1. Introduction

The Tekpower Mastech MAS343 is a high-performance digital multimeter designed for accurate measurement of various electrical parameters. It features an RS232 interface, enabling connection to a computer for real-time data monitoring, logging, and analysis. This device offers both auto-ranging and manual-ranging capabilities, making it versatile for different measurement needs. It is suitable for professional electricians, hobbyists, and educational purposes.

2. Safety Information

Always observe safety precautions when using any electrical testing equipment. Failure to do so may result in electric shock, injury, or damage to the meter or equipment under test.

General Safety Guidelines:

  • Read and understand this manual completely before use.
  • Do not exceed the maximum input values for any function.
  • Ensure the function switch is in the correct position for the desired measurement.
  • Inspect test leads for damage before each use. Replace if insulation is compromised.
  • Do not operate the meter if it appears damaged or is not functioning properly.
  • Use caution when working with voltages above 30V AC RMS, 42V peak, or 60V DC, as these pose a shock hazard.
  • Keep fingers behind the finger guards on the test probes during measurements.
  • Remove test leads from the circuit before changing functions.

Electrical Safety:

  • Never apply voltage to the meter when the function switch is set to current, resistance, capacitance, or diode/continuity test.
  • Be careful when making measurements on circuits with high voltage or current.
  • Always disconnect power to the circuit before measuring resistance, capacitance, or continuity.
  • Discharge all capacitors before measuring capacitance.

3. Package Contents

Verify that all items listed below are present in your package:

  • MAS343 Digital Multimeter: 1 unit
  • Test Leads: 1 pair (red and black)
  • 9V Battery: 1 (pre-installed or included)
  • Software CD: 1
  • RS232 Cable: 1
  • User Manual: 1
  • Storage Box: 1
Tekpower Mastech MAS343 Digital Multimeter and accessories in its storage box

Image showing the Tekpower Mastech MAS343 Digital Multimeter along with its included accessories: test leads, RS232 cable, software CD, and user manual, all neatly arranged within its storage box.

4. Product Overview

Controls and Display:

The MAS343 features a clear LCD with backlight and a rotary switch for function selection. Key components include:

  • LCD Display: 3999 counts digital display with 38-segment bar graph. Digit height is 15mm. Includes backlight for low-light conditions.
  • Function Rotary Switch: Selects measurement modes (Voltage, Current, Resistance, Capacitance, Frequency, Diode, Continuity, hFE).
  • Range Button: Toggles between auto-ranging and manual-ranging modes.
  • Data Hold Button: Freezes the current reading on the display.
  • Backlight Button: Activates/deactivates the display backlight.
  • Input Jacks:
    • COM Jack: Common input for all measurements (negative terminal).
    • VΩmA Jack: Input for Voltage, Resistance, Capacitance, Frequency, Diode, Continuity, and current up to 400mA (positive terminal).
    • 10A Jack: Input for current measurements up to 10A (positive terminal).
    • RS232C Interface: Port for connecting the meter to a computer using the provided RS232 cable.

5. Setup

5.1 Battery Installation:

The MAS343 uses one 9V (6F22) battery. If the battery is not pre-installed or needs replacement, follow these steps:

  1. Ensure the meter is turned off and test leads are disconnected.
  2. Locate the battery compartment cover on the back of the meter.
  3. Unscrew the retaining screw(s) and remove the cover.
  4. Connect the 9V battery to the battery clip, observing correct polarity.
  5. Place the battery into the compartment and replace the cover, securing it with the screw(s).

5.2 Connecting to a Computer (RS232 Interface):

The MAS343 can be connected to a computer for data logging and real-time monitoring.

  1. Install the provided software from the CD onto your computer. Follow the on-screen instructions.
  2. Connect one end of the RS232 cable to the RS232C port on the multimeter.
  3. Connect the other end of the RS232 cable to an available serial port on your computer. If your computer does not have an RS232 serial port, a USB-to-RS232 adapter (not included) may be required.
  4. Launch the installed software. Refer to the software's help documentation for specific instructions on data acquisition and analysis.

6. Operating Instructions

Before taking any measurement, ensure the meter is properly set up and test leads are connected correctly.

6.1 Measuring DC Voltage (V DC):

  1. Insert the red test lead into the VΩmA jack and the black test lead into the COM jack.
  2. Set the rotary switch to the desired DC Voltage range (e.g., 4V, 40V, 400V, 1000V). If the voltage is unknown, start with the highest range or use auto-ranging.
  3. Connect the test probes across the component or circuit to be measured.
  4. Read the voltage value on the display. Observe polarity.

6.2 Measuring AC Voltage (V AC):

  1. Insert the red test lead into the VΩmA jack and the black test lead into the COM jack.
  2. Set the rotary switch to the desired AC Voltage range (e.g., 4V, 40V, 400V, 750V).
  3. Connect the test probes across the component or circuit to be measured.
  4. Read the voltage value on the display.

6.3 Measuring DC Current (A DC):

  1. For currents up to 400mA: Insert the red test lead into the VΩmA jack and the black test lead into the COM jack.
  2. For currents up to 10A: Insert the red test lead into the 10A jack and the black test lead into the COM jack.
  3. Set the rotary switch to the appropriate DC Current range (e.g., 4mA, 400mA, 10A).
  4. Disconnect power to the circuit. Open the circuit where current is to be measured.
  5. Connect the meter in series with the circuit. Apply power.
  6. Read the current value on the display.
  7. Disconnect power, remove the meter, and close the circuit.

6.4 Measuring AC Current (A AC):

  1. Follow the same lead connection and series measurement procedure as for DC Current.
  2. Set the rotary switch to the appropriate AC Current range (e.g., 4mA, 400mA, 10A).
  3. Read the current value on the display.

6.5 Measuring Resistance (Ω):

  1. Insert the red test lead into the VΩmA jack and the black test lead into the COM jack.
  2. Set the rotary switch to the Resistance (Ω) function.
  3. Ensure the circuit or component is de-energized and all capacitors are discharged.
  4. Connect the test probes across the component.
  5. Read the resistance value on the display.

6.6 Measuring Capacitance (F):

  1. Insert the red test lead into the VΩmA jack and the black test lead into the COM jack.
  2. Set the rotary switch to the Capacitance (F) function.
  3. Ensure the capacitor is fully discharged before connecting the probes.
  4. Connect the test probes across the capacitor terminals.
  5. Read the capacitance value on the display.

6.7 Measuring Frequency (Hz):

  1. Insert the red test lead into the VΩmA jack and the black test lead into the COM jack.
  2. Set the rotary switch to the Frequency (Hz) function.
  3. Connect the test probes across the signal source.
  4. Read the frequency value on the display.

6.8 Diode Test:

  1. Insert the red test lead into the VΩmA jack and the black test lead into the COM jack.
  2. Set the rotary switch to the Diode symbol.
  3. Connect the red probe to the anode and the black probe to the cathode of the diode.
  4. A forward voltage drop (typically 0.5V to 0.8V for silicon diodes) will be displayed. Reverse the probes; the display should show "OL" (Open Loop) for a good diode.

6.9 Continuity Test:

  1. Insert the red test lead into the VΩmA jack and the black test lead into the COM jack.
  2. Set the rotary switch to the Continuity symbol.
  3. Connect the test probes across the circuit or component.
  4. If continuity exists (resistance below approximately 50Ω), the built-in buzzer will sound. The display will show the resistance value.

6.10 hFE Test (Transistor Gain):

  1. Set the rotary switch to the hFE function.
  2. Identify the type of transistor (NPN or PNP) and its Emitter, Base, and Collector leads.
  3. Insert the transistor leads into the corresponding sockets in the hFE test socket on the meter.
  4. Read the hFE value (DC current gain) on the display.

7. Specifications

Measurement FunctionRangeAccuracy
DC Voltage4V, 40V, 400V±0.5%
1000V±0.8%
AC Voltage4V, 40V, 400V±1.2%
750V±1.5%
DC Current4mA, 400mA±1.2%
10A±2.0%
AC Current4mA, 400mA±1.5%
10A±3.0%
Resistance400Ω, 4kΩ, 40kΩ, 400kΩ, 4MΩ±1.2%
40MΩ±3.0%
Capacitance4nF, 400nF±4.0%
Frequency40kHz±3.0%
hFE1-1000
Diode TestYes
Continuity TestYes (Buzzer < 50Ω)

General Characteristics:

  • Power: 6F22 (9V) battery × 1
  • Dimensions: 78 × 186 × 35 mm (approximately 3.07 x 7.32 x 1.38 inches)
  • Weight: Approximately 300g (0.66 lbs)
  • Display: Digital and Analog display, 3999 counts and 38-segment bar graph. Digit height 15mm.
  • Features: Auto Range, Manual Range, Data Hold, RS232C Interface, Backlight.

8. Maintenance

8.1 Battery Replacement:

When the battery symbol appears on the LCD, the battery voltage is low and needs to be replaced. Refer to section 5.1 for battery installation instructions.

8.2 Cleaning:

Wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Ensure the meter is completely dry before use.

8.3 Fuse Replacement:

If the current measurement functions stop working, the fuse may need replacement. This procedure should only be performed by qualified personnel.

  1. Ensure the meter is turned off and test leads are disconnected.
  2. Open the battery compartment cover (refer to 5.1).
  3. Carefully remove the old fuse(s). The MAS343 typically uses two fuses: one for the mA range and one for the 10A range.
  4. Replace with fuses of the exact same type and rating (e.g., F400mA/250V for mA, F10A/250V for 10A).
  5. Replace the battery compartment cover and secure it.

9. Troubleshooting

ProblemPossible CauseSolution
Meter does not power on.Dead or incorrectly installed battery.Check battery installation, replace if necessary.
"OL" (Overload) displayed.Measurement exceeds selected range or meter's maximum capacity.Select a higher range or ensure measurement is within meter's limits.
Incorrect readings.Incorrect function selected, poor test lead connection, or discharged battery.Verify function switch position, check test lead connections, replace battery.
Current measurement not working.Blown fuse.Replace the appropriate fuse (refer to 8.3).
RS232 connection issues.Incorrect cable connection, driver issues, or software settings.Ensure cable is securely connected. Reinstall software/drivers. Check software communication settings.

10. Warranty and Support

The Tekpower Mastech MAS343 Digital Multimeter comes with a one-year manufacturer's limited warranty from the date of purchase. This warranty covers defects in materials and workmanship under normal use. It does not cover damage caused by misuse, accident, unauthorized repair, or modification.

Additionally, a 30-day money-back guarantee is offered, allowing returns within 30 days of purchase if you are not satisfied with the product.

For warranty claims, technical support, or service inquiries, please contact Tekpower customer service through their official website or the retailer where the product was purchased. Please have your proof of purchase and product model number (MAS343) ready when contacting support.

© 2023 Tekpower. All rights reserved. | Model: MAS343