Macurco CM 6

Macurco CM-6-CH / CM-12 Carbon Monoxide CO Calibration Hood User Manual

Model: CM 6

1. Introduction

This user manual provides essential information for the safe and effective use of the Macurco CM-6-CH Carbon Monoxide (CO) Calibration Hood, designed for use with Macurco CM-6 and CM-12 fixed gas detectors. Please read this manual thoroughly before installation, operation, or maintenance to ensure proper functionality and safety.

2. Safety Information

IMPORTANT: Before use, read and follow the user instructions inside the box. Failure to do so may result in improper operation, inaccurate readings, or personal injury.

  • Always ensure the calibration hood is used in a well-ventilated area.
  • Do not paint the device, as this may interfere with its operation or sensor accuracy.
  • Handle with care to avoid damage to the hood or the detector.
  • Refer to the specific instructions for your Macurco CM-6 or CM-12 detector for detailed calibration procedures.

3. Product Overview

The Macurco CM-6-CH is a specialized calibration hood designed to facilitate the accurate calibration of Macurco CM-6 and CM-12 Carbon Monoxide fixed gas detectors. It ensures a controlled environment for applying calibration gas to the sensor, optimizing the detector's performance and accuracy.

Macurco CM-6-CH Carbon Monoxide Calibration Hood product box, front view

Figure 3.1: Front view of the Macurco CM-6-CH product packaging, showing the product name, brand, and key specifications for the CM-12 detector.

Close-up of Macurco CM-12 detector diagram and specifications on product box

Figure 3.2: Detailed view of the Macurco CM-12 detector diagram and technical specifications printed on the product box, including power, range, color, temperature, and certifications.

4. Setup

The CM-6-CH calibration hood is designed for easy attachment to compatible Macurco CO detectors. Follow these general steps for setup:

  1. Ensure your Macurco CM-6 or CM-12 detector is properly installed and powered.
  2. Carefully align the CM-6-CH calibration hood with the sensor area of the detector.
  3. Securely attach the hood to create a sealed environment around the sensor. Refer to your detector's manual for specific attachment points or methods.
  4. Connect the calibration gas cylinder to the inlet port on the calibration hood.

5. Operating the Calibration Hood

The calibration hood facilitates the application of a known concentration of CO gas to the detector's sensor for accurate calibration. Always follow the calibration procedure outlined in your Macurco CM-6 or CM-12 detector's specific user manual.

General Calibration Steps:

  1. With the calibration hood securely attached and the calibration gas connected, initiate the calibration sequence on your Macurco detector.
  2. Open the valve on the calibration gas cylinder to allow the gas to flow into the hood and over the sensor.
  3. Monitor the detector's display and indicators as per its manual until the calibration process is complete.
  4. Once calibration is confirmed, close the gas cylinder valve and remove the calibration hood.

The detector is designed to operate within a temperature range of 0 to 35°C (32 to 95°F) and a CO detection range of 0-200 PPM. Ensure your calibration environment adheres to these specifications for optimal results.

6. Maintenance

The CM-6-CH calibration hood requires minimal maintenance. Keep it clean and free from obstructions to ensure proper sealing during calibration.

  • Wipe the exterior with a soft, damp cloth. Do not use harsh chemicals or abrasive cleaners.
  • Ensure the internal surfaces that contact the detector are clean and free of debris.
  • Store the calibration hood in a clean, dry place when not in use to prevent damage.
  • As indicated on the product, do not paint the device.

7. Troubleshooting

If you encounter issues during calibration, consider the following:

  • Calibration Failure: Ensure the calibration hood is securely attached and creating a proper seal. Verify the calibration gas cylinder is not empty and the flow rate is correct.
  • Inaccurate Readings: Check the expiration date of your calibration gas. Ensure the ambient temperature is within the specified operating range for the detector.
  • Physical Damage: Inspect the calibration hood for any cracks or deformities that might prevent a proper seal. Replace if damaged.

For persistent issues, refer to the troubleshooting section of your Macurco CM-6 or CM-12 detector manual or contact Macurco technical support.

8. Specifications

FeatureSpecification
Model NumberCM-6-CH (for CM-6 / CM-12 detectors)
BrandMacurco
Item Weight0.96 ounces (27.22 g)
UPC808237728892
Power (CM-12 Detector)100-240 VAC (50 to 60 Hz)
CO Range (CM-12 Detector)0-200 PPM
Operating Temperature (CM-12 Detector)0 to 35°C (32 to 95°F)
Certifications (CM-12 Detector)UL 2075, UL 61010-1, CSA C22.2 No 61010-1
Top view of Macurco CM-6-CH product box showing UPC barcode

Figure 8.1: Top view of the Macurco CM-6-CH product packaging, displaying the UPC barcode (0783748811304) and "Made in the USA" label.

9. Warranty and Support

For warranty information and technical support regarding your Macurco CM-6-CH calibration hood or compatible detectors, please contact the manufacturer:

Manufacturer: Macurco (Manufactured by Aerionics Inc.)

Phone: 1-877-367-7891

Website: www.macurco.com

Please have your product model number (CM 6) and any relevant purchase information ready when contacting support.

© 2024 Macurco. All rights reserved. Information subject to change without notice.