1. Introduction
1.1 Product Overview
The Halogen Moisture Analyzer is designed for accurate and efficient determination of moisture content in various substances. It operates on the thermogravimetric analysis principle, utilizing a halogen heating element for rapid and uniform drying. This instrument is easy to operate, provides precise measurements, and is suitable for a wide range of applications including laboratories, agriculture, and food processing.
Key Features:
- Programmable color touch screen for intuitive operation.
- Durable, chemical-resistant structure.
- Ergonomic design with a large screen for easy readability.
- Simple menu navigation.
- Built-in multifunctional options for setting running mode, printing mode, etc.
- Selectable drying methods (Standard, Fast, Soft).
- Internal database capable of storing up to 100 moisture data sets.
- RS232 interface for data connectivity.
- Real-time display of all test data during the drying process.
- Optional external printer support.

Figure 1: Front view of the Halogen Moisture Analyzer.

Figure 2: Multiple drying and shutdown modes for ultimate flexibility.

Figure 3: Effortless touchscreen control with a 5-inch full-color display.
1.2 Interface Button Description
The device features an intuitive touchscreen interface with various buttons for control and navigation. Below is a description of the main interface elements:
| Key | Specific Operation |
|---|---|
| Settings | Enter the system settings interface. |
| When connecting to a printer, moisture data can be printed out. | |
| Save | Save the measurement data to the measurement record. |
| Start/Stop/Reset | Start, stop, or reset a drying test. |
| Convert | During moisture testing, convert and display data such as moisture content and moisture regain. |
| Tare | In the weighing state, reset the weight to zero. After moisture content testing, return to the weighing state. |
| OFF | Shut down the system. |
2. Setup
2.1 Unboxing and Standard Configuration Check
Carefully open the product packaging. Remove the instrument and all accessories. Verify that all standard components are present and undamaged. If any items are missing or damaged, contact your dealer immediately. Retain all packaging for future transportation or storage.
Standard Configuration:
| Serial Number | Standard Configuration Name | Quantity |
|---|---|---|
| 1 | Moisture analyzer host | 1 unit |
| 2 | Stainless steel sample tray | 2 of them |
| 3 | Stainless steel tripod stand | 1 piece |
| 4 | Stainless steel sample tray bracket | 1 piece |
| 5 | Stainless steel windproof cover | 1 piece |
| 6 | Calibration weight (100g) | 1 piece |
| 7 | Power cord | 1 piece |
| 8 | Product manual | 1 copy |
| 9 | Product Qualification Certificate | 1 copy |
| 10 | Warranty card | 1 copy |
| 11 | Stylus | 1 unit |
Video 1: Unboxing and initial setup of the moisture analyzer, showing included accessories and assembly.
2.2 Selection of Placement Location
The accuracy and repeatability of the moisture analyzer are highly dependent on its placement. Avoid locations with:
- Excessive airflow or rapid temperature changes (e.g., near doors, windows, air conditioners, radiators).
- Vibration, rotation, or reciprocating motion.
- Devices generating magnetic fields.
- Uneven operating surfaces.
- Insufficient space around the instrument, especially near heat sources.
Ensure the instrument is placed on a stable, level surface. Adjust the horizontal feet until the horizontal bubble (located under the front panel) is centered. Leveling is crucial for accurate measurements and should be performed if the instrument's position changes.
2.3 Installing Trays, Windshields, and Sample Tray Brackets
- Open the cover of the moisture analyzer.
- Install the stainless steel windproof cover on the tray (only one position is available).
- Place the guard (metal ring) onto the pins on the weighing platform.
- Assemble the sample tray bracket by pushing the pin into the frame and rotating it.
- Install the assembled sample disk bracket, rotating it until it is positioned and locked. The bracket should point towards the front of the instrument in a "Y" shape.
- Place a sample tray onto the holder.

Figure 4: Internal view of the moisture analyzer with the heating chamber open.

Figure 5: Included accessories for the moisture analyzer.
2.4 Power Connection
Warning: Electric shock hazard.
- Use only the provided 3-pin power cord with a grounding plug. Connect the power cord only to a 3-pin grounded socket.
- Only cables that comply with relevant standards and have grounding plugs are allowed to be used as extension cords.
Caution:
- Verify that the voltage on the instrument label matches your local ground voltage. Incorrect voltage can damage the halogen lamps.
- Connect the power cord to the power socket and the power outlet at the back of the instrument.
- After connecting power, turn on the device using the power switch and then press the power button on the display screen.
- Preheat the moisture analyzer for at least 30 minutes to adapt to environmental fluctuations. If installed in a cooler environment, several hours may be needed for stability.

Figure 6: Rear panel connections including power and RS232 port.
3. Operating Instructions
3.1 Testing Steps
The moisture analyzer offers various measurement methods. Users should select the appropriate testing program based on the sample type.
- Set Drying Program: Choose from STD (Standard), Mild, Fast (rapid temperature rise), or Slow (slow temperature rise).
- Set Dry Temperature: Adjust the temperature between 40°C and 160°C in 1°C increments, based on the sample's requirements.
- Set Shutdown Mode: Select Automatic (based on water loss per unit time), Timed (set a specific duration), or Manual (user-controlled stop).
- If the shutdown mode is Timed, the test will automatically stop after the set time. If Automatic, the stop condition can be configured in "Settings -> System Settings". If Manual, the user must press the stop button to end the test.
- Place the sample in the sample tray (refer to section 3.2 for sample weight selection), close the heating cover, and wait for the weighing to stabilize before pressing the start button. The machine will then measure the moisture content according to the set program.
- To view the drying data curve, click the dedicated button on the display. Click the numerical display button to return to the numerical view.
- Upon completion, a voice prompt will indicate the measurement is complete. You can then click the conversion button to view different result data (moisture content, solid content, moisture regain), or click save to store the data, or print to generate a report.
- After the instrument returns to room temperature, clean the sample tray, press the reset button to return to the weighing state, and prepare for the next measurement.
Note: When changing the drying mode, the corresponding drying temperature will automatically reset to its default value. You can manually adjust it if needed.
- Default drying temperature for Standard, Gentle, and Fast modes: 120°C.
- Default drying temperature for Slow mode: 105°C (suitable for temperature-sensitive samples).
3.2 Sample Weight Selection During Measurement
The amount of sample affects testing time and accuracy. Generally, a 10g sample is recommended for balanced results. A 3g sample may yield faster results but could compromise accuracy. A 20g sample provides consistent results but requires a longer testing time. The sample weight should not be less than 1g.
3.3 Modification of Sample Name and Sample Number
You can modify the sample name and sample number by clicking on them on the main interface. This allows for organized record-keeping when saving measurement data.
3.4 Function Settings
Click the "Settings" button on the main interface to access various configurations, including measurement records, calibration settings, date and time, printing settings, system settings, and machine information.
3.4.1 Measurement Records
This section allows you to view saved measurement history, including time, sample name, moisture content, solid content, moisture regain, wet weight, dry weight, and measurement duration. The device can store up to 100 records; new data will overwrite the oldest records once the limit is reached.
- Click the trash can icon to delete the currently selected record.
- Click the double trash can icon to clear all recorded data.
3.4.2 Calibration Settings
Weight Calibration:
Precautions before weight calibration:
- Power on and preheat the instrument for 30 minutes to ensure stability and avoid air convection.
- Maintain ambient temperature at 20°C ± 2.5°C to avoid vibrations in small indoor rooms.
- Prepare the calibration weight (100g, included) and adjust the horizontal position of the instrument.
- Remove all loads from the instrument before starting calibration.
Calibration Method:
- On the main interface, navigate to "Settings -> Calibration Settings -> Weight Calibration".
- Click "Start Calibration". The device will prompt for zero point calibration.
- Once prompted, place the 100g calibration weight gently on the weighing plate.
- The device will indicate that calibration is in progress until the weight stabilizes.
- Upon completion, the device will automatically return to the main interface.
Temperature Calibration:
- Set the drying temperature to 100°C in the main interface.
- Place more than 1g of sample into the sample tray and cover it with the heating cover.
- Click the start button and wait for the temperature to stabilize at 100°C.
- Click stop.
- Navigate to "Settings -> Calibration Settings -> Temperature Calibration".
- Enter the actual temperature displayed on the temperature component into the calibration box and click to start calibration.
3.4.3 Date and Time
This interface allows you to modify the system's date and time. Use the swipe up and down gestures to change the values and click the confirm setting button to save.
3.4.4 Printing Settings
The instrument supports external receipt printers via the RS232 interface. You can select the baud rate corresponding to your printer. The printed receipt includes sample name, sample number, date, time, dry weight, wet weight, moisture content, solid content, moisture regain, and drying temperature/time. You can define the print title and custom fields.
3.4.5 System Settings
This section allows configuration of dry stop conditions, backlight, sound, language, user management, and factory reset. A management password (default: 123456) is required for user management and factory reset.
- Dry Stop Conditions: This parameter defines the automatic shutdown criteria for moisture testing. The default setting is to stop when the weight change is less than 5mg within 30 seconds.
- Backlight and Sound Settings: Adjust backlight brightness and volume.
- Language: Switch between Chinese and English.
- User Management: Set a login username and password to restrict access.
- Restore Factory: Resets all settings to default. Requires management password verification.
4. Maintenance and Upkeep
Proper maintenance is essential for the longevity and accuracy of your moisture analyzer.
- Do not use sharp objects (e.g., pencils, ballpoint pens) on the touchscreen.
- Avoid dropping objects onto the weighing plate to prevent damage to the weighing mechanism.
- Do not expose the instrument to high humidity or dusty environments for extended periods.
- After use, cover the instrument to prevent dust accumulation.
- Keep the instrument clean and dry; wipe it frequently and remove dust.
- If used frequently, keep the instrument continuously powered on to reduce preheating time.
- Perform weight calibration before each precise measurement and regularly check electronic instruments.
4.1 Cleaning Precautions
- Before cleaning, always unplug the power cord.
- Do not attempt to pull out the weighing plate forcefully, as this can damage the sensor.
- Do not use corrosive cleaning agents or solvents. Use a lint-free soft cloth dampened with water and a neutral detergent (e.g., mild soap).
- Be careful not to let water drip into the instrument during cleaning. After cleaning, thoroughly dry the instrument with a clean, lint-free soft cloth.
- Avoid contact with strong magnetic objects.
- Keep the instrument clean to prevent dust from entering its interior.
5. Troubleshooting
This section provides general guidance for common issues. For more complex problems, please refer to the full user manual PDF or contact customer support.
- Inaccurate Readings: Ensure the instrument is perfectly level, calibrated, and preheated. Check for environmental factors like drafts or vibrations.
- Display Not Responding: Verify power connection. If the touchscreen is unresponsive, try restarting the device.
- Printer Connectivity Issues: Check RS232 cable connection and ensure the baud rate in settings matches your printer.
- Halogen Lamp Not Heating: Confirm correct voltage supply and proper power connection. If the issue persists, the lamp may need replacement (contact service).
6. Specifications
The following table details the technical parameters of the Halogen Moisture Analyzer ME Series:
| Parameter | GS110A (XY-110ME) | GS110B (XY-105ME) | GS110C (XY-101ME) |
|---|---|---|---|
| Capacity | 110 g | ||
| Readability | 0.01 g | 0.005 g | 0.001 g |
| Sensor | Load Cell | HBM | |
| Pan Size | Φ96mm (20pcs) | ||
| Display | 5 inch touch screen | ||
| Operating Temperature | 5°C - 35°C | ||
| Heating Source | Halogen lamp | ||
| Temperature Sensor | PT-100 | ||
| Temperature Range | 40°C - 160°C | ||
| Temperature Setting Increment | 1°C | ||
| Moisture Range | 90.00% - 10.00% | 95.00% - 5.00% | 100.00% - 0.00% |
| Moisture Readability | 0.1% | 0.05% | 0.01% |
| Drying Readability | 0.1% | 0.05% | 0.01% |
| Storage | 20 Groups | ||
| Analysis Curve | Real time dynamic curve | ||
| Testing Height | 30mm | ||
| Language | EN/CN | ||
| Power Supply | 220V 50Hz / 110V 60Hz, 450W | ||
| Dimensions (L×W×H) | 336 x 214 x 166 mm | ||
| Gross Weight / Net Weight | 4.0kg / 3.5kg | ||
| Communication Interface | RS232 | ||
| Display Results | Moisture value, solid content, sample weight, weight after drying, moisture curve graph | ||
7. User Tips
- For optimal accuracy, always allow the analyzer to preheat for at least 30 minutes before calibration or testing.
- When testing new samples, start with the recommended 10g sample weight. Adjust if faster results are needed (3g) or higher consistency (20g), understanding the trade-offs.
- Regularly clean the sample tray and heating chamber to prevent contamination and ensure consistent results.
- Utilize the data storage and RS232 interface to keep detailed records of your tests, especially for quality control or research purposes.
8. Warranty and Support
This product comes with a warranty card included in the packaging. Please refer to the warranty card for specific terms and conditions regarding coverage and duration.
For technical support, troubleshooting assistance, or inquiries about replacement parts, please contact your original dealer or the manufacturer's customer service. Ensure you have your product model and serial number ready when contacting support.
A detailed user manual in PDF format is available for download. This PDF contains additional information, diagrams, and advanced operational procedures. You can access the PDF manual here.