1. Product Overview
The OWON HDS25S and HDS25 are all-in-one handheld devices combining a digital oscilloscope, a multimeter, and a waveform generator (HDS25S only). These compact instruments are designed for ease of operation and portability, featuring a 3.5-inch LCD display and a standard USB Type-C interface for charging and PC communication.

Figure 1-1: OWON HDS25S highlighting its 3-in-1 functionality (Oscilloscope, Multimeter, Waveform Generator) and key specifications.
Key Features:
- Oscilloscope: 25MHz bandwidth, 250MSa/s sampling rate, 8K memory depth.
- Multimeter: 4.5-digit multimeter supporting standard voltage, current, resistance, capacitance, diode, and continuity tests.
- Waveform Generator (HDS25S only): 10 MHz output frequency, 125MSa/s sampling rate, 14-bit vertical resolution.
- Interface: Standard USB Type-C for power bank charging and PC communication.
- Display: 3.5-inch LCD display.
2. Safety Information
Before using this product, please read the safety information in advance to prevent personal injury and damage to the product or any connected devices.
Safety Terms and Symbols
Warning: Indicates conditions or practices that could result in injury or loss of life.
Caution: Indicates conditions or practices that could result in damage to this product or other property.
Symbols on the product:
- ⚠ Hazardous Voltage
- ⚠ Refer to Manual
- ⊗ Protective Earth Terminal
- ⊥ Chassis Ground
- ⏚ Test Ground
Safety Requirements
- Only a qualified person should perform internal maintenance.
- Check all Terminal Ratings: To avoid fire or shock hazard, check all ratings and markings.
- Do not operate without covers or panels removed.
- Avoid exposed circuits to prevent electric shock.
- Do not operate if any damage is suspected; have it inspected by qualified service personnel.
- Do not operate in damp conditions or an explosive atmosphere.
- Keep product surfaces clean and dry.
- Using the equipment not in accordance with the manufacturer's specified method may damage the protection provided.
Warning:
- To prevent electric shock or fire, use only the suitable power adapter dedicated to this product and approved for use in your country.
- The two channels of the oscilloscope are non-isolated. Ensure a common reference when measuring. Connecting ground wires of two probes to non-isolated places with different DC electrical levels may cause a short circuit.

Figure 2-1: Schematic diagram of internal ground connection when oscilloscope is connected with computer through the port.
Warning:
- If the input port is connected to a voltage higher than 42V (30Vrms) or a circuit with a peak value of more than 4800 VA, use only voltage probes, test wires, and adapters with proper insulation specified by OWON.
- Before use, check all multimeter test probes, oscilloscope probes, and accessories for mechanical damage. Replace if damaged.
- Remove all unused multimeter test probes, oscilloscope probes, and accessories (power adapter, USB, etc.).
- Firstly, plug the power adapter into the AC socket, and then connect it to the oscilloscope.
- When testing in a CAT II environment, do not connect a voltage higher than 400 V to any input port or a voltage with a voltage difference of more than 400 V to the isolated input port.
- Do not use an input voltage higher than the rated value of the instrument.
- Pay special attention when using 1:1 test wires, as the probe voltage will be directly transmitted to the oscilloscope.
- Do not touch the bare metal BNC or banana plug.
- Do not insert metal objects into the connector.
- Use the oscilloscope only in the specified way.
- Do not input any electric signal, voltage, or current into the GEN Out port of the signal generator.
3. General Inspection
Upon receiving your new oscilloscope, perform the following checks:
- Check for Transportation Damage: Inspect the packaging carton and foam protection. If serious damage is found, retain the packaging until the device and accessories pass electrical and mechanical tests.
- Check Accessories: Refer to the 'List of Accessories' in Appendix A to verify all items are present and undamaged. Contact your distributor if any items are missing or damaged.
- Check the Complete Instrument: If the instrument shows appearance damage, does not work normally, or fails performance tests, contact your distributor. If damage is due to transportation, keep the packaging.
4. Oscilloscope Operation
4.1. Front Panel and Keys

Figure 4-1: Front Panel of the Oscilloscope
- CH1 and CH2 input connectors.
- Waveform generator output connector (optional).
- Display area.
- F1 - F4 keys: Multi-function keys for menu selection.
- HOR key: Adjusts horizontal time base and displacement.
- Return key: Returns to the previous menu or closes the menu.
- Measurement menu key (oscilloscope) or range key (multimeter).
- Zoom or move keys (directional arrows): For waveform movement, time base changes, cursor movements, and trigger level changes.
- Automatic setting key (oscilloscope) or automatic range key (multimeter).
- Stop / run key (oscilloscope) or value hold key (multimeter) or turn on/off signal output (waveform generator).
- Input end of the multimeter.
- Trigger menu key (oscilloscope) or relative value key (multimeter).
- Power switch key.
- Save settings key.
- System settings key.
- Switch key for working state of oscilloscope and multimeter.
- CH1 / CH2 - channel switch key.
4.2. Side Panel

Figure 4-2: Side Panel of the Oscilloscope
- Probe compensation: 3.3V/1kHz square wave signal output.
- Charging or USB communication interface.
- Bracket.
4.3. User Interface Introduction
The oscilloscope display provides various indicators for trigger status, time base, trigger position, sampling rate, and channel information.
Trigger Status Indicators:
- Auto: Automatic mode, waveform collected without triggering.
- Trig: Trigger detected, post-trigger information collected.
- Ready: All pre-trigger data obtained, oscilloscope ready.
- Scan: Scan mode, continuously collects and displays waveform data.
- Stop: Stops collecting waveform data.
4.4. Functional Check
To verify proper operation:
- Press the power switch key (⑂) at the bottom left. The instrument performs self-checks and displays the startup screen.
- Set the oscilloscope probe switch to 10X and connect it to the CH1 channel. Align the probe slot with the BNC connector, insert, and turn to tighten. Connect the probe tip and ground clamp to the probe compensator (square terminal for signal, round for ground).
- Press the "Auto" key on the front panel. Within seconds, a square wave display (1kHz/3.3Vpp) should appear.
- Repeat steps 2 and 3 for the CH2 channel.

Figure 4-3: Automatic Setting displaying a square wave.
4.5. Probe Compensation
Adjust probe compensation to match the probe with the input channel for accurate measurements. Uncompensated probes lead to errors.
- Set the probe menu attenuation coefficient to 10X. Connect the probe tip to the signal output connector of the probe compensator and the reference wire clamp to the ground. Press the Auto key.
- Check the displayed waveform and adjust the probe until compensation is correct.

Figure 4-4: Display Waveform of Probe Compensation (Illustrative).
4.6. Probe Attenuation Coefficient Setting
Probes have various attenuation coefficients affecting the vertical position factor. To change this setting:
- Press the CH1/CH2 key to select the channel.
- Press F3 to select the desired attenuation coefficient.
Note:
- The preset attenuation coefficient is 10X. Ensure the probe switch matches the oscilloscope menu setting.
- When the attenuation switch is set to 1X, the oscilloscope bandwidth is limited to 5 MHz. For full bandwidth, set the switch to 10X.
4.7. Safe Use of Probe
A safety ring on the probe body protects against electric shocks.
Warning:
- Keep fingers behind the safety ring when using the probe.
- Do not touch the metal part of the probe head when connected to a voltage source.
4.8. Vertical System
Adjust vertical scale, position, and other channel settings. Each channel has a separate vertical menu.
Vertical Position:
Press CH1/CH2, then use directional keys to move the selected channel up/down. Press both up and down keys simultaneously to center the vertical position.
Vertical Volt/Div Setting:
Range: 10mV/div-10V/div (1X probe) up to 100V/div-100000V/div (10000X probe). Press CH1/CH2, then directional keys to change the volts/div setting.
| Function Menu | Setting | Description |
|---|---|---|
| Switch | ON / OFF | Turn on/off waveform display. |
| Coupling | DC / AC / GND | Normal sampling (AC/DC components), block DC component, disconnect input. |
| Probe | 1X, 10X, 100X, 1000X, 10000X | Choose based on probe attenuation factor for accurate vertical scale. |
| Bandwidth | Full band | The bandwidth of the oscilloscope. |
4.9. Horizontal System
Press the HOR key to enter the horizontal system setting menu. Use directional keys to change the horizontal scale (time base) and horizontal trigger position. Press both directional keys simultaneously to center the horizontal position.
| Function Menu | Setting | Description |
|---|---|---|
| Acqu Mode | Sample / Peak Detect | Normal sampling or detect interference glitches. |
| Length | 4K / 8K | Select record length. |
| XY Mode | ON / OFF | Turn XY mode on or off. |
| Refresh | High / Low | Set refresh rate. |
| Hor center | - | Set horizontal trigger position to screen center. |
| Counter | ON / OFF | Turn on/off counter. |
4.10. Measuring System
Automatic Measurement:
Press the Measure Range and F1 key to perform automatic measurements. Up to 9 types can be displayed. For bandwidth ≤100MHz models, automatic range types include Frequency, Period, Amplitude, Maximum, Minimum, Peak-to-Peak value, Mean, and RMS.
| Function Menu | Setting | Description |
|---|---|---|
| - | ON / OFF | Turn automatic measurement on or off. |
| Source | CH1 / CH2 | Set the source. |
| Measure | Freq (F), Period (T), Amp (Va), Min (Mi), Max (Ma), PK - PK (Vpp), Mean(V), RMS (RMS) | Add or delete selected measurement types. |
Automatic Measurement of Voltage Parameters:
The oscilloscope provides automatic voltage measurements including Mean, PK-PK, RMS, Max, Min, and Vamp.

Figure 4-5: Pulse waveform with voltage measurement points.
- Mean: Arithmetic mean over the entire waveform.
- PK-PK: Peak-to-Peak Voltage.
- RMS: True Root Mean Square voltage over the entire waveform.
- Max: Maximum amplitude.
- Min: Minimum amplitude.
- Vamp: Voltage between Vtop and Vbase.
Cursor Measurement:
Press Measure Range and F2 key to perform cursor measurement.
| Function Menu | Setting | Description |
|---|---|---|
| Type | CH1 Voltage, CH2 Voltage, Time, None | Selects measurement type or turns off cursor. |
| A | - | Moves cursor line A. |
| B | - | Moves cursor line B. |
| AB | - | Links A and B; moves both cursors simultaneously. |
Display Unknown Signals using Automatic Settings:
The auto-setting function optimizes position, range, timebase, and trigger for stable waveform display. Connect the probe to the signal and press the Auto key.
4.11. Trigger System
The trigger determines when the oscilloscope acquires and displays data. The trigger mode is edge trigger (rising or falling edge of input signal).
| Function Menu | Setting | Description |
|---|---|---|
| Source | CH1 / CH2 | Set Channel 1 or 2 as the source trigger signal. |
| Coupling | AC / DC | Set to prevent DC component or allow all components. |
| Type | Auto, Normal, Single | Acquire without trigger, acquire only when conditions met, or sample and stop. |
| Slope | Rising edge, Falling edge | Trigger on rising or falling edge. |
| Trigger center | - | Set trigger position to screen center. |
| Force | - | Forcibly generates a trigger signal. |
Trigger electrical level: The amplitude level the signal must cross. Press Trig/▲ key to enter the trigger menu, then use directional keys to move the trigger electrical level.
4.12. Save Settings
Press the Save key to store oscilloscope settings, reference waveforms, and files.
Configure:
Save and restore any settings. Use Object S1/S2 to set waveform name.
Reference Waveform:
Compare actual waveforms with stored reference waveforms. Press Save, then F2 to select reference waveform menu. Use Source CH1/CH2 to select waveform to be saved. Use Object R1-R4 to set waveform name. Display ON/OFF to show/hide stored waveforms.
File:
Save waveforms or images. Files can be read via USB data cable (MSC mode) or by selecting MSC in system settings.
| Function Menu | Setting | Description |
|---|---|---|
| Wave | File name (wave1-wave4), Source (CH1/CH2), Save | Select file name, waveform channel, save waveform to CSV. |
| Image | File name (image1-image4), Save | Select file name, save current screen image to BMP. |
4.13. System Settings
Press the System key to enter the system function menu.
Display Settings:
| Function Menu | Setting | Description |
|---|---|---|
| Brightness | 10% - 100% | Adjust screen backlight. |
| Backlight time | 30s, 60s, 120s, Unlimited | Set backlight luminance time. |
| Menu Time | 5s, 10s, 20s, 30s, 60s | Set menu display time. |
| Turned on | 00h: 00m | Display power-on duration. |
System Menu:
| Function Menu | Setting | Description |
|---|---|---|
| Language | - | Set the menu language. |
| Shutdown Time | 10 min, 30 min, 60 min, Unlimited | Set automatic shutdown time. |
| About | - | Display instrument model, serial number, version, and checksum. |
| Upgrade | - | Upgrade system firmware. |
Default Setting:
Press System, then F3 "Default setting". Confirm with F3 again to restore default settings.
USB Connection:
Press System, then F4 to enter the next page. Press F1 to select HID or MSC.
- MSC (Mass Storage Class): For USB to read files stored in built-in memory.
- HID (Human Interface Device): For PC control and communication.
Factory Settings:
Press System, then F4, then F2 "Factory setting". Confirm with F2 again to restore factory settings.
Automatic Calibration:
Disconnect all probes/wires. Press System, then F4, then F3 "Auto calibration". Confirm to execute. This program optimizes the oscilloscope for accurate measurements, especially if ambient temperature changes by ±5°C.
5. Multimeter Operation
This chapter details the multimeter function, including basic operations and menu usage.
5.1. Instrument Interface
The multimeter uses four 4-mm safety banana plug input ends: A, mA, COM, and VΩ→C.

Figure 5-1: Multimeter Interface
- Measurement type indication: DCV (DC voltage), ACV (AC voltage), DCA (DC current), ACA (AC current), Resist (Resistance), Diode, Cont (Continuity), Cap (Capacitance).
- Range indication: Manual (manual range), Auto (automatic range).
- Current measurement range.
- USB cable inserted indicator.
- Battery power indication.
- "Hold" function to keep the current reading.
- Measurement value and unit.
- Display of switching resistance, buzzer, diode, and capacitance measurement function.
- Selected range (V or mV for voltage; A or mA for current).
- AC or DC current measurement selection.
- AC or DC voltage measurement selection.
- Relative value measurement function display (only for DC current, DC voltage, and resistance).
6. Waveform Generator Operation (HDS25S only)
The HDS25S model provides 4 basic waveforms (sine, square, ramp, pulse) and 8 arbitrary waveforms.
6.1. Connect the Output
Press the Mode button to switch to the waveform generator interface. Ensure the screen shows "ON" in the upper left corner; if "OFF", press the Run/Stop key. Connect a BNC cable to the GEN Out port on top of the oscilloscope.
Warning:
BNC output ports are not allowed to input any signal (voltage, current, etc.), otherwise, it will be damaged.

Figure 6-1: Generator Output Ports
To observe the output, connect the other end of the BNC cable to the oscilloscope's signal input connector.
6.2. Set the Waveform
- Press the Mode button to switch to the waveform generator interface.
- Press F1 to select the desired waveform. The corresponding setting menu will appear.
- Set parameters using F2-F4 and directional keys.
6.3. Set the Load
- Press the System key.
- Press F4 to go to the next page.
- Press F3 to switch High Z / Ω (default is 50Ω). To change the load value, select *Ω, use directional keys to move the cursor and change the value. Load range: 1 Ω - 10 KΩ.
6.4. Output the Sine Waveform
Menu includes: Freq/Period, Amplitude/High level, Offset/Low level.
Set Frequency / Period:
Press F1 to enter the sine waveform setting. Use F3 or F4 to switch to Freq/Period. Use directional keys to set the value. Press F2 to switch between Freq / Period.
Set Amplitude / High level:
Use F3 or F4 to switch to Amplitude/High level. Use directional keys to set the value. Press F2 to switch between Amplitude / High level.
Set Offset / Low level:
Use F3 or F4 to switch to Offset/Low level. Use directional keys to set the value. Press F2 to switch between Offset / Low level.
6.5. Output the Square Waveform
Press F1 to enter the square waveform setting. Menu includes: Freq/Period, Amplitude/High level, Offset/Low level. Settings are similar to sine waveform.
6.6. Output the Ramp Waveform
Press F1 to enter the ramp waveform setting. Menu includes: Freq/Period, Amplitude/High Level, Offset/Low Level, Symmetry. Settings are similar to sine waveform. To set symmetry, use F3 or F4 to switch to Symmetry, then use directional keys.
6.7. Output the Pulse Waveform
Press F1 to enter the pulse waveform setting. Menu includes: Freq/Period, Amplitude/High level, Offset/Low level, Pulse width/Duty, Rise Time/Fall Time. Settings are similar to sine waveform. To set Pulse Width / Duty Cycle or Rise Time/Fall Time, use F3 or F4, then directional keys, and F2 to switch.
6.8. Output the Arbitrary Waveform
Press F1 to enter the arbitrary waveform setting. Menu includes: Freq/Period, Amplitude/High Level, Offset/Low Level, Type. Settings are similar to sine waveform.
Type (built-in waveform):
The system has 8 built-in waveforms. To select, use F3 or F4 to switch to Type, then F2 or directional keys to select the desired waveform.
| Name | Explanation |
|---|---|
| AmpALT | Attenuation oscillation curve |
| AttALT | Gain oscillation curve |
| StairDn | Stair-down waveform |
| StairUD | Stair-up and stair-down waveform |
| StairUp | Stair-up waveform |
| Besselj | Besselj function |
| Bessely | Bessely function |
| Sinc | Sinc function |
7. Communication with PC
The oscilloscope supports communication with a PC via USB for data storage, analysis, display, and remote control. Download the oscilloscope communication software from the official OWON website.
Connection Steps:
- Connection: Use a USB data cable to connect the USB Device port on the oscilloscope's right panel to a PC's USB port.
- USB Port Settings: On the oscilloscope, go to System → F4 → USB, and switch to HID (Human Interface Device).
- Software: Run the Oscilloscope software on your PC. The connection information in the bottom right corner of the software should turn green.

Figure 7-1: Oscilloscope connected with PC through USB port.
8. Troubleshooting
- Oscilloscope cannot be turned on: If the battery is exhausted, charge for 0.5-1 hour before attempting to turn on. If it still doesn't turn on, contact support.
- Oscilloscope turns off after a few seconds of startup: Battery may be exhausted. Check the battery symbol and charge.
- Multimeter displays 'E' after switching: Measurement type may not be selected. Press F4 to select. If 'E' persists, restart.
- Measured voltage amplitude is 10 times larger or smaller than actual: Check if the channel attenuation coefficient matches the actual probe ratio.
- Waveform displays but is unstable:
- Check if the source in the trigger mode menu matches the signal channel.
- Ensure the trigger electrical level is within the waveform range. Adjust or press Auto.
- Nothing displayed after pressing RUN/STOP: Check if trigger mode is normal or single, and if trigger electrical level is out of range. Center the trigger level or set to automatic. Press Auto to auto-set.
- Display speed becomes slower with average value sampling or longer duration: This is normal behavior.
9. Maintenance and Cleaning
General Maintenance
Do not store the instrument where the LCD screen will be exposed to direct sunlight for long periods.
Caution:
Do not let spray, liquid, or solvent touch the instrument or probe to prevent damage.
Cleaning:
Check the instrument and probe frequently. Clean external surfaces as follows:
- Wipe floating dust with a soft cloth. Be careful not to scratch the LCD protection screen.
- Wipe with a damp, non-dripping soft cloth. Disconnect power supply. Use soft detergent or water. Do not use abrasive chemical cleaning agents.
Warning:
Ensure the instrument is dry before re-energizing to avoid electrical short circuits or personal injury.
10. Battery Charging and Replacement
Battery Charging
If the device has been stored for a long time, the battery may be too low. Use the attached adapter to pre-charge for 0.5 to 1 hour before turning it on. Charge regularly to prevent over-discharge.
| Model | Charging Time | Discharging Time |
|---|---|---|
| For models < 100MHz | ≥4.5h | ≥4h |
| For models ≥ 100MHz | ≥4h | ≥3h |
Power supply and battery indicators:
- □ Power-on charging status.
- □ Battery power supply.
- □ Approximately five minutes of use time left. Charge immediately.
Charging Method
- Via Power Adapter: Connect the oscilloscope to a power socket using the USB data cable and power adapter.
- Via USB Interface: Connect the oscilloscope to a computer or other equipment via a USB data cable. Ensure the power supply equipment has sufficient load capacity.
Note:
To avoid battery overheating, ensure the ambient temperature does not exceed the allowable value in the technical specifications during charging.
Replacement of Lithium Battery
The battery generally does not need replacement. If necessary, it must be replaced by qualified personnel using only lithium batteries of the same specification.
11. Technical Specifications
All technical specifications apply with the probe attenuation switch set to 10X, and after the instrument has operated continuously for more than 30 minutes at the specified operating temperature. If the operating temperature changes by ±5°C, perform automatic calibration.
Oscilloscope Specifications
| Characteristics | Description |
|---|---|
| Bandwidth | 25 MHz |
| Vertical resolution | 8 bits |
| Channel | 2 |
| Sampling method | Sampling, peak detection |
| Real-time sampling rate | 125 MSa/s (Dual channel), 250 MSa/s (Single channel) |
| Waveform refresh rate | 10,000 wfms/s |
| Input coupling | DC, AC, ground |
| Input impedance (DC coupling) | 1 MΩ±2%, in parallel with 16 pF±10 pF |
| Probe attenuation | 1X, 10X, 100X, 1000X, 10000X |
| Maximum input voltage | 400 V (DC + AC, PK - PK) |
| Bandwidth limit | Full bandwidth |
| Sampling rate range | 0.25 Sa/s~250 MSa/s |
| Waveform interpolation | Sinx/x |
| Sweep speed range (s/div) | 5ns/div - 1000s/div, step by 1-2-5 way |
| Time base accuracy | ±100 ppm |
| Record length | 8K or 4K optional |
| Sensitivity (Volt/div) range | 10 mV/div~10 V/div |
| Displacement range | ±6 div |
| Analog bandwidth | 25MHz |
| Single bandwidth | Full bandwidth |
| Low frequency response (AC coupling, -3dB) | ≥10 Hz |
| Rise time (typical on BNC) | ≤ 14 ns |
| DC gain accuracy | 3% |
| Cursor | ∆V, ∆T |
| Automatic Measurement | Period, Frequency, Mean, PK-PK, Max, Min, Amplitude |
| Trigger Source | CH1, CH2 |
| Trigger Coupling | DC, AC |
| Trigger type | Auto, normal, single |
| Trigger electrical level range | ±4 div from the center of the screen |
| Trigger electrical level accuracy | ±0.3 div |
| Trigger displacement | According to Record length and time base |
| Edge triggering | Rising edge, falling edge |
Probe Compensator Output
| Characteristics | Description |
|---|---|
| Output voltage | 3.3Vpp, High-Z |
| Frequency (typical) | Square wave 1 kHz (±1%) |
Digital Multimeter Specifications
| Characteristics | Description |
|---|---|
| Digital display | 20,000 readings |
| Measurement type | Voltage, current, resistance, capacitance, continuity test, diode |
| Maximum Input voltage | AC: 750V DC: 1000V |
| Maximum Input current | AC: 10A DC: 10A |
| Basic function | Range | Minimum resolution | Accuracy |
|---|---|---|---|
| DC voltage | 200.00mV, 2.0000V, 20.000V, 200.00V, 1000.0V | 0.01mV, 0.1mV, 1mV, 0.01V, 0.1V | ±(0.3%+10dig) for mV, ±(0.3%+5dig) for V |
| AC voltage | 200.00mV, 2.0000V, 20.000V, 200.00V, 750.0V | 0.01mV, 0.1mV, 1mV, 0.01V, 0.1V | ±(0.8%+10dig) for mV/V, ±(1%+10dig) for 750V |
| DC current | 200.00mA, 10.000A | 0.01mA, 1mA | ±(0.8%+10dig) for mA, ±(2.5%+10dig) for A |
| AC current | 200.00mA, 10.000A | 0.01mA, 1mA | ±(1%+10dig) for mA, ±(2.8%+10dig) for A |
| Resistance | 200.00Ω, 2.0000KΩ, 2.0000MΩ, 20.000MΩ, 100.00MΩ | 0.01Ω, 0.1Ω, 0.1KΩ, 1KΩ, 0.01MΩ | ±(0.8%+10dig) for Ω/KΩ, ±(1%+3dig) for MΩ, ±(5%+10dig) for 100MΩ |
| Capacitance | 20.000nF, 200.00nF, 2.0000µF, 20.000µF, 200.00µF, 2.0000mF | 1pF, 10pF, 0.1nF, 1nF, 10nF, 0.1µF | ±(3.0%+10dig) |
| Others | Continuity test, Diode test, Auto range, TRMS | √ (<50Ω), √ (<0-2V), √, √ | - |
Note: When measuring AC voltage/current or capacitance, accuracy guarantee range is 5% to 100% of the range.
Arbitrary Waveform Generator Specifications (HDS25S only)
| Characteristics | Description |
|---|---|
| Waveform Frequency | Sine: 0.1Hz~10MHz, Square: 0.1Hz~2MHz, Ramp: 0.1Hz~1MHz, Pulse: 0.1Hz~5MHz, EXP: 0.1Hz~5MHz |
| Sampling | 125 MSa/s |
| Amplitude (50Ω) | 0.01Vpp~2.5Vpp |
| DC offset (High Z) | ±(2.5V - Amplitude Vpp/2) |
| Frequency Resolution | 0.01% |
| Channel | 1 |
| Waveform Depth | 8k |
| Vertical Resolution | 14 bits |
| Output Impedance | 50 Ω |
General Technical Specifications
| Characteristics | Description |
|---|---|
| Display type | 3.5-inch color LCD display |
| Display resolution | 320 horizontal × 240 vertical pixels |
| Display color | 65536 colors |
| Display Contrast | Adjustable |
| Power supply | 100-240 VACRMS, 50/60 Hz, CAT II; DC INPUT: 5VDC, 2A |
| Power consumption | <5 W |
| Battery | 2200mAh*2 (3.7V, 18650) |
| Working temperature | 0°C ~ 40°C |
| Storage temperature | -20°C ~ +60°C |
| Relative humidity | ≤90% |
| Operating Height | 3,000 meters |
| Non-operating Height | 15,000 meters |
| Cooling method | Natural cooling |
| Dimensions | 198 mm (length) × 96mm (height) × 38 mm (width) |
| Weight | Approx. 0.6 kg (instrument, without battery) |
| Calibration interval | One year (recommended) |
12. Accessories

Figure 12-1: Product Accessories
Appendix A: List of Accessories
- 1 USB cable
- 1 passive probe
- 1 crocodile clip cable
- 1 set of multimeter probes (one red and one black)
- 1 user manual
- 1 probe correction adjustment knife
13. User Tips
Based on common user feedback and questions, here are some tips for optimal use:
- Battery Life: For extended battery life, especially during long-term storage, ensure the device is charged periodically to prevent deep discharge.
- Probe Connection: Always ensure probes are securely connected and compensated correctly to avoid inaccurate readings.
- PC Communication: If experiencing issues with PC communication, double-check that the oscilloscope's USB mode is set to HID in the system settings.
- Troubleshooting 'E' on Multimeter: If the multimeter displays 'E', it often means the measurement type hasn't been selected. Press F4 to cycle through options.
- Automatic Calibration: Perform automatic calibration if the ambient temperature changes significantly (by 5°C or more) to maintain measurement accuracy.
14. Warranty and Support
General Warranty
The product is warranted to be free from defects in materials and workmanship for a period of 3 years from the date of purchase by the original purchaser. Accessories (probes, adapter) have a 12-month warranty. This warranty is non-transferable.
If the product is defective during the warranty period, OWON will repair or replace it. Replaced parts become the property of OWON.
This warranty does not cover defects or damage caused by improper use, inadequate maintenance, unauthorized repairs, connection to incompatible equipment, or use of non-OWON supplies. It also does not cover products modified or integrated with other products if such modification increases servicing difficulty.
Except for the after-sales services provided, OWON offers no other guarantees, express or implied, including merchantability or fitness for a particular purpose. OWON is not responsible for indirect, special, or consequential damages.
Product Support
For product support, visit: www.owon.com.hk/download
Fujian LILLIPUT Optoelectronics Technology Co., Ltd.
No. 19, Heming Road
Lantian Industrial Zone, Zhangzhou 363005 P.R. China
Tel: +86-596-2130430
Fax: +86-596-2109272
Web: www.owon.com.cn
E-mail: info@owon.com.cn
User manual (PDF) available for download: OWON HDS20 Dual Channel Series Handheld Oscilloscope User Manual