1. Introduction
The ALLOSUN EM480B is a versatile digital meter designed for accurate measurement of audio impedance, insulation resistance, AC voltage, and continuity. It is an essential tool for professionals working with speaker systems, PA systems, transformers, motors, cables, and electrical appliances. This manual provides detailed instructions for the safe and effective operation, setup, and maintenance of your EM480B tester.

Image 1.1: The ALLOSUN EM480B Audio Impedance Tester, including test leads, carrying case, and strap.
2. Safety Information
Please read and understand all safety instructions before operating the EM480B tester. Failure to follow these instructions may result in electric shock, fire, or damage to the instrument.
- Always ensure the instrument is in good working condition before use.
- Do not use the tester if it appears damaged or if the test leads are frayed.
- Observe all local and national safety codes.
- Avoid working alone when performing high-voltage tests.
- Do not apply voltage to the input terminals that exceeds the maximum rated voltage for the instrument.
- Replace batteries promptly when the low battery indicator appears to ensure accurate readings.
- Always disconnect the test leads from the circuit before changing functions or ranges.
3. Product Overview
3.1 Key Features
- True measurement of speaker system impedance at 1kHz.
- Three impedance test ranges: 20Ω, 200Ω, and 2000Ω.
- Insulation resistance test with selectable voltages: 250V, 500V, 1000V.
- Measures AC voltage and resistance (less than 200Ω).
- Continuity test function.
- Large LCD 3 1/2 digits (2000 counts) display with backlight.
- Data Hold function to freeze readings.
- Timer function for easier operation during tests (approx. 30 seconds after TEST button press).
- Automatic power-off after approximately 15 minutes of inactivity.
- Low battery indication.
- Protection circuit to prevent damage from discharge voltage.
- Comes with a carrying case and belt for portability.
3.2 Panel Layout
Familiarize yourself with the controls and indicators on the front panel of the EM480B tester.

Image 3.1: Front Panel of the EM480B Tester.
- TEST button: Activates the test function. Pressing this button initiates a test, and the unit enters test mode for approximately 30 seconds. Press again to exit test mode.
- LOCK key: For continuous measurement. Press this key while in test mode to lock the unit in test mode. The "LOCK" symbol appears on the display. Press TEST button again to unlock.
- HOLD key: Freezes the current reading on the display. Press again to exit data-hold mode.
- BACKLIGHT key: Toggles the LCD backlight on or off for improved visibility in low-light conditions.
- RANGE / FUNCTION switch: Rotary switch to turn the unit on/off and select the desired measurement function and range (e.g., 20Ω, 200Ω, 2000Ω, OFF).
- COM jack: Input jack for the black test lead.
- Ω jack: Input jack for the red test lead.
- High Voltage Indicator LED: Illuminates when high voltage is present during insulation resistance tests.
4. Setup
4.1 Battery Installation
The EM480B requires six (6) 1.5V AA batteries for operation. Batteries are not included with the unit.
- Ensure the unit is turned OFF.
- Carefully remove the four screws on the back cover of the tester.
- Gently open the back cover to access the battery compartment.
- Insert six 1.5V AA batteries, observing the correct polarity (+/-) as indicated inside the compartment.
- Replace the back cover and secure it with the four screws. Do not overtighten.

Image 4.1: Battery compartment with slots for six AA batteries.

Image 4.2: Illustration of removing the back cover for battery access.
4.2 Connecting Test Leads
Insert the black test lead into the COM jack and the red test lead into the Ω jack for impedance, resistance, ACV, and continuity measurements. For insulation resistance tests, refer to specific instructions in the operating section.
5. Operating Instructions
5.1 Powering On/Off
To power on the unit, rotate the RANGE / FUNCTION switch from the OFF position to any desired measurement range (e.g., 20Ω). To power off, rotate the switch back to the OFF position. The unit also features an auto power-off function after approximately 15 minutes of inactivity to conserve battery life.
5.2 Performing Audio Impedance Tests
The EM480B measures the actual impedance of speaker systems at 1kHz.
- Ensure the speaker system or individual speaker is disconnected from any power source or amplifier.
- Connect the black test lead to the COM jack and the red test lead to the Ω jack.
- Rotate the RANGE / FUNCTION switch to the appropriate impedance range (20Ω, 200Ω, or 2000Ω). For typical speakers (2Ω to 16Ω), use the 20Ω range. For higher impedance speakers or systems, use 200Ω or 2000Ω.
- Connect the test leads across the terminals of the speaker or speaker system.
- Press the TEST button. The reading will appear on the LCD. The unit will remain in test mode for about 30 seconds. For continuous measurement, press the LOCK key after pressing TEST.

Image 5.1: Measuring the impedance of a single speaker.

Image 5.2: Measuring the impedance of a Public Address (PA) system.
Troubleshooting Impedance Readings in PA Systems:
- If the measured impedance is lower than expected, check for short-circuited wires or faulty speakers/transformers.
- If the measured impedance is higher than expected, check for wiring issues or faulty components (speakers, transformers, or connections).
5.3 Performing Insulation Resistance Tests
The EM480B can test insulation resistance at 250V, 500V, and 1000V. This function is used to check the integrity of insulation in electrical systems and components.
- Ensure the circuit or device under test is completely de-energized and safely isolated.
- Connect the black test lead to the COM jack and the red test lead to the Ω jack.
- Rotate the RANGE / FUNCTION switch to the desired insulation test voltage (250V, 500V, or 1000V).
- Connect the test leads to the points where insulation resistance is to be measured.
- Press the TEST button. The High Voltage Indicator LED will illuminate, and the insulation resistance reading will be displayed.
- Exercise extreme caution during high-voltage tests.
5.4 Data Hold Function
To freeze the current reading on the display, press the HOLD key. The "H" symbol will appear on the LCD. Press the HOLD key again to release the reading and return to live measurement.
5.5 Backlight Function
To activate the LCD backlight for better visibility in dimly lit environments, press the BACKLIGHT key. Press it again to turn the backlight off.

Image 5.3: Display and function dial showing backlight and range options.
6. Maintenance
6.1 Battery Replacement
When the low battery indicator appears on the display, replace the batteries as described in Section 4.1. Always use fresh 1.5V AA batteries. Remove batteries if the unit will not be used for an extended period to prevent leakage.
6.2 Cleaning
Wipe the unit with a dry, soft cloth. Do not use abrasive cleaners or solvents. Ensure no moisture enters the casing.
6.3 Storage
Store the EM480B in its protective carrying case in a cool, dry place, away from direct sunlight and extreme temperatures. Remove batteries if storing for long periods.

Image 6.1: The EM480B tester stored in its protective carrying case.
7. Troubleshooting
If you encounter issues with your EM480B tester, consider the following common troubleshooting steps:
- No Display/Unit Not Powering On: Check battery installation and ensure batteries are fresh. Verify the RANGE / FUNCTION switch is not in the OFF position.
- Inaccurate Readings: Ensure test leads are properly connected and not damaged. Verify the correct measurement range is selected. Replace batteries if the low battery indicator is on.
- "H" Symbol Stuck: If the "H" symbol for Data Hold is stuck, press the HOLD key again.
- Test Mode Not Exiting: If the unit remains in test mode, press the TEST button again. If the LOCK key was pressed, press TEST to unlock.
- Low Battery Indication: Replace all six 1.5V AA batteries immediately.
For issues not resolved by these steps, please contact customer support.
8. Specifications
The ALLOSUN EM480B Audio Impedance Tester meets IEC 1010 (EN61010) standards and is rated for Installation CAT III 100V.
- Display: LCD 3 1/2 digits (2000 counts), with unit indication.
- Data Hold Indication: "H" symbol appears on the display.
- Power Supply: 9V DC (6 x 1.5V AA batteries).
- Auto Power Off: Approximately 15 minutes of non-use.
- Operating Temperature: 0°C to 40°C (32°F to 104°F).
- Storage Temperature: -10°C to 50°C (14°F to 122°F).
- Dimensions: 17 cm (L) x 6.5 cm (W) x 11 cm (H).
- Weight: Approximately 500g (including batteries).
8.1 Impedance Measurement
| Range | Resolution | Accuracy | Test Frequency |
|---|---|---|---|
| 20Ω | 10mΩ | ± (2% + 2) or ± 0.1Ω | 1kHz |
| 200Ω | 100mΩ | ± (2% + 2) | |
| 2000Ω | 1Ω | ± (2% + 2) |

Image 8.1: Dimensions and detailed specifications table.
8.2 Insulation Resistance Measurement
- Test Voltages: 250V, 500V, 1000V.
- Accuracy: Specified for a period of one year after calibration and at 18℃ to 28℃, with relative humidity. Accuracy specifications take the form of: ±([% of Reading] + [number of Least Significant Digits]).
9. Warranty and Support
For warranty information and technical support, please refer to the documentation provided with your purchase or contact the seller/manufacturer directly. Keep your purchase receipt as proof of purchase.
Manufacturer: Zhangzhou Eastern Intelligent Meter Co., Ltd