RuoShui VICTOR 05+ Loop Calibrator User Manual

High Precision LCD Voltage Current Simulation Transmitter Handheld Source Automatic Signal Generator

1. Product Overview

The RuoShui VICTOR 05+ Loop Calibrator is a high-precision handheld instrument designed for measuring and simulating DC current and DC voltage. It features a 5-digit LCD display for measured values and provides a 24V loop voltage supply, making it suitable for various industrial and scientific applications.

RuoShui VICTOR 05+ Loop Calibrator front view

Figure 1: Front view of the RuoShui VICTOR 05+ Loop Calibrator, showing the LCD display and control buttons.

Key Features:

  • Measurement accuracy: 0.02% (basic)
  • Output accuracy: 0.05% (basic)
  • 5-digit display for precise readings
  • Measures DC current and DC voltage
  • Provides 24V loop voltage supply with simultaneous current measurement
  • Analogue output for DC current and transmitter simulation
  • Current display in Am and Am% formats
  • Fast operation for step, zero, and full output
  • Auto ramp current output (4-20mA) with adjustable speeds
  • Adjustable output gain
  • Built-in self-tests and closed-case calibration

Package Contents:

Upon unboxing, verify that all the following items are included:

RuoShui VICTOR 05+ Loop Calibrator packaging and accessories

Figure 2: Contents of the product package, including the calibrator, test leads, and documentation.

  • Loop Calibrator (VICTOR 05+)
  • Packing Boxes
  • Alligator Clip
  • Industrial Test Leads
  • Cloth Bag
  • Product Manual (this document)
  • Test Report
  • Fuse

2. Components and Controls

Familiarize yourself with the various components and controls of the calibrator for effective operation.

RuoShui VICTOR 05+ Loop Calibrator labeled diagram

Figure 3: Labeled diagram of the calibrator's front panel, indicating the LCD display area, various keys, and input/output terminals.

  • LCD Display Area: Shows measured values, output settings, and unit information.
  • Power Switch Button: Turns the device on or off.
  • Output/Input Toggle Key: Switches between output and input modes.
  • Unit Selection Key: Changes the display unit (e.g., mA, mA%).
  • Single Step/Auto Switch Key: Toggles between manual step and auto step/ramp functions.
  • 25% / 100% / Start Keys: Used for quick output adjustments and initiating auto ramp.
  • Output Bit Selection Key: Selects which digit to adjust for output value setting.
  • Output Value Setting Key: Adjusts the numerical value of the output.
  • Input/Output Terminals: Connect test leads for measurement or signal output.
RuoShui VICTOR 05+ Loop Calibrator dimensions and accessories

Figure 4: Overview of the calibrator's dimensions (180mm x 90mm) and included accessories such as the protective case, pen tray, and bracket.

3. Setup

3.1. Battery Installation

The calibrator operates on AA batteries. Ensure correct polarity during installation.

  1. Locate the battery compartment on the rear of the device.
  2. Open the battery cover.
  3. Insert two AA batteries, observing the polarity markings (+/-).
  4. Close the battery cover securely.
RuoShui VICTOR 05+ Loop Calibrator battery compartment

Figure 5: Image showing the open battery compartment of the calibrator, indicating where two AA batteries are inserted.

3.2. Connecting Test Leads

Connect the appropriate test leads to the input/output terminals based on your measurement or simulation requirements.

  • For current measurement or output, connect leads to the 'mA' terminals.
  • For voltage measurement, connect leads to the 'V' and 'COM' terminals.
  • For loop power supply, use the '+LOOP' and '-mA' terminals.

4. Operating Instructions

4.1. Power On/Off

Press the green power button to turn the calibrator on. Press and hold the same button to turn it off.

4.2. Measurement Mode (Input)

To measure an external signal:

  1. Ensure the calibrator is in input mode by pressing the OUTPUT/INPUT toggle key. The display will indicate "INPUT".
  2. Connect the signal source to the appropriate input terminals (e.g., 'V' and 'COM' for voltage, 'mA' for current).
  3. Read the measured value on the LCD display.

4.3. Output Mode (Simulation)

To generate a signal:

  1. Ensure the calibrator is in output mode by pressing the OUTPUT/INPUT toggle key. The display will indicate "OUTPUT".
  2. Connect the device under test to the 'OUTPUT' terminals.
  3. Use the arrow keys and output bit selection key to set the desired output value.
  4. Press the UNIT key to switch between mA and mA% display if needed.
  5. Utilize the STEP/AUTO, 25%, 100%, and START keys for quick adjustments or auto ramp generation.

4.4. 24V Loop Power Supply

The calibrator can provide a 24V loop power supply while simultaneously measuring current.

  1. Connect the loop circuit to the '+LOOP' and '-mA' terminals.
  2. The calibrator will supply 24V and display the measured current.

5. Maintenance

5.1. Cleaning

Clean the calibrator with a soft, dry cloth. Do not use abrasive cleaners or solvents. Ensure the device is powered off before cleaning.

5.2. Storage

When not in use for extended periods, remove the batteries to prevent leakage. Store the calibrator in a cool, dry place, away from direct sunlight and extreme temperatures. The included protective case and pen tray can be used for organized storage.

Two RuoShui VICTOR 05+ Loop Calibrators with stands

Figure 6: Two calibrators shown with their integrated stands, demonstrating a stable storage or operational position.

5.3. Calibration

The calibrator features built-in self-tests and closed-case calibration. For professional calibration or service, contact authorized service centers. Do not attempt internal adjustments.

6. Troubleshooting

ProblemPossible CauseSolution
Device does not power on.Low or dead batteries; Incorrect battery installation.Replace batteries; Check battery polarity.
Inaccurate readings.Improper connection; Device needs calibration.Verify test lead connections; Contact service for calibration.
No output signal.Not in output mode; Incorrect settings; Faulty leads.Toggle to output mode; Check output value settings; Inspect test leads.
Display shows error message.Overload; Internal fault.Disconnect input/output and restart; If error persists, contact support.

7. Specifications

ParameterValue
BrandRuoShui
ModelVICTOR 05+
Basic Measurement Accuracy0.02%
Basic Output Accuracy0.05%
Display5-digit LCD
Loop Voltage Supply24V
Current Simulation RangeUp to 22mA DC
Power SourceAA Batteries (2x)
Dimensions (Approx.)180mm x 90mm
ASINB0CB5ZF6YS
UPC614166954031

8. Warranty and Support

RuoShui products are designed for reliability and performance. For warranty information, technical support, or service inquiries, please refer to the contact information provided with your purchase or visit the official RuoShui website.

Manufacturer: RuoShui

Brand Store: RuoShui Official Store on Amazon

For additional assistance, please contact your local distributor or the customer service department.

Related Documents - VICTOR 05+

Preview HTC CC-05 Multi-functional Signal Generator Operating Manual
Comprehensive operating manual for the HTC CC-05 Multi-functional Signal Generator, covering safety instructions, product description, button functions, signal types, wiring, maintenance, and production information.
Preview VICTOR Electronic Testing and Measuring Instruments Catalog | Digital Multimeters, Oscilloscopes, and More
Explore the comprehensive range of VICTOR electronic testing and measuring instruments, including digital multimeters, clamp meters, oscilloscopes, function generators, and more. High-quality, reliable equipment for professional use.
Preview GEYA GPS8-05 Voltage / Current Protector Instruction Manual
Instruction manual for the GEYA GPS8-05 Voltage / Current Protector, detailing its applications, features, technical specifications, parameter settings, wiring, and operational diagrams for household equipment protection.
Preview Digital Clamp Meter User Manual - Instructions, Specifications, and Safety
Comprehensive user manual for the Digital Clamp Meter (Model 00-05-3093), covering safety requirements, operation instructions, technical specifications, accuracy indicators, and maintenance. Learn how to measure AC/DC voltage, AC current, resistance, diode, and continuity.
Preview Teeppo PMB-05 Power Meter User Manual
User manual for the Teeppo PMB-05 Power Meter, detailing its features, operating modes, and specifications for monitoring electricity consumption.
Preview Dig Dug Schematic Package Supplement: Operation, Maintenance, and Service Manual
This document is a schematic package supplement for the Atari Dig Dug arcade game, providing detailed technical diagrams for its operation, maintenance, and service. It includes various wiring diagrams, PCB schematics (CPU, power supply, video, control panel), and component information.