1. Introduction
This manual provides essential instructions for the proper setup, operation, and maintenance of your Detecto 6560-C-AC Portable Digital Wheelchair Scale. Please read this manual thoroughly before using the scale to ensure safe and accurate measurements. This scale is designed for weighing individuals in wheelchairs, offering high capacity and portability for various medical environments.
2. Safety Information
- Always place the scale on a firm, level surface to ensure accurate readings and prevent tipping.
- Ensure the ramps are securely in place before a wheelchair is rolled onto or off the platform.
- Do not exceed the maximum weight capacity of 1,000 lb (450 kg).
- Keep the scale away from excessive moisture or extreme temperatures.
- Only use the provided AC adapter or approved power source.
- Regularly inspect the scale for any signs of damage or wear. Do not operate if damaged.
- Keep hands and feet clear of moving parts when folding or unfolding the scale.
3. Product Overview
The Detecto 6560-C-AC is a portable, low-profile digital wheelchair scale featuring a high capacity and a flexible remote display. Its design prioritizes ease of use and transport.
Key Features:
- MedVue Multifunctional Indicator with alphanumeric keypad.
- Simultaneous display of height, weight, and Body Mass Index (BMI).
- 1,000 lb (450 kg) capacity with 0.2 lb (0.1 kg) clinical accuracy.
- Two-way easy-access ramps for convenient wheelchair entry and exit.
- 2.2-inch (5.6 cm) thin platform for easy accessibility.
- Black rubber mat for enhanced traction.
- Folding column design with two guide handles and two wheels for vertical transport.
- Connectivity options: Bluetooth (BT), Wi-Fi, RS232, and USB ports.
Components:
The scale consists of a weighing platform, two integrated ramps, a folding column, and a MedVue indicator connected via a coiled cable.

Figure 1: The Detecto 6560-C-AC Wheelchair Scale shown in its fully assembled and operational state, ready for use. The ramps are extended, and the MedVue indicator is positioned at the top of the column.

Figure 2: A front view of the Detecto 6560-C-AC scale, highlighting the MedVue indicator unit positioned above the folding columns, ready to display measurements.
4. Setup
4.1 Unpacking
- Carefully remove all components from the packaging.
- Verify that all parts are present and undamaged. Contact your supplier immediately if any items are missing or damaged.
4.2 Assembly and Positioning
- Place the scale platform on a flat, stable surface.
- Unfold the vertical column from its transport position. Ensure it locks securely into place.
- Connect the MedVue indicator's coiled cable to the designated port on the scale platform.
- Position the MedVue indicator for optimal viewing. It can be placed on a tabletop or mounted on a wall using appropriate hardware (not included). The cable extends up to 14 ft (4.2 m).
- Extend the two-way ramps on both sides of the platform.

Figure 3: The Detecto 6560-C-AC scale shown in its folded, vertical transport configuration. This view highlights the compact design for storage and mobility.

Figure 4: A detailed view of the locking pin mechanism on the scale's column, demonstrating how to secure or release the folding mechanism for setup or transport.
4.3 Power Connection
- Connect the supplied AC adapter to the power input port on the scale or indicator.
- Plug the AC adapter into a suitable electrical outlet.
- The scale may also operate on internal rechargeable batteries (if equipped). Ensure batteries are charged before first use if operating without AC power.
5. Operation
5.1 Powering On/Off
- To power on, press the ON/OFF button on the MedVue indicator.
- To power off, press and hold the ON/OFF button until the display turns off.
5.2 Taking a Measurement
- Ensure the scale is powered on and displays '0.0' or '0.00' before use. If not, press the ZERO button.
- Carefully roll the wheelchair with the patient onto the scale platform using the ramps. Ensure the wheelchair is centered and fully on the platform.
- Once the weight stabilizes, the reading will be displayed on the MedVue indicator.
- To subtract the wheelchair's weight (tare), press the TARE button. The display will show '0.0' and indicate that the tare function is active. The subsequent reading will be the patient's net weight.
- To calculate BMI, input the patient's height using the alphanumeric keypad on the MedVue indicator. The scale will then display height, weight, and BMI simultaneously. Refer to the MedVue indicator manual for detailed BMI calculation steps.
- After measurement, carefully roll the wheelchair off the platform.
5.3 Connectivity (Optional)
The MedVue indicator supports Bluetooth (BT), Wi-Fi, RS232, and USB connectivity for data transfer. Refer to the MedVue indicator's specific manual for instructions on configuring and utilizing these features for electronic medical record (EMR) integration or data logging.
6. Maintenance and Cleaning
6.1 Cleaning
- Disconnect the scale from the power source before cleaning.
- Wipe down the scale platform, ramps, and indicator with a soft cloth dampened with a mild, non-abrasive cleaner or disinfectant.
- Do not use harsh chemicals, abrasive pads, or excessive liquid, as these can damage the scale's components.
- Ensure no liquid enters the indicator or other electronic parts.
6.2 Storage
- For storage or transport, fold the column into its vertical position and secure it with the locking pin.
- Store the scale in a dry, temperate environment.
7. Troubleshooting
- Scale does not power on: Check AC adapter connection. Ensure the outlet is functional. If using battery power, ensure batteries are charged.
- Inaccurate readings: Ensure the scale is on a level, stable surface. Verify nothing is touching the platform or ramps. Recalibrate if necessary (refer to MedVue indicator manual for calibration procedures).
- Display shows 'OVERLOAD': The weight on the platform exceeds the scale's maximum capacity. Remove weight immediately to prevent damage.
- Connectivity issues: Refer to the MedVue indicator manual for specific troubleshooting steps for Bluetooth, Wi-Fi, RS232, or USB connections.
8. Specifications
- Model Number: 6560-C-AC
- Capacity: 1,000 lb x 0.2 lb / 450 kg x 0.1 kg
- Platform Dimensions: 32 in x 36 in (81 cm x 91 cm)
- Product Dimensions (Operational): 44 x 15 x 46 inches (approximate, varies with display position)
- Weight: 126 Pounds
- Indicator: MedVue Multifunctional Indicator
- Connectivity: Bluetooth, Wi-Fi (optional), RS232, USB
- Power: AC Adapter (included)
- Manufacturer: Cardinal Detecto Inc.
9. Warranty and Support
For warranty information, technical support, or service inquiries, please contact Cardinal Detecto Inc. or your authorized Detecto dealer. Keep your purchase receipt and model number (6560-C-AC) readily available when contacting support.