Dwyer 2205
Contents
Dwyer® Magnehelic® Differential Pressure Gage User Manual
Model: 2205
1. Introduction
This manual provides essential information for the proper installation, operation, and maintenance of your Dwyer® Magnehelic® Differential Pressure Gage, Model 2205. This precision instrument is designed for high accuracy measurement of low air or non-corrosive gas pressures, including positive, negative (vacuum), or differential pressures. Its robust design resists shock, vibration, and over-pressures, offering a reliable and fluid-free alternative to manometers.
2. Safety Information
Always observe the following safety precautions to prevent personal injury and damage to the instrument.
- Do not exceed the specified pressure limits of -20 in Hg to 15 psig.
- Ensure the gage is used only with air and non-combustible, compatible gases.
- Avoid exposing the gage to temperatures outside the range of 20 to 140°F (-6.67 to 60°C).
- Always depressurize the system before connecting or disconnecting the gage.
- Handle the instrument with care to avoid physical damage.
3. Product Overview
The Dwyer Magnehelic Gage Model 2205 features a clear dial for easy reading and robust construction for reliable performance.

Figure 3.1: Front view of the Magnehelic Gage, Model 2205. The dial clearly displays pressure readings in pounds per square inch (PSI) from 0 to 5. A red needle indicates the current pressure. The face also includes a caution label regarding maximum pressure and temperature limits.

Figure 3.2: Angled view of the Magnehelic Gage, Model 2205, highlighting the side pressure connection ports. Also shown are the included brass barb fittings for hose connections and threaded plugs for sealing unused ports, demonstrating the versatility of its process connections.
Key Features:
- High accuracy: Guaranteed within 2% of full scale.
- Frictionless Magnehelic gage movement for quick indication.
- Resistant to shock, vibration, and over-pressures.
- No manometer fluid required, eliminating evaporation, freezing, or toxic issues.
- Process Connections: 1/8" female NPT duplicate high and low pressure taps - one pair side and one pair back.
4. Setup and Installation
Mounting:
The standard mounting orientation for the Magnehelic gage is with the diaphragm in a vertical position. For other mounting orientations, consult the manufacturer for specific instructions to ensure accurate readings.
Process Connections:
The gage is equipped with 1/8" female NPT (National Pipe Taper) taps for pressure connections. There are two sets of taps: one pair on the side and one pair on the back.
- Identify the high and low pressure ports. These are typically marked on the gage.
- Connect your pressure lines to the appropriate ports using 1/8" NPT fittings. Ensure all connections are sealed properly to prevent leaks.
- Use the side taps or back taps as convenient for your installation. Unused taps should be sealed with the provided plugs.
- For differential pressure measurement, connect the higher pressure source to the "High" port and the lower pressure source to the "Low" port.
5. Operating Instructions
Once installed, the Dwyer Magnehelic Gage operates automatically.
- Ensure the system is pressurized within the gage's operating limits.
- The gage's needle will move to indicate the current pressure reading on the dial.
- Read the pressure directly from the scale. The gage is calibrated in pounds per square inch (PSI).
- For zero adjustment, if necessary, gently turn the external zero adjustment screw located on the face of the gage until the needle aligns with the zero mark when no pressure is applied.
6. Maintenance
The Dwyer Magnehelic Gage is designed for minimal maintenance.
- Cleaning: Wipe the exterior of the gage with a soft, damp cloth. Do not use abrasive cleaners or solvents.
- Inspection: Periodically inspect the pressure connections for any signs of leaks or damage.
- Calibration: While the gage is highly accurate, periodic calibration by a qualified technician is recommended to ensure continued precision, especially in critical applications.
- Storage: When not in use, store the gage in a clean, dry environment within its specified temperature limits.
7. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| No reading or erratic reading |
|
|
| Inaccurate reading |
|
|
| Needle sticks or moves slowly |
|
|
8. Specifications
| Parameter | Value |
|---|---|
| Service | Air and non-combustible, compatible gases (Natural Gas option available) |
| Pressure Limits | -20 in Hg. to 15 psig |
| Temperature Limits | 20 to 140°F (-6.67 to 60°C) |
| Accuracy | ±2% of full scale |
| Mounting Orientation | Diaphragm in vertical position (consult factory for others) |
| Process Connections | 1/8" female NPT duplicate high and low pressure taps - one pair side and one pair back |
| Material | Aluminum |
| Item Weight | 1.2 Pounds |
| Package Dimensions | 6.2 x 5.5 x 2.9 inches |
9. Warranty and Support
For warranty information, technical support, or service inquiries, please contact Dwyer Instruments, Inc. directly. Refer to your purchase documentation for specific warranty terms and conditions.
Manufacturer: Dwyer Instruments, Inc.
Contact Information: Please visit the official Dwyer Instruments website or refer to the contact details provided with your product packaging for the most current support information.
Documents and downloads
Ask a question about this manual
Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.