Introduction

This manual provides essential information for the safe and effective use of the Laerdal Compact Suction Unit 4 (LCSU 4), 800 mL model. The LCSU 4 is designed as a portable and robust suction device, crucial for emergency medical applications. Please read this manual thoroughly before operating the device to ensure proper function and patient safety.

Package Contents

Upon unpacking, verify that all components listed below are present and undamaged:

Laerdal LCSU 4 Compact Suction Unit with 800 mL canister and tubing

Image: The Laerdal LCSU 4 Compact Suction Unit, showing the main yellow unit, the clear 800 mL collection canister, and connected tubing. A blue wire stand supports the canister, and a blue handle is visible on top of the main unit.

Setup Instructions

  1. Battery Installation: Ensure the rechargeable battery is properly inserted into the battery compartment of the LCSU 4 unit. Charge the battery fully before first use using the provided AC/DC adapter.
  2. Canister Assembly: Place the 800 mL disposable canister into the wire stand. Ensure the canister lid is securely fastened.
  3. Tubing Connection:
    • Connect one end of the vacuum tube to the vacuum port on the LCSU 4 unit.
    • Connect the other end of the vacuum tube to the "VACUUM" port on the canister lid.
    • Connect the patient tube to the "PATIENT" port on the canister lid.
  4. Power Connection: For continuous operation or charging, connect the AC/DC adapter to the unit's power input and then to a suitable electrical outlet.

Operating the LCSU 4

  1. Power On: Press the power button to turn on the unit. The indicator light will illuminate.
  2. Vacuum Adjustment: Use the vacuum regulator knob to set the desired vacuum level. The vacuum gauge will display the current suction pressure.
  3. Patient Connection: Attach the appropriate suction catheter (not included) to the free end of the patient tube.
  4. Suction Procedure: Follow standard medical protocols for suctioning. Ensure the catheter is properly inserted before applying suction.
  5. Power Off: After use, press the power button to turn off the unit.

Important Considerations:

Maintenance and Care

Troubleshooting Guide

ProblemPossible CauseSolution
No Suction / Low Suction
  • Unit not powered on.
  • Loose tubing connections.
  • Canister lid not sealed properly.
  • Canister full.
  • Blocked patient tube/catheter.
  • Vacuum regulator set too low.
  • Battery low or depleted.
  • Turn on unit.
  • Check and secure all connections.
  • Ensure lid is tightly sealed.
  • Replace canister.
  • Clear blockage or replace.
  • Adjust vacuum regulator.
  • Charge battery or connect to AC power.
Unit is unusually noisy
  • Normal operating noise.
  • Internal component issue.
  • Some operational noise is normal.
  • If excessive or new, contact Laerdal support.

Technical Specifications

Model Number88005101
Canister Volume800 mL
Dimensions (Approx.)12 x 12 x 12 inches (Package)
Weight (Approx.)10 Pounds (Package)
ManufacturerLaerdal Medical Corporation
ASINB01MT48JMR
First Available DateDecember 29, 2016

Warranty and Support

For warranty information, technical support, or service inquiries regarding your Laerdal LCSU 4 unit, please contact Laerdal Medical Corporation directly.

Manufacturer Contact: (845) 297-7770

Please have your model number (88005101) and purchase date available when contacting support.

Related Documents - 88005101

Preview Laerdal Little Anne PLUS Important Product Information
Comprehensive product information for the Laerdal Little Anne PLUS CPR training manikin, covering setup, maintenance, safety, specifications, and regulatory compliance.
Preview Laerdal Heartstart 911 Technical Manual and Circuit Diagrams
Comprehensive technical manual and circuit diagrams for the Laerdal Heartstart 911 defibrillator, detailing specifications, functional descriptions, service procedures, and mechanical construction for maintenance and repair.
Preview CPRmeter 2 Battery Hatch and Lanyard Replacement Guide | Laerdal
Comprehensive guide for replacing the battery hatch and optional lanyard on the Laerdal CPRmeter 2, providing step-by-step instructions and visual descriptions.
Preview Laerdal SimPad System User Guide
User guide for the Laerdal SimPad System, including SimPad, Link Box, SimPad PLUS, and Link Box PLUS. Covers setup, operation in automatic and manual modes, QCPR feedback, simulated patient monitor integration, network connectivity, software, sounds, batteries, and features.
Preview Laerdal Multi-Venous IV Training Arm User Guide | Medical Simulation
User guide for the Laerdal Multi-Venous IV Training Arm, a lifelike reproduction designed for peripheral intravenous therapy training. Covers adult male, adult female, and pediatric versions.
Preview Little Baby QCPR: Important Product Information and Usage Guide
Learn about the Laerdal Little Baby QCPR, an infant CPR training manikin. This guide provides essential product information, usage instructions, maintenance, and specifications.