HACH 2106769
Contents
Hach SulfaVer 4 Sulfate Reagent Powder Pillows Instruction Manual
Model: 2106769
Introduction
This manual provides essential instructions for the safe and effective use of Hach SulfaVer 4 Sulfate Reagent Powder Pillows, Model 2106769. These reagents are designed for the determination of sulfate in water samples, specifically for wastewater analysis, and are accepted by the USEPA for reporting results using Hach Method 8051.
Product Overview
The Hach SulfaVer 4 Sulfate Reagent is supplied in convenient powder pillow form, ensuring accurate dosing for each test. Each pack contains 100 individual powder pillows, suitable for 100 tests. This reagent is formulated for use with a 10 mL sample size and is capable of measuring sulfate concentrations up to 70 mg/L.

Figure 1: Hach SulfaVer 4 Sulfate Reagent Powder Pillows. This image displays a pack of Hach SulfaVer 4 Sulfate Reagent Powder Pillows. The pillows are individually sealed in foil packets and contained within a larger clear, resealable plastic bag. One individual foil packet is shown outside the main bag, illustrating the product's packaging. The Hach logo and 'PERMACHEM® REAGENTS' are visible on a white label inside the bag.
Safety Precautions
Always handle chemical reagents with care. Wear appropriate personal protective equipment (PPE), including safety glasses and gloves, when handling SulfaVer 4 reagent. Refer to the Safety Data Sheet (SDS) for complete safety information. Do not ingest. Keep out of reach of children.
Setup
- Gather Equipment: Ensure you have all necessary equipment, including a Hach colorimeter or spectrophotometer (e.g., DR-900), sample cells, a timer, and a 10 mL graduated cylinder or pipette for accurate sample measurement.
- Prepare Sample: Collect a representative water sample. The sample size required for Hach Method 8051 is 10 mL.
- Ensure Cleanliness: Use clean glassware and sample cells to prevent contamination, which can affect test results.
Operating Instructions (Hach Method 8051)
Follow these steps for accurate sulfate determination using Hach Method 8051:
- Prepare Sample Cell: Fill a clean sample cell with 10 mL of the water sample. For a blank, fill another clean sample cell with 10 mL of deionized water.
- Add Reagent: Open one Hach SulfaVer 4 Powder Pillow. Add the entire contents of the powder pillow to the 10 mL sample cell.
- Mix: Cap the sample cell and shake vigorously for 30 seconds to ensure the reagent dissolves completely and mixes thoroughly with the sample.
- Reaction Time: Allow the sample to react for a specific time, typically 5 minutes. Refer to your Hach instrument's manual for the exact reaction time for Method 8051. During this time, a turbidity will develop if sulfate is present.
- Measure: After the reaction time, insert the prepared sample cell into your Hach colorimeter or spectrophotometer. Follow the instrument's instructions to select Hach Method 8051 and measure the sulfate concentration.
- Record Results: Record the sulfate concentration displayed by the instrument.
Note: Always refer to your specific Hach instrument's operation manual for detailed instructions on performing Method 8051, as procedures may vary slightly between models.
Maintenance
- Storage: Store Hach SulfaVer 4 Powder Pillows in their original packaging in a cool, dry place, away from direct sunlight and moisture. Ensure the resealable bag is properly closed after each use to protect unused pillows.
- Expiration Date: Check the expiration date on the packaging. Do not use expired reagents, as they may yield inaccurate results.
- Disposal: Dispose of used reagents and samples according to local environmental regulations. Consult your facility's waste disposal guidelines.
Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| Inaccurate or inconsistent results | Expired reagent, improper mixing, contaminated glassware, incorrect sample volume, instrument calibration issue. | Check reagent expiration date. Ensure vigorous mixing for 30 seconds. Use clean glassware. Verify 10 mL sample volume. Calibrate instrument as per manufacturer's instructions. |
| No turbidity development | Sulfate concentration is below detection limit, reagent not added, or reagent is expired. | Confirm sulfate presence with another method if possible. Ensure reagent was added. Check reagent expiration date. |
| Excessive turbidity or precipitate | High sulfate concentration exceeding range, interfering substances, or improper mixing. | Dilute sample if sulfate concentration is expected to be high and retest. Check for known interferences. Ensure proper mixing. |
Specifications
- Product Name: SulfaVer 4 Sulfate Reagent Powder Pillows
- Brand: HACH
- Model Number: 2106769
- Method: Hach Method 8051 (USEPA accepted for wastewater)
- Measurement Range: Up to 70 mg/L Sulfate
- Sample Size: 10 mL
- Quantity: 100 tests per pack
- Form: Powder
- Item Weight (Pack): Approximately 1.12 ounces (31.75 grams)
- Package Dimensions: Approximately 6 x 4 x 0.4 inches
Warranty and Support
For specific warranty information and technical support, please contact Hach customer service directly or visit the official Hach website. Product support resources, including Safety Data Sheets (SDS) and detailed method procedures, are available through Hach's official channels.
Hach Company World Headquarters: P.O. Box 389, Loveland, CO 80539 U.S.A.
Website: www.hach.com
Documents and downloads
Ask a question about this manual
Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.