1. Introduction
Thank you for choosing the Yieryi 5 in 1 Water Quality Tester. This versatile digital meter is designed to accurately measure pH, Total Dissolved Solids (TDS), Electrical Conductivity (EC), Salinity, and Temperature in various water sources. It is an essential tool for monitoring water quality in applications such as aquaculture, drinking water, swimming pools, aquariums, food processing, and laboratories.
This manual provides detailed instructions for the proper setup, operation, calibration, and maintenance of your device to ensure optimal performance and longevity.

2. Key Features
- 5-in-1 Functionality: Measures pH, TDS, EC, Salinity, and Temperature.
- High Accuracy: Provides precise readings for critical water parameters.
- Automatic Temperature Compensation (ATC): Ensures accurate readings across varying temperatures.
- Backlight Display: Clear, easy-to-read digital display (available on some models).
- IP67 Waterproof Rating: Protection against dust penetration and short soaking.
- Replaceable Electrode: Allows for easy maintenance and extended device lifespan.
- Automatic Shut-off: Conserves battery life by automatically turning off after 5 minutes of inactivity.
- 3-Point Calibration: For pH, and multi-point calibration for EC/TDS/Salinity.
3. Product Components
The Yieryi 5 in 1 Water Quality Tester consists of the following main parts:
- Protective Cap: Covers the electrode when not in use.
- Electrode: The sensing element that comes into contact with the water sample.
- Digital Display: Shows measurement values, units, and indicators.
- Control Buttons:
- Power Button (⏻): Turns the device ON/OFF.
- HOLD/TEMP Button (▲): Press to hold the current reading; long press to switch between Celsius (℃) and Fahrenheit (℉).
- MODE/CAL Button (▼): Press to cycle through measurement modes (pH, EC, TDS, Salinity); long press to enter calibration mode.
- Battery Compartment: Located at the top, secured by a screw cap.



4. Specifications
| Parameter | Range | Resolution | Accuracy |
|---|---|---|---|
| pH | 0.01 - 14.00 pH | 0.01 pH | ±0.05 pH |
| EC | 0 - 10000 µS/cm 10.01 - 19.99 mS/cm 20.1 - 400 mS/cm | 1 µS/cm 0.01 mS/cm 0.1 mS/cm | ±2% F.S. |
| TDS | 0 - 1000 ppm 1000 - 10000 ppm 10.1 - 200.0 ppt | 1 ppm 10 ppm 0.1 ppt | ±2% F.S. |
| Salinity | 0.00 - 25.00 % | 0.01 % | 0.01-5.00% (±0.1%) 5.10-25.00% (±1%) |
| Temperature | 0.1 - 60.0 ℃ / 32.0 - 140.0 ℉ | 0.1 ℃ / 0.1 ℉ | ±0.5 ℃ |
- Automatic Temperature Compensation (ATC): Yes
- Power Supply: 3 x 1.5V LR44 button cell batteries (not included)
- Automatic Shut-off: 5 minutes
- Waterproof Rating: IP67
- Backlight: Yes (on models EZ-9909A, EZ-9909SP)
5. Setup and Battery Installation
- Unpack: Carefully remove the tester and accessories from the packaging.
- Battery Installation:
- Unscrew the battery compartment cap at the top of the device.
- Insert three 1.5V LR44 button cell batteries, ensuring correct polarity (+/-).
- Securely screw the cap back on.
- Initial Cleaning: Before first use, rinse the electrode with distilled water and gently blot dry with a clean tissue.
- Remove Protective Cap: Always remove the protective cap before use.
6. Operation
6.1 Power On/Off
Press the Power button (⏻) to turn the device ON or OFF.
6.2 Temperature Unit Conversion
- Turn on the device.
- Press the HOLD/TEMP button (▲) briefly to hold the reading.
- Long press the HOLD/TEMP button (▲) to switch between Celsius (℃) and Fahrenheit (℉).

6.3 Mode Switching
Press the MODE/CAL button (▼) briefly to cycle through the measurement modes: pH → EC → TDS → Salinity → Temperature.

6.4 Measurement Procedure
- Turn on the meter.
- Immerse the electrode into the target solution, ensuring the liquid level is below the immersion line.
- Gently stir the solution and wait for the reading to stabilize on the display.
- Press the HOLD/TEMP button (▲) to lock the reading if desired.
- Record the measurement.
- After use, rinse the electrode with distilled water and replace the protective cap.




7. Calibration
Regular calibration ensures the accuracy of your meter. Calibration solutions are typically required and may need to be purchased separately.
7.1 pH Calibration (3-Point Automatic)
The meter supports 3-point pH calibration (6.86, 4.00, 9.18) with automatic buffer recognition.
- Prepare Solutions: Have pH 6.86, pH 4.00, and pH 9.18 buffer solutions ready.
- Clean Electrode: Rinse the electrode with distilled water and blot dry.
- Enter Calibration Mode: Turn on the meter, switch to pH mode, and long press the MODE/CAL button (▼) for 6 seconds until 'CAL' appears.
- Calibrate pH 6.86:
- Immerse the electrode into the pH 6.86 buffer solution.
- Wait for the reading to stabilize. The display will flash '6.86' three times and then show the stable reading. This indicates successful calibration.
- Calibrate pH 4.00:
- Rinse the electrode thoroughly with distilled water.
- Immerse the electrode into the pH 4.00 buffer solution.
- Wait for the reading to stabilize. The display will flash '4.00' three times and then show the stable reading.
- Calibrate pH 9.18:
- Rinse the electrode thoroughly with distilled water.
- Immerse the electrode into the pH 9.18 buffer solution.
- Wait for the reading to stabilize. The display will flash '9.18' three times and then show the stable reading.
- Exit Calibration: The meter will automatically exit calibration mode after successful 3-point calibration.

7.2 EC/TDS/Salinity Calibration
The meter supports calibration for EC/TDS/Salinity using standard solutions (e.g., 1413 µS/cm and 12.88 mS/cm).
- Prepare Solutions: Have the appropriate EC/TDS/Salinity standard solutions ready.
- Clean Electrode: Rinse the electrode with distilled water and blot dry.
- Enter Calibration Mode: Turn on the meter, switch to EC/TDS/Salinity mode, and long press the MODE/CAL button (▼) for 6 seconds until 'CAL' appears.
- Calibrate:
- Immerse the electrode into the standard solution (e.g., 1413 µS/cm).
- Wait for the reading to stabilize. The display will flash the standard value three times and then show the stable reading. This indicates successful calibration.
- Repeat with other standard solutions if necessary (e.g., 12.88 mS/cm).
- Exit Calibration: The meter will automatically exit calibration mode after successful calibration.


8. Maintenance
- Electrode Cleaning: Always rinse the electrode with distilled water after each use to prevent contamination and ensure accurate future readings.
- Electrode Storage: Store the electrode with the protective cap on. If the electrode becomes dry, soak it in electrode storage solution or pH 4.00 buffer solution for at least 30 minutes before use.
- Battery Replacement: Replace batteries when the low battery indicator appears on the display. Use three 1.5V LR44 button cell batteries.
- General Care: Keep the device clean and dry. Avoid exposing it to extreme temperatures or direct sunlight for prolonged periods.
- Electrode Lifespan: Electrodes are consumable parts. Their lifespan depends on usage frequency and maintenance. Replace the electrode if readings become consistently inaccurate despite proper calibration.
9. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| Inaccurate readings | Dirty electrode, expired calibration, dry electrode, faulty electrode. | Clean the electrode, recalibrate the meter, soak the electrode in storage solution, replace the electrode. |
| Meter does not turn on | Dead batteries, incorrect battery installation. | Replace batteries, check battery polarity. |
| Display shows 'CAL' but doesn't calibrate | Incorrect calibration solution, unstable reading. | Ensure correct buffer solution is used, wait for reading to stabilize before confirming. |
| Electrode cannot be replaced | Incorrect procedure. | Unscrew the electrode collar carefully to replace. Refer to product video for visual guide. |
10. User Tips
- Calibration Solutions: Ensure you have fresh and correct calibration solutions for pH and EC/TDS/Salinity. These are crucial for maintaining accuracy and may not be included with the meter.
- Electrode Care: The accuracy of your meter heavily relies on the condition of its electrode. Always keep it clean and properly stored.
- Automatic Shut-off: The meter will automatically power off after 5 minutes of inactivity to save battery life.
- Backlight: Some models feature a backlight for easier reading in low-light conditions. This is a model-specific feature.
- Model Variants: Models EZ-9902 and EZ-9909 refer to the same product, with EZ-9909 being the product code for better identification.
11. Warranty and Support
For any issues or questions regarding your Yieryi 5 in 1 Water Quality Tester, please refer to the official user manual (PDF) available for download or contact Yieryi customer support. The PDF manual can be referenced at User manual (PDF).
Please note that electrodes are consumable parts and their lifespan is not typically covered under standard warranty against wear and tear.
12. Product Video Guide
Watch this video for a visual guide on product features, operation, and calibration procedures.