Owon 01SDS6062

OWON SDS6062 60MHz Deep Memory Digital Storage Oscilloscope User Manual

Model: 01SDS6062 | Brand: Owon

1. Introduction

This manual provides comprehensive instructions for the safe and effective operation, maintenance, and troubleshooting of your OWON SDS6062 Digital Storage Oscilloscope. Please read this manual thoroughly before using the device to ensure proper functionality and to prevent damage.

The OWON SDS6062 is a 60 MHz Deep Memory Digital Storage Oscilloscope featuring 10 M points of memory and 2 channels. It is equipped with LAN and VGA interfaces for enhanced connectivity and display options. This package includes the oscilloscope, two probes, a power cable, a USB cable, this manual, and a CD with software.

2. Product Overview

2.1 Key Features

2.2 What's in the Box

Upon unpacking, verify that all items listed below are present and in good condition:

2.3 Product Views

Front view of the OWON SDS6062 Digital Storage Oscilloscope

Figure 2.1: Front view of the OWON SDS6062 Digital Storage Oscilloscope, showing the large 8-inch display and control panel with various knobs and buttons for measurement, acquisition, horizontal, vertical, and trigger settings.

Rear view of the OWON SDS6062 Oscilloscope with power input

Figure 2.2: Rear view of the OWON SDS6062 Oscilloscope, highlighting the power input socket and the main power switch. The included power cable is also shown connected.

Side view of the OWON SDS6062 Oscilloscope showing connectivity ports and probes

Figure 2.3: Side view of the OWON SDS6062 Oscilloscope, illustrating the USB (Device and Host), VGA, LAN, and Probe Compensation ports. Two oscilloscope probes and a USB cable are also displayed, indicating the included accessories.

Dimensional representation of the OWON SDS6062 Oscilloscope

Figure 2.4: A silhouette showing the approximate dimensions of the OWON SDS6062 Oscilloscope, indicating a height of approximately 7.0 inches (17 cm) for scale.

3. Safety Information

Always observe the following safety precautions to prevent injury and damage to the instrument:

4. Setup

4.1 Unpacking and Inspection

Carefully remove the oscilloscope and all accessories from the packaging. Inspect for any signs of damage during transit. If any damage is found, contact your dealer immediately.

4.2 Power Connection

  1. Ensure the oscilloscope's power switch (located on the rear panel, refer to Figure 2.2) is in the OFF position.
  2. Connect the provided power cord to the AC power input socket on the rear panel of the oscilloscope.
  3. Plug the other end of the power cord into a grounded AC power outlet.

4.3 Probe Connection and Compensation

  1. Connect the BNC connector of a probe to one of the input channels (CH1 or CH2) on the front panel.
  2. Attach the probe ground clip to the ground terminal on the oscilloscope's front panel (typically near the probe compensation output).
  3. Connect the probe tip to the probe compensation output terminal (PROBE COMP) on the front panel (refer to Figure 2.3).
  4. Power on the oscilloscope.
  5. Adjust the probe compensation screw on the probe until a flat-top square wave is displayed on the screen. This ensures accurate measurements.

4.4 Connecting to External Devices (LAN/VGA/USB)

The SDS6062 offers multiple connectivity options:

5. Operating Instructions

5.1 Power On/Off

To power on the oscilloscope, press the power switch on the rear panel to the ON position. To power off, press the switch to the OFF position.

5.2 Understanding the Display

The 8-inch TFT display provides a clear view of waveforms and measurement parameters. The screen is divided into a grid for easy amplitude and time measurement. Key information such as channel status, trigger settings, and measurement values are displayed around the waveform area.

5.3 Basic Waveform Acquisition

  1. Connect Probe: Ensure your probe is correctly connected to a channel (CH1 or CH2) and compensated.
  2. Connect to Circuit: Connect the probe tip to the test point in your circuit and the ground clip to the circuit's ground.
  3. Auto Scale: Press the Auto button on the front panel. The oscilloscope will automatically adjust the vertical and horizontal settings to display a stable waveform. This is useful for quickly viewing an unknown signal.
  4. Manual Adjustments:
    • Vertical Position (VOLTS/DIV): Use the vertical position knob for the selected channel to move the waveform up or down. Use the VOLTS/DIV knob to adjust the vertical scale (voltage per division).
    • Horizontal Position (SEC/DIV): Use the horizontal position knob to move the waveform left or right. Use the SEC/DIV knob to adjust the horizontal scale (time per division).

5.4 Triggering

Triggering stabilizes repetitive waveforms and captures single-shot events. The trigger controls are located on the right side of the front panel.

5.5 Measurements

The SDS6062 offers various automatic measurement functions:

5.6 Pass/Fail Function

The Pass/Fail function allows for automated testing against predefined waveform masks. If a waveform falls outside the specified mask, it is flagged as a "Fail".

5.7 Saving and Recalling Data

Waveforms, setups, and screenshots can be saved to internal memory or a USB flash drive.

6. Maintenance

6.1 General Care and Cleaning

To maintain the performance and longevity of your oscilloscope:

6.2 Storage

When not in use for extended periods, store the oscilloscope in a cool, dry environment, away from direct sunlight and extreme temperatures. Use the original packaging or a suitable carrying case to protect it from physical shock and dust.

7. Troubleshooting

This section addresses common issues you might encounter with your OWON SDS6062 oscilloscope. If the problem persists after following these steps, contact customer support.

ProblemPossible CauseSolution
No display after power on.Power cord not connected; power switch off; power outlet fault.Ensure power cord is securely connected. Verify power switch is ON. Test power outlet with another device.
No waveform displayed.Probe not connected; signal too small/large; incorrect vertical/horizontal settings; trigger issues.Verify probe connection to channel and circuit. Press Auto. Adjust VOLTS/DIV and SEC/DIV. Check trigger level and mode.
Unstable or rolling waveform.Incorrect trigger settings; signal noise.Adjust trigger level. Change trigger mode (e.g., to Edge). Select appropriate trigger source. Ensure stable signal input.
VGA/LAN/USB connectivity issues.Cable not connected; driver issues; software incompatibility; port malfunction.Ensure cables are securely connected. Install/reinstall drivers from the included CD. Verify software compatibility with your operating system. Test with different cables/ports if possible.
Inaccurate measurements.Probe not compensated; incorrect probe attenuation setting; signal distortion.Perform probe compensation (Section 4.3). Ensure probe attenuation setting on oscilloscope matches the probe (e.g., 1X, 10X). Check signal integrity.

8. Specifications

ParameterValue
Bandwidth60 MHz
Channels2
Sample Rate500 MS/s (Real-time)
Record Length10 M points
Display8-inch Color TFT LCD (800 x 600 pixels)
ConnectivityLAN, VGA, USB (Host & Device)
FunctionsAuto Scale, Pass/Fail, Multiple Measurement Types
Dimensions (L x W x H)14 x 3 x 7 inches (approx.)
Weight5.5 Pounds (approx.)
ManufacturerOWON

9. Warranty and Support

For information regarding warranty coverage, technical support, or service, please refer to the documentation included with your product or visit the official OWON website. Keep your purchase receipt as proof of purchase for warranty claims.

Manufacturer: OWON

Related Documents - 01SDS6062

Preview Manuel d'utilisation des oscilloscopes numériques OWON Série SDS
Manuel d'utilisation complet pour les oscilloscopes numériques OWON Série SDS, couvrant les modèles SDS6062(V) à SDS9302, avec des instructions d'opération, de sécurité et des spécifications.
Preview OWON TAO3000 Series Handheld Oscilloscopes User Manual
User manual for the OWON TAO3000 Four-Channel Series Handheld Oscilloscopes, detailing operation, features, technical specifications, and troubleshooting.
Preview OWON HDS200 Series 2-Channel Portable Oscilloscope Quick Start Guide
This quick start guide provides essential information for operating the OWON HDS200 Series 2-channel portable oscilloscopes, including models HDS272, HDS242, HDS2102, and HDS2202. It covers safety precautions, general inspection, oscilloscope and multimeter usage, signal generator functions, software connectivity, troubleshooting, and maintenance.
Preview Multimeter SCPI Protocol Guide
Explore the Standard Commands for Programmable Instruments (SCPI) protocol for OWON multimeters. This guide details command syntax, rules, and common commands for programming test equipment.
Preview OWON HDS200 Series Handheld Oscilloscope User Manual - Operation and Specifications
This comprehensive user manual provides detailed instructions for operating the OWON HDS200 Dual Channel Series handheld oscilloscopes, including models HDS272(S) and HDS242(S). It covers safety information, general inspection, oscilloscope, multimeter, and optional waveform generator functions, along with technical specifications and maintenance guidelines.
Preview OWON FDS Dual-Channel Series Quick Guide: Oscilloscope, Function Generator, Power Supply, Multimeter
This quick guide provides essential information for operating the OWON FDS Dual-Channel Series, a versatile instrument featuring an oscilloscope, arbitrary waveform/function generator, power supply, and multimeter. Learn about safety, basic functions, and operational tips from Fujian LILLIPUT Optoelectronics Technology Co., Ltd.