1. Introduction
The Klein Tools MM420 is an auto-ranging, True Root Mean Square (TRMS) digital multimeter designed for precise electrical measurements. This device is capable of measuring AC/DC voltage up to 600V, AC/DC current up to 10A, and resistance up to 50 MOhms. Additionally, it offers functions for measuring temperature, capacitance, frequency, duty-cycle, and performing diode and continuity tests. This manual provides essential information for the safe and effective operation of your MM420 multimeter.

Figure 1: Front view of the Klein Tools MM420 Digital Multimeter, displaying its screen, rotary dial, and function buttons.
2. Safety Information
Always adhere to safety precautions when using electrical testing equipment. Failure to follow these instructions can result in serious injury or death. This multimeter is rated for CAT III 600V, indicating its suitability for measurements in distribution panels, circuit breakers, wiring, and lighting systems in large buildings.
- Inspect Test Leads: Before each use, check test leads for damaged insulation, exposed metal, or loose connections. Replace immediately if any damage is found.
- Proper Terminal Connection: Ensure test leads are connected to the correct input terminals for the desired measurement function. The MM420 features Lead-Alert protection, where LEDs illuminate to indicate proper test lead placement, enhancing accuracy and safety.
- Voltage Limits: Do not exceed the maximum rated input values for any function. The maximum voltage for this device is 600V AC/DC.
- Personal Protective Equipment: Always wear approved eye protection and other personal protective equipment (PPE) when working with electrical circuits.
- Working Environment: Do not use the meter in wet conditions or in the presence of explosive gas or dust.

Figure 2: Close-up of the MM420's input terminals, highlighting the Lead Alert Protection feature where LEDs indicate correct test lead insertion for safe operation.
3. Product Features and Components
The MM420 multimeter is equipped with several features designed for convenience and accuracy:
- Auto-Ranging: Automatically selects the correct measurement range, simplifying operation.
- True Root Mean Square (TRMS): Provides accurate measurements of AC voltage and current, especially for non-sinusoidal waveforms.
- Backlit LCD Display: Ensures clear readings in low-light conditions.
- Data Hold: Freezes the displayed reading for easy recording.
- Min/Max Measurements: Records minimum and maximum values during a measurement session.
- Auto Power-Off: Conserves battery life by automatically shutting off after a period of inactivity.
- Integrated Kickstand: Allows for hands-free operation by propping the meter at an angle.
- Test Lead Holders: Conveniently store test leads on the back of the meter.
- Optional Magnetic Hanger: Mount for magnetic hanger (Cat. Nos. 69445 or 69417, sold separately) for additional hands-free use.

Figure 3: The rear of the MM420, illustrating the integrated test lead storage, the foldable kickstand, and the mounting point for an optional magnetic hanger.
4. Setup
4.1. Unpacking and Inspection
Carefully remove the multimeter and all accessories from the packaging. Inspect for any signs of damage. The package should include the MM420 Digital Multimeter, test leads, 2 AAA batteries, and a K-type thermocouple.

Figure 4: The MM420 multimeter displayed with its standard accessories, including test leads, AAA batteries, and a K-type thermocouple for temperature measurements.
4.2. Battery Installation
- Ensure the multimeter is turned OFF.
- Locate the battery compartment cover on the back of the meter.
- Use a screwdriver to loosen the screw securing the cover.
- Remove the cover and insert two (2) AAA batteries, observing correct polarity (+/-).
- Replace the cover and tighten the screw securely.
4.3. Connecting Test Leads
The included test leads are essential for most measurements.
- For voltage, resistance, continuity, capacitance, frequency, and diode tests, insert the black test lead into the 'COM' (common) jack and the red test lead into the 'VΩHzCAP' jack.
- For current measurements up to 10A, insert the black test lead into the 'COM' jack and the red test lead into the '10A MAX' jack.
- For current measurements up to 500mA, insert the black test lead into the 'COM' jack and the red test lead into the '500mA MAX' jack.

Figure 5: The red and black test leads provided with the MM420 multimeter, used for making electrical connections during measurements.
5. Operating Instructions
To begin using the MM420, rotate the main dial to the desired measurement function. The auto-ranging feature will automatically select the appropriate range for most measurements. For specific functions, use the 'SEL' button to cycle through options (e.g., AC/DC voltage, diode/continuity).
5.1. Voltage Measurement (AC/DC)
- Set the rotary dial to the 'V' position. Use the 'SEL' button to switch between AC (~) and DC (–) voltage.
- Connect the black test lead to the 'COM' jack and the red test lead to the 'VΩHzCAP' jack.
- Apply the test probes across the circuit or component to be measured.
- Read the voltage value on the display.
5.2. Current Measurement (AC/DC)
- Set the rotary dial to the '10A' or 'mA/µA' position. Use the 'SEL' button to switch between AC (~) and DC (–) current.
- Connect the black test lead to the 'COM' jack. For measurements up to 10A, connect the red test lead to the '10A MAX' jack. For measurements up to 500mA, connect the red test lead to the '500mA MAX' jack.
- Open the circuit and connect the meter in series with the load.
- Read the current value on the display.
5.3. Resistance Measurement
- Set the rotary dial to the 'Ω' position.
- Connect the black test lead to the 'COM' jack and the red test lead to the 'VΩHzCAP' jack.
- Ensure the circuit is de-energized. Apply the test probes across the component.
- Read the resistance value on the display.
5.4. Continuity Test
- Set the rotary dial to the 'Ω' position and press 'SEL' until the continuity symbol (speaker icon) is displayed.
- Connect the black test lead to 'COM' and the red test lead to 'VΩHzCAP'.
- Ensure the circuit is de-energized. Apply the test probes across the component or wire.
- An audible tone indicates continuity (low resistance).
5.5. Temperature Measurement
The MM420 includes a K-type thermocouple for temperature measurements.
- Set the rotary dial to the '°C/°F' position. Use the 'SEL' button to switch between Celsius and Fahrenheit.
- Insert the K-type thermocouple into the 'VΩHzCAP' and 'COM' jacks, observing polarity.
- Place the thermocouple tip on or near the object whose temperature is to be measured.
- Read the temperature value on the display.

Figure 6: The K-type thermocouple, an accessory for the MM420, used to measure temperature by connecting it to the multimeter's input jacks.
5.6. Capacitance, Frequency, Duty Cycle, and Diode Test
These functions are accessed by rotating the dial to the appropriate position and using the 'SEL' button to cycle through the options. Refer to the on-screen symbols for guidance.

Figure 7: An electrician utilizing the Klein Tools MM420 Digital Multimeter to perform measurements within an electrical panel, demonstrating its application in real-world scenarios.
6. Maintenance
6.1. Cleaning
Wipe the meter with a dry, lint-free cloth. Do not use abrasive cleaners or solvents. Ensure the meter is off and disconnected from any circuits before cleaning.
6.2. Battery Replacement
When the battery indicator appears on the display, replace the batteries as described in Section 4.2. Prompt battery replacement ensures accurate readings and proper device function.
6.3. Storage
When not in use, store the multimeter in a dry, cool environment, away from direct sunlight and extreme temperatures. If storing for extended periods, remove the batteries to prevent leakage.
7. Troubleshooting
- No Display/Power: Check battery installation and ensure batteries are not depleted. Replace if necessary.
- Incorrect Readings: Verify test lead connections are correct for the selected function. Ensure the rotary dial is set to the appropriate measurement. Check for damaged test leads.
- 'OL' or 'OVERLOAD' on Display: The measured value exceeds the meter's range for the current setting. Select a higher range if available, or ensure the input is within the meter's maximum specifications.
- No Continuity Tone: Ensure the circuit is de-energized. Check if the component or wire has an open circuit.
8. Specifications
| Specification | Value |
|---|---|
| Brand | KLEIN TOOLS |
| Model Number | MM420 |
| Measurement Type | Digital Multimeter, TRMS, Auto-Ranging |
| AC/DC Voltage | Up to 600V |
| AC/DC Current | Up to 10A |
| Resistance | Up to 50 MOhms |
| Additional Functions | Temperature, Capacitance, Frequency, Duty Cycle, Diode Test, Continuity |
| Safety Rating | CAT III 600V |
| Power Source | 2 AAA batteries (included) |
| Item Weight | 14.6 ounces (approx. 413 grams) |
| Product Dimensions | 9.33 x 5.08 x 3.82 inches (approx. 23.7 x 12.9 x 9.7 cm) |
| Special Features | Lead-Alert, Auto-Range, Min/Max, Backlight, Auto Power-Off, Data-Hold |

Figure 8: An overview of the MM420's key specifications and features, including True RMS technology, auto-ranging capability, and measurement ranges for voltage, current, resistance, and temperature.
9. Warranty and Support
Specific warranty details for the Klein Tools MM420 Digital Multimeter are typically provided with the product packaging or can be found on the official Klein Tools website. For technical support, service, or warranty claims, please refer to the contact information provided by Klein Tools. Always ensure you have your product model number (MM420) and proof of purchase available when contacting support.