Dwyer DM-1107

Dwyer DigiMag Series DM-1000 Differential Digital Pressure Gauge User Manual

Model: DM-1107

1. Introduction

This manual provides detailed instructions for the installation, operation, and maintenance of the Dwyer DigiMag Series DM-1000 Differential Digital Pressure Gauge. This device is designed to accurately monitor the pressure of air and compatible gases.

The DM-1000 series features a 4-digit LCD for displaying readings in common English and metric units, eliminating the need for manual conversions. Its simplified four-button operation facilitates quick setup and calibration, including digital push-button zero and span adjustments.

2. Safety Information

Always observe standard safety precautions when installing or servicing this device. Failure to do so may result in injury or damage to equipment.

  • Ensure the device is used only with air and non-combustible, compatible gases.
  • Disconnect power before performing any maintenance or installation procedures.
  • Refer to local codes and regulations for proper electrical and pressure connections.
  • Do not exceed the specified pressure range of the gauge.

3. Product Overview

The Dwyer DigiMag Series DM-1000 is a robust digital differential pressure gauge designed for precise measurements. Key components include the digital display, control buttons, and pressure ports.

Dwyer DigiMag Series DM-1000 Digital Pressure Gauge, angled view showing display and buttons.

Figure 3.1: Angled view of the Dwyer DigiMag DM-1000 showing the digital display and control buttons.

Front view of the Dwyer DigiMag Series DM-1000 Digital Pressure Gauge display.

Figure 3.2: Front view of the Dwyer DigiMag DM-1000 display, showing the "INWC" unit and a reading of "0.25".

3.1 Package Contents

Upon unpacking, verify that the following items are included:

  • Digital Differential Pressure Gauge (DM-1000 Series)
  • Instruction Manual (this document)
  • Mounting hardware (if applicable, consult factory for specifics)

Note: The product package typically includes the Digital Gauge, Differential, and is rated for 5 In WC.

4. Setup and Installation

Proper installation is crucial for accurate readings and long-term performance.

4.1 Mounting

The DM-1000 gauge is designed for panel mounting. Ensure adequate space for connections and ventilation. Refer to the product dimensions for cutout requirements.

  • Product Dimensions: 6.7 x 3.8 x 7.7 inches (17.0 x 9.7 x 19.6 cm)
  • Weight: 1.28 Pounds (0.58 kg)

4.2 Electrical Connections

The device features a removable terminal block for electrical connections. Use appropriate wiring gauges.

  • Wire Gauge: 16 to 26 AWG
  • Cable Gland: Compatible with 0.114 to 0.250 inches (2.9 to 6.4 mm) diameter cable.

Ensure all electrical connections comply with local electrical codes.

4.3 Pressure Connections

Connect the pressure lines to the appropriate ports on the gauge. Ensure connections are secure and leak-free. The gauge is designed for use with air and non-combustible, compatible gases.

5. Operating Instructions

The DM-1000 features a user-friendly interface with four buttons for operation.

5.1 Powering On/Off

Connect the device to a suitable power source. The unit will power on automatically and display the current pressure reading.

5.2 Display Units

The 4-digit LCD can display readings in various common English and metric units. Refer to the full product specifications or consult the factory for a complete list of supported units and how to cycle through them.

5.3 Zero and Span Calibration

The DM-1000 allows for simplified digital push-button zero and span calibration. This feature helps maintain accuracy over time.

  1. Zero Adjustment: With no pressure applied (or at a known zero reference), press the designated zero button (e.g., 'MENU' or 'E' in combination with others, refer to on-screen prompts or detailed product sheet).
  2. Span Adjustment: Apply a known full-scale pressure (or a specific calibration pressure) and use the span adjustment function (e.g., 'MENU' or 'E' in combination with others) to match the display to the applied pressure.

Note: For precise calibration procedures, it is recommended to consult the detailed product manual available from Dwyer Instruments or contact their technical support.

6. Maintenance

The Dwyer DigiMag DM-1000 is designed for minimal maintenance. Regular checks ensure optimal performance.

  • Cleaning: Clean the exterior of the gauge with a soft, damp cloth. Do not use abrasive cleaners or solvents.
  • Pressure Connections: Periodically inspect pressure connections for leaks or damage.
  • Recalibration: Although factory calibrated, periodic recalibration may be necessary depending on application and environmental conditions. Consult Dwyer Instruments for recommended recalibration intervals.

7. Troubleshooting

This section provides solutions to common issues encountered with the DM-1000 gauge.

ProblemPossible CauseSolution
No display/Power offNo power supply; Loose connectionCheck power source; Secure electrical connections.
Inaccurate readingsGauge needs zero/span calibration; Pressure leaks; Incorrect unit settingPerform zero/span calibration; Check pressure connections for leaks; Verify display unit setting.
Display shows "OVER" or "UNDER"Pressure exceeds/falls below rangeEnsure applied pressure is within the gauge's specified range (e.g., 0-5"WC for some models).

If the problem persists after attempting these solutions, contact Dwyer Instruments technical support.

8. Specifications

Detailed technical specifications for the Dwyer DigiMag Series DM-1000 Differential Digital Pressure Gauge.

  • Model: DM-1107
  • Manufacturer: Dwyer Instruments
  • Product Dimensions: 6.7 x 3.8 x 7.7 inches (17.0 x 9.7 x 19.6 cm)
  • Item Weight: 1.28 Pounds (0.58 kg)
  • Media Compatibility: Air and non-combustible, compatible gases
  • Display: 4-digit LCD
  • Electrical Connections: Removable terminal block for 16 to 26 AWG wire
  • Cable Gland Compatibility: 0.114 to 0.250 inches (2.9 to 6.4 mm) diameter cable
  • Housing Material: Glass filled plastic
  • First Available Date: May 26, 2012

Note: Specific pressure ranges (e.g., 0-5"WC) are factory calibrated and may vary by specific model variant. Consult the product label or original purchase documentation for your exact range.

9. Warranty and Support

For warranty information, technical support, or service inquiries, please contact Dwyer Instruments directly. Details regarding warranty periods and terms are typically provided with the product documentation or can be found on the official Dwyer Instruments website.

Manufacturer: Dwyer Instruments

For further assistance, visit the Dwyer Instruments Official Website.

© 2024 Dwyer Instruments. All rights reserved.

This manual is for informational purposes only. Specifications are subject to change without notice.