1. Product Overview
The Mobiclinic Mobi 2 Anti-Decubitus Mattress system is designed to prevent and treat pressure sores (decubitus ulcers) of Grade I and II. It features an alternating air pressure system with 20 individual air cells that inflate and deflate cyclically, providing continuous pressure redistribution. This movement helps to stimulate blood circulation and reduce prolonged pressure on specific body areas, enhancing patient comfort and therapeutic effectiveness. The system includes a quiet, compact compressor for easy operation and adjustable pressure settings.
The mattress is constructed from high-quality, fire-retardant nylon and PVC medical-grade material, ensuring durability and ease of cleaning. It is suitable for individuals requiring long-term bed rest and supports a maximum weight of 140 kg.

Image 1.1: The Mobiclinic Mobi 2 Anti-Decubitus Mattress with its accompanying compressor unit. The mattress is blue with alternating air cells, and the compressor is a compact white unit with control knobs.

Image 1.2: An illustration depicting the characteristics of Grade I and Grade II pressure sores, which this mattress is designed to prevent and treat.
2. Safety Information
- Always consult a healthcare professional regarding the appropriate use of this mattress for specific medical conditions.
- Ensure the mattress is placed on a stable and flat bed surface.
- Do not use near open flames or excessive heat sources. The mattress material is fire-retardant, but caution is advised.
- Keep the compressor unit away from water or moisture to prevent electrical hazards.
- Do not puncture the mattress cells with sharp objects. A repair kit is included for minor punctures.
- Ensure the patient's weight does not exceed the maximum supported weight of 140 kg.
- Regularly inspect the mattress and compressor for any signs of damage or wear.
3. Setup Instructions
- Unpack Components: Carefully remove the mattress and compressor from their packaging.
- Place Mattress: Lay the deflated mattress flat on the bed frame. Ensure the air tubes are positioned at the foot end of the bed.
- Connect Air Tubes: Connect the two air tubes from the mattress securely to the air outlets on the compressor unit. Ensure a snug fit to prevent air leakage.
- Position Compressor: Place the compressor unit on a stable surface, either on the floor next to the bed or hanging from the footboard using its integrated hooks. Ensure it is not obstructed and has adequate ventilation.
- Power Connection: Plug the compressor's power cord into a standard electrical outlet (AC 220V 50 Hz).
- Initial Inflation: Turn the compressor ON using the power switch. The mattress will begin to inflate. Allow approximately 20-30 minutes for the mattress to fully inflate and the alternating pressure cycle to stabilize before placing the patient on it.
- Adjust Pressure: Once inflated, adjust the pressure setting on the compressor using the rotary knob. The ideal pressure should allow a hand to be comfortably slipped between the patient's body and the mattress when the cells beneath are deflated. Refer to the compressor's markings for guidance (e.g., weight-based settings).

Image 3.1: Close-up of the compressor unit, showing the power switch, pressure adjustment knob, and air tube connection ports. The compressor is designed for quiet operation.
4. Operating Instructions
The Mobiclinic Mobi 2 mattress operates automatically once set up. The compressor continuously inflates and deflates alternating air cells to provide dynamic pressure relief.
- Power On/Off: Use the ON/OFF switch on the compressor to start or stop the system.
- Pressure Adjustment: The rotary knob on the compressor allows you to fine-tune the air pressure within the mattress cells. Adjust this based on the patient's weight and comfort, ensuring effective pressure redistribution.
- Alternating Cycle: The system automatically cycles air through the mattress cells every 6 minutes, ensuring that different areas of the patient's body are relieved of pressure periodically.

Image 4.1: An infographic illustrating how the alternating air cells provide pressure relief, showing areas of the body being supported and relieved in sequence.
5. Maintenance and Cleaning
- Cleaning the Mattress: The mattress is made of PVC and nylon, which is easy to clean. Wipe the surface with a damp cloth and a mild disinfectant solution. Do not submerge the mattress in water. Ensure the mattress is completely dry before use.
- Compressor Care: Keep the compressor unit clean and free from dust. Wipe with a dry cloth. Do not use liquid cleaners directly on the compressor.
- Storage: If storing the mattress, ensure it is deflated, clean, and dry. Fold it neatly and store it in a cool, dry place away from direct sunlight and sharp objects.
- Repair Kit: A repair kit is included for minor punctures. To repair, locate the puncture, clean the area, apply adhesive from the kit, and cover with a patch. Allow sufficient drying time before re-inflating.

Image 5.1: The Mobiclinic Mobi 2 mattress folded, demonstrating its portability and ease of storage after cleaning and deflation.
6. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| Mattress not inflating or deflating properly. |
|
|
| Compressor is unusually noisy. |
|
|
| Mattress feels too hard or too soft. |
|
|
7. Product Specifications
| Feature | Detail |
|---|---|
| Model | Mobi 2 |
| Mattress Dimensions (L x W x H) | 200 x 86 x 9.5 cm |
| Mattress Weight | 3.3 kg |
| Maximum Supported Weight | 140 kg |
| Number of Air Cells | 20 |
| Material | Medical-grade Nylon and PVC, fire-retardant |
| Recommended for Pressure Sore Grade | I and II |
| Compressor Dimensions (L x W x H) | 23 x 11.5 x 10 cm |
| Compressor Weight | 1 kg |
| Power Supply | AC 220V 50 Hz, 0.1A |
| Air Flow | 6-8 liters/minute |
| Applied Pressure | 40-100 mmHg |
| Cycle Duration | 6 minutes |
| Noise Level | 33-34 dB |

Image 7.1: Visual representation of the Mobiclinic Mobi 2 mattress dimensions, highlighting its length (200 cm), width (86 cm), and height (9.5 cm).
8. Warranty and Support
Mobiclinic is committed to providing high-quality products and customer satisfaction. For any questions, technical assistance, or warranty claims, please contact Mobiclinic customer support through their official website or the retailer where the product was purchased.
Please retain your proof of purchase for warranty purposes. The product is designed for durability and includes a repair kit for minor issues, ensuring long-term usability.
Note: Anti-decubitus mattresses cannot be returned after being inflated, unless the original packaging is intact for return processing.