1. Introduction
Thank you for choosing the yieryi Digital Water Quality Meter. This device is designed to accurately measure pH, Total Dissolved Solids (TDS), Electrical Conductivity (EC), and temperature of various water sources. It is an essential tool for maintaining optimal water quality in pools, drinking water, aquariums, hydroponics, and more. Please read this manual thoroughly before use to ensure proper operation and longevity of the device.
2. Safety Information
- Keep the device out of reach of children.
- Do not immerse the entire device in water; only the electrode part is designed for submersion.
- Handle calibration powders and solutions with care. Avoid contact with skin and eyes.
- Ensure batteries are inserted correctly according to polarity markings.
- Do not attempt to disassemble or modify the device, as this will void the warranty.
3. Package Contents
Upon opening your package, please verify that all items are present and in good condition:
- yieryi Digital Water Quality Meter
- pH Buffer Powder (pH 4.00, pH 6.86, pH 9.18)
- Instruction Manual
Note: Batteries (LR44) are typically not included and must be purchased separately for some models. EC calibration solution (e.g., 1413 µS/cm) may also need to be purchased separately.


4. Product Overview
The yieryi Digital Water Quality Meter features a clear digital display and intuitive buttons for easy operation. Depending on the model, it may include functions for pH, TDS, EC, Temperature, ORP, and Salinity.

Key Components:
- LCD Display: Shows measurement readings (pH, TDS, EC, Temperature, etc.) and battery status.
- ON/OFF Button: Powers the device on or off.
- TEMP/CAL Button: Switches between Celsius and Fahrenheit, and initiates calibration mode.
- MODE Button: Cycles through different measurement modes (pH, TDS, EC, etc.).
- HOLD Button: Freezes the current reading on the display.
- Electrode: The sensing probe that is immersed in the water.
- Battery Compartment: Located at the top, for LR44 batteries.
5. Specifications
| Parameter | Range | Accuracy | Resolution |
|---|---|---|---|
| pH | 0.00 - 14.00 pH | ±0.05 pH | 0.01 pH |
| TDS | 0 - 19999 ppm | ±2% F.S. | 1 ppm |
| EC | 0 - 19999 µS/cm | ±2% F.S. | 1 µS/cm |
| Temperature | 0 - 60°C (32 - 140°F) | ±0.5°C | 0.1°C |
| Smart Device | Yes (for BLE models) | ||
| Backlight | Yes | ||
| Battery | 3x 1.5V LR44 button cells (not included) | ||
| Product Calibration | Before delivery | ||
6. Setup
6.1. Battery Installation
- Unscrew the battery compartment cap at the top of the meter.
- Insert three LR44 button cell batteries, ensuring correct polarity (+/-).
- Securely screw the battery compartment cap back on.
6.2. Preparing Calibration Solutions
The meter comes with pH buffer powders. You will need distilled water to prepare the solutions.
- Empty one packet of pH 4.00 buffer powder into a clean, dry container (e.g., a small cup).
- Add 250ml of distilled water to the container and stir until the powder is completely dissolved.
- Repeat this process for the pH 6.86 and pH 9.18 buffer powders, using separate clean containers for each.
- Label each solution clearly.

7. Operating Instructions
7.1. Basic Operation
- Press the ON/OFF button to turn on the meter.
- Press the MODE button to cycle through the different measurement modes (pH, EC, TDS, etc.).
- Immerse the electrode into the sample solution, ensuring the liquid level is below the device's body.
- Wait for the reading to stabilize on the display.
- Press the HOLD button to freeze the current reading. Press again to release.
- After use, rinse the electrode with distilled water and turn off the meter by pressing ON/OFF.


7.2. Temperature Unit Switching
While in any measurement mode, press the TEMP/CAL button to switch between Celsius (°C) and Fahrenheit (°F).
7.3. Bluetooth Connectivity (for Smart Models)
For models with Bluetooth functionality (e.g., BLE-9909, BLE-C600, BLE-9908), you can connect the meter to a smartphone app for data logging and advanced features.
- Download the official yieryi app from your smartphone's app store.
- Ensure Bluetooth is enabled on your smartphone.
- Long press the ON/OFF button on the meter to activate Bluetooth pairing mode.
- Open the app and follow the on-screen instructions to connect to your device. You may need to grant location permissions for the app to function correctly.

8. Calibration
Calibration is crucial for accurate readings. It is recommended to calibrate the meter regularly, especially after prolonged storage or if readings appear inconsistent.
8.1. pH Calibration
- Prepare the pH 4.00, pH 6.86, and pH 9.18 buffer solutions as described in Section 6.2.
- Turn on the meter and switch to pH mode.
- Rinse the electrode with distilled water and blot dry.
- Immerse the electrode into the pH 6.86 buffer solution.
- Press and hold the TEMP/CAL button until the display shows 'CAL' and then the pH value starts flashing.
- Wait for the reading to stabilize and automatically confirm calibration. The display will show 'END' briefly.
- Rinse the electrode with distilled water.
- Repeat steps 4-7 for pH 4.00 solution, then for pH 9.18 solution. Always calibrate in ascending or descending order (e.g., 6.86 -> 4.00 -> 9.18 or vice versa).
8.2. EC/TDS Calibration
EC/TDS calibration typically uses a 1413 µS/cm standard solution (not always included). If your model measures EC/TDS, follow these steps:
- Obtain a standard 1413 µS/cm EC calibration solution.
- Turn on the meter and switch to EC or TDS mode.
- Rinse the electrode with distilled water and blot dry.
- Immerse the electrode into the 1413 µS/cm standard solution.
- Press and hold the TEMP/CAL button until the display shows 'CAL' and then the EC/TDS value starts flashing.
- Wait for the reading to stabilize and automatically confirm calibration. The display will show 'END' briefly.
- Rinse the electrode with distilled water.
9. Maintenance
9.1. Cleaning the Electrode
After each use, always rinse the electrode thoroughly with distilled water to remove any residue. Gently blot dry with a soft tissue. Do not rub the electrode, as this can damage the sensitive sensor.
9.2. Storage
Proper storage is essential for extending the life of the electrode. Never store the electrode dry or in distilled water.
- For short-term storage (days to weeks), keep the electrode cap filled with a few drops of pH 4.00 buffer solution or a dedicated electrode storage solution (e.g., KCl 3M solution).
- For long-term storage (weeks to months), ensure the electrode is stored in a dedicated electrode storage solution.
- Always replace the protective cap securely after use and storage.
9.3. Electrode Replacement
The electrode is a consumable part and may need replacement over time due to wear or damage. If the meter consistently shows unstable readings, fails to calibrate, or gives inaccurate results despite proper cleaning and calibration, the electrode may need to be replaced. Replacement electrodes can be purchased separately.
10. Troubleshooting
- Inaccurate/Unstable Readings:
- Ensure the electrode is clean and free of debris.
- Recalibrate the meter using fresh buffer solutions.
- Check battery level; replace if low.
- Ensure the electrode is properly stored in solution when not in use.
- Calibration Error ('ERR'):
- Verify that the correct buffer solution is being used for calibration.
- Ensure the buffer solutions are fresh and correctly prepared with distilled water.
- The electrode might be damaged or at the end of its lifespan; consider replacement.
- Bluetooth Connectivity Issues:
- Ensure Bluetooth is enabled on your phone and the meter is in pairing mode.
- Grant the app necessary permissions, including location access, as some apps require this for Bluetooth functionality.
- Restart both the meter and your smartphone.
- Display Not Turning On:
- Check battery installation and ensure batteries are new or fully charged.
11. User Tips
- Regular Calibration: Calibrate your meter frequently, especially if you need high accuracy or if the device has been stored for a long time. Some users recommend checking calibration weekly.
- Proper Storage: Always store the electrode in a pH 4.00 buffer solution or a dedicated KCl storage solution. Never store it dry or in distilled water, as this can significantly shorten the electrode's lifespan.
- Submersion Depth: Only immerse the electrode part of the device into the water. Avoid getting the main body wet to prevent internal damage.
- Stable Readings: Allow a few seconds for the readings to stabilize before recording. If readings fluctuate, press the 'HOLD' button to capture the current value.
- Bluetooth Setup: For smart models, ensure your phone's geolocation is enabled and the app has location permissions for successful Bluetooth connection.
- Battery Life: The device uses LR44 batteries. Keep spares on hand, as they are not always included.
12. Warranty and Support
Your yieryi Digital Water Quality Meter is covered by a manufacturer's warranty against defects in materials and workmanship. For specific warranty details, technical support, or to inquire about replacement parts, please contact the seller or refer to the warranty information provided at the time of purchase.