ANENG AN8009

ANENG AN8009 True-RMS Digital Multimeter Instruction Manual

1. Introduction

The ANENG AN8009 is a compact, auto/manual ranging True-RMS digital multimeter designed for accurate measurement of various electrical parameters. It is suitable for use in automotive maintenance, electrical engineering, and educational laboratories. This manual provides detailed instructions for the safe and effective operation, setup, and maintenance of your AN8009 multimeter.

2. Safety Information

To ensure safe operation and service of the meter, please read all instructions carefully before use. Failure to observe safety warnings can result in electric shock, fire, or damage to the meter.

  • Always disconnect test leads from the circuit before changing functions.
  • Do not apply more than the rated voltage, as marked on the meter, between the terminals or between any terminal and earth ground.
  • Use extreme caution when working with voltages above 30V AC RMS, 42V peak, or 60V DC. Such voltages pose a shock hazard.
  • Keep your fingers behind the probe barriers during measurements.
  • Do not use the meter if it appears damaged or if the insulation on the test leads is compromised.
  • Ensure the battery cover is securely closed before operating the meter.
  • Replace batteries immediately when the low battery indicator appears to ensure accurate readings.

3. Product Overview

The ANENG AN8009 multimeter features a large, backlit LCD display for clear readings in various lighting conditions. It includes a rotary function switch, input jacks, and various buttons for mode selection and data hold.

ANENG AN8009 Digital Multimeter front view

Figure 3.1: Front View of ANENG AN8009 Digital Multimeter. This image displays the multimeter's front panel, showing the large digital display, rotary function switch, input jacks (VΩHz, COM, A, mA), and various function buttons. The orange casing is visible around the black main body.

ANENG AN8009 Digital Multimeter with accessories

Figure 3.2: ANENG AN8009 Digital Multimeter with Test Leads and Accessories. This image shows the ANENG AN8009 multimeter alongside its standard accessories, including red and black test leads, alligator clips, and various probe tips. A user manual is also visible in the background.

3.1 Key Components

  • LCD Display: Shows measurement readings, units, and function indicators.
  • Function Switch: Rotary dial to select measurement modes (Voltage, Current, Resistance, etc.).
  • Input Jacks: Terminals for connecting test leads (VΩHz, COM, A, mA).
  • Buttons: For features like RANGE, AUTO POWER OFF, SEL/HOLD, and backlight activation.

4. Setup

4.1 Battery Installation

The ANENG AN8009 requires two 1.5V AAA batteries (not included) for operation.

  1. Ensure the multimeter is powered off.
  2. Locate the battery compartment on the back of the meter.
  3. Use a screwdriver to open the battery compartment cover.
  4. Insert two 1.5V AAA batteries, observing the correct polarity (+ and -).
  5. Replace the battery compartment cover and secure it with the screw.

4.2 Connecting Test Leads

Always connect the black test lead to the COM (common) jack. Connect the red test lead to the appropriate input jack based on the desired measurement:

  • VΩHz: For voltage, resistance, frequency, capacitance, diode, and continuity measurements.
  • mA: For current measurements up to 600mA.
  • A: For current measurements up to 10A.

5. Operating Instructions

Turn the rotary switch to the desired function. The meter typically defaults to auto-ranging. Press the 'SEL' button to switch between functions within a single rotary switch position (e.g., AC/DC voltage, diode/continuity).

5.1 DC/AC Voltage Measurement (V, mV)

  1. Connect the black lead to COM and the red lead to VΩHz.
  2. Turn the rotary switch to the V (Voltage) position. The meter will typically default to DCV. Press 'SEL' to switch to ACV if needed.
  3. Connect the test probes in parallel to the circuit or component under test.
  4. Read the voltage value on the display.

5.2 DC/AC Current Measurement (A, mA, uA)

Caution: Never connect the meter in parallel to a voltage source when measuring current. This can blow the fuse or damage the meter. Always connect in series with the load.

  1. Connect the black lead to COM. Connect the red lead to mA for currents up to 600mA, or to A for currents up to 10A.
  2. Turn the rotary switch to the A (Current) position. Press 'SEL' to switch between DC/AC current or different ranges (uA, mA, A) if necessary.
  3. Open the circuit where current is to be measured and connect the meter in series.
  4. Read the current value on the display.

5.3 Resistance Measurement (Ω)

  1. Connect the black lead to COM and the red lead to VΩHz.
  2. Turn the rotary switch to the Ω (Resistance) position.
  3. Ensure the circuit is de-energized before measuring resistance. Connect the test probes across the component.
  4. Read the resistance value on the display.

5.4 Capacitance Measurement (nF, uF, mF)

  1. Connect the black lead to COM and the red lead to VΩHz.
  2. Turn the rotary switch to the Capacitance position.
  3. Discharge the capacitor completely before measurement. Connect the test probes across the capacitor terminals.
  4. Read the capacitance value on the display.

5.5 Frequency Measurement (Hz, KHz, MHz)

  1. Connect the black lead to COM and the red lead to VΩHz.
  2. Turn the rotary switch to the Hz (Frequency) position.
  3. Connect the test probes in parallel to the signal source.
  4. Read the frequency value on the display.

5.6 Diode Test

  1. Connect the black lead to COM and the red lead to VΩHz.
  2. Turn the rotary switch to the Diode/Continuity position. Press 'SEL' until the diode symbol appears.
  3. Connect the red probe to the anode and the black probe to the cathode of the diode.
  4. A forward voltage drop will be displayed for a good diode. Reverse the probes; the display should show 'OL' (open loop).

5.7 Continuity Test

  1. Connect the black lead to COM and the red lead to VΩHz.
  2. Turn the rotary switch to the Diode/Continuity position. Press 'SEL' until the continuity symbol appears.
  3. Connect the test probes across the circuit or component.
  4. If continuity exists (resistance below approximately 50Ω), the buzzer will sound.

5.8 Temperature Measurement

  1. Connect the temperature probe (K-type thermocouple) to the VΩHz and COM jacks, observing polarity.
  2. Turn the rotary switch to the Temp position.
  3. Place the probe tip on or near the object whose temperature is to be measured.
  4. Read the temperature value on the display (in °C or °F).

5.9 Non-Contact Voltage (NCV) Detection

  1. Turn the rotary switch to the NCV position.
  2. Move the top end of the meter close to the conductor being tested.
  3. If AC voltage is detected, the meter will emit an audible beep and the NCV indicator will light up.

5.10 Data Hold Function

Press the 'HOLD' button to freeze the current reading on the display. Press it again to release the hold function.

5.11 Backlight

Press the backlight button (often combined with 'HOLD' or 'SEL') to turn the display backlight on or off. The backlight typically turns off automatically after a short period to conserve battery life.

6. Maintenance

6.1 Cleaning

Wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Keep the input terminals free of dirt and moisture.

6.2 Battery Replacement

When the low battery indicator appears on the display, replace the batteries as described in Section 4.1. Using the meter with a low battery can lead to inaccurate readings.

6.3 Fuse Replacement

If the current measurement function fails, the fuse may need replacement. The AN8009 typically uses two fuses: a 200mA/250V fast-blow fuse for the mA input and a 10A/250V fast-blow fuse for the A input. Refer to the markings on the meter for exact fuse specifications. Fuse replacement should only be performed by qualified personnel.

  1. Ensure the meter is powered off and test leads are disconnected.
  2. Open the battery compartment and then the main casing screws (if applicable, consult the meter's physical design).
  3. Carefully remove the old fuse and replace it with a new fuse of the identical type and rating.
  4. Reassemble the meter, ensuring all screws are tightened.

6.4 Storage

If the meter is not to be used for an extended period, remove the batteries to prevent leakage and damage. Store the meter in a cool, dry place, away from direct sunlight and extreme temperatures.

7. Troubleshooting

  • No display or faint display: Check battery installation and charge. Replace batteries if necessary.
  • 'OL' (Overload) displayed: The input value exceeds the selected range or the maximum measurement capability of the meter. Select a higher range or ensure the input is within specifications.
  • Incorrect readings: Check battery level, ensure test leads are properly connected, and verify the correct function is selected. Clean input jacks if dirty.
  • Current measurement not working: Check the fuse for the respective current input (mA or A). Replace if blown.

8. Specifications

ParameterSpecification
Operating Temperature0 - 40 °C
Operating Humidity75% RH
Storage Condition-20 ~ 60 °C
Storage Humidity80% RH
Operating ModeAuto/Manual Ranging
Measuring Temperature Range-20 ~ 1000 °C / -4 ~ 1832 °F
Measuring Capacitance Range9.99nF / 99.99nF / 999.9nF / 9.99uF / 99.99uF / 999.9uF / 9.999mF
Measuring Voltage Range (DC/AC)999.9mV / 9.999V / 99.99V / 999.9V (DC) / 750V (AC)
Measuring Current Range (DC/AC)99.99uA / 999.9uA / 999.9mA / 9.999A
Measuring Resistance Range99.99Ω / 999.9Ω / 9.999kΩ / 99.99kΩ / 999.9kΩ / 99.99MΩ
Frequency Range99.99Hz / 999.9Hz / 9.999KHz / 99.99KHz / 999.9KHz / 9.999MHz
Square Wave Output50Hz - 1000Hz (various steps)
Display TypeDigital Display (9999 Counts)
Diode TestYes
ContinuityYes
Duty Cycle1% - 99%
Sample Rate3 times per second
Power2 * 1.5V AAA batteries

9. Warranty and Support

For warranty information and technical support, please refer to the documentation provided with your purchase or contact the seller/manufacturer directly. Keep your purchase receipt as proof of purchase.

Related Documents - AN8009

Preview ANENG 681 Digital Multimeter User Manual - Features, Specifications, and Operation Guide
Comprehensive user manual for the ANENG 681 Digital Multimeter. Learn about safety, specifications, test ranges, operation instructions, maintenance, and troubleshooting for this versatile measuring tool.
Preview ANENG SZ08 Digital Multimeter Operating Instructions
Operating instructions and technical specifications for the ANENG SZ08 digital multimeter, covering safety precautions, measurement methods, electrical specifications, maintenance, and battery replacement.
Preview Aneng SZ02 Digital Multimeter User Manual and Specifications
Comprehensive user manual and technical specifications for the Aneng SZ02 digital multimeter by ANENG, covering device features, usage instructions, measurement capabilities, and warranty information.
Preview ANENG 616 Digital Intelligent Multimeter: Operating Instructions and Specifications
Comprehensive operating instructions and detailed specifications for the ANENG 616 digital intelligent multimeter. Learn how to measure voltage, current, resistance, capacitance, frequency, and temperature safely and effectively.
Preview ANENG 683 Digital Multimeter User Manual and Specifications
Comprehensive user manual for the ANENG 683 digital multimeter, covering safety instructions, technical specifications, operating procedures, maintenance, and package contents. Learn how to accurately measure voltage, current, resistance, capacitance, and temperature.
Preview ANENG 683 Touch Meter User Manual: Features, Specs & Operation Guide
Comprehensive user manual for the ANENG 683 digital multimeter. Covers safety, specifications, measurement functions (voltage, current, resistance, etc.), and operational guidance for electrical testing.