1. Introduction
The VAKIND BLE-YC01 is a versatile 6-in-1 Smart Floating Water Quality Tester designed for comprehensive analysis of various water sources. This device accurately measures pH, Electrical Conductivity (EC), Total Dissolved Solids (TDS), Oxidation-Reduction Potential (ORP), Chlorine (CL), and Temperature. With Bluetooth connectivity, it allows for remote monitoring and data management via a dedicated mobile application.
Figure 1: VAKIND BLE-YC01 Smart Water Quality Tester with Mobile App Interface
Its waterproof design and backlit LCD display ensure ease of use in various environments, from swimming pools and aquariums to laboratories and drinking water analysis.
Figure 2: Wide Range of Applications for the Water Quality Tester
2. Safety Information
- Read all instructions carefully before operating the device.
- Keep the device out of reach of children.
- Do not attempt to disassemble or modify the device, as this will void the warranty and may cause damage.
- Ensure the battery compartment is properly sealed after battery installation to maintain waterproof integrity.
- Avoid dropping the device or subjecting it to severe impacts.
- Use only specified AAA batteries (not included).
3. Package Contents
Please check the package for the following items:
- 1 x VAKIND BLE-YC01 Water Quality Tester
- (Calibration powders and AAA batteries are typically sold separately)
Figure 3: Typical Package Contents
4. Specifications
| Parameter | Value/Range | Resolution | Accuracy |
|---|---|---|---|
| Origin | Mainland China | ||
| Size | 160.00 x 90.00 x 26.00mm (6.3 x 3.54 x 1.02 inches) | ||
| Power Supply | 3*AAA batteries (not included) | ||
| pH | 0.1-14.0pH | 0.1pH | ±0.1pH |
| Chlorine (CL) | 0.1-4.0mg/L | 0.1mg/L | N/A |
| Total Dissolved Solids (TDS) | 0-999ppm, 1000-9990ppm | 1ppm, 10ppm | ±2%FS |
| ORP | -999 to +999mV | 1mV | 15mV |
| Temperature | 0.1℃-80℃ (32.1℉-176.0℉) | 0.1℃ | ±0.5℃ |
| pH Calibration | 6.8/4.0/9.2 or 7.0/4.0/1.0 | ||
| ORP Calibration | 256mV | ||
| ATC Temperature | 0.1℃-60℃ | ||
| Environment | 0.1℃-80℃, RH: max 90% | ||
| TDS Factor | 0.1-3.0 (default 0.5) | ||
| EC Temp. Compensation | 0.1-3.0 %/℃ (default 2.0 %/C) | ||
| Bluetooth | Version 5.2 | ||
| Waterproof Rating | IP67 | ||
| Data Storage | Up to 1000 data records | ||
Figure 4: Device Dimensions and Model Number (BLE-YC01)
5. Setup
5.1. Battery Installation
- Unscrew the battery compartment cover located at the top of the device.
- Insert three (3) AAA batteries, ensuring correct polarity (+/-).
- Securely replace the battery compartment cover to maintain the device's waterproof seal.
5.2. App Installation and Bluetooth Connection
- Download the official mobile application for the BLE-YC01 from your smartphone's app store (iOS or Android).
- Enable Bluetooth on your smartphone.
- Turn on the water quality tester by pressing the 'ON/OFF' button.
- Open the installed app on your smartphone. The app should automatically detect and connect to the BLE-YC01 device via Bluetooth.
- Once connected, all test results from the meter will be conveniently displayed on the app.
6. Operation
6.1. Taking Measurements
- Ensure the device is clean and calibrated (see Section 7 for calibration).
- Turn on the device using the 'ON/OFF' button.
- Immerse the electrode of the tester into the water solution to be tested. Ensure the electrode is fully submerged.
- The device's LCD screen will display the current readings.
- Simultaneously, the data will be synchronized to the mobile application, allowing you to view multiple parameters at once.
Figure 5: Data Synchronization and Backlight Control
6.2. App Features
- Multi-parameter Display: View pH, EC, TDS, ORP, CL, and Temperature readings simultaneously.
- Backlight Control: Turn the LCD backlight on or off directly from the app for better readability in different lighting conditions.
- Customizable Alerts: Set high or low limits for pH, EC, and ORP. If a measurement exceeds or falls below the specified range, the corresponding value will be displayed in red on the app, signaling an alert.
Figure 6: Setting Measurement Limits in the App
- Data Storage and Sharing: The app supports automatic and manual saving of up to 1000 data records. Saved data can be exported to Excel and shared.
Figure 7: Automatic Data Storage and Sharing Features
- Language Switch: The app allows switching between English and Chinese interfaces.
Figure 8: Overview of App Features
6.3. Device Controls
Figure 9: Device Controls and High Precision Probe
- LCD Display: Shows current readings.
- HOLD/TEMP Switch Key: Used to hold the current reading or switch to temperature display.
- ON/OFF/Bluetooth Switch Key: Powers the device on/off and controls Bluetooth connectivity.
- MODE/CAL Switch Key: Used to switch between measurement modes and initiate calibration.
7. Calibration
The BLE-YC01 supports automatic 5-point pH calibration and ORP calibration. Calibration powders are required and typically sold separately.
7.1. pH Calibration
- Prepare pH buffer solutions according to the standards you wish to use (Asia standard: pH 4.00, pH 6.86, pH 9.18; EU standards: pH 4.00, pH 7.00, pH 10.01).
- Ensure the electrode is clean and dry before calibration.
- Turn on the device and open the app. Navigate to the 'Calibrate' section in the app.
- Select the appropriate pH buffer standard (Asia or EU) in the app.
- Immerse the electrode into the first buffer solution (e.g., pH 6.86 or 7.00).
- Follow the on-screen instructions in the app to complete the calibration for each point. The app will guide you through immersing the electrode in subsequent buffer solutions.
- The app will record the calibration history.
7.2. ORP Calibration
The device supports ORP calibration at 256mV. Refer to the app's calibration section for specific instructions on ORP calibration using a standard ORP buffer solution.
Figure 10: pH Calibration Process with Buffer Powders
8. Maintenance
8.1. Cleaning the Electrode
- After each use, rinse the electrode thoroughly with distilled or deionized water to remove any residue.
- Gently wipe the electrode with a soft cloth. Do not scrub or use abrasive materials.
- For stubborn deposits, consult the app or manufacturer's guidelines for appropriate cleaning solutions.
8.2. Storage
- Store the device in a cool, dry place away from direct sunlight.
- Ensure the electrode is kept moist by storing it in a protective cap with a small amount of electrode storage solution or pH 4 buffer solution. Do not store dry.
- If storing for an extended period, remove the AAA batteries to prevent leakage.
9. Troubleshooting
- Inaccurate Readings:
- Ensure the device is properly calibrated.
- Clean the electrode thoroughly.
- Check battery levels.
- Ensure the electrode is fully immersed in the sample.
- Device Not Turning On:
- Check if batteries are installed correctly and have sufficient charge.
- Ensure the battery compartment is securely closed.
- Bluetooth Connection Issues:
- Ensure Bluetooth is enabled on your smartphone.
- Restart both the device and the app.
- Make sure the device is within range of your smartphone.
- Water Ingress:
- If water is observed inside the device, immediately remove batteries and allow to dry. This may indicate a compromised seal.
- Ensure the battery cap is always tightly sealed before submerging the device.
10. User Tips
- Prevent Water Ingress: Always double-check that the battery compartment cover is securely tightened and sealed before submerging the device. Even with an IP67 rating, improper sealing can lead to water damage.
- Regular Maintenance: Consistent cleaning and proper storage of the electrode will extend the lifespan and accuracy of your tester.
- Battery Life: If the device is not used for extended periods, remove the batteries to prevent potential leakage and damage to the internal components.
11. Warranty and Support
For warranty information, technical support, or service inquiries, please contact the retailer or manufacturer directly. Keep your purchase receipt as proof of purchase.