HACH 2100Q01

Hach 2100Q Portable Turbidimeter Instruction Manual

Model: 2100Q01

Introduction

The Hach 2100Q Portable Turbidimeter is designed for accurate and reliable turbidity measurements in various environments, including field testing, plant monitoring, and water quality analysis for regulatory compliance. This manual provides essential information for the proper setup, operation, and maintenance of your 2100Q Turbidimeter.

Hach 2100Q Portable Turbidimeter

Image: The Hach 2100Q Portable Turbidimeter, a compact device with a digital display showing a turbidity reading of 0.10 NTU.

Key Features:

  • EPA Compliant Design: Meets USEPA Method 180.1 criteria for regulatory reporting.
  • Accurate Measurement: Measures turbidity across a 0-1000 NTU range.
  • Guided Calibration: On-screen assisted calibration, including Rapidcal™ option.
  • Data Logging: Stores up to 500 measurements with detailed information.
  • Portable and Durable: Rugged design suitable for field and plant operations.

Setup

1. Unpacking and Inspection

Carefully unpack all components from the carrying case. Verify that all items listed in the packing slip are present and undamaged. Report any missing or damaged items to Hach customer service immediately.

2. Battery Installation

The Hach 2100Q requires four (4) AA batteries for operation.

  1. Locate the battery compartment cover on the back of the instrument.
  2. Open the cover by sliding or unlatching it according to the indicator.
  3. Insert four AA batteries, ensuring correct polarity (+/-) as indicated inside the compartment.
  4. Securely close the battery compartment cover.

3. Initial Power On

Press the power button (usually indicated by a circle with a vertical line) to turn on the turbidimeter. The display should illuminate, and the device will perform a self-check.

Operating Instructions

1. Calibration

Regular calibration ensures accurate turbidity measurements. The 2100Q features on-screen guided calibration.

  1. Turn on the turbidimeter.
  2. Navigate to the "Calibrate" or "Calibration" menu option using the arrow keys.
  3. Follow the on-screen prompts to insert the required calibration standards (e.g., 20 NTU, 100 NTU, 800 NTU).
  4. For quick low-level calibration, utilize the Rapidcal™ option if available.
  5. Press "Read" or "Done" as prompted after each standard is inserted and measured.
  6. Ensure the calibration is successfully completed and saved.
Hach 2100Q Turbidimeter Calibration Screen

Image: Close-up of the Hach 2100Q screen during calibration, showing "Cal: StablCal® 99.4 NTU" and options for different NTU standards.

Inserting a sample vial into Hach 2100Q Turbidimeter

Image: A hand inserting a sample vial into the measurement cell of the Hach 2100Q Turbidimeter for calibration or measurement.

2. Taking a Measurement

The 2100Q measures turbidity in Nephelometric Turbidity Units (NTU).

  1. Ensure the turbidimeter is calibrated.
  2. Prepare your sample according to standard laboratory practices, ensuring the sample cell is clean and free of air bubbles or scratches.
  3. Insert the sample cell into the measurement compartment, aligning any marks as instructed by the device or manual.
  4. Close the lid of the measurement compartment.
  5. Press the "Read" button. The device will display the turbidity reading in NTU.
Hach 2100Q Turbidimeter held in hand showing a reading

Image: A hand holding the Hach 2100Q Turbidimeter, displaying a turbidity reading of 103 NTU on its screen.

3. Data Logging and Transfer

The 2100Q automatically stores up to 500 measurements. Each entry includes a timestamp, operator ID, and calibration status.

  • Viewing Data: Access the "Data Log" or "History" menu option to review stored measurements.
  • Data Transfer: For transferring data to a computer, an optional USB+Power module is available. Connect the module to the turbidimeter and your computer, then follow the software instructions for data export.

Maintenance

1. Cleaning the Instrument

Keep the exterior of the turbidimeter clean using a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure the measurement cell compartment is free of dust and debris.

2. Sample Cell Care

Clean sample cells thoroughly before each use. Fingerprints, scratches, or residue on the cell can affect measurement accuracy. Use a lint-free cloth to wipe the exterior of the cell.

3. Battery Replacement

Replace batteries when the low battery indicator appears on the display. Refer to the "Battery Installation" section under Setup for instructions.

4. Optical System Care

The 2100Q uses a two-detector optical system. Avoid touching the optical components inside the measurement compartment. If cleaning is necessary, consult the full Hach 2100Q user manual for specific instructions to prevent damage.

Troubleshooting

This section addresses common issues you might encounter with your Hach 2100Q Turbidimeter.

ProblemPossible CauseSolution
Instrument does not power on.Dead or incorrectly installed batteries.Replace batteries, ensuring correct polarity.
Inaccurate or inconsistent readings.
  • Dirty or scratched sample cell.
  • Instrument out of calibration.
  • Air bubbles in sample.
  • Clean sample cell thoroughly.
  • Perform a full calibration.
  • Degas sample if necessary; tap cell gently to remove bubbles.
"Error" message on display.Specific error code will indicate the issue.Refer to the comprehensive Hach 2100Q user manual for specific error code interpretations and solutions.

If you encounter issues not listed here or if the suggested solutions do not resolve the problem, please contact Hach Technical Support.

Specifications

Below are the technical specifications for the Hach 2100Q Portable Turbidimeter (Model 2100Q01).

FeatureDetail
Model Number2100Q01
Measurement Range0 - 1000 NTU
ComplianceMeets USEPA Method 180.1 criteria
Light SourceTungsten Filament Lamp
Power Source4 AA batteries required
Data Logging CapacityUp to 500 measurements
Optical SystemTwo-detector system for color and stray light compensation
Dimensions (Approx.)16 x 10 x 8 inches (Package Dimensions)
Weight (Approx.)4 Pounds (Package Weight)

Warranty and Support

Warranty Information

Specific warranty terms and conditions for the Hach 2100Q Portable Turbidimeter are provided with the product documentation at the time of purchase. Please refer to the warranty card or the full user manual for detailed information regarding coverage, duration, and claims procedures. For the most current warranty details, contact Hach Company directly.

Technical Support

For technical assistance, troubleshooting guidance, or service inquiries, please contact Hach Company customer support.

  • Manufacturer: HACH COMPANY
  • Website: www.hach.com (Please visit the official Hach website for contact information, support resources, and additional documentation.)

Additional Resources

No official product videos from the seller were found in the provided data. For further operational guidance or visual demonstrations, please refer to the official Hach Company website or contact their support channels.

Technician performing water quality testing with a portable device

Image: A technician in a hard hat and safety glasses using a portable water quality testing device in an industrial setting, demonstrating typical field use.

Related Documents - 2100Q01

Preview Hach 2100Q and 2100Qis Portable Turbidimeters User Manual
This user manual provides comprehensive instructions for operating and maintaining the Hach 2100Q and 2100Qis portable turbidimeters. Learn about specifications, installation, user interface, advanced operation, and troubleshooting.
Preview HACH TL2360 Basic User Manual - Turbidity Measurement Guide
User manual for the HACH TL2360 laboratory turbidimeter. Learn about installation, operation, calibration, and maintenance for accurate water turbidity measurements.
Preview Hach Model 2100N Laboratory Turbidimeter Instruction Manual
This instruction manual provides comprehensive guidance for the Hach Model 2100N Laboratory Turbidimeter, covering its operation, specifications, measurement procedures, calibration, maintenance, and troubleshooting for accurate turbidity analysis in laboratory environments.
Preview Hach Methods Quick Reference Guide: Chemical Analysis Instruments and Reagents
A comprehensive guide to Hach's chemical analysis methods, detailing test parameters, EPA approvals, instrument compatibility, and product numbers for water quality testing.
Preview Hach sension6 Portable Dissolved Oxygen Meter Instruction Manual
Comprehensive instruction manual for the Hach sension6 Portable Dissolved Oxygen Meter, covering setup, operation, calibration, maintenance, and troubleshooting for accurate dissolved oxygen measurements in aqueous solutions.
Preview HACH EZ4041 Total Hardness Analyser - Water Quality Monitoring
Discover the HACH EZ4041 Total Hardness Analyser, an online, automatic colorimetric titrator designed for industrial and environmental water monitoring. Features include a wide measuring range, smart automatic functions, flexible titration methods, and multiple stream analysis capabilities.