Hanna Instruments HI1285-6

Hanna Instruments HI1285-6 pH/EC/TDS Multiparameter Probe User Manual

Model: HI1285-6

1. Introduction

The Hanna Instruments HI1285-6 is a multiparameter probe designed for simultaneous measurement of pH, Electrical Conductivity (EC), and Total Dissolved Solids (TDS). This probe is specifically engineered as a replacement for the GroChek HI9813 meter, ensuring accurate and reliable readings for various applications, including hydroponics, agriculture, and environmental monitoring. This manual provides essential information for the proper setup, operation, and maintenance of your HI1285-6 probe.

Hanna Instruments HI1285-6 pH/EC/TDS Multiparameter Probe

Figure 1: Hanna Instruments HI1285-6 pH/EC/TDS Multiparameter Probe. This image displays the white cylindrical probe with a black connector at the top and a clear sensing tip at the bottom, showing the internal components.

2. Setup and Installation

Proper setup is crucial for accurate measurements. Follow these steps to install your HI1285-6 probe:

  1. Unpacking: Carefully remove the probe from its packaging. Inspect for any visible damage.
  2. Connector: Ensure the probe's connector is clean and dry. Align the connector with the corresponding port on your compatible meter (e.g., GroChek HI9813).
  3. Secure Connection: Gently push and twist the connector until it is firmly seated. Avoid overtightening.
  4. Pre-conditioning: Before first use, or after prolonged storage, immerse the probe tip in HI70300 storage solution for at least 30 minutes to hydrate the pH glass bulb.
  5. Rinsing: Rinse the probe tip with deionized water before and after each measurement.

3. Operating Instructions

The HI1285-6 probe is designed for use with compatible Hanna Instruments meters. Refer to your meter's instruction manual for specific operational procedures. General guidelines are provided below:

  • Calibration: Regular calibration is essential for accurate readings. Calibrate the pH, EC, and TDS functions according to your meter's instructions using fresh calibration solutions. The CAL Check feature, if supported by your meter, assists in verifying probe condition.
  • Measurement:
    1. Immerse the probe tip into the sample solution, ensuring the sensing elements are fully submerged.
    2. Gently stir the probe to remove any air bubbles and allow the reading to stabilize.
    3. Record the stabilized readings from your meter.
  • Temperature Compensation: The probe features built-in temperature compensation for accurate EC/TDS readings. Ensure your meter is set to the correct temperature compensation mode.

4. Maintenance and Storage

Proper maintenance extends the life and accuracy of your probe:

  • Cleaning:
    • After each use, rinse the probe thoroughly with deionized water.
    • If deposits are visible, clean the probe with a suitable cleaning solution (e.g., Hanna Instruments HI7061 for general purpose).
    • Never scratch or rub the glass bulb or platinum sensors.
  • Storage:
    • Always store the pH/EC/TDS probe with its tip immersed in HI70300 storage solution. Never store in deionized water.
    • Ensure the protective cap is filled with storage solution and securely placed over the probe tip when not in use.
  • Electrolyte Level: For refillable probes (if applicable to your specific HI1285-6 variant), ensure the electrolyte level is maintained above the reference junction.

5. Troubleshooting

If you encounter issues with your HI1285-6 probe, consider the following common problems and solutions:

ProblemPossible CauseSolution
Unstable/Erratic Readings
  • Dirty probe tip
  • Air bubbles on sensor
  • Damaged probe
  • Insufficient hydration
  • Clean probe with appropriate solution.
  • Gently stir to dislodge bubbles.
  • Inspect for physical damage; replace if necessary.
  • Re-hydrate in storage solution for several hours.
Slow Response Time
  • Dehydrated pH glass
  • Clogged reference junction
  • Aged probe
  • Soak in storage solution overnight.
  • Clean with specialized cleaning solution.
  • Consider probe replacement.
Incorrect Readings (after calibration)
  • Expired/contaminated calibration solutions
  • Improper calibration procedure
  • Damaged probe
  • Use fresh calibration solutions.
  • Repeat calibration carefully following meter instructions.
  • Inspect probe; replace if damaged.

If problems persist, contact Hanna Instruments technical support or your local distributor for further assistance.

6. Specifications

Key specifications for the Hanna Instruments HI1285-6 probe:

  • Model: HI1285-6
  • Type: pH/EC/TDS Multiparameter Probe
  • Compatibility: Designed as a replacement probe for Hanna Instruments GroChek HI9813 meters.
  • Dimensions (Package): Approximately 10.2 x 2.68 x 2.05 inches (25.9 x 6.8 x 5.2 cm)
  • Weight (Item): Approximately 7.6 ounces (215 grams)
  • Manufacturer: Hanna Instruments

7. Warranty and Support

Hanna Instruments products are warranted against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The specific warranty period for the HI1285-6 probe may vary. Please refer to the warranty information provided with your original meter or contact Hanna Instruments customer service for details.

For technical support, service, or to order replacement parts and solutions, please visit the official Hanna Instruments website or contact their customer support department. Always provide the model number (HI1285-6) and any relevant meter information when seeking support.

Hanna Instruments Official Website: www.hannainst.com

Related Documents - HI1285-6

Preview Hanna Instruments HI9813-51 & HI9813-61 Portable pH/EC/TDS/Temperature Meters Instruction Manual
Comprehensive instruction manual for Hanna Instruments HI9813-51 and HI9813-61 portable meters, covering specifications, operation, calibration, measurement, maintenance, and troubleshooting.
Preview Hanna HI5522 Research Grade Meter: pH, ORP, ISE, EC, TDS, Salinity, Resistivity, and Temperature
Detailed overview of the Hanna HI5522 research-grade benchtop meter, highlighting its advanced features for measuring pH, ORP, ISE, EC, TDS, Salinity, Resistivity, and Temperature. Includes information on calibration, data logging, incremental methods, and USP compliance.
Preview Hanna HI9814 GrePro pH/EC/TDS/Temperature Meter Instruction Manual
Instruction manual for the Hanna HI9814 GrePro, detailing its features for measuring pH, EC, TDS, and temperature. Covers setup, operation, calibration, maintenance, specifications, and warranty.
Preview Hanna Instruments Dissolved Oxygen & Multiparameter Meters Catalog
Explore Hanna Instruments' comprehensive catalog of advanced Dissolved Oxygen (DO) meters and versatile multiparameter water quality analysis instruments. Discover detailed product specifications, features, and applications for professional environmental and laboratory use.
Preview Manual de Instrucciones Hanna HI98330 Tester de Nutrientes GroLine® CE/TDS
Manual de instrucciones detallado para el Tester de Nutrientes Hanna HI98330 GroLine® CE/TDS. Cubre especificaciones, operación, mantenimiento, limpieza y garantía para aplicaciones hidropónicas y agrícolas.
Preview Hanna Instruments HI98195 Multiparameter Waterproof Meter Specifications and Features
Detailed specifications, features, and technical information for the Hanna Instruments HI98195 waterproof portable logging multiparameter meter, covering pH, ORP, EC, TDS, Resistivity, Salinity, and Temperature measurements.