1. Introduction
The I-TECH Mag 2000 Premium is a low-frequency, high-intensity magnetotherapy device designed for therapeutic applications. This manual provides essential information for the safe and effective operation, maintenance, and troubleshooting of your device. Please read this manual thoroughly before first use and retain it for future reference.
2. Safety Information
Important Safety Precautions:
- Consult a healthcare professional before using this device, especially if you have pre-existing medical conditions, are pregnant, or have implanted electronic devices (e.g., pacemakers).
- Do not use the device near water or in damp environments.
- Keep the device out of reach of children.
- Use only the power adapter and accessories supplied by the manufacturer.
- Do not open or attempt to repair the device yourself. Refer all servicing to qualified personnel.
- Discontinue use and consult a physician if you experience any adverse reactions or discomfort during treatment.
3. Package Contents
Verify that your package contains the following items:
- I-TECH Mag 2000 Premium device
- Power adapter
- Applicator(s) / Solenoid(s)
- User Manual
- 4 non-standard batteries (pre-installed or separate)
4. Device Overview
The Mag 2000 Premium features a clear display screen and intuitive controls for easy program selection and intensity adjustment. It is equipped with two output channels, allowing for simultaneous treatment of different body areas.
- Display Screen: Shows program information, treatment time, and intensity levels.
- Control Buttons: Include 'OK' for confirmation, and arrow buttons for navigation and adjustment.
- Output Channels: Two ports for connecting applicators/solenoids.
- Power Button: For turning the device on and off.
5. Setup
- Power Connection: Connect the power adapter to the device and then to a suitable electrical outlet. The device can also operate on its internal batteries.
- Applicator Connection: Plug the applicator(s) into the output channels on the device. Ensure a secure connection.
- Initial Power-On: Press and hold the power button to turn on the device. The display will illuminate, and the main menu will appear.
6. Operating Instructions
6.1 Program Selection
The Mag 2000 Premium offers a variety of pre-set programs:
- 20 Pre-set Programs: Designed for common conditions such as osteoporosis, osteoarthritis, cervical arthritis, joint pain, epicondylitis, lower back pain, knee osteoarthritis, and periarthritis.
- 14 Pre-set Programs with Variable Frequencies: These programs operate with different frequencies ranging from 1 to 100 Hertz.
- 1 Autoscan Program: Features continuous frequency variation for treating both soft and hard tissues within the same therapy session, suitable for more complex conditions.
Use the arrow buttons to navigate through the program list on the display and press 'OK' to select a program.
6.2 Intensity Adjustment
The intensity of the magnetic field can be adjusted up to 300 Gauss in total. When using both channels, the intensity is distributed, with a maximum of 150 Gauss per channel. Adjust the intensity using the arrow buttons as indicated on the display.
6.3 Dual Channel Operation
The device features two independent output channels, allowing you to treat two different body areas simultaneously with the same selected therapy program.
6.4 Battery Information
The device includes 4 non-standard batteries, providing an approximate battery life of 8 weeks under typical usage conditions. Ensure batteries are properly installed and charged for optimal performance.
7. Maintenance
- Cleaning: Clean the device and applicators with a soft, damp cloth. Do not use abrasive cleaners or immerse the device in water.
- Storage: Store the device in a cool, dry place, away from direct sunlight and extreme temperatures.
- Battery Replacement: If the device is not used for an extended period, remove the batteries to prevent leakage. Replace batteries as needed.
8. Troubleshooting
If you encounter issues with your Mag 2000 Premium, refer to the following common solutions:
- Device not turning on: Check the power adapter connection or ensure batteries are correctly installed and charged.
- No output from applicators: Verify that applicators are securely connected to the device and that the program is active. Check intensity settings.
- Display issues: If the display is blank or unreadable, try restarting the device. If the problem persists, contact customer support.
For problems not listed here, or if troubleshooting steps do not resolve the issue, please contact customer support.
9. Specifications
| Model Name | Mag 2000 Premium |
| Model Number | 8019781011077 |
| Manufacturer | I.A.C.E.R. S.r.l. |
| Dimensions (L x W x H) | 18 x 11 x 5 cm |
| Weight | 1.74 kilograms |
| Color | White |
| Batteries | 4 non-standard batteries (included) |
| Battery Life | Approximately 8 weeks |
| Intensity | Up to 300 Gauss (150 Gauss per channel) |
| Frequency Range | 1 to 100 Hertz (for specific programs) |
10. Warranty and Support
The I-TECH Mag 2000 Premium device is covered by a manufacturer's warranty. Please refer to the warranty card included with your product for specific terms and conditions. For technical support, service, or warranty claims, please contact your authorized dealer or the manufacturer directly.
Contact Information: Please refer to the product packaging or the manufacturer's official website for the most current support contact details.