Mastech MAS830L

MASTECH MAS830L Digital Multimeter User Manual

Model: MAS830L

1. Introduction

This manual provides detailed instructions for the safe and effective use of your MASTECH MAS830L Digital Multimeter. This device is a portable, battery-powered digital multimeter designed for measuring DC/AC voltage, DC current, resistance, diode, and continuity. It features a 2000-count display and a low battery indicator.

Please read this manual thoroughly before operating the multimeter and keep it for future reference.

2. Safety Information

Always observe the following safety precautions when using the multimeter:

  • Do not exceed the maximum input values for any range.
  • Use extreme caution when working with voltages above 30V AC RMS, 42V peak, or 60V DC. These voltages pose a shock hazard.
  • Before changing functions, disconnect the test leads from the circuit under test.
  • Do not use the meter if it appears damaged or if the insulation on the test leads is compromised.
  • Ensure the battery cover is securely closed before operation.
  • Adhere to local and national safety codes.

This device complies with CE, UL, and RoHS standards.

3. Product Overview and Components

MASTECH MAS830L Digital Multimeter front view

Figure 3.1: Front view of the MASTECH MAS830L Digital Multimeter, showing the display, rotary switch, and input jacks.

Multimeter test leads and probes

Figure 3.2: Close-up view of the test leads and probes, essential accessories for measurement.

Key Features:

  • 2000 Counts Display
  • Diode Open Voltage: 3.0V
  • Data Hold Function
  • Low Battery Display Indicator
  • Continuity Buzzer: < 70 ±30Ω
  • Display Backlight

Package Contents:

Multimeter package contents and applications

Figure 3.3: Illustration of the multimeter's main applications and included accessories.

  • MASTECH MAS830L Digital Multimeter
  • Test Leads (Red and Black)
  • 1x 9V 6F22 Battery (included)
  • Calibration Certificate
  • Quick Start Guide

4. Setup

4.1 Battery Installation

The MAS830L multimeter is powered by one 9V 6F22 battery. To install or replace the battery:

  1. Ensure the multimeter is turned OFF and disconnect all test leads.
  2. Locate the battery compartment on the back of the unit.
  3. Unscrew the retaining screw(s) and remove the battery cover.
  4. Insert the 9V battery, observing the correct polarity (+ and -).
  5. Replace the battery cover and secure it with the screw(s).

A low battery indicator will appear on the display when the battery needs replacement.

4.2 Connecting Test Leads

Connect the red test lead to the "VΩmA" input jack and the black test lead to the "COM" input jack for most measurements. For high current measurements (up to 10A), connect the red test lead to the "10A" input jack.

5. Operating Instructions

Before taking any measurement, ensure the rotary switch is set to the correct function and range.

5.1 DC Voltage Measurement (V—)

  1. Set the rotary switch to the desired DC Voltage range (e.g., 200mV, 2V, 20V, 200V, 600V).
  2. Connect the red test lead to the positive (+) side of the circuit and the black test lead to the negative (-) side.
  3. Read the voltage value on the display.

5.2 AC Voltage Measurement (V∼)

  1. Set the rotary switch to the desired AC Voltage range (e.g., 200V, 600V).
  2. Connect the test leads across the AC voltage source.
  3. Read the voltage value on the display.

5.3 DC Current Measurement (A—)

  1. Set the rotary switch to the desired DC Current range (e.g., 2mA, 20mA, 200mA, 10A).
  2. For measurements up to 200mA, connect the red lead to "VΩmA". For 10A, connect the red lead to "10A".
  3. Break the circuit and connect the multimeter in series with the load.
  4. Read the current value on the display.

5.4 Resistance Measurement (Ω)

  1. Ensure the circuit is de-energized before measuring resistance.
  2. Set the rotary switch to the desired Resistance range (e.g., 200Ω, 2kΩ, 20kΩ, 200kΩ, 2MΩ).
  3. Connect the test leads across the component to be measured.
  4. Read the resistance value on the display.

5.5 Diode Test (→|→)

  1. Set the rotary switch to the Diode Test position.
  2. Connect the red test lead to the anode and the black test lead to the cathode of the diode.
  3. The display will show the forward voltage drop. Reverse the leads to check for open circuit (OL).

5.6 Continuity Test (♥)

  1. Set the rotary switch to the Continuity Test position.
  2. Connect the test leads across the circuit or component.
  3. If the resistance is less than 70 ±30Ω, the built-in buzzer will sound, indicating continuity.

5.7 Data Hold Function (HOLD)

Press the "HOLD" button to freeze the current reading on the display. Press it again to release the hold and resume live readings.

6. Maintenance

6.1 Cleaning

Wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Ensure the meter is dry before use.

6.2 Battery Replacement

Refer to Section 4.1 for battery installation and replacement instructions. Replace the battery promptly when the low battery indicator appears to ensure accurate readings.

6.3 Fuse Replacement

The multimeter is protected by internal fuses. If the current measurement function stops working, the fuse may need replacement. This procedure should only be performed by qualified personnel. Use only fuses of the specified type and rating (e.g., MAX 200mA/250V FUSED, MAX 10A/250V FUSED).

7. Troubleshooting

Problem Possible Cause Solution
No display or dim display Low or dead battery Replace the 9V battery.
"OL" (Overload) displayed Input value exceeds selected range or meter's maximum capacity. Select a higher range or ensure the input is within the meter's specifications.
Incorrect readings Incorrect range selected, poor test lead connection, or damaged test leads. Verify range, check lead connections, or replace leads if damaged.
Current measurement not working Blown fuse. Replace the appropriate fuse (refer to Maintenance section).

8. Specifications

Detailed specifications table for MASTECH multimeters including MAS830L

Figure 8.1: Comparative specifications table for various MASTECH digital multimeters, including the MAS830L.

  • Display: 2000 Counts
  • DC Voltage: Up to 600V
  • AC Voltage: Up to 600V
  • DC Current: Up to 10A
  • Resistance: Up to 2MΩ
  • Continuity Buzzer: < 70 ±30Ω
  • Diode Test: Yes (Open Voltage 3.0V)
  • Data Hold: Yes
  • Low Battery Display: Yes
  • Power Source: 1x 9V 6F22 Battery (included)
  • Dimensions (L x W x H): 20 x 15 x 25 cm (approx. 5.7" x 3" x 1.57")
  • Weight: 1 Kilogram (approx. 0.44lb)
  • Operating Temperature: Up to 50°C
  • Safety Rating: CAT III 600V
  • Compliance: CE, UL, RoHS
  • Model Number: YQ-MAS830L

9. Support and Warranty

For technical support or inquiries, please contact your retailer or the manufacturer, Mastech.

Spare Parts Availability: 1 Year (EU: 1 Year).

Please retain your purchase receipt for warranty purposes.

Related Documents - MAS830L

Preview MASTECH® Control and Measuring Instruments Catalog 2015
Explore the comprehensive range of MASTECH® control and measuring instruments, including digital multimeters, clamp meters, insulation testers, cable testers, and more. This catalog provides detailed specifications and product information for precision electronic measurement equipment.
Preview MASTECH Catalogue 2025: Digital Multimeters, Clamp Meters & Testing Instruments
Explore the MASTECH Catalogue 2025, featuring a comprehensive range of digital multimeters, clamp meters, and electrical testing instruments. Discover detailed product specifications and selection guides for professionals and DIYers.
Preview MASTECH MAS830L Digital Multimeter Quick Start Guide
This guide provides quick start instructions for the MASTECH MAS830L Digital Multimeter, covering its specifications, safety precautions, and basic usage for voltage, current, and resistance measurements.
Preview Mastech MAS830B Digital Multimeter Quick Start Guide
A quick start guide for the Mastech MAS830B Digital Multimeter, covering its specifications, safety precautions, and basic usage for measuring voltage, current, and resistance.
Preview MASTECH MAS830B Digital Multimeter Quick Start Guide
A quick start guide for the MASTECH MAS830B Digital Multimeter, covering safety precautions, specifications, and basic measurement procedures for voltage, current, and resistance.
Preview MASTECH MAS838 Digital Multimeter Quick Start Guide
A quick start guide for the MASTECH MAS838 Digital Multimeter, covering its specifications, safety precautions, and basic usage instructions for various measurements.