CHICIRIS ZD-2A Handheld Digital Turbidity Meter: User Manual

Model: ZD-2A

1. Introduction

The CHICIRIS ZD-2A Handheld Digital Turbidity Meter is a portable instrument designed for accurate measurement of turbidity in various water samples. It utilizes a 90° scattered light principle to provide reliable results for applications such as swimming pools, laboratories, water treatment plants, and environmental monitoring. This manual provides essential information for the proper setup, operation, maintenance, and troubleshooting of your device.

Image showing various applications of the turbidity meter, including water quality detection, liquor making, environmental protection, scientific research, pharmaceutical industry, and sewage disposal.

Figure 1.1: Wide range of applications for the ZD-2A Turbidity Meter.

Key Features:

  • Easy to Use: Designed for straightforward operation with accurate measurement results.
  • LED Display: Backlit LED display for clear data readability in various lighting conditions.
  • Durable Construction: Made from ABS material for stable performance and longevity.
  • Portable Design: Lightweight and compact for easy transport and field use.
  • Automatic Shutdown: Features an automatic power-off function to conserve battery life.

2. Safety Information

Please read and understand all safety instructions before operating the device to prevent injury or damage. Keep this manual for future reference.

  • Do not immerse the device in water or other liquids.
  • Use only the specified power adapter and batteries.
  • Avoid dropping the device or subjecting it to strong impacts.
  • Do not attempt to open or repair the device yourself. Refer to qualified personnel for service.
  • Keep the sample bottles and optical components clean to ensure accurate readings.
  • Dispose of batteries and electronic components according to local regulations.

3. Package Contents

Carefully unpack the box and verify that all items listed below are present and undamaged. If any items are missing or damaged, contact your supplier immediately.

  • 1 x CHICIRIS ZD-2A Turbidimeter
  • 1 x Power Adapter (US plug 100-240V 50/60Hz/0.1A/DC5V)
  • 3 x Sample Bottles
  • 1 x Tube (for sample transfer)
  • 1 x Storage Box
  • 1 x User Manual

Note: 5 x AA dry batteries are required but not included in the package.

Image showing the CHICIRIS ZD-2A Turbidity Meter along with its accessories: power adapter, three sample bottles, a tube, and a storage box.

Figure 3.1: Turbidimeter and included accessories.

4. Product Overview

The ZD-2A Turbidity Meter features a compact design with an intuitive interface for ease of use. Familiarize yourself with the main components:

  • LED Display: Shows measurement readings, menu options, and status indicators.
  • Key Panel: Contains buttons for power, menu navigation, measurement, and data storage.
  • Sample Compartment: Area where the sample bottle is placed for measurement.
  • Integrated Blackout Cover: A sliding cover that ensures ambient light does not interfere with measurements.
  • Power Input: Port for connecting the AC adapter.
  • Battery Compartment: Located on the underside for AA battery installation.
Image showing the CHICIRIS ZD-2A Turbidity Meter with its digital display showing 'Measuring' and 'Enter to Menu'.

Figure 4.1: Digital display of the portable turbidimeter.

Image showing the CHICIRIS ZD-2A Turbidity Meter with its integrated blackout cover open, revealing the sample compartment.

Figure 4.2: Integrated blackout cover design.

Detailed view of the turbidimeter showing the LED display, AC/DC power option, and key panel with buttons for Query, Menu, Enter, Switch, and Power.

Figure 4.3: Detail display of the turbidimeter components.

5. Setup

5.1 Power Supply

The ZD-2A Turbidity Meter can be powered by either AA batteries or the provided AC adapter.

  1. Battery Installation: Open the battery compartment on the underside of the device. Insert 5 x AA dry batteries, ensuring correct polarity. Close the compartment securely.
  2. AC Adapter Connection: Connect the provided AC adapter to the power input port on the device and plug it into a suitable power outlet (100-240V AC). The device supports dual AC/DC power.

5.2 Initial Calibration

For accurate measurements, the device requires calibration. The ZD-2A supports 5 calibration points. Refer to the detailed calibration procedure in the full manual for precise steps. Generally, this involves using standard turbidity solutions and following the on-screen prompts via the 'Menu' and 'Enter' buttons.

6. Operating Instructions

6.1 Preparing a Sample

  1. Ensure the sample bottle is clean and dry.
  2. Fill the sample bottle with the water sample to be tested, avoiding air bubbles.
  3. Wipe the exterior of the sample bottle with a lint-free cloth to remove any fingerprints or smudges.

6.2 Taking a Measurement

  1. Press the power button to turn on the device.
  2. Open the integrated blackout cover and carefully insert the prepared sample bottle into the sample compartment.
  3. Close the blackout cover completely.
  4. Navigate to the 'Measure' option using the 'Menu' and 'Switch' buttons, then press 'Enter' to initiate the measurement.
  5. The turbidity reading will be displayed on the LED screen in NTU (Nephelometric Turbidity Units).
  6. To save the reading, press the 'Store' button. Use the 'Query' button to review stored data.
  7. After measurement, open the cover, remove the sample bottle, and clean the compartment if necessary.

7. Maintenance

7.1 Cleaning

  • Device Exterior: Wipe the exterior of the turbidimeter with a soft, damp cloth. Do not use abrasive cleaners or solvents.
  • Sample Compartment: Use a soft, lint-free cloth to gently clean the inside of the sample compartment, especially the optical surfaces, to ensure accurate readings.
  • Sample Bottles: Wash sample bottles thoroughly with distilled water after each use and allow them to air dry.

7.2 Storage

When not in use, store the turbidimeter in its storage box in a cool, dry place, away from direct sunlight and extreme temperatures. If storing for an extended period, remove the AA batteries to prevent leakage.

8. Troubleshooting

If you encounter issues with your ZD-2A Turbidity Meter, refer to the following common problems and solutions:

ProblemPossible CauseSolution
Device does not power onLow or dead batteries; AC adapter not connected; Power button malfunctionReplace batteries; Ensure AC adapter is securely connected; Contact support if issue persists.
Inaccurate or inconsistent readingsDirty sample bottle; Air bubbles in sample; Sample compartment dirty; Device out of calibrationClean sample bottle; Remove air bubbles; Clean sample compartment; Perform calibration.
Display is dim or unreadableLow battery power; Display settingsReplace batteries; Check menu for display brightness settings.

If the problem persists after attempting these solutions, please contact customer support.

9. Specifications

The following are the technical specifications for the CHICIRIS ZD-2A Handheld Digital Turbidity Meter:

Image illustrating the key specifications of the turbidimeter: Range 0-200 NTU, Accuracy 0.1 NTU, and 90° scattered light principle.

Figure 9.1: Key measurement specifications.

ParameterValue
Item TypeTurbidimeter
Model NumberZD-2A (CHICIRISg4s79vg62b7954)
MaterialABS
Product SizeApprox. 21 x 7.5 cm / 8.3 x 3 in
Measuring Principle90° Scattered light
Minimum Indication0.1 NTU
Measuring Range0-200 NTU
Calibration Points5
Indication Error±6%(±2%F.S)
Repeatability≤0.5%
Zero Drift±0.5%F.S
Power SupplyDC 1.5V 5 x AA dry batteries (not included)
Adapter VoltageUS plug 100-240V 50/60Hz/0.1A/DC5V
Power ModesAC and DC dual purpose
Automatic ShutdownYes
WeightApprox. 1210g / 42.7oz
Image showing the CHICIRIS ZD-2A Turbidity Meter with its dimensions indicated as approximately 21 x 7.5 cm / 8.3 x 3 inches.

Figure 9.2: Product dimensions.

10. Warranty and Support

For specific warranty details, please refer to the documentation provided at the time of purchase or contact your retailer. Keep your proof of purchase for any warranty claims.

For technical support, troubleshooting assistance beyond this manual, or inquiries about replacement parts, please contact the manufacturer, CHICIRIS, or your authorized seller. Contact information can typically be found on the product packaging or the seller's website.

© 2023 CHICIRIS. All rights reserved.