Acogedor YH.YX11A

YH.YX11A Buoy Type Oxygen Inhaler User Manual

Brand: Acogedor

1. Introduction

This manual provides comprehensive instructions for the safe and effective use of the YH.YX11A Buoy Type Oxygen Inhaler. This device is primarily designed for use within medical center oxygen supply systems to facilitate oxygen inhalation for patients requiring oxygen therapy, including those with hypoxic conditions. It functions by decompressing high-pressure oxygen from a cylinder into a stable, breathable low-pressure flow.

Acogedor Buoy Type Oxygen Inhaler banner showing stable pressure, accurate flow, and clear scale.

Figure 1.1: Overview of the Acogedor Buoy Type Oxygen Inhaler, highlighting its key features such as stable pressure, accurate flow, and clear scale for precise oxygen delivery.

2. Safety Information

Please read all safety warnings and instructions carefully before operating the oxygen inhaler. Failure to do so may result in injury or damage to the equipment.

3. Components and Parts

The YH.YX11A Buoy Type Oxygen Inhaler comes with the following main components:

Diagram showing labeled components of the oxygen inhaler: Flowmeter, Pressure Gauge, Flow Control Valve, G5/8 BSPP connection, Safety Valve, Oxygen Outlet, Water Level Scale, Built-in Muffler.

Figure 3.1: Detailed diagram illustrating the various components of the oxygen inhaler, including the flowmeter, pressure gauge, flow control valve, and humidifier bottle.

4. Setup Instructions

Follow these steps for proper setup of your oxygen inhaler:

  1. Prepare the Humidifier Bottle: Unscrew the humidifier bottle from the main unit. Fill it with distilled water up to the 'PREFERRED' or 'UPPER' level indicated on the bottle. Do not overfill.
  2. Attach Humidifier Bottle: Carefully screw the filled humidifier bottle back onto the main regulator unit, ensuring it is securely tightened to prevent leaks.
  3. Connect to Oxygen Cylinder: Ensure the oxygen cylinder valve is closed. Connect the G5/8 BSPP connector of the oxygen inhaler to the outlet valve of the oxygen cylinder. Tighten the connection firmly using an appropriate wrench, ensuring a gas-tight seal.
  4. Attach Nasal Oxygen Tube: Connect one end of the nasal oxygen tube to the oxygen outlet on the humidifier bottle.
  5. Verify Connections: Double-check all connections for tightness and proper seating before opening the oxygen cylinder valve.
Oxygen inhaler attached to a blue oxygen cylinder.

Figure 4.1: The oxygen inhaler securely connected to an oxygen cylinder, ready for operation.

5. Operating Instructions

Once the device is set up, follow these steps to operate the oxygen inhaler:

  1. Open Oxygen Cylinder Valve: Slowly open the main valve of the oxygen cylinder. The pressure gauge on the inhaler will show the output pressure (0.2 MPa).
  2. Adjust Flow Rate: Turn the flow control valve (blue knob) to adjust the desired oxygen flow rate. The buoy in the flowmeter will rise to indicate the current flow in L/min.
Close-up of the flowmeter and flow control knob on the oxygen inhaler.

Figure 5.1: Detailed view of the flowmeter, showing the buoy indicating the flow rate, and the blue flow control knob for adjustment.

Close-up of the pressure gauge on the oxygen inhaler, displaying pressure in MPa.

Figure 5.2: Close-up of the pressure gauge, which indicates the regulated output pressure of the oxygen.

  • Patient Connection: Once the flow rate is set, connect the nasal oxygen tube to the patient as instructed by a healthcare professional.
  • To Stop Oxygen Flow: First, close the main valve on the oxygen cylinder. Then, allow the oxygen in the system to deplete until the pressure gauge reads zero and the flowmeter buoy drops. Finally, close the flow control valve on the inhaler.
  • 6. Maintenance and Care

    Proper maintenance ensures the longevity and safe operation of your oxygen inhaler:

    7. Troubleshooting

    This section addresses common issues you might encounter with the oxygen inhaler:

    ProblemPossible CauseSolution
    No oxygen flow / Buoy not risingOxygen cylinder valve closed; Flow control valve closed; Cylinder empty; Leak in connection.Open cylinder valve; Open flow control valve; Replace cylinder; Check and tighten all connections.
    Pressure gauge reads zero after opening cylinder valveCylinder empty; Regulator not properly connected; Faulty gauge.Replace cylinder; Re-connect regulator securely; Contact support if gauge is suspected faulty.
    Water bubbling excessively in humidifier bottleFlow rate set too high.Reduce the oxygen flow rate using the control valve.
    Unusual noise during operationLoose connection; Internal component issue.Check all connections for tightness. If noise persists, discontinue use and contact support.

    8. Technical Specifications

    Detailed specifications for the YH.YX11A Buoy Type Oxygen Inhaler:

    The Acogedor YH.YX11A Buoy Type Oxygen Inhaler with its humidifier bottle and nasal oxygen tube.

    Figure 8.1: The complete YH.YX11A Buoy Type Oxygen Inhaler unit, including the regulator, flowmeter, humidifier, and nasal cannula.

    9. Warranty and Support

    Acogedor stands behind the quality of its products. For specific warranty details and terms, please refer to the warranty card included with your purchase or visit the official Acogedor website. If you encounter any issues not covered in the troubleshooting section or require further assistance, please contact Acogedor customer support through the vendor's contact information provided at the point of purchase or on the official brand store page.

    For additional information and support, you may visit the Acogedor Store on Amazon.

    Related Documents - YH.YX11A

    Preview VEVOR Air Compressor Filter Regulator AC3010-03/AC4010-04 User Manual and Specifications
    Comprehensive user manual and specifications for VEVOR Air Compressor Filter Regulator models AC3010-03 and AC4010-04. Includes safety guidelines, installation instructions, limited warranty details, and a parts breakdown.
    Preview Comfytemp Portable Red Light Belt Pro K4052 User Manual
    Comprehensive user manual for the Comfytemp Portable Red Light Belt Pro (Model K4052), detailing its functions, usage instructions, specifications, maintenance, troubleshooting, and warnings for red light and vibration therapy.
    Preview VEVOR Air Compressor Filter Regulator AW+AF+AF3000-03 User Manual
    User manual for the VEVOR Air Compressor Filter Regulator model AW+AF+AF3000-03, detailing specifications, safety guidelines, installation instructions, limited warranty, and package contents.
    Preview VEVOR Air Compressor Filter Regulator AW+AF+AFD3000 3/8" User Manual and Specifications
    Comprehensive user manual and technical specifications for the VEVOR Air Compressor Filter Regulator, Model AW+AF+AFD3000 3/8". Includes unpacking instructions, safety guidelines, installation procedures, warranty details, parts list, and compliance information.
    Preview VEVOR Air Compressor Filter Regulator AF4+AF4+AR4+AF5 1/2" User Manual
    User manual for the VEVOR Air Compressor Filter Regulator, model AF4+AF4+AR4+AF5 1/2". Includes specifications, safety guidelines, installation instructions, and warranty information.
    Preview Comfytemp K4032 Portable Red Light Therapy Belt User Manual
    Comprehensive user manual for the Comfytemp K4032 portable red light therapy belt, covering product functions, operation, specifications, maintenance, troubleshooting, and safety guidelines.