yieryi PH-110

Contents

Yieryi Industrial PH/ORP Controller Meter

Model: PH-110

1. Introduction

The Yieryi Industrial PH/ORP Controller Meter (Model PH-110) is a high-precision digital instrument designed for continuous online monitoring and control of pH and ORP levels in various industrial applications. Its robust design, clear LCD display, and advanced features make it suitable for urban sewage treatment plants, chemical industries, printing and dyeing, papermaking, pharmaceutical, electroplating, and environmental protection sectors.

Yieryi Industrial PH/ORP Controller Meter with probe

Figure 1: Yieryi Industrial PH/ORP Controller Meter with its probe.

2. Safety Information

  • Always ensure the power supply matches the device's requirements (AC 220V ± 10% 50Hz).
  • Avoid exposing the device to extreme temperatures or humidity beyond specified environmental conditions.
  • Handle probes with care to prevent damage.
  • Do not attempt to open or repair the device yourself. Refer to qualified personnel for service.
  • Keep calibration solutions and other chemicals out of reach of children.

3. Product Overview

The PH-110 controller features a plastic case with a waterproof and dustproof design, ensuring durability in industrial environments. It is equipped with silicone buttons tested for 10,000 presses without damage. The LCD liquid crystal display offers a wide viewing angle of up to 120 degrees for clear readability.

Close-up of the controller showing waterproof and dustproof design

Figure 2: Controller close-up highlighting waterproof and dustproof design.

Controller LCD display with 120-degree viewing angle

Figure 3: Controller LCD display with wide viewing angle.

3.1. Main Components

  • Controller Meter: The main unit with LCD display and control buttons.
  • PH/ORP Probe: An electrode for measuring pH or ORP levels. The electrode housing is made of acid-resistant ABS material, suitable for long-term detection in harsh wastewater.
  • Power Cable: For connecting the unit to an AC 220V power source.
Electrode housing made of acid-resistant ABS material

Figure 4: Electrode housing for durability in harsh environments.

4. Specifications

Technical Parameters table for the PH/ORP controller

Figure 5: Technical Parameters.

Technical Parameters
ParameterValue
Measuring RangepH (0 – 14 pH); ORP (-1900 – +1900 mV)
Accuracy± 0.02 pH; ± 1 mV
Resolution0.01 pH; 1 mV
Stability≤ 0.02 pH / 24 hours; ≤ 3 mV / 24 hours
PH Standard Solution4.00 / 6.86 / 9.18
Display Mode128 * 64 dot matrix LCD
Power SupplyAC 220V ± 10% 50Hz
Power Consumption≤ 3W
Temperature Compensation0–100 °C manual/automatic (NTC10K)
Signal Output4-20mA isolation protection output Maximum loop resistance 300Ω
Alarm OutputOne set of high and low limit alarm contacts (3A/250 V AC), normally open contact relay
Environmental Conditions(1) Temperature 0 to 60 °C (2) Humidity ≤ 85% RH
Dimensions96 × 96 × 110mm (height × width × depth)
Opening Size92 × 92mm (height × width)

5. Setup

5.1. Physical Installation

The controller is designed for panel mounting. Ensure an opening of 92 × 92mm (height × width) is prepared for installation. The overall dimensions of the unit are 96 × 96 × 110mm.

Dimensions and labeled parts of the controller and probe

Figure 6: Controller and probe dimensions with labeled display values and buttons.

  • Main measured value display: Shows the primary pH or ORP reading.
  • Temperature display value: Shows the current temperature.
  • Current transmission value: Displays the 4-20mA output.
  • Shift key (right shift key): Used for navigating menu options.
  • Enter (OK) key: Confirms selections.
  • Return (ESC) key: Exits current menu or cancels an action.
  • High/Low buttons: Used for adjusting values or navigating.

5.2. Electrical Connections

Connect the device to an AC 220V ± 10% 50Hz power supply. Refer to the wiring diagram on the unit or in the detailed product images for correct connections of the power, probe, signal output (4-20mA), and alarm output terminals.

5.3. Probe Connection

Connect the pH or ORP probe to the designated input terminals on the controller. Ensure a secure connection. Note that different probes (e.g., BNC pH probe, ORP probe, pH probe ATC, RS-485 PH/ORP Meter, PH Meter Glass) may have different connectors. Verify your probe type matches the controller's input.

6. Operating Instructions

6.1. Powering On

Once all connections are secure, switch on the power supply. The device will perform a self-test and then display the current pH/ORP reading, temperature, and current transmission value.

6.2. Menu Navigation

Use the 'ESC', 'OK', 'HIGH' (up arrow), and 'LOW' (down arrow) buttons to navigate through the settings menu. The video demonstrates accessing the setting menu, which includes options for measurement parameters, control parameters, communication settings, backlight settings, alarm voice, factory parameters, and electrode calibration.

6.3. Electrode Calibration

Calibration is crucial for accurate measurements. The device supports calibration using standard pH solutions (4.00, 6.86, 9.18). The video demonstrates a 3-point calibration process:

  1. Access the 'Electrode calibration' option from the setting menu.
  2. The display will prompt to 'Insert the first standard solution'. Place the probe into the first standard solution (e.g., pH 6.86). Press 'OK'.
  3. Once the reading stabilizes, the display will prompt to 'Insert the second standard solution'. Rinse the probe with distilled water, then place it into the second standard solution (e.g., pH 4.00). Press 'OK'.
  4. After stabilization, the display will prompt to 'Insert the third standard solution'. Rinse the probe, then place it into the third standard solution (e.g., pH 9.18). Press 'OK'.
  5. The device will complete the calibration process.

Video 1: pH/ORP Meter Calibration Demonstration.

6.4. Measurement

After successful calibration, the device is ready for measurement. Immerse the probe into the sample solution. The LCD will display the real-time pH or ORP value, temperature, and 4-20mA output.

6.5. Alarm Settings

The device includes a high and low limit alarm function. These can be configured in the settings menu to trigger an alarm (normally open contact relay) when measured values exceed or fall below the set thresholds.

6.6. Signal Output

The 4-20mA isolation protection output provides a current signal proportional to the measured value, allowing integration with other control systems. The maximum loop resistance for this output is 300Ω.

7. Maintenance

7.1. Probe Care

  • Cleaning: Regularly clean the probe with distilled water to remove any residue that might affect accuracy. For stubborn deposits, use a mild cleaning solution recommended for pH/ORP electrodes.
  • Storage: When not in use, store the probe in a suitable storage solution (e.g., KCL solution) to keep the glass bulb hydrated. Do not store dry, as this can damage the electrode and lead to unstable readings.
  • Replacement: Probes have a limited lifespan and may need replacement if they become slow to respond or cannot be calibrated accurately.

7.2. General Unit Cleaning

Wipe the controller unit with a soft, damp cloth. Do not use abrasive cleaners or solvents. Ensure the unit remains waterproof and dustproof by keeping all covers and seals intact.

8. Troubleshooting

Common Issues and Solutions
ProblemPossible CauseSolution
Unstable pH/ORP readingsDirty or damaged probe; dry probe; incorrect calibration.Clean the probe; re-calibrate the device; ensure probe is stored in solution when not in use; replace probe if damaged.
Inaccurate readings after calibrationExpired or contaminated calibration solutions; improper calibration procedure.Use fresh calibration solutions; repeat calibration carefully following instructions.
No display/powerNo power supply; faulty power cable; internal fault.Check power connection; verify power outlet; contact support if issue persists.
Probe connector mismatchIncorrect probe type ordered or supplied.Verify the connector type (e.g., BNC vs. two-pronged) and ensure it matches your controller. Contact seller for correct probe if necessary.

9. User Tips

  • **Probe Storage:** Always ensure your pH/ORP probe is stored correctly in a hydrating solution to prevent it from drying out, which can lead to unstable and inaccurate readings.
  • **Calibration Solutions:** Use fresh and accurate buffer solutions for calibration. Old or contaminated solutions can lead to incorrect calibration and unreliable measurements.
  • **Temperature Sensor:** Note that a separate temperature sensor is not included with all models. The device uses NTC10K for temperature compensation.
  • **Probe Compatibility:** When ordering replacement probes, double-check the connector type (e.g., BNC, two-pronged) to ensure compatibility with your controller.

10. Warranty and Support

For warranty information, technical support, or service inquiries, please contact the seller or manufacturer directly. Keep your purchase records for reference.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.