Sper Scientific 850099

Sper Scientific Manometer SD Card Logger Model 850099 Instruction Manual

Model: 850099

1. Introduction

The Sper Scientific Manometer SD Card Logger Model 850099 is designed for precise measurement of gauge and differential pressure in non-corrosive and non-ionized air and gas environments. This device records pressure data along with date and time directly onto standard SD memory cards, facilitating easy data transfer and analysis without specialized software.

Sper Scientific Manometer SD Card Logger 850099 front view

Image 1: Front view of the Sper Scientific Manometer SD Card Logger 850099, showing the display and control buttons.

2. Package Contents

Verify that all items listed below are present and in good condition upon unpacking:

  • Sper Scientific Manometer SD Card Logger (Model 850099)
  • Carrying Case
  • 6 AA Batteries
  • Tubing and Pipe Attachment Lugs
  • SD Card
  • Instruction Manual
Sper Scientific Manometer 850099 with included accessories in carrying case

Image 2: The manometer and its accessories, including the carrying case, batteries, tubing, and SD card.

3. Product Features

  • Measures gauge and differential pressure up to 101 PSI.
  • Accuracy: ±2% full scale.
  • Records data onto standard SD memory cards with date and time stamps.
  • Direct data upload to Excel without additional software or cables.
  • Supports 10 units of measure (e.g., PSI, kPa, mbar).
  • Push-button zero offset for accurate readings.
  • Min/Max function to capture extreme values.
  • Hold function to freeze current readings.
  • Automatic power-off feature to conserve battery life.
  • Low battery indicator.
  • Large, easy-to-read backlit LCD display.
  • Integrated fold-out easel back for convenient viewing.
  • Compatible with optional USB Power Cable 840058 or AC Adapter 840097.
Sper Scientific Manometer 850099 with fold-out easel stand

Image 3: Rear view of the manometer showing the fold-out easel stand for desktop use.

4. Setup

4.1 Battery Installation

  1. Locate the battery compartment on the back of the device.
  2. Open the battery compartment cover.
  3. Insert 6 AA batteries, ensuring correct polarity.
  4. Close the battery compartment cover securely.

4.2 SD Card Insertion

  1. Locate the SD card slot on the side of the device.
  2. Insert the provided SD card into the slot until it clicks into place.
  3. To remove, push the SD card gently inward until it springs out.

4.3 Tubing Connection

Connect the provided tubing to the P1 and P2 differential input ports located at the top of the device. P1 is for positive pressure, and P2 is for negative pressure. Ensure connections are secure to prevent leaks.

Sper Scientific Manometer 850099 top view showing P1 and P2 differential input ports

Image 4: Top view of the manometer, highlighting the P1 and P2 differential input ports.

5. Operation

5.1 Power On/Off

Press the POWER button to turn the device on. Press and hold the POWER button to turn it off. The device features an auto power-off function to conserve battery life after a period of inactivity.

5.2 Zero Offset

Before taking measurements, ensure the device is at ambient pressure and press the ZERO button to calibrate the reading to zero. This ensures accurate differential pressure measurements.

5.3 Unit Selection

Press the UNIT button to cycle through the 10 available units of measure (e.g., PSI, kPa, mbar, In H²O). The selected unit will be displayed on the LCD.

5.4 Hold Function

Press the HOLD button to freeze the current reading on the display. Press it again to release the hold and resume live measurements.

5.5 Min/Max Function

The device can display the minimum and maximum pressure values recorded during a measurement session. Refer to the full instruction manual for detailed steps on accessing and resetting Min/Max readings.

5.6 Backlight

The LCD display features a backlight for improved visibility in low-light conditions. Activate the backlight by pressing the light bulb icon button.

6. Data Logging

6.1 Recording Data

To begin recording pressure data, ensure an SD card is inserted. Press the REC button. The device will log measurements along with the date and time at predefined intervals. Press REC again to stop recording.

6.2 Data Transfer to PC

The recorded data is stored on the SD card in a format compatible with Microsoft Excel. To transfer data:

  1. Remove the SD card from the manometer.
  2. Insert the SD card into a computer's memory card slot or an external SD card reader.
  3. Open the data files directly in Microsoft Excel for analysis. No special software or drivers are required.

Additional SD cards can be used for unlimited data storage. Ensure the date and time are set correctly on the device for accurate data logging timestamps.

7. Specifications

  • Model: 850099
  • Measurement Range: Up to 101 PSI (differential and gauge pressure)
  • Accuracy: ±2% full scale
  • Units of Measure: 10 selectable units (e.g., PSI, kPa, mbar, In H²O, mm Hg)
  • Data Storage: SD Card (up to 16GB, user-supplied or included)
  • Display: Large backlit LCD
  • Power Source: 6 AA batteries (included) or optional 9V DC adapter/USB cable
  • Dimensions: 7" × 2¾" × 1¾" (178 × 70 × 44 mm)
  • Weight: 1 lb. (0.45 kg)
  • Operating Environment: Non-corrosive and non-ionized air and gas only. Not for use with liquids.

7.1 Detailed Measurement Ranges and Resolution

Table showing detailed specifications for Sper Scientific Manometer models 850097, 850098, and 850099, including range and resolution for various pressure units.

Image 5: Detailed specifications table for Manometer models, highlighting ranges and resolutions for various pressure units. Model 850099 is listed with a maximum pressure of 101 PSI.

8. Maintenance

  • Cleaning: Wipe the device with a soft, damp cloth. Do not use abrasive cleaners or solvents.
  • Battery Replacement: Replace batteries when the low battery indicator appears on the display to ensure continuous operation and data integrity.
  • Storage: When not in use for extended periods, remove the batteries and store the device in its carrying case in a cool, dry place.
  • SD Card Care: Handle SD cards carefully. Avoid bending or exposing them to extreme temperatures or moisture.

9. Troubleshooting

  • Device does not power on:
    • Check if batteries are inserted correctly with proper polarity.
    • Replace with fresh batteries.
    • Ensure the power button is pressed firmly.
  • Inaccurate readings:
    • Perform a zero offset calibration before measurement.
    • Verify that tubing connections are secure and free of leaks.
    • Ensure the device is used within its specified operating environment (non-corrosive, non-ionized air/gas).
  • Data not recording to SD card:
    • Ensure the SD card is properly inserted and not full.
    • Try a different SD card.
    • Check if the REC button was pressed to start recording.
  • Display is dim or unreadable:
    • Activate the backlight.
    • Replace batteries if the low battery indicator is on.

10. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the contact details provided in the original product packaging or visit the official Sper Scientific website. Keep your purchase receipt as proof of purchase for warranty claims.

Manufacturer: Sper Scientific

Website: www.sperscientific.com

Related Documents - 850099

Preview Sper Scientific Non-Contact Dual Use Infrared Thermometer TE-94 - Precision Temperature Measurement
Discover the Sper Scientific TE-94 NIST Certified Non-Contact Dual Use Infrared Thermometer. Offers precision, convenience, and advanced non-contact measurement for body and surface temperatures. Ideal for home, office, and industrial use.
Preview Sper Scientific 850045 Dissolved Oxygen Pen User Manual and Specifications
User manual and specifications for the Sper Scientific 850045 Dissolved Oxygen Pen. Learn about its features, how to calibrate and measure dissolved oxygen and temperature, probe maintenance, battery replacement, technical specifications, and warranty information.
Preview UVAB Light Meter 850032 Instruction Manual
Instruction manual for the Sper Scientific 850032 UVAB Light Meter, detailing its features, operation, specifications, and safety precautions for measuring UVA and UVB irradiance in industrial, scientific, and safety applications.
Preview Sper Scientific 800130 Infrared Thermometer Instruction Manual
Comprehensive instruction manual for the Sper Scientific 800130 infrared thermometer, covering operation, safety, maintenance, and specifications for accurate non-contact temperature measurement.
Preview Sper 800133 USB Humidity/Temperature Datalogger - Specifications and Features
Detailed overview and technical specifications for the Sper Scientific 800133 USB Humidity/Temperature Datalogger, designed for accurate ambient temperature monitoring and cold chain documentation.
Preview Sper Scientific Radiation Monitors 840007, 840026 Instruction Manual
This instruction manual provides detailed information on the Sper Scientific Radiation Monitors models 840007 and 840026. It covers operation, battery replacement, interpreting radiation readings, technical specifications, and basic principles of ionizing radiation. Learn how to use and understand your Geiger counter for accurate measurements.