Welch Allyn WEL7670-02

Welch Allyn 7670-02 767 Wall Aneroid with 8' Coiled Tubing User Manual

Introduction

The Welch Allyn 767 Wall Aneroid is a blood pressure monitor designed for accurate and reliable readings in clinical settings. Its improved ergonomics and modern design enhance usability. This manual provides essential information for the proper setup, operation, and maintenance of your device.

Key Features

Product Overview

Welch Allyn 767 Wall Aneroid blood pressure monitor mounted on a wall, showing the dial, inflation bulb, and coiled tubing.

Figure 1: Welch Allyn 767 Wall Aneroid Blood Pressure Monitor.

This image displays the Welch Allyn 767 Wall Aneroid blood pressure monitor. It features a large, square black dial with white markings for blood pressure readings, housed in a light-colored square casing. Below the dial, a black inflation bulb and a long, black coiled tubing are visible, connecting the unit to a cuff (not shown). The entire unit is designed for wall mounting.

Components:

Setup Instructions

  1. Choose Location: Select a suitable wall location, preferably near the examination area, ensuring it is at eye level for easy reading.
  2. Mounting: Securely attach the wall mount bracket to a sturdy wall stud or use appropriate wall anchors. Ensure the bracket is level before tightening screws.
  3. Attach Aneroid: Carefully slide the Welch Allyn 767 Aneroid onto the mounted bracket until it clicks securely into place.
  4. Connect Tubing: Connect the coiled tubing from the aneroid to the blood pressure cuff. Ensure all connections are snug to prevent air leaks.
  5. Initial Check: Before first use, perform a quick visual inspection to ensure all components are intact and properly connected.

Operating Instructions

Follow these steps for accurate blood pressure measurement:

  1. Patient Positioning: Have the patient sit comfortably with their arm supported at heart level. Ensure they are relaxed and have not exercised or consumed caffeine/nicotine recently.
  2. Cuff Placement: Select the appropriate size blood pressure cuff. Wrap the cuff firmly around the patient's bare upper arm, ensuring the artery marker on the cuff aligns with the brachial artery.
  3. Inflation: Close the air release valve on the inflation bulb. Rapidly inflate the cuff by squeezing the bulb until the gauge reads approximately 20-30 mmHg above the patient's usual systolic pressure, or until the radial pulse is no longer palpable.
  4. Deflation and Reading: Slowly open the air release valve to deflate the cuff at a rate of 2-3 mmHg per second.
    • Systolic Pressure: Note the reading on the aneroid gauge when you first hear the Korotkoff sounds (a tapping sound).
    • Diastolic Pressure: Note the reading when the Korotkoff sounds disappear or become muffled.
  5. Release Air: Fully open the air release valve to deflate the cuff completely.
  6. Repeat Measurement: If necessary, wait at least 1-2 minutes before taking another reading on the same arm.

Maintenance

Proper maintenance ensures the longevity and accuracy of your Welch Allyn 767 Wall Aneroid.

Troubleshooting

ProblemPossible CauseSolution
Inaccurate ReadingsCuff size incorrect, improper cuff placement, air leak, device needs calibration.Ensure correct cuff size and placement. Check for leaks in tubing/cuff. Schedule professional calibration.
Cuff Does Not InflateAir release valve open, leak in tubing/cuff, faulty inflation bulb.Close valve. Check all connections for leaks. Replace faulty components if necessary.
Gauge Needle Jumps ErraticallyRapid cuff deflation, patient movement, internal gauge issue.Deflate cuff slowly (2-3 mmHg/sec). Ensure patient stillness. Contact service if issue persists.

Specifications

Warranty and Support

For warranty information, technical support, or service inquiries regarding your Welch Allyn 7670-02 Wall Aneroid, please contact the manufacturer directly.

Manufacturer Contact:

Always refer to the official Welch Allyn website or documentation for the most current warranty terms and support procedures.

Related Documents - WEL7670-02

Preview Welch Allyn Connex Spot Monitor Upute za uporabu
Ovaj dokument pruža detaljne upute za uporabu Welch Allyn Connex Spot Monitor uređaja, uključujući postavljanje, rad, upravljanje podacima o pacijentu, alarme i rješavanje problema.
Preview Welch Allyn HScribe: Uživatelská příručka pro analýzu Holterova monitorování
Tato uživatelská příručka poskytuje podrobné informace o systému Welch Allyn HScribe pro analýzu Holterova monitorování, včetně instalace, provozu, údržby a řešení problémů.
Preview Welch Allyn HScribe Holter Analiz Sistemi Kullanıcı Kılavuzu
Welch Allyn HScribe Holter Analiz Sistemi için kapsamlı kullanıcı kılavuzu. Tıbbi profesyoneller için kurulum, kullanım, bakım ve sorun giderme bilgilerini içerir.
Preview Welch Allyn® HScribe™ System analizy holterowskiej Instrukcja obsługi
Instrukcja obsługi systemu analizy holterowskiej Welch Allyn HScribe, zawierająca informacje o instalacji, konfiguracji, obsłudze, analizie danych EKG i raportowaniu dla profesjonalistów medycznych. Dokument obejmuje funkcje systemu, bezpieczeństwo użytkowania oraz rozwiązywanie problemów.
Preview Welch Allyn Blood Pressure Cuffs and Aneroids Catalog
Comprehensive catalog of Welch Allyn blood pressure cuffs, including FlexiPort™ reusable and disposable options, EcoCuff™ single-patient use cuffs, and various aneroid sphygmomanometers like the Tycos® 767-Series and Durashock™ models. Features size guides, product specifications, and benefits for accurate blood pressure measurement.
Preview Welch Allyn Wall Mount Sphygmomanometer and Integrated Charger User Manual
Comprehensive user manual for the Welch Allyn Wall Mount Sphygmomanometer with integrated Android tablet/PC charger and aneroid gauge. Covers operation, specifications, maintenance, and warranty for models 5000 series.