MESTEK IR05/IR05A/IR05B

Contents

Mestek IR05 Series Infrared Thermometer User Manual

Models: IR05, IR05A, IR05B

1. Introduction

The Mestek IR05 Series Infrared Thermometer is a high-precision, non-contact temperature measurement device designed for industrial use. It features dual laser positioning, adjustable emissivity, and the ability to measure a wide range of temperatures, including high temperatures up to 2000°C (model dependent). This manual provides essential information for safe and effective operation of your thermometer.

2. Safety Instructions

  • Warning: Do not aim the laser directly at human eyes or reflective surfaces. The laser is Level 2, <1mW, 620-690nm.
  • This device cannot measure through transparent surfaces like glass or plastic.
  • Avoid direct contact with smoke or steam when measuring boiling liquids. Maintain a measurement angle of approximately 45° to 60° in the temperature sensing area.
  • If the ambient temperature changes significantly or suddenly, allow the thermometer to acclimatize for at least 30 minutes before use to ensure accurate readings.
  • Ensure proper battery installation and replacement as described in the manual.

3. Product Overview

The Mestek IR05 Series Infrared Thermometer is equipped with industrial-grade components for reliable performance.

Diagram showing internal components of the infrared thermometer, including high precision laser indication, high precision sensors, aluminum alloy sensing housing, and a high speed intelligent chip.
Figure 3.1: Internal Components

The device features a clear LCD display with various indicators and functions:

Diagram of the thermometer's LCD panel with numbered labels pointing to different functions: 1. Emissivity, 2. Measurement indication, 3. Laser indication, 4. Battery undervoltage indication, 5. Data locking, 6. Irradiation measurement temperature value, 7. Temperature unit, 8. Maximum lock, 9. Minimum lock, 10. Average, 11. Temperature difference value, 12. Alarm, 13. Hi high alarm, 14. Low alarm, 15. K-type thermocouple temperature.
Figure 3.2: LCD Panel Functions
  1. Emissivity
  2. Measurement indication
  3. Laser indication
  4. Battery undervoltage indication
  5. Data locking
  6. Irradiation measurement temperature value
  7. Temperature unit
  8. Maximum lock
  9. Minimum lock
  10. Average
  11. Temperature difference value
  12. Alarm
  13. Hi high alarm
  14. Low alarm
  15. K-type thermocouple temperature

4. Setup

4.1 Battery Installation

The thermometer requires 2x 1.5V AA batteries (not included).

  1. Open the battery compartment located at the handle of the device.
  2. Install two No. 5 (AA) batteries, ensuring correct polarity.
  3. Close the battery compartment securely.
Image showing the quick measurement steps: 1. Opening the battery compartment and installing 2 AA batteries. 2. Pulling the trigger to start the machine. 3. Starting measurement.
Figure 4.1: Battery Installation and Initial Use

5. Operating Instructions

5.1 Basic Temperature Measurement

  1. Point the thermometer at the target object.
  2. Pull and hold the trigger to activate the laser and begin measurement. The temperature will be displayed on the LCD.
  3. Release the trigger to hold the measurement on the display.
Image showing a hand holding the infrared thermometer, pointing it at a welding area, with the display showing a temperature reading. Text indicates 'One click temperature measurement' and 'Quick response', 'No-touch', 'Safety measurement'.
Figure 5.1: One-Click Temperature Measurement

5.2 Unit Conversion (°C/°F)

To switch between Celsius (°C) and Fahrenheit (°F):

  • Long press the 'SET' button until the C/F unit flashes.
  • Use the Up/Down arrow buttons to select your desired unit.
  • Short press the trigger to exit setup mode.
Image showing the thermometer display with 'C/F switching' text, indicating that two temperature measurement modes are available for real-time demand with one-click switch.
Figure 5.2: Celsius/Fahrenheit Switching

5.3 Advanced Measurement Modes (MIN/MAX/AVG/DIFF)

The thermometer can display maximum, minimum, average, and temperature difference values. These modes can be toggled by pressing the MIN/MAX button.

  • While measuring (trigger held), press the MIN/MAX button to cycle through Max, Min, Avg, and Diff readings.
  • The selected mode will be indicated on the LCD.
Image showing the thermometer display with 'Measurement value mode' text, explaining that the default maximum value can be toggled between minimum, average, and temperature difference values by lightly pressing the MIN/MAX button.
Figure 5.3: Measurement Value Modes

5.4 Emissivity Adjustment

Emissivity (ε) is crucial for accurate infrared temperature measurement. Different materials have different emissivities. To adjust:

  1. Long press the SET button until the emissivity value (ε) flashes.
  2. Use the Up/Down arrow buttons to adjust the emissivity value (range 0.10-1.00).
  3. Short press the trigger to exit setup mode.

5.5 High/Low Temperature Alarm Setting

You can set high and low temperature alarms:

  1. Long press the SET button to enter the setup menu.
  2. Press SET repeatedly until 'Alarm Hi' flashes. Use Up/Down arrows to set the high alarm threshold.
  3. Press SET again until 'Alarm Low' flashes. Use Up/Down arrows to set the low alarm threshold.
  4. Short press the trigger to exit setup mode.
Four images demonstrating thermometer parameter settings: 1. Long pressing SET to enter high temperature alarm setting. 2. Gently pressing SET again to enter low temperature alarm setting. 3. Pressing SET again to enter emissivity setting. 4. Pressing SET again to enter C/F switching mode. Short press trigger to exit.
Figure 5.4: Thermometer Parameter Settings

5.6 K-type Thermocouple Probe Usage

For contact temperature measurement, connect the K-type thermocouple probe to the designated port on the thermometer. The K-type thermocouple temperature will be displayed on the LCD (item 15 in Figure 3.2).

Image showing the infrared thermometer with a K-type probe connected, highlighting its capability for high temperature measurement.
Figure 5.5: K-type Probe in Use

5.7 Distance to Spot Ratio (D:S)

The D:S ratio for this thermometer is 50:1. This means that at a distance of 50 units from the target, the measurement spot diameter will be 1 unit. For accurate readings, ensure the target is larger than the spot size. The further you are from the target, the larger the measurement area.

Diagram illustrating the 50:1 Distance to Spot ratio, showing how the measurement spot diameter increases with distance from the target.
Figure 5.6: Distance to Spot Ratio (D:S)

5.8 Water/Oil Temperature Measurement Method

When measuring the temperature of boiling liquids, it is important to avoid direct contact with steam or smoke, as this can affect accuracy. Position the thermometer at an angle of approximately 45° to 60° to the surface of the liquid, ensuring the temperature sensing area is not directly exposed to steam.

Image demonstrating the correct and incorrect methods for measuring water/oil temperature. The incorrect method shows the thermometer directly over steam, while the correct method shows it angled to avoid steam contact.
Figure 5.7: Water/Oil Temperature Measurement

5.9 Sleep Mode

To conserve battery life, the thermometer will automatically enter a sleep state after 35 seconds of inactivity.

5.10 Video Demonstration

Watch this video for a quick overview of the IR05 Series Infrared Thermometer's features and basic operation.

Video 5.1: IR05 Series Infrared Thermometer Overview

6. Maintenance

6.1 Battery Replacement

When the battery undervoltage indicator appears on the LCD, replace the batteries as described in Section 4.1. Always use new 1.5V AA batteries.

6.2 Cleaning

  • Clean the lens with a soft, damp cloth. Do not use abrasive cleaners or solvents.
  • Wipe the thermometer's body with a damp cloth.
  • Ensure the device is dry before storage.

6.3 Storage

Store the thermometer in a dry environment, away from extreme temperatures and direct sunlight. The recommended storage temperature is -20°C to 60°C.

7. Troubleshooting

  • No Display: Check battery installation and ensure batteries are not depleted. Replace if necessary.
  • Inaccurate Readings:
    • Verify emissivity setting matches the target material.
    • Ensure the D:S ratio is appropriate for your measurement distance and target size.
    • Allow the thermometer to acclimatize to ambient temperature if it has been moved from a significantly different temperature environment.
    • Avoid steam or smoke interference when measuring liquids.
  • Laser Not Working: Short press the laser control button (part of the Up/Down arrow cluster) to toggle the laser on/off.

8. Specifications

The following table details the specifications for the IR05, IR05A, and IR05B models:

Table comparing specifications for models IR05, IR05A, and IR05B. Details include Temperature range, Accuracy, K-type range, K-type accuracy, Display, D:S, Spectral response, Laser, Response time, Automatic shutdown, Usage environment, Storage temperature, Size, Source, and Weight.
Figure 8.1: Product Specifications Comparison
Feature IR05 IR05A IR05B
Temperature range -40°C~1200°C (-40°F~2192°F) -40°C~1600°C (-40°F~2912°F) -40°C~2000°C (-40°F~3632°F)
Accuracy -40°C~800°C (±1.5%), 800°C~2000°C (±3%)
K-type range -10°C~500°C (14°F~932°F)
K-type accuracy ±(1.5%+2°C/4°F)
Display VA reverse display color screen
D:S 50:1
Spectral response 8~14um
Laser <1mW /620~690nm Level 2
Response time <0.5S
Automatic shutdown About 35 seconds
Usage environment 0~50°C / 10%~90% RH
Storage temperature -20°C~60°C
Size 180x120x60mm
Source 2x1.5V AA battery
Weight Approximately 265g (excluding battery)

9. User Tips

  • For best accuracy, always ensure the target object fills the measurement spot indicated by the laser.
  • Remember to adjust the emissivity setting for different materials (e.g., shiny metals have lower emissivity than painted surfaces).
  • The K-type probe is ideal for precise contact measurements, especially for liquids or surfaces where non-contact measurement might be affected by environmental factors.
  • Keep the lens clean to prevent inaccurate readings.

10. Warranty and Support

For warranty information, technical support, or service inquiries, please contact the seller or refer to the official product page. Keep your purchase receipt as proof of purchase.

A PDF version of the user manual is available for download. Please refer to the product listing for the link to the PDF document.