1. Safety Information
1.1 General Safety Warnings
- The GL090 Diode Laser Hair Removal Machine emits high-intensity infrared laser beams. Always wear suitable protection glasses for both the operator and patient to protect eyes.
- Do not point the laser directly at the eyes or skin.
- Do not power down the machine directly during operation.
- Familiarize yourself with the machine's structure and this manual before operation.
- This device should only be used by individuals who have received adequate safety and operation training.
- Turn off the cooling function when the machine is in standby mode to prevent probe damage.
- Do not use the machine in environments with flammable anesthetic or oxidizing gases (e.g., N2O, O2). Ensure combustible solutions used for cleaning and disinfection have evaporated before operation.
- Operating the control part, regulating, or using the system in an unauthorized manner can lead to dangerous radiation exposure.
- The 808nm diode laser hair removal equipment emits high-intensity infrared laser radiation harmful to tissues, especially ocular tissue. Ensure the treatment room's doors and windows are covered with high-density shading material. Display a warning sign outside the treatment room.
- Do not repair anything not explicitly described in this instruction book. Maintenance should be performed by the manufacturer or authorized distributor.
- Do not make laser points at any areas not intended for treatment. Ensure the output end points in a safe direction.
- Post "laser warning" or "danger" signs in conspicuous places.
- Post "Authorized Personnel Only" outside the treatment room.
- Remove any reflective materials from the treatment room to prevent laser reflection.
1.2 Electrical Safety
- Ensure the ground connection is properly established. Do not power on the machine if it is not working correctly.
- High voltage is present inside the machine and laser handpiece. Do not open them unless trained and authorized.
- Strong electric current is present during operation. Do not touch electrodes when powered on.
- The machine uses 220V AC single-phase power (2000VA). Ensure the treatment room has appropriate power supply.
- Operators must have basic knowledge of electrical safety and correct machine operation.
1.3 Fire Safety
- Do not pile sundries around the machine.
- Do not place explosive or flammable materials in the laser path.
- Ensure flammable liquids used for cleaning and sterilization evaporate completely before use.
- Keep fire extinction appliances readily available.
1.4 Contraindications
- History of keloid scarring or photosensitivity.
- Any inflammatory skin condition (e.g., eczema, active Herpes Simplex) at the treatment site.
- Waxing, plucking, sugaring, or threading in the previous 6 weeks.
- Allergy to hydroquinone or other bleaching cream.
- Pregnancy, or until periods return and end of breastfeeding.
- Skin cancer or any other cancer.
- Use of Isotretinoin in the previous 6 months.
- Tanned skin (sun exposure or tanning bed use) in the previous 30 days.
- Note: The device is not effective for white hair.
2. Product Overview
The Myoslim 4-in-1 Titanium Ice Laser Hair Removal and Tattoo Removal Device is a versatile beauty machine combining four advanced technologies: Diode Laser, Pico Laser, Radiofrequency (RF), and Intense Pulsed Light (IPL). It is designed for permanent hair removal, effective tattoo removal, and comprehensive skin rejuvenation treatments.
2.1 Key Features
- 4-in-1 Multifunctional: Integrates Diode Laser, Pico Laser, RF, and IPL for diverse aesthetic treatments.
- Permanent Hair Removal: Utilizes 755nm, 808nm, 940nm, and 1064nm wavelengths for painless and effective hair removal across various skin and hair types.
- Tattoo Removal: Q-switch Pico Laser effectively targets and eliminates tattoos of all colors, as well as eyebrow/eyeliner tattoos, lip lines, freckles, moles, and birthmarks.
- Skin Rejuvenation: RF and IPL technologies reduce fine lines, tighten skin, treat acne, remove pigments (age spots, Ota spots), and promote skin whitening.
- Advanced Cooling System: Features TEC air cooling, water cooling, and semiconductor refrigeration for a comfortable and continuous treatment experience.
- User-Friendly Interface: Equipped with a 15.6-inch color LCD touch screen for intuitive operation.
- High Power Output: Whole machine output power ranges from 4000W to 5000W.
- Long Lifespan: Diode laser offers 20 million shots, and Pico laser offers over 60 million shots.
2.2 Package Contents
Upon receiving your device, please verify that all the following components are included:
- User Manual (Instructions)
- Power Line
- Laser Treatment Handle
- Foot Switch
- 1064nm Laser Head
- 532nm Laser Head
- Black Doll Treatment Head
- Funnel
- Plastic Hose
- Operator Protective Glasses
- Treated Subjects Goggles
3. Specifications
3.1 General Technical Parameters
| Parameter | Value |
|---|---|
| Power Supply | ~220V/110V, 50Hz/60Hz (110V±22V) |
| Input Current | 35A |
| Whole Machine Output Power | 4000W-5000W |
| Screen | 15.6 inch color LCD touch screen |
| Material | ABS |
| Use Areas | BODY, Underarm, Bikini, Whole Face |
| Warranty | 1 Year |
| Certification | RoHs, CE |
| Temperature Alarm | 10~60°C adjustable |
| Operation Interface | 16.5 inch |
| Power Type | Electric |
| Number of Lamp Heads | 5 or more pieces |
| High-concerned chemical | None |
| Efficacy Principle of Epilator | Laser/IPL/Radio Frequency |
| Origin | Mainland China |
| Item Type | Epilator |
3.2 Diode Laser Specifications
| Parameter | Value |
|---|---|
| Laser Type | Semiconductor laser 800W |
| Wavelengths | 755nm, 808nm, 940nm, 1064nm |
| Treatment Spot | 12*30mm |
| Pulse Width | 10-400ms (continuous adjustable) |
| Power Density | 1-140J/cm² (continuous adjustable) |
| Pulse Frequency | 1-20Hz |
| Cooling System | Sealed inner circled water cooling system, treatment head temperature -4~4°C |
| Shots Life | 20 million shots |
3.3 Pico Laser Specifications
| Parameter | Value |
|---|---|
| Laser Type | PICO Laser |
| Function | Tattoo removal, pigment removal, skin whitening, carbon peeling |
| Pico Treatment Head | 532nm, 1064nm, 1320nm, No scars 1064nm, No scars 532nm |
| Maximum Energy | 2500J |
| Shots | Above 60,000,000 shots |
| Frequency | 1-10Hz |
| Cooling System | Water cycle + wind + TEC air cooling |
3.4 Working Environment and Conditions
- Ambient Temperature: 5°C ~ 40°C
- Relative Humidity: ≤80%
- Atmospheric Pressure Range: 860hPa ~ 1060hPa
3.5 Shipping and Storage Conditions
- Temperature Range: -20°C ~ 60°C
- Relative Density Range: ≤100%
- Atmospheric Pressure Range: 500hPa ~ 1060hPa
4. Setup
4.1 Installation and Debugging
Installation and debugging must be performed by professional, trained staff.
- Place the device on a stable, appropriate surface.
- Connect the handpiece and insert the power wire into a functional power supply.
- Professional staff will test the machine. Always wear protective eyewear during testing to prevent eye damage.
4.2 Turn On Procedure
- Ensure the handpiece is securely plugged into the fiber optic connector.
- Verify that the remote interlock switch is inserted into the designated "Remote interlock" port and the foot switch is connected to the "Foot Switch" port.
- Plug the three-core power cord into a compatible AC220V power outlet.
- Open the KCD switch on the back of the machine and turn the key from "OFF" to "ON" clockwise. Ensure the emergency stop switch is in the released state.
- Allow a few minutes for the machine to complete its automatic detection procedure. The device will then enter the standby screen, indicating successful startup. If there is a cooling system issue, the machine will not enter standby mode.



5. Operating Instructions
5.1 Host Machine Operation
- The touch screen is the primary interface for operation and display.
- Indicator Lights: Green (POWER) indicates power on. Yellow (READY) indicates the instrument is ready. Blue (WORK) indicates the machine is working, and a red light flashing on the handpiece indicates laser output.
- Emergency Stop Switch: The red button with a yellow circle. Press to immediately disconnect power and laser output. Turn right to reset.
- Footswitch/Button Switch: Used to trigger laser output when the machine is in the READY state.
- KCD Switch: Main power switch located at the back of the machine. Turn to "OFF" and remove the key when not in use to prevent unauthorized operation.
5.2 System Control
Parameter settings can only be adjusted when the machine is in standby mode.
- Setting Energy Density:
Method 1: Tap "Energy" on the touch screen, then use the blue triangle buttons to adjust the energy in increments/decrements of 1 J/cm². Range: 1-160 J/cm².
Method 2: Tap the numerical display area (white part) to bring up a keypad. Enter the desired energy density and press "Enter". Use "ESC" to exit without saving or "CLR" to clear and re-enter. - Turning On/Off the Cooling System:
The cooling system is off by default. Tap the "COOLING" button to activate it (button turns orange) or deactivate it (button turns blue). - Pulse Width Setting:
Method 1: Tap "Pulse width", then use the blue triangle buttons to adjust the pulse width in 1ms steps. Range: 10-1400ms.
Method 2: Tap the numerical display area to use the keypad for direct input. - Setting Frequency:
Method 1: Tap "Rep Rate", then use the blue triangle buttons to adjust the frequency in 1Hz steps. Range: 0.5-10Hz.
Method 2: Tap the numerical display area to use the keypad for direct input.
5.3 Ready State
- The "ST/RY" button controls the machine's state. When "ST" is orange and "RY" is blue, the machine is in standby. When "ST" is blue and "RY" is orange, the machine is in ready state.
- In the ready state, the yellow indicator light is on, and system parameters cannot be adjusted.
- To initiate laser output, ensure the machine is in the ready state, then press the foot pedal or the button switch. The red light on the handpiece will flash, and a buzzer will sound during output.
- Release the foot switch or button to stop the laser output.
- To return to standby from ready state, tap the "ST/RY" button. The yellow light will turn off, and parameters can be adjusted again.
5.4 Expert Reserves
The device allows saving up to four sets of commonly used parameters for quick recall. Access this feature by tapping "USER PROG" in standby mode. Parameters can be modified via the digit keypad and saved using "SAVE".
5.5 Intelligent Pulse Counter
The system tracks the number of light pulses. The "Count" tab shows the current cumulative count, and "Total Count" records the total optical pulses. The current pulse count resets on reboot or by tapping the "RESET" button.
5.6 Treatment Method
5.6.1 Treatment Preparation
- For individuals with darker skin, protect the skin from sun exposure (using appropriate clothing and sunblock) for 4-6 weeks prior to the first treatment.
- Ensure the treatment area is clean and dry. Shave hair closely before treatment.
5.6.2 Anesthesia
Local anesthesia is optional and depends on the patient's sensitivity and the treatment area.
5.6.3 Treatment Procedure
- Both the operator and patient must wear protective glasses.
- Apply a thin layer of gel to the treatment site to reduce skin temperature and provide lubrication.
- Set the treatment parameters (energy, pulse width, frequency) according to the desired treatment. Always perform a test treatment first.
- During treatment, press the handpiece firmly against the skin to ensure good contact. Move the handpiece to the next spot after a brief pause (0.25-0.5s).
- Ensure appropriate overlapping (under 10%) to avoid skin damage.
- Regularly clean the sapphire tip of the handpiece to maintain optimal performance and prevent skin damage.
- Treatment intervals vary by body area due to different hair growth cycles: head (12-16 weeks, 30-day interval), body/limbs (12-24 weeks, 60-day interval). The best time for treatment is when hair begins to re-grow.
- After treatment, apply an antimicrobial ointment (e.g., chlorotetracycline eye ointment, erythrocin) or gently wash the treated area to prevent local infection. Use sunblock (SPF>15: PA+++) to avoid direct sun exposure and pigmentation.
5.7 Visual Guides
Video 5.1: Demonstration of the device's functions, parameter adjustments, and laser output on a black card.










6. Maintenance
6.1 Routine Maintenance
- Power Fuse Tube: To check or replace the 10A/250VAC fuse, turn off all power, unplug the cord, use a small flat screwdriver to open the fuse light jacket (counterclockwise), remove the old fuse, and insert a new one. Reconnect power and check operation.
- Mainframe: Keep the machine in a clean, dry, dust-free environment with an ambient temperature between 4°C and 40°C. If not in use for extended periods, shield the equipment. If temperatures drop below 0°C or for transportation, the cooling water must be drained by professional operators.
- Laser Handles: Avoid dropping, hitting, or unauthorized opening of the laser handles. The optical crystal face should be checked and cleaned before each use. Avoid extreme bending of the handle's transmission system.
- Cooling System: Regularly check the cooling fan for proper function. Inspect the water storage tank, cooling capacity, and water quality. Maintenance personnel should add or recharge cooling water as needed.
- Laser Handpiece Cleaning: After each treatment, clean the laser head, especially the sapphire tip, with a soft, cleanser tissue.
7. Troubleshooting
7.1 Malfunction Alarm System
The device features a multi-function protection and alarm system for safety interlock switches, flow switches, water temperature, fiber interlock switch, and current over-current. Indicator lights on the touch screen display the current situation (green for normal, red for failure). If an alarm persists more than three times, contact after-sales support.
- Safety Interlock Switch Alarm: Check if the machine shell is opened, the safety interlock switch is loose, or the remote interlock is open.
- Flow Switches Alarm: Check the water tank for sufficient water level and inspect water pipes/radiators for leaks.
- Water Temperature Alarm: Reduce or stop laser shooting for a while, then restart.
- Fiber Interlock Switch Alarm: Verify that the fibers are correctly connected to the optical output port.
- Current Over-current Alarm: Stop current operation, turn down the machine, and restart.
7.2 Shutdown Procedure
- Tap the shutdown button on the touch screen. A "Shutdown Protection Pattern" will appear. Wait until the message "the machine can be shut safely now" is displayed.
- Rotate the key switch to "OFF".
- Close the hull switch on the rear chassis and disconnect the external power supply.
- Note: A forced shutdown (by pressing the emergency switch or turning off the key switch) will result in the loss of current treatment parameters.
8. User Tips
- For optimal results, consult with a professional or the seller for personalized advice on treatment plans, especially if you are new to using such devices.
- Consistency is key for visible results. For hair removal, aim for 2-3 sessions per week over 4-8 weeks.
- Take advantage of the device's multi-functional capabilities (hair removal, tattoo removal, skin rejuvenation) for comprehensive skin care.
9. Frequently Asked Questions (FAQ)
Q: How long will it take to see results?
A: We recommend using it 2-3 times a week. Improvement should be visible after 4-8 weeks of continuous use.
Q: Is it safe?
A: Yes, it has passed CE and RoHS certifications. It is designed as a Body Sculpting Machine and EMS HIEMT Body Slimming Sculpting Machine.
Q: How do I clean it?
A: After use, please wipe the handle with a damp paper towel.
10. Warranty and Support
10.1 Warranty Information
The device comes with a one-year warranty. During this period, repairs are provided free of charge.
10.2 Technical Support and Repair Centers
Technical advice is available at all times. We offer local repair centers in Mexico and Europe for faster after-sales service, reducing international shipping costs and repair times. Our professional technical team provides online diagnostics and local repair support.
For further assistance, please refer to the User Manual (PDF) available at User Manual (PDF).