OBEST 3 in1
OBEST 8-in-1 Digital Water Quality Tester Instruction Manual
Model: 3 in1
1. Introduction
This manual provides detailed instructions for the operation and maintenance of your OBEST 8-in-1 Digital Water Quality Tester. This device is designed to accurately measure multiple water quality parameters including Salinity, Specific Gravity (SG), pH, Hydrogen (H2), Total Dissolved Solids (TDS), Oxidation-Reduction Potential (ORP), Electrical Conductivity (EC), and Temperature. It is suitable for various applications such as testing drinking water, tap water, aquariums, ponds, swimming pools, and laboratory use.
2. Product Overview
Familiarize yourself with the components of your water quality tester:

Figure 2.1: Product Components. The image displays the main body of the tester with labels pointing to the battery compartment, backlit digital display, power switch, data hold/temperature switch, function switch, and the protective pen cap. An inset shows the coin cell battery installation.
- Battery Compartment: Located at the top, houses the coin cell batteries.
- Display: Backlit LCD screen showing measurement readings and indicators.
- Switch (Power Button): Turns the device on or off.
- Data Hold/Temperature Switch: Short press to hold data, long press to switch temperature units.
- Function Switch (MODE/CAL Button): Press to cycle through measurement modes, long press for calibration.
- Pen Cap: Protective cover for the electrode.
3. Setup
3.1 Battery Installation
- Unscrew the battery compartment cap at the top of the device.
- Insert three Lithium-metal coin cell batteries, ensuring correct polarity.
- Securely replace the battery compartment cap.
3.2 Initial Power On
Press the Power Button to turn on the device. The display will illuminate, showing the default measurement mode and temperature in Celsius.
4. Operating Instructions
4.1 General Operation
- Ensure the electrode is clean before each use.
- Remove the protective pen cap.
- Immerse the electrode into the water sample. Ensure the water level does not exceed the electrical limit or the height of the protective cover to prevent short circuits.
- Wait for the reading to stabilize on the display.
4.2 Mode Switching
Press the MODE/CAL button to cycle through the 8 measurement indicators: Salinity, SG, pH, H2, TDS, ORP, EC, and Temperature.
4.3 Temperature Unit Conversion

Figure 4.1: Temperature Unit Conversion. The image illustrates how to switch between Celsius (°C) and Fahrenheit (°F) by long-pressing the TEMP button.
To switch between Celsius (°C) and Fahrenheit (°F), press and hold the HOLD/TEMP button for 3 seconds. The device defaults to Celsius upon power-on.
4.4 Data Hold Function
To temporarily hold the current reading on the display, short press the HOLD/TEMP button. Press it again to release the hold and resume live readings.
4.5 Specific Measurement Parameters
4.5.1 pH Measurement

Figure 4.2: pH Testing. The image shows the tester measuring pH in a water sample, alongside a pH scale from 0 to 14, highlighting the drinking water standard of 6.5-9.5.
The pH measurement indicates the acidity or alkalinity of the water. For drinking water, a pH standard of 6.5-9.5 is generally recommended.
4.5.2 H2 (Hydrogen) Measurement

Figure 4.3: H2 Testing. The image displays the tester measuring hydrogen content in water, with a measurement range of 0-2400 ppb.
Measures the concentration of dissolved hydrogen in water. The measurement range is 0-2400 ppb.
4.5.3 EC (Electrical Conductivity) Measurement

Figure 4.4: EC Testing. The image shows the tester immersed in a glass of water, displaying an EC reading. The measuring range is 0-200000 us/cm with an accuracy of ±2% F.S.
EC measures the electrical conductivity of total ions in water. It is primarily used for testing agricultural watering water and liquid fertilizer. The measuring range is 0-200000 us/cm with an accuracy of ±2% F.S.
4.5.4 ORP (Oxidation-Reduction Potential) Measurement

Figure 4.5: ORP and TDS Testing. The image displays the tester measuring ORP and TDS. For ORP, it explains that a lower value indicates dirtier water, with a drinking water reference of +150-550mV. For TDS, it states that a higher value indicates more dissolved substances, with a measurement range of 0-100000ppm.
ORP indicates the presence of organic matter and microbiological conditions in water. A lower ORP value generally suggests poorer water quality. For drinking water, a reference ORP value is typically +150-550mV.
4.5.5 TDS (Total Dissolved Solids) Measurement
TDS reflects the concentration of minerals and other dissolved substances in the water. A higher TDS value indicates a greater amount of dissolved substances. The measurement range is 0-100000ppm.
4.5.6 Salinity Measurement

Figure 4.6: Salinity and Specific Gravity Testing. The image shows the tester measuring salinity in a glass of water, with a range of 0-20%. It also depicts the tester in seawater for Specific Gravity (S.G) testing, with a range of 1.000-1.222.
Salinity measures the salt content in water. It is a crucial physicochemical factor in aquaculture environments. The measuring range is 0-20%.
4.5.7 SG (Specific Gravity) Measurement
Specific Gravity measures the density of dissolved salts in water, particularly relevant for seawater. The measurement range is 1.000-1.222.
5. Calibration
Regular calibration ensures accurate readings. Calibration powder is included with your device.

Figure 5.1: pH Calibration Process. This diagram illustrates the five steps for calibrating the pH meter using pH 4.00, pH 6.86, and pH 9.18 buffer powders and distilled water.
5.1 pH Calibration Procedure
- Prepare buffer solutions: Empty each pack of pH 4.00, pH 6.86, and pH 9.18 powder into separate 250ml containers of distilled water. Stir gently until the powder is completely dissolved.
- Power on the water quality tester and select the pH measurement mode.
- Immerse the electrode into the pH 6.86 standard solution. Gently stir and wait for the reading to stabilize.
- Press and hold the MODE/CAL button for 6 seconds. The value will flash. Once the 6.86 reading stops flashing, the calibration for this point is complete. Rinse the probe with distilled water before proceeding.
- Repeat steps 3 and 4 for the pH 4.00 and pH 9.18 solutions. After calibrating all three solutions, the pH calibration is finished.
Note: If the measured value is unstable, ensure the probe is clean. Rinse it with clean water to remove impurities and scale before attempting calibration.
6. Maintenance
- Cleaning: After each use, rinse the electrode thoroughly with clean water to remove any residue.
- Storage: Always replace the protective pen cap after use to keep the electrode moist and protected. Store the device in a cool, dry place.
- Battery Life: The device features an automatic shutdown function after 5 minutes of inactivity to conserve battery power. Replace batteries when the low battery indicator appears on the display.
- Electrode Care: Do not allow the electrode to dry out completely. If the pH value reads 0, remove the electrode, air dry it, and reinsert it before performing a new test.
7. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| Unstable readings | Dirty electrode; need for calibration. | Rinse the probe with clean water to remove impurities. Perform calibration as described in Section 5. |
| pH reading shows 0 | Electrode issue or temporary malfunction. | Remove the electrode, air dry it completely, and then reinsert it before attempting another test. |
| Device does not power on | Dead batteries; incorrect battery installation. | Check battery polarity and ensure they are correctly installed. Replace batteries if necessary. |
| Inaccurate readings | Device requires calibration. | Perform a full calibration for the relevant parameters (e.g., pH calibration). |
8. Specifications
| Feature | Detail |
|---|---|
| Model Number | 3 in1 |
| Measurement Parameters | Salinity, SG, pH, H2, TDS, ORP, EC, Temperature |
| pH Range | 0-14 |
| EC Measuring Range | 0-200000 us/cm |
| EC Accuracy | ±2% F.S |
| H2 Measuring Range | 0-2400 ppb |
| TDS Measuring Range | 0-100000 ppm |
| Salinity Measuring Range | 0-20% |
| SG Measuring Range | 1.000-1.222 |
| Batteries | 3 x Lithium-metal (Coin Cell) |
| Material | Plastic |
| Dimensions | 21.1 x 7.9 x 4.9 cm |
| Weight | 140 g |
| IP Rating | IP67 Waterproof |
| Temperature Compensation | Automatic |
| Included Components | Water tester, Calibration powders |
| Recommended Uses | Pools, spas, laboratories, agricultural, drinking water, aquariums |

Figure 8.1: Product Packaging and Contents. The image displays the product box, the water quality tester, calibration buffer powders, and the operation manual, along with dimensions.
9. Warranty and Support
Specific warranty information is not provided in the product details. Please refer to the retailer or manufacturer's website for warranty terms and conditions.
For technical support or inquiries, please contact OBEST customer service through the contact information provided on their official website or the product packaging.