RUISHAN RSTP2002-2kg/0.01g SP-zfh
RUISHAN Lab Precision Analytical Balance User Manual
Model: RSTP2002-2kg/0.01g SP-zfh
Introduction
This user manual provides comprehensive instructions for the setup, operation, maintenance, and troubleshooting of your RUISHAN Lab Precision Analytical Balance. Designed for high accuracy and reliability, this electronic weighing scale is suitable for a wide range of applications in laboratories, chemistry, jewelry, and other fields requiring precise measurements.
Key Features
- High-sensitivity built-in sensor for fast and accurate response.
- Large 170x170mm stainless steel weighing pan.
- Multiple unit conversions: grams (g), ounces (oz), carats (ct), pounds (lb).
- High accuracy precision: 0.01g.
- Built-in level and adjustable leveling feet for precise setup.
- Overload alarm and error indication for safe operation.
- External calibration capability.
- Large, easy-to-read high-definition LED display.
- Durable ABS body and aluminum alloy base.
- RS232 serial port for data output (data cable not included).
Setup
1. Unpacking and Inspection
Carefully remove the balance from its packaging. Inspect all components for any signs of damage. Ensure all accessories, including the power adapter and calibration weights, are present.
2. Placement and Leveling
Place the balance on a stable, level surface free from vibrations, drafts, and direct sunlight. Use the adjustable leveling feet at the bottom of the unit to ensure the built-in bubble level indicates a perfectly level position.

Adjust the four screws on the bottom to ensure the scale is placed steadily and level.
3. Power Connection
Connect the provided AC power adapter to the balance's adapter interface and plug it into a suitable power outlet. The balance can also operate on batteries (not included). Please refer to the detailed user instructions for battery model and installation method.

The balance connected to its power adapter, ready for use.
Operation
1. Basic Weighing
- Power On/Off: Press the ON/OFF button to turn the balance on or off.
- Zeroing/Taring: Place an empty container on the weighing pan. Press the TARE/CAL button (short press) to zero out the weight of the container. The display will show "0.00" or "0.000" depending on the precision.

The control panel showing ON/OFF, Print/Menu, PCS/UNIT, and Tare/Cal buttons.
2. Unit Conversion
To convert between different weighing units (g, oz, ct, lb), long press the PCS/UNIT button. The display will cycle through the available units.
Demonstration of unit conversion on the RSTP series electronic balance. Long press the PCS/UNIT button to switch between grams, ounces, carats, and pounds.
3. Counting Function
To use the counting function, short press the PCS/UNIT button. Follow the on-screen prompts to set the sample quantity (e.g., 10, 20, 50, 100 pieces) and then place the bulk items on the pan for counting.
Demonstration of the counting function on the RSTP series electronic balance. This feature allows for quick counting of multiple small items.
4. Calibration
Regular calibration ensures the accuracy of your balance. Calibration weights are included in the package.
- Ensure the scale is placed stably and level.
- Press and hold the TARE/CAL button until the display shows "CAL".
- When the balance displays a twinkling weight, place the suitable calibration weight (e.g., 500g) on the pan. The balance will automatically recognize and calibrate the current weight.
- Wait approximately 3 seconds; the number will stop flashing. Remove the calibration weight. The display will show "0.00" (or "0.000"), indicating that calibration is complete. If the balance is not reset to zero, repeat the operation.

Visual guide for the calibration process.
Demonstration of the calibration process for the RSTP series electronic balance.
Connectivity (RS232)
The balance is equipped with an RS232 serial port, allowing connection to a printer or computer for data output. Please note that the RS232 data cable is not included and must be purchased separately.

The RS232 interface for connecting to external devices.
1. Connecting to a Printer
Connect a compatible printer to the RS232 port. Short press the Print/Menu button to send weighing data to the printer.
Demonstration of connecting the electronic balance to a printer and outputting weighing data.
2. Connecting to a Computer
Connect the balance to a computer via the RS232 port. Data can be transmitted to software applications for logging and analysis. The balance supports button-triggered, stable, continuous, or R/r commands for data output.
Demonstration of connecting the precision balance to a computer and outputting weighing data, including to an Excel spreadsheet.
Maintenance
1. Cleaning
To clean the balance, use a soft, damp cloth with a mild detergent. Do not use abrasive cleaners or immerse the unit in water. The stainless steel weighing pan can be removed for easier cleaning.
2. Storage
When not in use, store the balance in a clean, dry environment away from extreme temperatures, humidity, and direct sunlight. Ensure the weighing pan is clean and free of debris.
3. Regular Checks
Periodically check the leveling of the balance and recalibrate as needed, especially if the balance has been moved or if accuracy is critical for your application.
Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| Inaccurate readings | Not level, needs calibration, unstable surface, drafts. | Adjust leveling feet, perform calibration, place on stable surface, avoid drafts. |
| Display shows "EEEE" or "OUEr" | Overload. | Remove excess weight from the pan. Do not exceed maximum capacity. |
| Display is blank or flickering | Power issue, low battery. | Check power adapter connection, replace batteries if applicable. |
| Cannot switch units or modes | Incorrect button press duration, software glitch. | Ensure correct long/short press duration. Turn off and restart the balance. |
Specifications
| Feature | Detail |
|---|---|
| Product Dimensions | 11 x 7.5 x 3 inches |
| Item Model Number | RSTP2002-2kg/0.01g SP-zfh |
| Weight Capacity | 2000g (2kg) |
| Accuracy | 0.01g |
| Pan Dimensions | 170 x 170 mm |
| Pan Material | Stainless Steel |
| Body Material | Durable ABS with Aluminum Alloy Base |
| Display Type | LED |
| Connectivity | RS232 Port |
| Preheating Time | 5-30 minutes |
| Stabilization Time | ≤5 seconds |
| Linearity Error | ±0.03g |
| Special Feature | Auto Shut Off |
Warranty and Support
RUISHAN stands behind the quality of its instruments. This product comes with a 30-day return policy and a 1-year warranty. For detailed user instructions or any technical assistance, please contact our knowledgeable customer support team.
You can also refer to the official user manual PDF for more information: Download User Manual (PDF)