1. Introduction

The BRIGHTWIN LB02A Process Calibrator is a versatile multifunction instrument designed for precise measurement and generation of various electrical signals. It covers essential functions for PLC, DCS, ESD, field instrumentation, and valve maintenance. Featuring a color LCD with backlight, it offers a compact and portable design with simple operation. The LB02A supports a wide range of signal types including Voltage, Millivolt, Current (mA), Resistance, and Frequency, and includes Modbus master/slave and Hart functions for advanced applications.

Front view of the LB02A Process Calibrator displaying current output.
Figure 1: LB02A Process Calibrator displaying current output.

2. Safety Precautions

Please read and understand the following safety precautions before operating the BRIGHTWIN LB02A Process Calibrator. Failure to comply may result in personal injury or damage to the instrument.

  • This instrument incorporates multiple internal self-recovery fuses and a quick one-time fuse. While these provide protection, no measures can guarantee 100% complete and reliable protection. Always prioritize user safety.
  • The instrument can withstand up to 30V DC or 20V AC for 5 seconds during an error operation. Do NOT connect the instrument to 220V mains power under any circumstances.
  • During operation, the instrument may generate some internal heat, which can affect the accuracy of internal temperature measurement components. For critical temperature measurements, it is recommended to use an external Pt100 temperature module to minimize errors caused by ambient temperature, active output current, load resistance, and operator hand temperature.
  • When using resistance output, the accuracy is related to the external excitation current. If the current is too low, the output resistance may have a degree of error. Use a suitable multimeter for verification.

3. Product Features

  • Multifunctionality: Supports measurement and output for Voltage (V, mV), Current (mA), Resistance (Ω), Frequency (Hz), Thermocouple (TC), and Resistance Temperature Detectors (RTD).
  • Enhanced Safety: Built-in fuses protect against damage from incorrect connections, particularly accidental connection to 220V mains.
  • High Accuracy & Stability: Utilizes high-precision, temperature-stable AD, DA, reference voltage, and operational amplifier chips to ensure long-term performance and accuracy.
  • Optimized Circuit Design: Reduced sensitivity of thermal resistance and resistance output to external excitation current, improving output precision. Compatible with constant current and non-constant current resistance measurement methods.
  • User-Friendly Interface: Compact and portable design with a clear color LCD and intuitive panel layout for simple operation.
  • Advanced Communication: Includes Modbus master/slave and Hart functions for integration into complex industrial systems.
  • Battery Included: Yes.
  • Display Size: 3.0 - 4.9 Inches.
  • Certifications: CE, RoHS.

4. Specifications

General Specifications

Attribute Value
Model NumberBRT LB02A
Brand NameBRIGHTWIN
DIY SuppliesElectrical
OriginMainland China
CertificationCE, RoHS
Display Size3.0 - 4.9 Inches
Size122x78x27mm
DisplayColor LCD with back light
Accuracy0.025% TYP
ColorBlack and Grey
Battery IncludedYes
High-concerned chemicalNone
VIP NoteNo MODBUS Hart functions (This note seems contradictory to other features, please refer to the Modbus/Hart sections for specific model capabilities.)

Output Functions (LB02A)

Function Output Type LB02A
DC VoltageV✓
DC MillivoltmV✓
OhmΩ✓
DC CurrentmA✓
Thermocouple TCR, S, B, K, E, J, T, N✓
Resistance Temperature Detector RTDPt100, Cu50, Pt1000✓
24V Loop PowerV✓
FrequencyHz✓
Modbus MasterX
Modbus SlaveX
Hart HandheldX
Firmware Upgradable✓

Measurement Functions (LB02A)

Function Measurement Type LB02A
DC VoltageV✓
DC MillivoltmV✓
OhmΩ✓
DC CurrentmA✓
Thermocouple TCR, S, B, K, E, J, T, N✓
Resistance Temperature Detector RTDPt100, Cu50, Pt1000✓
FrequencyHz✓
Range ConversionV, mV, Ω, mA, Hz✓

5. Setup

5.1 Initial Inspection

Upon receiving your LB02A Process Calibrator, carefully unpack all components and inspect for any signs of damage. Ensure all accessories are present:

  • LB02A Process Calibrator unit
  • Test leads (red and black)
  • USB charging cable
  • Power adapter
  • Protective carrying case
  • User Manual (physical copy)
LB02A Process Calibrator, user manual, and accessories laid out.
Figure 2: Included accessories with the LB02A Calibrator.

5.2 Charging the Battery

The LB02A uses a lithium polymer battery. Before first use, or if the battery is low, charge the unit using the provided USB cable and power adapter. The charging process typically takes 5-6 hours.

  • Connect the USB charging cable to the calibrator and the power adapter.
  • Plug the power adapter into a suitable power outlet.
  • The battery can be charged whether the calibrator is turned on or off.
  • When charging while the unit is ON, the battery icon on the display will flash, indicating charging in progress.
  • When charging while the unit is OFF, there will be no display indication, but charging will proceed. The unit will automatically stop charging once the battery is full.
  • Avoid completely draining the battery. For long-term storage, ensure the battery is fully charged.

6. Operating Instructions

6.1 Power On/Off

Press the red power button (ن) to turn the calibrator ON or OFF.

6.2 Basic Operation: Measure and Output Modes

The LB02A can operate in two primary modes for each function: Measure mode (←) and Output mode (→). To switch between these modes for a specific function (e.g., Voltage, Current, mV, etc.):

  1. Press the function key corresponding to the desired signal type (e.g., V for Voltage). The unit will initially enter the Measure mode for that function.
  2. Press the same function key again to switch to the Output mode for that function (if both input and output capabilities exist for that function).

Example: mV Measurement

Front view of the LB02A Process Calibrator displaying mV measurement.
Figure 3: Conventional mV measurement mode display.

6.3 Full-Function Mode and Digital Keyboard Input

For more precise adjustments or to access advanced parameters, activate the Full-Function mode. In this mode, a digital keyboard and special function keys become active. When Full-Function mode is active, parts of the screen will have a light teal background.

  • Small Range Adjustment: Use the ↗ and ↘ keys for small incremental adjustments.
  • Large Range Adjustment: Use the ↱ and ↰ keys for larger incremental adjustments.
  • Parameter Selection: Press the → key (next to 'RTD') to cycle through available parameters.
  • Digital Delete: Press the Del key to delete digits from right to left.
  • Confirm: Press the Enter key to confirm input parameters.
  • Enter/Exit Full-Function Mode: Use the F-Fn key.
Video 1: Demonstration of adjusting current output using the LB02A Process Calibrator.

6.4 Range Conversion

The LB02A supports range conversion for various signals. The formula for range conversion (e.g., for current) is:

CNT = (mA_in * (C_H - C_L)) / (mA_H - mA_L)

Where:

  • CNT: Count (unit of range conversion)
  • C-H: Upper limit of range conversion
  • C-L: Lower limit of range conversion
  • V-H/V-L: Voltage upper/lower limit
  • mA-H/mA-L: Current upper/lower limit
  • mV-H/mV-L: mV upper/lower limit
  • R-H/R-L: Resistance upper/lower limit
  • Hz-H/Hz-L: Frequency upper/lower limit
  • C-T: Time of a complete cycle for current program output
  • W-T: Waiting time for current program output to reach limits
  • Hz-V: Threshold value for frequency measure/output (input: signal must exceed to count; output: peak of waveform)
  • Hz-DC: PWM Duty cycle setting for frequency output

6.5 Wire Resistance Compensation (for RTD)

When measuring thermal resistance (RTD), wire resistance compensation can be applied:

  • Manual Compensation (WR-M): Allows manual setting of wire resistance compensation within a range of 0-50 Ω.
  • Automatic Compensation (WR-A): If the measured resistance value is within 0-10 Ω, the calibrator will automatically set this value as the wire resistance compensation, eliminating the need for manual adjustment.

6.6 Thermocouple (TC) and RTD Functions

The LB02A supports various types of Thermocouples (TC) and Resistance Temperature Detectors (RTD) for both measurement and output. TC types (R, S, B, K, E, J, T, N) and RTD types (Pt100, Cu50, Pt1000) are available.

  • CJ_M: Manual set value for cold junction temperature when TC measure or output.
  • CJ_A: Automatic set value for cold junction temperature when TC measure or output.

6.7 Modbus Function (LB06 specific, not available on LB02A)

Note: The LB02A model does not support Modbus functions. This feature is available on the LB06 model.

6.8 Hart Handheld Function (LB06 specific, not available on LB02A)

Note: The LB02A model does not support Hart functions. This feature is available on the LB06 model.

6.9 Transmitter Calibration (4mA and 20mA)

This function is used for calibrating transmitters. This operation should not be performed while the transmitter is in active use, as it requires resetting the transmitter to its initial state.

  • 4mA_calibrate (0-point calibration): Press 'Start' (or equivalent key) to send a 0-point calibration command. The transmitter will be forced to output 4mA. If the displayed External_Current is not 4mA, input the correct value into the 4mA_calibrate digit box, press 'Enter', and wait until External_Current is equal to or very close to 4mA.
  • 20mA_calibrate (Full-scale calibration): Press 'Start' (or equivalent key) to send a full-scale calibration command. The transmitter will be forced to output 20mA. If the displayed External_Current is not 20mA, input the correct value into the 20mA_calibrate digit box, press 'Enter', and wait until External_Current is equal to or very close to 20mA.

Range Setting Parameters:

  • Set_to_Pv_L_Range: Sets the current measurement value as the lower limit of the user range.
  • Set_to_Pv_H_Range: Sets the current measurement value as the upper limit of the user range.
  • Set_to_Pv_Zero: Sets the current measurement value as the 0-point of the user range.
  • Set_Square_Root: Indicates whether the signal is square root (1 for ON, 0 for OFF).

7. Maintenance

7.1 Fuse Replacement

The instrument uses a 5x20mm 500mA (216 series) fast-acting fuse. If a fuse needs replacement, it is crucial to use a fuse of the same type and rating (250V/500mA) from a reputable brand like Littelfuse to ensure proper protection and functionality.

For detailed instructions on accessing and replacing the fuse, please refer to the physical user manual or contact customer support.

8. Troubleshooting

  • Automatic Shutdown: The calibrator is designed to automatically shut down after approximately 30 minutes of inactivity (no key presses) to conserve battery. This feature is disabled when the unit is charging. If you need to keep the unit on for extended periods without interaction, connect it to the charger.
  • Accuracy Issues: If you experience unexpected accuracy issues, especially with temperature measurements, consider using an external Pt100 temperature module. Ensure sufficient current is supplied for resistance output to avoid errors.
  • Frequency Output Range: Note that the frequency output range starts from 100Hz, not 0Hz.
  • "VIP Note" Discrepancy: The product specifications mention "No MODBUS Hart functions" under "VIP Note", which contradicts other descriptions. Please rely on the detailed function tables (Section 4) and the Modbus/Hart sections (Section 6.7, 6.8) for accurate information regarding your specific LB02A model's capabilities.

9. User Tips

  • This calibrator is an excellent tool for simulating various signals, including thermocouples, thermistors, 0-10V, and 4-20mA.
  • The device generally offers good accuracy, with parameters corresponding to manufacturer data when tested with laboratory equipment.
  • Be aware that precision might decrease at very small mV values, and some units may exhibit a slight offset across the range. Consider external calibration if extreme precision is required at low values.
  • The 24V output can power sensors when measuring 4-20mA signals, providing a convenient solution for loop testing.

10. Frequently Asked Questions (FAQ)

Q: What is this device for?
A: The BRIGHTWIN LB02A is a process calibrator used for both measuring and generating various electrical signals (Voltage, Current, Resistance, Frequency, Thermocouple, RTD) in industrial and laboratory settings. It's essential for testing, calibrating, and troubleshooting control systems and instrumentation.
Q: Does it have Hart and Modbus Protocol or does it only work in Hart Protocol?
A: The LB02A model, as per its specifications, does NOT support Hart or Modbus functions. These advanced communication protocols are typically available on the LB06 model.
Q: How long does the battery last when working with illumination?
A: The exact battery life with backlight illumination can vary based on usage and settings. The device uses a lithium polymer battery and is designed for several hours of operation. For specific battery life figures, please refer to the full technical specifications or contact customer support.
Q: Can it power a sensor from its own 24V source when measuring 4-20mA?
A: Yes, the LB02A has a 24V output function, which can be used to power a sensor or transmitter while simultaneously measuring its 4-20mA output, facilitating loop testing.

11. Warranty and Support

BRIGHTWIN products are manufactured to high-quality standards. For warranty information, technical support, or service inquiries, please contact your original point of purchase or the BRIGHTWIN customer service department. Please have your product model number (BRT LB02A) and purchase details ready when contacting support.

A PDF version of the user manual is available for download. You can find it at: BRIGHTWIN LB02A User Manual (PDF)

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.