1. Introduction

This Benetech GM63B Vibration Meter utilizes the piezoelectric effect of polarized ceramics for its design. It is engineered for monitoring various vibrating mechanical facilities, particularly for rotating and reciprocating machinery. The device measures acceleration, velocity, and displacement, making it suitable for applications in mechanical manufacturing, electric power metallurgy, and general aviation.

2. Features

  • LCD display shows measurement results and conditions directly.
  • Measures acceleration (m/s² peak), velocity (mm/s rms), and displacement (mm p-p).
  • Offers selective vibration characteristics (Hi-Low frequency selection).
  • Incorporates a high-sensitivity vibration sensor for accurate measurements.
  • Equipped with two probes (standard and magnetic) to adapt to different measurement requirements.
  • Provides a magnetic probe for convenient use in situations where manual holding is difficult.
  • Includes a low battery indication.
  • Features an auto turn-off function.
  • LCD backlight function for visibility in various lighting conditions.
  • Maximum value hold function.
  • Temperature unit selection (°C / °F).
  • A/C output for external connection.

3. Package Contents

The Benetech GM63B Vibration Meter package typically includes the following items:

  • Benetech GM63B Vibration Meter Unit
  • Standard Vibration Probe
  • Magnetic Vibration Probe
  • Instruction Manual
  • Protective Carrying Case
Benetech GM63B Vibration Meter kit in its protective carrying case.
The complete Benetech GM63B Vibration Meter kit, including the meter, probes, and manual, neatly organized within its protective carrying case.
Benetech GM63B Vibration Meter with standard and magnetic probes.
The Benetech GM63B Vibration Meter shown with its standard probe connected and the magnetic probe alongside, ready for various measurement applications.

4. Product Overview

The Benetech GM63B Vibration Meter consists of a main unit with an LCD display and control buttons, and a detachable vibration sensor (probe). The sensor connects to the main unit via a cable.

Benetech GM63B Vibration Meter connected to the standard vibration sensor.
The Benetech GM63B Vibration Meter main unit with the standard vibration sensor connected, illustrating the device's compact design.
Close-up view of the Benetech GM63B Vibration Meter's LCD display and control buttons.
A detailed view of the Benetech GM63B's LCD screen, displaying measurement values, and the clearly labeled control buttons for operation.

5. Setup

5.1 Battery Installation

  1. Locate the battery compartment on the back of the meter.
  2. Open the battery compartment cover.
  3. Insert a 9V Alkaline battery, ensuring correct polarity.
  4. Close the battery compartment cover securely.

5.2 Sensor Connection

  1. Identify the sensor input port on the top or side of the main unit.
  2. Connect the desired vibration probe (standard or magnetic) to the sensor input port. Ensure a firm and secure connection.

6. Operating Instructions

6.1 Power On/Off

Press the ON/MEAS button to power on the device. The meter will automatically turn off after 60 seconds of inactivity to conserve battery life. To manually turn off, press and hold the ON/MEAS button.

6.2 Measurement Mode Selection

Press the A/V/D button to cycle through the measurement modes: Acceleration (m/s²), Velocity (mm/s), and Displacement (mm). The selected mode will be indicated on the LCD display.

6.3 Frequency Range Selection

Press the HI/LO button to switch between high and low frequency measurement ranges, as required by your application.

6.4 Maximum Value Hold

Press the MAX button to hold the maximum measured value on the display. Press it again to exit this function.

6.5 Temperature Unit Selection

Press the °C/°F button to toggle the temperature display between Celsius and Fahrenheit.

6.6 Taking Measurements

  1. Ensure the correct probe is attached and the meter is powered on.
  2. Select the desired measurement mode (Acceleration, Velocity, or Displacement) and frequency range (Hi/Lo).
  3. Firmly place the tip of the probe against the surface of the object to be measured.
  4. The measurement will be displayed on the LCD screen.

7. Maintenance

7.1 Cleaning

Wipe the meter and probes with a soft, dry cloth. Do not use abrasive cleaners or solvents, as they may damage the device.

7.2 Storage

When not in use, store the Benetech GM63B Vibration Meter and its accessories in the provided protective carrying case. Store in a dry environment, away from extreme temperatures and direct sunlight.

7.3 Battery Replacement

When the low battery indicator appears on the display, replace the 9V battery promptly to ensure accurate measurements and proper device function. Refer to section 5.1 for battery installation instructions.

8. Troubleshooting

  • No display or device does not power on:
    Check if the 9V battery is installed correctly and has sufficient charge. Replace the battery if necessary.
  • Inaccurate or inconsistent readings:
    Ensure the vibration probe is securely connected to the main unit. Verify that the probe is firmly placed against the object being measured. Check if the correct measurement mode (A/V/D) and frequency range (Hi/Lo) are selected for the application.
  • LCD backlight not working:
    The backlight function is typically automatic or activated with specific button presses. If it fails, ensure the battery has sufficient power.

9. Specifications

The following table details the technical specifications of the Benetech GM63B Vibration Meter:

Table detailing the technical specifications of the Benetech GM63B Vibration Meter.
A comprehensive table outlining the technical parameters and specifications of the Benetech GM63B Vibration Meter.
ParameterSpecification
Vibration pickupPiezoelectric ceramic accelerometer (shear-type)
Measurement range of acceleration0.1 ~ 199.9 m/s² peak
Measurement range of velocity0.1 ~ 1999.9 mm/s rms
Measurement range of displacement0.001 ~ 1.999 mm p-p (Velocity and displacement range is limited by acceleration 199.9 m/s²)
Measurement accuracy±5% ±2 digits
Measurement frequency range of acceleration10 Hz ~ 1 KHz (LO), 1 KHz ~ 15 KHz (HI)
Measurement frequency range of velocity10 Hz ~ 1 KHz (LO)
Measurement frequency range of displacement10 Hz ~ 1 KHz (LO)
Displays update cycle1 second
LCD display3 1/2 digits display
Single outputAC output 2 V peak (display full scale), Load impedance 10 KΩ or more, earphones can be connected
Power supply9V battery
Stand-by current≤ 15 µA
Operating current≤ 25 mA
Battery life20 H continuous use
Auto power-off functionTurns off automatically after 60 seconds
LCD backlight function7 seconds
Operating temperature range0 ~ 40°C
Operating humidity range30 ~ 90% RH
Low battery indication6.4 V ± 0.2 V
Dimensions72 x 35 x 145 mm
Weight229 g (not including battery)
Temperature range-10 ~ 80°C
Temperature accuracy±2°C

10. Warranty and Support

For technical assistance, warranty claims, or any product-related inquiries, please contact Benetech customer service. Refer to the official Benetech website or your purchase documentation for contact information and detailed warranty terms.