Introduction
This manual provides instructions for the proper use and maintenance of the Greisinger GE 101 Penetration Electrode. This electrode is designed for precise pH and temperature measurements in various applications, including industrial, agricultural, pharmaceutical, cosmetic, and food processing sectors.

Image of the Greisinger GE 101 Penetration Electrode, a slender probe with a cable connection, suitable for direct insertion into various materials.
Product Overview
The Greisinger GE 101 is a robust penetration electrode suitable for measuring pH values from 0 to 12 and temperatures from 0 to 80 °C. Its design allows for direct insertion into semi-solid or viscous substances, making it versatile for diverse measurement needs.
Key Features:
- pH measurement range: 0 to 12 pH
- Temperature measurement range: 0 to 80 °C
- Designed for penetration applications
- Compatible with digital pH/mV meters (e.g., GPHR 1400)
Setup
Unpacking: Carefully remove the electrode from its packaging. Inspect for any visible damage before proceeding.
Connection: Connect the electrode's cable to a compatible pH/mV meter. Ensure the connection is secure and properly seated to prevent signal loss.
Initial Calibration: Before first use, and regularly thereafter, calibrate the electrode according to your pH meter's instructions using standard buffer solutions. This step is crucial for ensuring accurate and reliable readings.
Operating Instructions
Preparation: Ensure the electrode tip is clean and free from residues. Rinse with distilled water if necessary. Avoid touching the glass membrane with bare hands.
Measurement Procedure:
- Gently insert the electrode tip into the substance to be measured. Ensure good contact with the material and that the sensing part is fully immersed.
- Allow sufficient time for the reading to stabilize on your connected pH meter. Stabilization time may vary depending on the sample and electrode condition.
- Record the pH and temperature values as displayed on the meter.
Post-Measurement: After each measurement, rinse the electrode thoroughly with distilled water. Store the electrode properly in a suitable storage solution to prevent drying out and maintain its performance.
Maintenance
Cleaning:
- Regularly clean the electrode tip to remove any accumulated samples or contaminants. Use a soft cloth or brush and a mild cleaning solution if needed, followed by a thorough rinse with distilled water.
- Avoid abrasive materials that could damage the glass membrane, as this can impair accuracy and lifespan.
Storage:
- When not in use, store the electrode in a dedicated storage solution (e.g., 3 mol/L KCl solution or a pH 4 buffer with KCl).
- Do not store the electrode dry, as this can lead to irreversible damage to the pH-sensitive glass membrane and reference junction.
Calibration: Regular calibration is crucial for accurate measurements. The frequency depends on usage, required precision, and the nature of the samples being measured.
Troubleshooting
Inaccurate Readings:
- Cause: Electrode not calibrated or improperly calibrated. Solution: Calibrate the electrode with fresh, certified buffer solutions.
- Cause: Electrode tip dirty or damaged. Solution: Clean the electrode carefully or inspect for damage. Replace if the glass membrane or reference junction is visibly compromised.
- Cause: Electrode stored dry. Solution: Rehydrate the electrode in storage solution for several hours. If performance does not improve, replacement may be needed.
Slow Response Time:
- Cause: Electrode aging or contamination of the reference junction. Solution: Clean the electrode. If the issue persists, the electrode may be nearing the end of its operational lifespan and require replacement.
Specifications
| Property | Value |
|---|---|
| Model | GE 101 |
| Manufacturer | GREISINGER |
| pH Range | 0 - 12 pH |
| Temperature Range | 0 - 80 °C |
| Item Weight | 9.07 grams |
| ASIN | B003A60L6E |
| Model Number | 600767 |
| International Article Code | 04016138033281 |
Warranty and Support
For detailed warranty information and technical support, please refer to the official GREISINGER website or contact your local authorized distributor. Keep your purchase receipt and product model number (600767) handy for any inquiries or service requests.
Online Resources: www.greisinger.de