Heltec HT-CC20ABP HT-CC40ABP

Contents

Heltec 9-99V VRLA Lithium Battery Charge-Discharge Tester User Manual

Models: HT-CC20ABP, HT-CC40ABP

1. Introduction

The Heltec HT-CC20ABP/HT-CC40ABP battery capacity tester is a sophisticated device designed for testing various battery types. It integrates high-precision capacity series discharge detection with high-precision charging functions. This tester is capable of evaluating cadmium-nickel batteries, lithium battery packs, lead-acid batteries, and other battery types within a 9-99V range.

It is an ideal tool for electric vehicle dealers and battery manufacturers to assess battery performance. Key functionalities include battery capacity testing, charge and discharge cycling, and the ability to set work steps via a connected computer. The device accurately collects real-time battery voltage signals, stores data, and processes operations, ensuring high precision, timeliness, and ease of use.

Key Features:

  • Automatic Collection and Analysis: Automatically measures and analyzes battery voltage, continuously outputting a constant current or power.
  • Software Control: Connects to a host computer for convenient and efficient work, allowing for software-based control.
  • Adjustable Current: The current is adjustable, with the HT-CC40ABP model capable of a peak discharge current of up to 40A, and the HT-CC20ABP model up to 20A.
  • Reverse Polarity Protection: Ensures battery safety by protecting against incorrect positive and negative terminal connections.
  • Simplified Operation Interface: Real-time display of voltage, capacity, current, and other parameters, supporting both Chinese and English languages.
  • Powerful PC Software: Offers comprehensive functions including data management, exporting, analysis, and report statistics.
  • Intelligent Cooling Fan: Features double heat sinks for faster heat dissipation and improved performance.

Product Application Range:

This tester is suitable for various applications, including testing lead-acid batteries in electric bikes, notebook batteries, and battery packs for small and medium-sized electronic products like radios.

Video: Overview of the Heltec battery tester, demonstrating its features and connection to a laptop for data analysis.

2. Safety Instructions

  • Grounding: The device must be properly grounded. High voltage is present inside; only qualified professionals should open the casing.
  • Battery Connection: Always connect the battery with the correct polarity. Ensure a secure and stable connection.
  • Environment: Keep the device away from flammable and explosive materials. Operate and store in a cool, well-ventilated area.
  • Ventilation: Do not block the air inlet and outlet of the device to ensure proper cooling.
  • Handling: Store the device properly and avoid violent actions such as collisions during use and transportation.
  • Instruction Manual: Read this instruction manual carefully before operation. Contact customer service if you have any questions or difficulties.

3. Product Overview

The Heltec battery tester is designed for ease of use and robust performance. Below are diagrams illustrating the device's components and interface.

Heltec 9-99V VRLA/Lithium Battery Charge-Discharge Tester Machine

Image: Heltec 9-99V VRLA/Lithium Battery Charge-Discharge Tester Machine, showing the device connected to a battery and a laptop.

Front and Back Panel Diagram

Image: Front and Back panel diagram with labels for ports and controls. The front panel includes the LCD display, Start/Stop button, Set button, and Output port. The back panel includes the LAN port, Status Light, Power Socket, and Switch.

Detailed LCD Display Interface

Image: Detailed LCD display interface showing various parameters: Real-time voltage (V), Real-time current (A), Charging/Discharging Running Time (H), Charging/Discharging Capacity (AH), Voltage Setup (V), Mode Selection, and Loop Number. Buttons for Setup, Charge, Discharge, Recovery, Formation, Run, and Stop are also visible.

4. Setup

4.1. Physical Setup

  1. Connect the power cord to the device's power socket and a suitable AC220V 50HZ power outlet.
  2. Connect the clamping lines (red for positive, black for negative) to the battery terminals. Ensure correct polarity and a firm connection.
  3. Connect the data collection line from the device to your computer's USB port.
  4. (Optional) Connect the temperature collection line to the device and place the sensor near the battery for real-time temperature monitoring.

4.2. Software Installation and Connection

The device can be controlled and monitored via PC software. Follow these steps to install and connect the software:

  1. Download Software: Obtain the installation package for the upper computer software. You can scan the QR code provided below or contact customer service for the software link.
  2. Installation: Open the installation software. Choose an installation location (preferably not the C drive to avoid potential issues).
  3. Initial Run: After installation, open the software. During the first run, select the network type (ensure public network is checked if applicable, especially if a firewall is enabled).
  4. Connect Device: In the software interface, click the "Connect" button (usually in the top left corner). Then click "Search" to find available devices. The IP address of your device should appear. Select your device (it will turn blue) and click "Connect Device". A window for "IP Settings" may appear; click "OK" by default.
Upper Computer Download QR Code

Image: Upper Computer Download section with a QR code for software installation. After installation, connect the instrument to the computer using a communication cable, click 'Connect', then 'Search' to find the device's IP address, and click 'Connect' to control the instrument.

Software Interface Device Connection Settings

Image: Software interface showing device connection settings, including options for Serial Port, Device, Connect Method, IP, and Status. This interface is used to establish communication between the tester and the PC.

5. Operation

5.1. Device Interface Operation

Once the device is powered on, you can use the knob and buttons on the front panel to operate it directly.

  • Start/Stop Button: Press to start a test. Press again while running to pause. Pressing it again will resume the operation.
  • Set Button: Used to enter parameter setting mode and confirm settings.
  • Display Screen: Shows real-time parameters such as current, voltage, capacity, and time.

5.2. Working Modes

The device features 8 working modes, selectable via the setting knob:

Mode Description
00Historical data query
01Capacity testing (discharge)
02Charging mode
03Start with discharge and end with charge (1-50 cycles)
04Start with charging and end with charging (1-50 cycles)
05Start with discharge and end with discharge (1-50 cycles)
06Start with charging and end with discharging (1-50 cycles)
07Networking mode (automatically activated when started by computer)

5.3. Parameter Setting (Device Interface)

After selecting a mode, you need to set its corresponding parameters. Generally, you rotate the black setting knob to adjust values and press it to confirm or move to the next parameter. The current display will show "0.00" upon successful setting.

  • Discharge Mode (Mode 01): Adjust minimum discharge cut-off voltage and discharge current.
  • Charging Mode (Mode 02): Adjust charging cut-off voltage, charging current, and charging cut-off current.
  • Cycle Mode (Mode 03-06): Both charging and discharging parameters need to be adjusted.
  • Inquire Mode (Mode 00): Used to query historical data. Rotate the knob to view data for the last single charge/discharge or specific loop charges/discharges.

5.4. PC Software Operation

The PC software provides advanced control and data management capabilities.

  • Step Settings: Click the "Step Settings" button to create, edit, and manage test steps. You can define charging and discharging parameters, cycles, and other settings for specific battery types.
  • Work Step Setting Guidelines: The manual provides detailed guidelines for setting work steps for different battery chemistries (e.g., ternary lithium, lithium iron phosphate, lead-acid) based on nominal voltage, recommended voltage settings (Vmax, Vmin), and current settings (charge/discharge current, cut-off current). For example, for a 10-string ternary lithium battery pack (37V, 10.4AH), charging/discharging current should be less than half of total capacity (e.g., 5A), cut-off current 0.02 times total capacity (e.g., 0.2A), max charging cut-off voltage 42V, min discharging cut-off voltage 30V.
  • Data Storage Conditions: Configure how data is saved to your computer. You can set minimum time, voltage, and current intervals for data recording.
  • Data Analysis: Use the analysis function to retrieve record files and launch the data curve analyzer. The software supports 3Y axes, single timeline, panning, and zooming for detailed data review.
  • Data Import/Export: Import existing data files or export test data as an XLS report for further analysis.
  • Multiple Curve Comparison: Compare data from multiple record files or a single file.
  • Right-Click Functions: Right-clicking on a device in the software allows you to Setup/Start tests, Restore/Stop operations, and view Channel Info (current work step, status, etc.).

6. Maintenance

  • Heat Dissipation: After each charge or discharge cycle, allow the battery to rest for a period to dissipate heat before further testing.
  • Cleaning: Keep the device clean and free from dust. Use a soft, dry cloth for cleaning.
  • Storage: Store the device in a cool, dry, well-ventilated place when not in use.
  • Firmware Updates: Periodically check for software and firmware updates from Heltec to ensure optimal performance and new features.

7. Troubleshooting

This section provides solutions to common problems you might encounter.

Encountered Problem Possible Cause Solution
Power on, LCD screen doesn't light up
  • Power plug not properly connected
  • Fuse in power socket blown
  • Fuse in circuit blown
  • Flat cable between LCD and mainboard loose
  • Switch power supply not working
  • Re-connect power plug
  • Replace with 5A fuse
  • Replace with 1.5A fuse
  • Plug flat cable tightly
  • Return to factory for repair/replacement
Power on, LCD lights up but displays nothing
  • Flat cable between LCD and mainboard loose
  • LCD damaged
  • Communication issue between SCM and LCD
  • Plug flat cable tightly
  • Replace LCD
  • Return to factory for repair/replacement
Setting knob does not work
  • Flat cable between knob and mainboard loose
  • Knob pressed too deeply/tightly
  • Encoder damage
  • Plug flat cable tightly
  • Pull out and reinstall knob
  • Replace encoder
Device has abnormal noise
  • Foreign matter in fan
  • Fan not rotating properly
  • Open case, remove foreign matter
  • Refuel/replace fan if damaged
No voltage displays after test cables connect to battery
  • Poor connection between test cables and battery
  • Voltage sampling flat cable damaged
  • SCM cannot detect battery
  • Clear clamps/cathode tab
  • Reinsert/replace flat cable
  • Return to factory for repair/replacement
Press start button, fail to start
  • Flat cable of start button loose
  • Start button damaged
  • SCM cannot detect voltage
  • Reinsert flat cable
  • Replace start button
  • Return to factory for repair/replacement
Voltage displays, but can't charge/discharge after start
  • Connection of four core wire damaged
  • SCM cannot detect current or switch power supply damaged
  • Heat wire loose
  • MOS tube damaged
  • Reconnect wire/replace quad
  • Return to factory for repair/replacement
  • Tighten heat wire
  • Return to factory for repair/replacement
Software connection issues
  • Software installed on C drive
  • Public network firewall blocking connection
  • Run software as administrator or reinstall on another drive
  • Disable public network firewall or modify connection mode
Machine stops running unexpectedly
  • Connecting cable came off
  • Ensure cable is firmly connected. Press red "Start" button for 5 seconds to resume.

8. Specifications

Detailed specifications for the Heltec Battery Charge-Discharge Tester.

Parameter HT-CC20ABP HT-CC40ABP
Model NumberHT-CC20ABPHT-CC40ABP
TypeElectronic Device Battery Tester
DIY SuppliesElectrical
Brand NameHeltecbms
OriginMainland China
CertificationCE
Battery IncludedNo
High-concerned chemicalNone
Power SourceNone
Scope of ApplicationLead-acid/lithium/nickel-cadmium or other batteries within 9-99V
Charging Range9-99V, 0.1V continuously adjustable, 0.5-10A (HT-CC20ABP) / 0.5-20A (HT-CC40ABP), 0.1A adjustable
Discharge Range9-99V, 0.1V continuously adjustable, 0.5-20A (HT-CC20ABP) / 0.5-40A (HT-CC40ABP), 0.1A adjustable
Operation StepCharge/discharge/rest/cycle (up to 99 times)
Upper Computer CommunicationSystems above WIN XP, supports both Chinese and English
Protection FunctionBattery reverse polarity protection
Measuring ErrorVoltage ±0.05V, current ±0.03A (accuracy guarantee period is within one year from purchase)
AttentionProhibited to test batteries with voltage exceeding 99V
Equipment Power SupplyAC220V 50HZ
Dimensions (L x W x H)60cm x 57cm x 27cm
Net Weight15Kg
Preset Charging Capacity (end of cycle)0-99AH (Setting 0 indicates no preset)
Product Specifications Table

Image: Product specifications table for the HT-CC20ABP and HT-CC40ABP models, detailing parameters such as voltage range, current range, operation steps, communication, protection, and dimensions.

9. Warranty and Support

Warranty Regulations:

  • Warranty Period: The warranty service period is 1 year from the date of purchase.
  • Coverage: Warranty service is limited to normal use.
  • Exclusions: The warranty does not cover man-made damage, self-disassembly, modification, repair by unauthorized personnel, use not in accordance with instructions, or damage caused by external force majeure factors.
  • Accessories: Consumable accessories such as test fixtures are not covered by warranty.
  • Service Process: For warranty service, please contact your dealer. If you cannot reach the dealer, contact Heltec directly via email or phone.
  • International Service: There are no after-sales service stations outside China. For international warranty service, you will need to send the product to Heltec for free repair, but you are responsible for the round-trip freight.

Support:

For software downloads or technical assistance, please refer to the QR code in the setup section or contact Heltec customer service.

You can find the user manual PDF for this product here.

10. User Tips

  • Accuracy vs. Speed: The effectiveness of capacity testing is influenced by both time and speed. Faster testing speeds can sometimes lead to a decrease in measured capacity (known as the Pickett effect). For improved accuracy, it is recommended to use a smaller discharge current value.
  • Battery Condition: This capacity tester is designed for functional batteries. It cannot accurately test damaged batteries, including those with increased internal resistance, terminal voltage below discharge cut-off, or internal short/open circuits. Ensure batteries are in reasonable condition before testing.
  • Software Administration: If you encounter issues with the PC software (e.g., errors when entering username/password or adding processing steps), try running the software as an administrator. If the problem persists, reinstalling the software on a different drive (not C drive) might resolve it.
  • Network Configuration: For PC software connectivity, ensure your network firewall is not blocking the connection. You may need to temporarily disable the public network firewall or adjust its settings to allow the software to connect.

11. Frequently Asked Questions (FAQ)

Q: Can I test batteries with voltage outside the 9-99V range?
A: No, it is strictly prohibited to test batteries with voltage exceeding 99V, as stated in the specifications. Testing outside the specified range can damage the device.
Q: What should I do if the device stops unexpectedly during a test?
A: First, check if all connecting cables are firmly attached. If a cable came loose, re-secure it. Then, press the red "Start" button for 5 seconds; the machine should resume its operation from where it left off.
Q: Is it necessary to set parameters every time I use the device?
A: No, after the system completes parameter settings, it automatically saves them. The next time you start the device, you only need to click the start button to run it directly, unless you need to change the test parameters.
Q: How do I know if the PC software is connected to the device?
A: In the PC software, after clicking "Connect" and "Search", the IP address of your device should appear. Select it and click "Connect Device". The software interface will display the connection status, and the green network indicator light on the device will be constantly on when the software is connected.
Documents and downloads

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.