Oakton WD-35805-04

Oakton WD-35805-04 Laboratory pH Electrode Instruction Manual

Model: WD-35805-04

Introduction

This manual provides comprehensive instructions for the proper setup, operation, and maintenance of your Oakton WD-35805-04 Laboratory pH Electrode. Adhering to these guidelines will ensure optimal performance and longevity of the device.

Product Overview

The Oakton WD-35805-04 is a double-junction, glass body, refillable pH electrode designed for laboratory applications. Its robust construction and double-junction design make it suitable for a wide range of chemical measurements, offering enhanced durability and resistance to aggressive chemicals. This electrode features an Ag/AgCl reference cell for stable and accurate readings.

Oakton WD-35805-04 Laboratory pH Electrode

Image 1: Oakton WD-35805-04 Laboratory pH Electrode. This image shows the full electrode, featuring a black top housing with an "OPEN" indicator, a clear glass body revealing internal components, and a blue spherical tip. A black cable extends from the top.

Key Features:

Setup

  1. Unpacking: Carefully remove the electrode from its packaging. Inspect for any visible damage.
  2. Electrolyte Filling (if necessary): This electrode is refillable. If the electrolyte level is low, open the fill hole (usually indicated by an "OPEN" arrow or cap) and carefully add the appropriate reference electrolyte solution (e.g., 3M KCl) until the level is just below the fill hole. Close the fill hole securely.
  3. Rinsing: Rinse the electrode tip with distilled or deionized water to remove any storage solution. Do not wipe the glass bulb, as this can create static charges.
  4. Conditioning: For optimal performance, soak the electrode in a pH 7 buffer solution for at least 30 minutes (or longer, up to several hours) before first use or after prolonged dry storage. This rehydrates the glass membrane.
  5. Connecting to Meter: Connect the electrode's BNC connector (or other specified connector type) to the corresponding input on your pH meter. Ensure a secure connection.
  6. Calibration: Calibrate your pH meter with the electrode using at least two, preferably three, fresh pH buffer solutions (e.g., pH 4.01, 7.00, and 10.00). Follow your pH meter's specific calibration procedure.

Operating Instructions

  1. Preparation: Ensure the electrode is properly conditioned and calibrated.
  2. Sample Measurement:
    • Rinse the electrode with distilled or deionized water between samples.
    • Immerse the electrode tip into the sample solution, ensuring the glass bulb and junction are fully submerged.
    • Stir the solution gently to ensure homogeneity and faster response.
    • Allow the reading to stabilize before recording the pH value. Stability is typically indicated by a stable reading on the pH meter for a few seconds.
  3. Temperature Compensation: If your pH meter has automatic temperature compensation (ATC), ensure the temperature probe is also immersed in the sample. If not, manually input the sample temperature into the meter.
  4. Post-Measurement: After each measurement, rinse the electrode thoroughly with distilled or deionized water.

Maintenance

Troubleshooting

ProblemPossible CauseSolution
Slow Response / Unstable Readings
  • Dehydrated glass membrane
  • Clogged junction
  • Low electrolyte level
  • Contaminated buffer solutions
  • Temperature not stable
  • Recondition electrode in pH 7 buffer.
  • Clean electrode as per Maintenance section.
  • Refill electrolyte.
  • Use fresh buffer solutions.
  • Allow sample/electrode to reach thermal equilibrium.
Inaccurate Readings
  • Improper calibration
  • Dirty electrode
  • Expired or contaminated buffers
  • Electrode nearing end of life
  • Recalibrate with fresh buffers.
  • Clean electrode.
  • Replace buffers.
  • Consider replacing electrode if issues persist after cleaning/reconditioning.
No Reading / "Error" on Meter
  • Loose connection
  • Broken electrode
  • Meter malfunction
  • Check electrode connection to meter.
  • Inspect electrode for visible damage (cracks in glass).
  • Test meter with a known good electrode or consult meter manual.

Specifications

AttributeDetail
Model NumberWD-35805-04
Electrode TypeLaboratory pH Electrode, Double-Junction
Body MaterialGlass
Reference TypeRefillable Ag/AgCl
Package Dimensions10.9 x 4.1 x 2.2 inches
Item Weight6.38 ounces
ManufacturerOAKTON Instruments
Date First AvailableJune 2, 2016
ASINB01GIRDK64
Best Sellers RankSee Top 100 in Industrial & Scientific

Warranty and Support

For specific warranty information regarding your Oakton WD-35805-04 Laboratory pH Electrode, please refer to the documentation included with your purchase or visit the official Oakton Instruments website. For technical support, troubleshooting assistance beyond this manual, or to inquire about replacement parts and accessories, please contact Oakton Instruments customer service directly.

Oakton Instruments Contact Information: (Please refer to product packaging or official website for current contact details.)

Related Documents - WD-35805-04

Preview OAKTON 410 Timer with Alarm: Operating Instructions and Specifications
Comprehensive operating instructions and specifications for the OAKTON 410 Timer with Alarm (Model WD-35002-00), detailing its features, button functions, timing modes, and alarm settings.
Preview OAKTON TEMP TC Acorn Series Thermocouple Thermometer Operating Instructions
Operating instructions and specifications for the OAKTON TEMP TC Acorn Series Thermocouple Thermometer, including details on measurement, calibration, troubleshooting, and accessories.
Preview OAKTON ION Acorn Series Meters: Operating Instructions and Guide
Comprehensive operating manual for OAKTON ION 5 and ION 6 Acorn Series pH/Ion Meters. Covers setup, calibration, measurement, maintenance, and troubleshooting for accurate water quality analysis.
Preview OAKTON Single-Event 220 Stopwatch with Alarm Operating Instructions
Detailed operating instructions for the OAKTON Single-Event 220 Stopwatch with Alarm (Model WD-35002-10), covering stopwatch functions, time/calendar settings, alarm features, specifications, and warranty information.
Preview Oakton DO250/DO260 Dissolved Oxygen Meter Instruction Manual
Comprehensive instruction manual for the Oakton DO250 and DO260 Dissolved Oxygen Meters, covering product overview, basic operations, calibration, data management, setup, maintenance, troubleshooting, and safety precautions.
Preview OAKTON 300 Series Handheld Water Quality Meters Instruction Manual
This instruction manual provides comprehensive guidance for operating the OAKTON 300 series handheld water quality meters. It covers product overview, setup, calibration, measurement, data management, maintenance, and troubleshooting for accurate water quality analysis.