1. Introduction
The Mastech MS8301B is a compact, battery-powered digital multimeter designed for measuring AC/DC voltage, AC/DC current, resistance, capacitance, frequency, duty cycle, temperature, diode, and continuity. It features a 4000-count display, backlight, and a non-contact voltage (NCV) detector, making it suitable for various electrical and electronic testing applications.
This manual provides essential information for the safe and effective operation, maintenance, and troubleshooting of your MS8301B digital multimeter. Please read it thoroughly before use.
2. Safety Information
WARNING: To avoid electrical shock, remove test leads before opening the case or battery cover. For your safety, always adhere to the following precautions:
- Always ensure the multimeter is in good working condition before use.
- Do not apply voltage or current that exceeds the maximum specified limits for each range.
- Use caution when working with voltages above 60V DC or 30V AC RMS, as they pose a shock hazard.
- Never connect the test leads across a voltage source when the rotary switch is set to current, resistance, or continuity mode.
- Ensure the test leads are properly seated in the correct input jacks for the desired measurement.
- Replace batteries promptly when the low battery indicator appears to ensure accurate readings.
- Do not operate the meter in explosive atmospheres or in the presence of flammable gases or dust.
- Always disconnect power to the circuit under test before measuring resistance, capacitance, or continuity.

3. Product Overview
The Mastech MS8301B digital multimeter is designed for ease of use and reliability. Key features include:
- Display: 4000 Counts for precise readings.
- Backlight: Improves visibility in low-light conditions.
- Non-contact Voltage Detector (NCV): For quick and safe detection of AC voltage without direct contact.

Components:
- LCD Display: Shows measurement readings, units, and function indicators.
- Rotary Switch: Selects the desired measurement function (e.g., V∼, V−, Ω, A∼, A−, TEMP, Diode, Continuity, NCV).
- Function Buttons:
- FUNC: Toggles between different functions within a single rotary switch position (e.g., AC/DC, Diode/Continuity).
- HZ%: Activates frequency or duty cycle measurement.
- NCV: Activates the Non-Contact Voltage detection mode.
- Input Jacks:
- COM: Common terminal for all measurements (connect black test lead).
- VΩTEMP: Input for voltage, resistance, capacitance, frequency, duty cycle, and temperature measurements (connect red test lead).
- mAμA: Input for milliampere and microampere current measurements (connect red test lead).
- A: Input for 10A current measurements (connect red test lead).
4. Setup
Before using your multimeter, ensure the batteries are installed and test leads are correctly connected.
4.1. Battery Installation
The MS8301B requires 2 LR44 batteries (included). To install or replace batteries:
- Ensure the multimeter is turned OFF and disconnect all test leads.
- Locate the battery compartment on the back of the unit.
- Use a screwdriver to open the battery cover.
- Insert the 2 LR44 batteries, observing correct polarity (+/-).
- Replace the battery cover and secure it with the screw.
4.2. Connecting Test Leads
Always connect the black test lead to the COM jack. Connect the red test lead to the appropriate input jack based on the measurement type:
- For voltage, resistance, capacitance, frequency, duty cycle, diode, continuity, and temperature measurements, connect the red test lead to the VΩTEMP jack.
- For milliampere or microampere current measurements, connect the red test lead to the mAμA jack.
- For 10A current measurements, connect the red test lead to the A jack.
5. Operating Instructions
This section details how to perform various measurements with your MS8301B multimeter.
5.1. DC Voltage Measurement (V−)
- Set the rotary switch to the V− position.
- Connect the black test lead to COM and the red test lead to VΩTEMP.
- Connect the test leads in parallel to the DC voltage source or component.
- Read the voltage value on the display.
5.2. AC Voltage Measurement (V∼)
- Set the rotary switch to the V∼ position.
- Connect the black test lead to COM and the red test lead to VΩTEMP.
- Connect the test leads in parallel to the AC voltage source or component.
- Read the voltage value on the display.
5.3. Resistance Measurement (Ω)
- Set the rotary switch to the Ω position.
- Connect the black test lead to COM and the red test lead to VΩTEMP.
- Ensure the circuit or component is de-energized before connecting the test leads.
- Connect the test leads across the component to be measured.
- Read the resistance value on the display.
5.4. Capacitance Measurement
- Set the rotary switch to the Capacitance position (often shared with Ω or Diode/Continuity, use FUNC button to select).
- Connect the black test lead to COM and the red test lead to VΩTEMP.
- Ensure the capacitor is fully discharged before connecting the test leads.
- Connect the test leads across the capacitor.
- Read the capacitance value on the display.
5.5. DC Current Measurement (A−, mA−, μA−)
- Set the rotary switch to the appropriate current range (A−, mA−, or μA−).
- Connect the black test lead to COM.
- For 10A measurements, connect the red test lead to the A jack. For mA/μA measurements, connect the red test lead to the mAμA jack.
- Disconnect power to the circuit. Open the circuit where current is to be measured.
- Connect the multimeter in series with the circuit.
- Apply power to the circuit and read the current value on the display.
5.6. AC Current Measurement (A∼, mA∼, μA∼)
- Set the rotary switch to the appropriate current range (A∼, mA∼, or μA∼).
- Connect the black test lead to COM.
- For 10A measurements, connect the red test lead to the A jack. For mA/μA measurements, connect the red test lead to the mAμA jack.
- Disconnect power to the circuit. Open the circuit where current is to be measured.
- Connect the multimeter in series with the circuit.
- Apply power to the circuit and read the current value on the display.
5.7. Frequency (Hz) and Duty Cycle (%) Measurement
- Set the rotary switch to a voltage or current range where frequency/duty cycle is applicable (e.g., V∼ or A∼).
- Connect the test leads as you would for voltage or current measurement.
- Press the HZ% button to toggle between frequency and duty cycle display.
- Read the frequency or duty cycle value on the display.
5.8. Diode Test (→|▶)
- Set the rotary switch to the Diode/Continuity position. Press FUNC if necessary to select Diode Test.
- Connect the black test lead to COM and the red test lead to VΩTEMP.
- Ensure the diode is disconnected from the circuit.
- Connect the red test lead to the anode and the black test lead to the cathode of the diode. The display shows the forward voltage drop.
- Reverse the test leads. The display should show 'OL' (Open Loop) for a good diode.
5.9. Continuity Test (♫)
- Set the rotary switch to the Diode/Continuity position. Press FUNC if necessary to select Continuity Test.
- Connect the black test lead to COM and the red test lead to VΩTEMP.
- Ensure the circuit or component is de-energized.
- Connect the test leads across the circuit path or component.
- If the resistance is less than approximately 50 Ω, the buzzer will sound, indicating continuity. The display will show the resistance value.
5.10. Temperature Measurement (TEMP °C/°F)
- Set the rotary switch to the TEMP position.
- Connect the K-type thermocouple (included) to the VΩTEMP and COM jacks, observing polarity.
- Place the thermocouple probe on or near the object whose temperature is to be measured.
- Read the temperature value on the display.
5.11. Non-Contact Voltage (NCV) Detection
- Set the rotary switch to the NCV position.
- Move the top end of the multimeter near a live AC voltage source (e.g., an electrical outlet or wire).
- The NCV indicator light will illuminate, and the buzzer will sound, with increasing intensity as the meter gets closer to the voltage source.
6. Maintenance
Proper maintenance ensures the longevity and accuracy of your multimeter.
6.1. Cleaning
Wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Keep the input jacks free of dust and debris.
6.2. Battery Replacement
Refer to Section 4.1 for instructions on battery replacement. Always replace batteries when the low battery indicator appears on the display.
6.3. Fuse Replacement
The current input jacks (mAμA and A) are protected by fuses. If the multimeter fails to measure current, the fuse may be blown. To replace a fuse:
- Ensure the multimeter is turned OFF and disconnect all test leads.
- Open the multimeter case (refer to the safety warnings regarding opening the case).
- Locate the blown fuse and carefully remove it.
- Replace with a fuse of the exact same type and rating (e.g., F1: 400mA/250V, F2: 10A/250V). Using an incorrect fuse can damage the meter or pose a safety hazard.
- Close the case and secure it.
7. Troubleshooting
If your multimeter is not functioning as expected, check the following common issues:
- No display or dim display: Check battery charge. Replace batteries if necessary.
- Incorrect readings:
- Ensure test leads are correctly connected to the appropriate input jacks.
- Verify the rotary switch is set to the correct measurement function and range.
- Check for damaged test leads.
- No current measurement: Check the fuses (refer to Section 6.3).
- 'OL' (Overload) displayed: The measured value exceeds the selected range. Select a higher range or ensure the circuit is within the meter's capabilities.
8. Specifications
The following table outlines the technical specifications for the Mastech MS8301B Digital Multimeter.

| Measurement | Range | Resolution | Accuracy |
|---|---|---|---|
| DC Voltage | 400mV | 0.1mV | ±(0.5% of reading + 3 digits) |
| 4V | 0.001V | ||
| 40V | 0.01V | ||
| 600V | 1V | ||
| AC Voltage | 4V | 0.001V | ±(0.8% of reading + 5 digits) |
| 40V | 0.01V | ||
| 400V | 0.1V | ||
| 600V | 1V | ||
| Resistance | 400Ω | 0.1Ω | ±(0.8% of reading + 5 digits) |
| 4kΩ | 0.001kΩ | ||
| 40kΩ | 0.01kΩ | ||
| 400kΩ | 0.1kΩ | ||
| Capacitance | 50nF | 0.01nF | ±(3.0% of reading + 5 digits) |
| 500nF | 0.1nF | ||
| 5μF | 0.001μF | ||
| 50μF | 0.01μF | ||
| DC Current | 400μA | 0.1μA | ±(1.0% of reading + 5 digits) |
| 4000μA | 1μA | ||
| 40mA | 0.01mA | ||
| 400mA | 0.1mA | ||
| AC Current | 400μA | 0.1μA | ±(1.2% of reading + 5 digits) |
| 4000μA | 1μA | ||
| 10A | 0.01A | ±(2.5% of reading + 10 digits) | |
| Frequency | 5.000Hz | 0.001Hz | ±(1.0% of reading + 5 digits) |
| 50.00Hz | 0.01Hz | ||
| 500.0Hz | 0.1Hz | ||
| 5.000kHz | 0.001kHz | ||
| Duty Ratio | 0.1%~99.9% | 0.1% | ±3% |
| Temperature | -4~1832°F | 1°F | ±(3.0% of reading + 3 digits) |
| Diode Test | 1.5V | 0.001V | Display approximate diode forward voltage drop value. |
| Continuity Test | When built-in buzzer sounds, the resistance is less than 50Ω. | Open circuit voltage: about 0.4V | |
| Overload Protection | 250V DC or AC (RMS) | ||
General Specifications:
- Display: 4000 Counts
- Power Source: Battery Powered (2 LR44 batteries, included)
- Product Dimensions: 20 x 15 x 25 cm
- Item Weight: 1 Kilogram
- Safety Rating: CAT II 600V
- Operating Temperature: 0°C to 50°C (32°F to 122°F)
- Storage Temperature: -10°C to 50°C (14°F to 122°F)
- Humidity: <80% RH (non-condensing)
- Certifications: CE, RoHS, UL
9. Warranty and Support
Mastech products are designed for reliability and performance. This product comes with a 1-year EU Spare Part Availability Duration, indicating support for replacement parts within this period.
For technical support, warranty claims, or service inquiries, please contact your retailer or the manufacturer directly. Keep your purchase receipt as proof of purchase.