Introduction
The Yieryi C-600 is a versatile 7-in-1 water quality tester designed for accurate measurement of pH, Total Dissolved Solids (TDS), Electrical Conductivity (EC), Oxidation-Reduction Potential (ORP), Salinity, Specific Gravity (S.G), and Temperature. This device is ideal for various applications including drinking water analysis, aquariums, swimming pools, hydroponics, and laboratories.
Package Contents
- 1 x 7-in-1 Water Quality Tester (C-600)
- 1 x English User Manual
- 3 x pH Buffer Powder (pH 4.00, pH 6.86, pH 9.18)

Image: The C-600 tester, English manual, and three packets of pH buffer powder.
Product Overview

Image: Diagram showing the main components of the C-600 water quality tester.
The C-600 water quality tester consists of the following main components:
- Battery Compartment: Located at the top, houses 3 x 1.5V LR44 batteries.
- Power Button (ON/OFF): To turn the device on or off.
- Mode Key / Calibration Key (MODE/CAL): To switch between measurement modes and initiate calibration.
- Screen Display: LCD screen to show readings and indicators.
- Data Hold / Temperature Key (HOLD/TEMP): To freeze readings or switch temperature units.
- Protective Cap: Covers the electrode when not in use.
- High Precision Probe: The sensing element for measurements.
Specifications
| Parameter | Range | Resolution | Accuracy |
|---|---|---|---|
| pH | 0.00-14.00pH | 0.01pH | ±0.05pH |
| TDS | 0-10000ppm, 10.1-200.0ppt | 0.1ppm; 0.1ppt | ±2%F.S |
| EC | 0-10000µS/cm, 10.1-19.99mS/cm, 20.1-400.0mS/cm | 1µS/cm; 0.1mS/cm | ±2%F.S |
| Salinity | 0.01%-25.00%, 0-10000ppm, 10.1-200.0ppt | 1ppm; 0.1ppt; 0.01% | ±2%F.S (0.01-5.00%: ±0.1%; 5.1-25.00%: ±1%) |
| ORP | ±999mV | 1mV | ±2mV |
| S.G | 1.000-1.222 | N/A | N/A |
| Temperature | 0.1-60.0°C (32.0-140°F) | 0.1°C/°F | ±0.5°C |
Additional Specifications:
- Waterproof Level: IP67
- Automatic Temperature Compensation (ATC): 0-60°C
- Operation Environment: 0-60.0℃ (32.0-140°F); RH 100%
- TDS Factor: 0.5
- Automatic Shut-off: 5 minutes
- Display: LCD Digital Display
- Backlight: Optional (C-600A model)
- Battery: 3 x 1.5V (LR44) (Not included)
- Screen Size: 33 x 24mm
- Dimensions/Weight: 183 x 37 x 37mm (7.2 x 1.5 x 1.5in) / 90g (3.2oz)
Setup
Battery Installation
The device requires 3 x 1.5V LR44 button batteries (not included). To install:
- Unscrew and remove the battery compartment cap at the top of the device.
- Insert the batteries according to the polarity markings (+/-) indicated inside the compartment.
- Replace the cap and screw it on securely.
Initial Preparation
- Remove the protective cap from the electrode.
- If the electrode appears dry, rinse it with distilled water before first use.
Operating Instructions
Power On/Off
Press the 'ON/OFF' button to turn the device on or off.
Mode Switching
Press the 'MODE/CAL' button to cycle through the different measurement modes: EC, TDS, Salinity, pH, ORP, S.G, and Temperature.

Image: Key features of the C-600 tester including data retention, IP67 waterproof rating, automatic temperature compensation, and temperature unit conversion.

Image: Visual representation of the seven measurement functions: EC, Salinity, pH, S.G, ORP, Temperature, and TDS.
Data Hold
Press the 'HOLD/TEMP' button briefly to freeze the current reading on the display. Press again to release the hold function.
Temperature Unit Conversion
While the device is on, press and hold the 'HOLD/TEMP' button for 5 seconds to switch the temperature unit between Celsius (°C) and Fahrenheit (°F).
Understanding Readings
Immerse the electrode into the sample solution. Stir gently and wait for the reading to stabilize before recording.
pH Value Reference

Image: pH scale from 0 to 14, indicating acidic, neutral, and alkaline ranges with common examples.
TDS Value Reference

Image: TDS value reference chart for water quality: Pure water (0-9ppm), Mountain spring water (10-60ppm), Purified water (60-100ppm), Tap water (100-300ppm), Polluted water (>300ppm).
ORP Value Reference

Image: ORP value reference table correlating ORP ranges (mV) with breath type and microbial species.
Calibration
pH Calibration
It is recommended to calibrate the pH function regularly for accurate readings. Use the provided pH buffer powders (pH 4.00, pH 6.86, pH 9.18).
Video: Official video demonstrating the pH calibration process using buffer solutions.
- Prepare pH 4.00, pH 6.86, and pH 9.18 buffer solutions by dissolving each powder packet in 250ml of distilled water. Stir gently.
- Turn on the meter.
- Immerse the electrode into the pH 6.86 buffer solution.
- Press and hold the 'MODE/CAL' button for 5 seconds. The display will flash 'CAL' and then '6.86'. It will flash 3 times to indicate completion of 6.86 calibration.
- Rinse the electrode thoroughly with distilled water.
- Immerse the electrode into the pH 4.00 buffer solution.
- Press and hold the 'MODE/CAL' button for 5 seconds. The display will flash 'CAL' and then '4.00'. It will flash 3 times to indicate completion of 4.00 calibration.
- Rinse the electrode thoroughly with distilled water.
- Immerse the electrode into the pH 9.18 buffer solution.
- Press and hold the 'MODE/CAL' button for 5 seconds. The display will flash 'CAL' and then '9.18'. It will flash 3 times to indicate completion of 9.18 calibration.
- Rinse the electrode thoroughly with distilled water. pH calibration is now complete.
EC/TDS/Salinity Calibration
The meter can automatically recognize standard solutions for EC/TDS/Salinity calibration (e.g., 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm).
- Prepare the desired standard solution (e.g., 1413 µS/cm).
- Turn on the meter and switch to EC mode.
- Immerse the electrode into the standard solution.
- Press and hold the 'MODE/CAL' button for 5 seconds. The display will flash 'CAL' and then the value of the standard solution (e.g., '1413'). It will flash 3 times to indicate completion.
- Rinse the electrode thoroughly with distilled water.
- Repeat for other standard solutions as needed.
- If the meter cannot identify the solution, the electrode may be faulty and require replacement, or contact customer support.
Maintenance
Electrode Care
- Always clean the electrode with pure (distilled) water after each use and gently wipe it dry.
- When taking measurements, stir the electrode gently several times to ensure no air bubbles are trapped around the sensor, which can affect accuracy.
- Store the electrode with its protective cap on to prevent damage and keep it from drying out excessively.
- If the electrode is damaged or consistently fails to identify calibration solutions, it may need replacement.
Battery Replacement
When the low battery indicator flashes on the display, it's time to replace the batteries. Follow the battery installation steps outlined in the Setup section, using 3 new 1.5V LR44 button batteries.
Troubleshooting
pH Reset
If you experience persistent issues with pH readings or calibration, you can perform a pH reset:
- Turn off the meter.
- Press and hold the 'HOLD/TEMP' and 'MODE/CAL' buttons simultaneously.
- While holding these two buttons, press the 'ON/OFF' button.
- Release all three buttons. The device is now reset to factory pH settings.
General Issues
- Inaccurate or fluctuating pH readings: Ensure the electrode is clean. Perform a 2-point or 3-point calibration. If calibration fails repeatedly, the electrode might be faulty.
- ORP readings fluctuate: ORP measurements are highly sensitive to changes in the solution and can naturally fluctuate. If the glass bulb of the ORP sensor is visibly broken, the sensor is faulty and needs replacement.
- No backlight (if applicable): Verify that you have the C-600A model, which includes a backlight. If you have the C-600A and the backlight is not working, check batteries or contact support.
- Readings not changing significantly or stuck: Ensure the electrode is fully immersed and clean. Try recalibrating the specific parameter.
- Device not turning on: Check that the batteries are correctly installed and are not depleted. Replace with new batteries if necessary.
- No instruction manual in your language: The product typically comes with an English manual. For other languages, please refer to the seller or manufacturer's website for digital versions.
User Tips
- Always use distilled water for preparing calibration solutions to ensure accuracy.
- If you suspect inaccurate pH readings, perform a 2-point or 3-point calibration using buffer solutions that bracket your expected measurement range.
- Ensure the electrode is thoroughly cleaned with distilled water after each use to maintain accuracy and prolong its lifespan.
- When ordering, confirm the model number (C-600 vs. C-600A) if a backlight feature is important to you.
- For ORP measurements, understand that readings can naturally fluctuate due to the sensitivity of the parameter.
Warranty and Support
Warranty
The Yieryi C-600 water quality tester is warranted to be free from defects in material and workmanship for a period of three years from the date of purchase. This warranty covers normal operational defects but does not extend to man-made damage or damage resulting from misuse. Proof of purchase is required for all warranty repair claims. The warranty becomes void if the meter has been disassembled or tampered with.
Return Authorization
Before returning any item for warranty service or any other reason, authorization must be obtained from the supplier. When requesting a return authorization, please provide detailed information regarding the defective reason or issue. Ensure the meter is returned with adequate protective packaging to prevent any damage during shipment. It is recommended to insure the shipment against potential loss or damage.