Drive Medical CECOMINOD048509

Drive Medical Oxygen Humidifier Bubbler Bottle Instruction Manual

Model: CECOMINOD048509

Introduction

The Drive Medical Oxygen Humidifier Bubbler Bottle is designed to prevent patient airways from becoming dry when using an oxygen concentrator. This humidifier features a 500 ml water capacity, allowing for extended use between refills. It is specifically designed for use with continuous flow oxygen concentrators and is not compatible with pulse-only or on-demand settings.

Product Features

  • Multi-threaded Lid: Ensures a secure fit, preventing leaks.
  • Audible Back Pressure Alarm: Activates at 6 psi to indicate potential issues.
  • No Aluminum Fitting in Stem: Designed for safety and compatibility.
  • Translucent Jar: Allows for easy monitoring of water levels.
  • 500 ml Water Capacity: Provides longer intervals between refills.
Oxygen Humidifier Bubbler Bottle highlighting benefits like improved patient comfort and enhanced hydration.

Image: The humidifier bottle with text indicating benefits such as improved patient comfort, enhanced hydration, moisture addition, and easily visible refill lines.

Oxygen Humidifier Bubbler Bottle highlighting reasons to use it, such as preventing mucosal dryness and reducing infection risk.

Image: The humidifier bottle with text explaining why to use a bubble humidifier, including preventing mucosal dryness, reducing risk of infection, improving oxygen absorption, and providing a simple solution.

Setup Instructions

1. Filling the Humidifier Bottle

  1. Unscrew the top of the humidifier bottle.
  2. Fill the bottle with distilled water up to the "MAXIMUM" line. Ensure hands are clean to prevent contamination.
  3. Carefully place the top back onto the bottle, ensuring the long tube extends into the water. Screw the lid on tightly to create a good seal and prevent leaks.
Oxygen Humidifier Bubbler Bottle with text 'JUST ADD STERILE WATER'.

Image: The humidifier bottle with a clear indication to add sterile water.

2. Attaching to the Oxygen Concentrator

  1. Locate the Velcro strap on your oxygen concentrator, typically designed to hold the humidifier bottle. Slide the filled humidifier bottle into this strap.
  2. Identify the oxygen outlet port on your concentrator. This is where the oxygen tubing from the humidifier will connect.
  3. Connect the short tube from the humidifier bottle's lid to the oxygen outlet port on the concentrator. Ensure a secure, tight connection.
  4. Connect the patient's oxygen tubing (e.g., nasal cannula) to the remaining port on the humidifier bottle's lid.
Oxygen Humidifier Bubbler Bottle attached to an oxygen concentrator.

Image: The humidifier bottle securely attached to an oxygen concentrator, ready for use.

Video: How to Attach an O2 Concentrator Humidity Bottle

Video: This short video demonstrates the process of filling the humidifier bottle with distilled water and attaching it to an oxygen concentrator, including connecting the oxygen tubing.

Operating Instructions

  1. Ensure all connections are secure before turning on the oxygen concentrator.
  2. Turn on your oxygen concentrator and adjust the flow rate as prescribed by your healthcare provider.
  3. Observe the humidifier bottle for bubbling, which indicates proper operation and humidification.
  4. Checking for Leaks: To verify a proper seal, temporarily block the patient's oxygen tubing outlet on the humidifier lid. If the audible back pressure alarm on the concentrator activates within 3-4 seconds, the seal is good. Release the blockage immediately after the alarm sounds.

Maintenance

  • Water Replacement: The water in the humidifier bottle should be changed daily. Always use distilled water to prevent mineral buildup and ensure optimal performance. If distilled water is unavailable, tap water can be boiled and allowed to cool before use.
  • Cleaning: Regularly clean the humidifier bottle and lid with mild soap and water. Rinse thoroughly and allow to air dry completely before refilling.
  • Inspection: Periodically inspect the bottle for cracks or damage, and the tubing for kinks or blockages. Replace any damaged components immediately.

Troubleshooting

  • No Bubbling in Bottle:
    • Check all tubing connections for tightness.
    • Ensure the oxygen concentrator is on and set to the correct flow rate.
    • Verify the humidifier lid is screwed on securely.
    • Inspect tubing for kinks or blockages.
  • Audible Alarm Sounds Continuously:
    • This indicates a blockage in the oxygen flow. Check the patient's tubing for kinks or obstructions.
    • Ensure the humidifier bottle is not overfilled, which can impede airflow.
    • If the alarm persists, consult your oxygen concentrator's manual or contact your healthcare provider.
  • Water Leakage:
    • Ensure the lid is tightly screwed onto the bottle.
    • Check all tubing connections for secure fit.
    • Inspect the bottle for cracks or damage.

Specifications

AttributeDetail
BrandDrive Medical
Model NumberCECOMINOD048509
Water Capacity500 ml
Alarm Activation6 psi (audible back pressure)
CompatibilityContinuous flow oxygen concentrators ONLY
Package Dimensions7.8 x 4.45 x 3.31 inches
Item Weight0.01 ounces
Date First AvailableDecember 20, 2013

Warranty and Support

For specific warranty information and customer support, please refer to the documentation included with your product or contact Drive Medical directly. Contact details for Lincare, a related company, are provided on the product label for reference:

Lincare, A Linde company
561.266.1011
800.940.8989
Fax 561.243.1717
2885 S. Congress Ave. Ste. G/H, Delray Beach, FL 33445

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.