Input Connector: For connecting the power cord to the AC power supply.
- Power Switch: Turns the device on or off.
- Air Outlet: Provides cooling for the internal components.
3.2 Dimensions
The MUXWELL DH-801 features a compact design suitable for desktop use.

Image: Side view of the MUXWELL DH-801 Programmable DC Electronic Load with dimensions indicated: approximately 7.5 inches (length), 2.2 inches (width), and 5.1 inches (height).
4. What's in the Box
Upon unpacking, ensure all the following items are present:

Image: The MUXWELL DH-801 DC Electronic Load Tester shown with its included accessories: a power cable and a set of red and black test leads.
- MUXWELL DH-801 DC Electronic Load Tester
- AC Power Cord
- Test Leads (Red and Black)
- User Manual (this document)
5. Setup
- Placement: Place the electronic load on a stable, flat surface with adequate ventilation around the air inlets and outlets.
- Power Connection: Connect the supplied AC power cord to the AC Input Connector (10) on the rear panel and then to a suitable AC power outlet.
- Device Under Test (DUT) Connection: Connect the positive (+) terminal of your DUT to the Load Input Terminal + (6) and the negative (-) terminal of your DUT to the Load Input Terminal - (8) using appropriate test leads. Ensure correct polarity.
- Power On: Flip the Power Switch (11) on the rear panel to the 'ON' position. The display area (1) should illuminate.
6. Operating Modes
The MUXWELL DH-801 supports four primary operating modes. Press the MODE key (2) to cycle through these options. Use the ROTARY/PUSH knob (7) to adjust parameters within each mode.

Image: The MUXWELL DH-801 display indicating various operating modes and measurement capabilities.
6.1 Constant Current (CC) Mode
In CC mode, the electronic load draws a constant current from the DUT, regardless of the input voltage (within the device's operating limits). This mode is ideal for testing power supplies, battery discharge, and current limiting circuits.
- Press MODE until 'CC' is displayed.
- Push the ROTARY/PUSH knob to enter parameter setting.
- Rotate the knob to set the desired current value.
- Push the knob again to confirm and exit setting.
- Press LOAD to activate the electronic load.
6.2 Constant Voltage (CV) Mode
CV mode maintains a constant voltage across the input terminals of the electronic load. This is useful for testing current sources, battery charging circuits, and voltage regulators.
- Press MODE until 'CV' is displayed.
- Push the ROTARY/PUSH knob to enter parameter setting.
- Rotate the knob to set the desired voltage value.
- Push the knob again to confirm and exit setting.
- Press LOAD to activate the electronic load.
6.3 Constant Resistance (CR) Mode
In CR mode, the electronic load simulates a fixed resistance. The current drawn will vary proportionally with the input voltage (Ohm's Law). This mode is suitable for testing power supplies under varying load conditions.
- Press MODE until 'CR' is displayed.
- Push the ROTARY/PUSH knob to enter parameter setting.
- Rotate the knob to set the desired resistance value.
- Push the knob again to confirm and exit setting.
- Press LOAD to activate the electronic load.
6.4 Constant Power (CP) Mode
CP mode maintains a constant power dissipation by adjusting the current drawn inversely to the input voltage. This is useful for testing power supplies and batteries under constant power conditions.
- Press MODE until 'CP' is displayed.
- Push the ROTARY/PUSH knob to enter parameter setting.
- Rotate the knob to set the desired power value.
- Push the knob again to confirm and exit setting.
- Press LOAD to activate the electronic load.
6.5 Short-Circuit Test Mode
The SHORT key (4) enables or disables a short-circuit test mode, allowing you to test the short-circuit protection of power sources.
7. Measurement Functions
The DH-801 provides real-time measurements and integral timing capabilities for comprehensive testing.
- Real-time Display: The LCD screen shows real-time voltage (V) and current (A) readings. Rotate the ROTARY/PUSH knob (7) in the main menu to cycle through other display items like power (W).
- Watt-hours (WH) Measurement: Measures the total energy consumed over time.
- Amp-hours (AH) Measurement: Measures the total charge capacity over time, ideal for battery testing.
- Test Duration Time: The device can track the duration of a test.
To use the integral timing capability for WH and AH measurements, press the INTEGRATOR key (3) to start or stop the calculation.
8. Specifications
Detailed technical specifications for the MUXWELL DH-801 electronic load:
| Parameter | Value |
|---|---|
| Input Voltage on Load | 0 ~ 150V DC |
| Input Current on Load | 0 ~ 20A DC |
| Input Power on Load | 0 ~ 120W |
| Timer | 1 Second ~ 99 Minutes 59 Seconds |
| Power Supply | 90 ~ 264V AC, 50/60Hz |
| Size (L x W x H) | 190 x 55 x 130 mm (7.48 x 2.17 x 5.12 inches) |
| Weight | 907 g (approx. 2 lbs) |
| Measurement Accuracy | ±0.1% (Voltage, Current), ±0.2% (Power) |
| Operating Temperature | 0 ~ 40°C |
| Certifications | CE, RoHS |
Load Accuracy
| Mode | Range | Accuracy | Resolution |
|---|---|---|---|
| Constant Current Mode | 0~20A | ±0.1% | 10mA |
| Constant Voltage Mode | 1~150V | ±0.1% | 100mV |
| Constant Resistance Mode | 0.1~999.9Ω | ±Vin/Rset * 0.2% | 12bit |
| Constant Power Mode | 0~120W | ±0.2% | 100mW |
Measurement Accuracy
| Parameter | Range | Accuracy | Resolution |
|---|---|---|---|
| Current | 0~20A | ±0.1% | 1mA (3.5A below), 10mA (3.5A above) |
| Voltage | 1~150V | ±0.1% | 10mV (35V below), 100mV (35V above) |
| Power | 0~120W | ±0.2% | 100mW |
9. Maintenance
- Cleaning: Use a soft, dry cloth to clean the exterior of the device. Do not use abrasive cleaners or solvents.
- Ventilation: Ensure the air inlets and outlets are free from dust and obstructions to maintain proper cooling.
- Storage: When not in use, store the device in a cool, dry place away from direct sunlight and extreme temperatures.
- Inspection: Periodically inspect the power cord and test leads for any signs of damage. Replace if necessary.
10. Troubleshooting
10.1 Device Does Not Power On
- Check if the power cord is securely connected to both the device and the AC outlet.
- Ensure the power switch on the rear panel is in the 'ON' position.
- Verify that the AC outlet is functional by plugging in another device.
10.2 Incorrect Readings
- Ensure test leads are properly connected to the DUT and the electronic load, with correct polarity.
- Verify that the selected operating mode and parameters are appropriate for your test.
- Check if the DUT is functioning correctly.
10.3 Load Does Not Activate
- Ensure the LOAD key (5) has been pressed to activate the electronic load function.
- Check if the input voltage or current from the DUT is within the specified operating range of the DH-801.