OWON SDS220S Digital Storage Oscilloscope User Manual

Model: SDS220S | Brand: OWON

1. Introduction and Safety Information

This manual provides essential information for the safe and effective operation of your OWON SDS220S Digital Storage Oscilloscope. Please read it thoroughly before use.

1.1 General Safety Requirements

To prevent bodily injury and damage to the product, observe the following safety precautions:

  • Use Proper Power Cord: Only use the power cord supplied with the product and certified for use in your country.
  • Connect or Disconnect Correctly: When connecting or disconnecting probes/test leads to a voltage source, ensure proper procedure.
  • Product Grounded: The instrument is grounded through the power cord. Ensure proper grounding to avoid electric shock.
  • AC Power Sources: Do not measure AC power sources directly when the instrument is powered by AC, as the testing ground and power cord ground conductor are connected, which can cause a short circuit.
  • Check Terminal Ratings: Verify all ratings and markings before connecting to the instrument.
  • Do Not Operate Without Covers: Do not operate the instrument with covers or panels removed.
  • Use the Proper Fuse: Use only the specified type and rating fuse.
  • Avoid Exposed Circuitry: Exercise caution when working on exposed circuits.
  • Do Not Operate if Damaged: If damage is suspected, have it inspected by service personnel.
  • Well-Ventilated Area: Use the oscilloscope in a well-ventilated area.
  • Environmental Conditions: Do not operate in damp conditions or an explosive atmosphere. Keep product surfaces clean and dry.

1.2 Safety Terms and Symbols

Warning: Indicates conditions or practices that could result in injury or loss of life.

Caution: Indicates conditions or practices that could result in damage to this product or other property.

Symbols on the product:

  • ⚠ Hazardous Voltage
  • ⚠ Refer to Manual
  • ☎ Protective Earth Terminal
  • ☎ Chassis Ground
  • ☎ Test Ground

Warning: The two channels of the oscilloscope are not electrically isolated. They share a common ground during measuring. To prevent short circuits, the 2 probe grounds must not be connected to 2 different non-isolated DC levels.

Warning: To avoid fire or electrical shock, when the oscilloscope input signal is connected to more than 42V peak (30Vrms) or on circuits of more than 4800VA, please note:

  • Only use accessory insulated voltage probes and test leads.
  • Check accessories for damage before use.
  • Remove probes, test leads, and other accessories immediately after use.
  • Remove USB cable connecting oscilloscope and computer.
  • Do not apply input voltages above the instrument's rating.
  • Do not use exposed metal BNC or banana plug connectors.
  • Do not insert metal objects into connectors.

2. Product Overview

The OWON SDS220S is a professional dual-channel digital storage oscilloscope with a 200MHz bandwidth, 1GSa/s real-time sampling rate, and 20M record length. It also features a built-in 25MHz arbitrary waveform generator.

OWON SDS220S Oscilloscope with key features
Figure 2.1: OWON SDS220S Oscilloscope highlighting 200MHz Bandwidth, 1GS/s Sampling Rate, 20M Record Length, and 30 Automatic Measurements. Also shows USB Host/Device, SCPI, and LabVIEW support.

2.1 Front Panel

The front panel features a 7-inch TFT LCD display, knobs, and function buttons for controlling the oscilloscope. Menu selection buttons on the right side of the display allow setting options for the current menu. Other function buttons provide direct access to specific applications.

Front Panel Layout of SDS220S Oscilloscope
Figure 2.2: Front Panel Layout. Key areas include the Button and Knob Control Area, Power Switch, Function Keys, USB Host Interface, Signal Generator Output Port, Oscilloscope Signal Input Ports (CH1, CH2), and Probe Compensation output.

2.2 Rear Panel

The rear panel includes essential ports and features for power and connectivity.

Rear Panel Layout of SDS220S Oscilloscope
Figure 2.3: Rear Panel Layout. Features include a Foldable handle, Ventilation Holes, Power socket, USB Device Interface (for PC connection), and a Stand.

2.3 Control Area Overview

The control area is divided into Horizontal, Trigger, and Vertical sections, each with dedicated buttons and knobs for precise adjustments.

  • Horizontal Control Area: Includes the 'HOR' button and 'Horizontal Position' and 'Horizontal Scale' knobs for time base and trigger position control.
  • Trigger Control Area: Features the 'Trigger Level' knob to adjust trigger voltage and buttons for trigger system settings.
  • Vertical Control Area: Contains 'Single', 'Run/Stop', 'Math', 'CH1', 'CH2' buttons, and 'Vertical Position' and 'Scale' knobs for each channel. The 'M' knob (Multipurpose knob) is used for menu selection and value setting.

2.4 User Interface Introduction

The 7-inch TFT LCD display provides a comprehensive view of waveforms and measurement data. Key elements include:

  • Waveform Display Area: Shows the captured signals.
  • Run/Stop Indicator: Shows the acquisition status.
  • Trigger State: Indicates Auto, Trig, Ready, Scan, or Stop.
  • Cursor Measurement Lines: Blue dotted lines for vertical cursor measurements.
  • T Pointer: Indicates horizontal trigger position.
  • Trigger Position Pointer: Shows trigger position in record length.
  • Triggering Value: Displays current trigger value and window site in internal memory.
  • USB Disk Connection Indicator: Appears when a USB disk is connected.
  • Channel Identifiers: Shows current menu channel (CH1, CH2).
  • Right Menu: Displays options for current settings.
  • Current Trigger Type: Indicates the active trigger mode.

The display also shows various measurement readings, including Period, Frequency, Peak-to-Peak Voltage (Vpp), RMS, Max, Min, Amplitude, Rise/Fall Time, Duty Cycle, and more. Coupling modes (DC, AC, GND) and bandwidth limits are also indicated.

3. Setup and Initial Configuration

3.1 General Inspection

Upon receiving your oscilloscope, perform the following checks:

  1. Check for Transportation Damage: Inspect the packaging for serious damage. If found, retain packaging until product functionality is verified.
  2. Check Accessories: Verify all supplied accessories against the enclosure list (Appendix A). Contact your distributor if any are missing or damaged.
  3. Check Complete Instrument: Inspect the instrument for appearance damage. If it doesn't work normally, contact your distributor.

3.2 Function Inspection

To quickly verify normal operation:

  1. Connect the power cord and press the power button (⑁) on the bottom left. The instrument will perform self-checks and display the Boot Logo.
  2. Set the probe switch to 10X and connect to CH1.
  3. Push the Autoset button on the front panel. A 1 KHz, 5V peak-to-peak square wave should be displayed.
Oscilloscope screen showing auto-set square wave
Figure 3.1: Example of an auto-set square wave display.

3.3 Probe Compensation

Adjust probe compensation for accurate measurements:

  1. Set the probe attenuation coefficient to 10X (both on the probe switch and in the oscilloscope menu). Connect the probe to CH1.
  2. Connect the probe tip to the signal connector of the probe compensator and the ground clamp to its ground wire connector.
  3. Push the Autoset button.
  4. Check the displayed waveforms. Adjust the probe until a correctly compensated square wave is achieved.
Probe Compensation Waveforms
Figure 3.2: Examples of Overcompensated, Compensated Correctly, and Undercompensated waveforms.
Probe Adjustment Diagram
Figure 3.3: Diagram illustrating how to adjust the probe for compensation.

3.4 Setting Probe Attenuation Coefficient

To ensure accurate vertical scale readings, match the probe's physical attenuation setting with the oscilloscope's menu setting.

  1. Push the function menu button of the used channel (CH1 or CH2).
  2. Select Probe in the right menu. Turn the M knob to select the correct value (e.g., 1X, 10X, 100X, 1000X) corresponding to your probe.

Caution: The default attenuation coefficient is 10X. Ensure the probe's physical switch matches the oscilloscope's setting.

Probe Attenuation Switch
Figure 3.4: Probe attenuation switch, typically offering 1X and 10X settings.

3.5 Using the Probe Safely

Always keep your finger behind the safety guard ring of the probe to prevent electric shock.

Probe with Safety Guard Ring
Figure 3.5: Probe with a visible safety guard ring.

3.6 Self-calibration

Perform self-calibration to optimize measurement accuracy, especially if the ambient temperature changes by 5°C or more. Disconnect all probes/wires from the input connector before starting.

  1. Push the Utility button, select Function in the right menu.
  2. Select Adjust in the left menu, then Self Cal in the right menu.
  3. Run the program after ensuring everything is ready.

4. Operating Instructions

4.1 Vertical System Settings

The Vertical Controls include CH1, CH2, and Math buttons, along with Vertical Position and Vertical Scale knobs for each channel.

  • Turning Waveforms On/Off: Press CH1, CH2, or Math buttons to toggle waveform display and their respective menus.
  • Vertical Position Knob: Adjusts the vertical display position of the signal. Pushing the knob resets the position to 0.
  • Vertical Scale Knob: Regulates the vertical resolution (Volts/div).
  • Channel Coupling: Set to DC (passes both AC/DC), AC (blocks DC), or Ground (disconnects signal).
  • Inverted: Toggles inverted waveform display.
  • Probe: Matches the probe attenuation factor.
  • MeasCurr (Measure Current): For current measurements using voltage drop across a resistor.
  • Limit: Limits channel bandwidth to 20MHz to reduce display noise.

4.2 Horizontal System Settings

The Horizontal Controls consist of the Horizontal HOR button and Horizontal Position and Horizontal Scale knobs.

  • Horizontal Scale Knob: Changes the horizontal time base setting (time/div).
  • Horizontal Position Knob: Adjusts the horizontal position of the signal. Pushing the knob resets the position to 0.
  • Horizontal HOR Button: Switches between normal mode and wave zoom mode.

4.3 Trigger System Settings

The Trigger Controls include a knob (Trigger Level) and buttons (Trigger Menu, Force).

  • Trigger Level Knob: Adjusts the trigger voltage. Pushing the knob sets the trigger level to the vertical midpoint of the signal amplitude.
  • Trigger Menu Button: Displays the trigger menu for settings.
  • Force Button: Forces a trigger signal, mainly used in 'Normal' and 'Single' trigger modes.

Trigger Types:

  • Single Trigger: Captures stable waveforms in two channels simultaneously based on a trigger level. Includes Edge Trigger (rising/falling edge) and Video Trigger (NTSC, PAL, SECAM).
  • Alternate Trigger: Triggers on non-synchronized signals, used for observing two unrelated signals.

4.4 Sampling and Display Settings

Access these settings via the Acquire button.

  • Acqu Mode: Sample (normal), Peak Detect (captures max/min samples), Average (reduces noise).
  • Length: Selects the record length (e.g., 1000, 10K, 100K, 1M, 10M, 20M).
  • Persist: Sets persistence time (OFF, 1 Second, 2 Seconds, 5 Seconds, Infinity) to simulate picture tube oscilloscope display.
  • XY Mode: Turns on/off XY display function (Channel 1 on horizontal, Channel 2 on vertical).
  • Counter: Turns on/off the 6-digit single-channel frequency counter (2Hz to full bandwidth).

4.5 Mathematical Manipulation (Math Function)

The Math button allows operations between two channels (addition, subtraction, multiplication, division) or FFT analysis.

  • Type: Select Math (for channel operations) or FFT.
  • Factor1/Factor2: Select signal sources (CH1, CH2).
  • Sign: Select operation (+, -, *, /).
  • FFT Function: Converts time-domain waveforms into frequency components. Useful for analyzing signal frequencies.
  • FFT Window Types: Hamming, Rectangle, Blackman, Hanning, Kaiser, Bartlett. Each has trade-offs in frequency resolution and magnitude accuracy.

4.6 Waveform Generator (Optional)

The instrument can generate 4 basic waveforms (sine, square, ramp, pulse) and 16 arbitrary waveforms.

  1. Connect the BNC cable to the Output port on the oscilloscope.
  2. Press the AFG Output button to display the waveform setting menu.
  3. Select the desired waveform type (e.g., Sine, Square, Ramp, Pulse, Arbitrary).
  4. Adjust parameters like Freq/Period, Amplitude/High Level, Offset/Low Level, Pulse Width/Duty Cycle, Rise/Fall Time using the M knob.
  5. For arbitrary waveforms, select from built-in types (e.g., AmpALT, StairDn, Besselj, Sinc, Gauss).
  6. Press the AFG Output button again to start generation.

5. Measurements

5.1 Automatic Measurements

The oscilloscope provides 30 auto-measurement parameters, displayed on the bottom left of the screen. Access via the Measure button.

  • Time Parameters: Period, Frequency, Rise Time, Fall Time, +PulseWidth, -PulseWidth, +Duty Cycle, -Duty Cycle, Delay A→B, Screen Duty, Phase.
  • Voltage Parameters: Mean, PK-PK, RMS, Max, Min, Top, Base, Amplitude, Overshoot, Preshoot, Cycle RMS, Cursor RMS.
  • Other: Area, Cycle Area, +PulseCount, -PulseCount, RiseEdgeCnt, FallEdgeCnt.

To add a measurement:

  1. Push the Measure button.
  2. Select Add CH1 or Add CH2.
  3. Turn the M knob to select the desired measurement type.
  4. Press Add CH1/CH2 again to add it.

5.2 Cursor Measurements

Use cursors for precise measurements of voltage and time. Push the Cursor button to toggle cursors on/off and display the cursor menu.

  • Type: Voltage, Time, Time&Voltage, AutoCursr.
  • Line Type: Freq, Vamp (for FFT mode).
  • Line (a, b, ab): Use the M knob to move individual cursors or linked pairs.
  • Window: Main or Extension (for zoom mode).
Oscilloscope screen showing X-Y mode Lissajous figure
Figure 5.1: Lissajous Graph in X-Y mode, useful for phase difference analysis.

6. Communication with PC

The oscilloscope supports communication with a PC via USB for data storage, analysis, display, and remote control.

  1. Set USB Device Protocol: On the oscilloscope, navigate to Utility → Function → Configure → Device and set to PC.
  2. Connect: Use a USB data cable to connect the oscilloscope's USB Device port to a PC's USB port.
  3. Software Port Setting: Run the Oscilloscope software on your PC. In the software, go to 'Communications' → 'Ports-Settings' and select 'Connect using' as 'USB'. The connection status will turn green upon success.

6.1 USB Disk Requirements

For saving data to a USB disk:

  • Format: USB 2.0 or below, FAT16 or FAT32, allocation unit size no more than 4KB, max capacity 64GB.
  • If the USB disk doesn't work, format it using your computer's system tools or dedicated formatting software.

To format a USB disk using system-provided function:

  1. Connect the USB disk to the computer.
  2. Right-click 'Computer' → 'Manage' to enter 'Computer Management'.
  3. Click 'Disk Management'.
  4. Right-click the USB disk area, choose 'Format', and confirm.
  5. Set File System to FAT32, Allocation unit size to 4096. Check 'Perform a quick format'. Click OK and confirm.

7. User Tips

  • Maximizing Sample Rate: The SDS220S offers a 1GSa/s sample rate. If you need this high sample rate on both channels simultaneously for short captures, reduce the memory depth (e.g., to 1000 points). This allows the 1GSa/s rate to be maintained across both channels, whereas with full memory depth, the sample rate is halved when both channels are active.
  • Firmware Updates: Regularly check the OWON website for new firmware versions to ensure optimal performance and access to the latest features.
  • Signal Generator Output: When using the arbitrary waveform generator, connect the output to an oscilloscope input (e.g., CH1 or CH2) to visualize and verify the generated waveform.
Play video
Watch video online
Video 7.1: Demonstration of memory depth and sampling rate behavior when switching between single and dual-channel modes, highlighting how to maintain 1GSa/s with both channels active by adjusting memory depth.

8. Maintenance

8.1 General Care

  • Do not store the instrument where the LCD will be exposed to direct sunlight for long periods.
  • Caution: To avoid damage, do not expose the instrument or probes to sprays, liquids, or solvents.

8.2 Cleaning

  1. Wipe dust from the instrument and probe surfaces with a soft cloth. Avoid scuffing the LCD screen.
  2. Disconnect power before cleaning. Use a wet soft cloth (not dripping water) with mild detergent or fresh water. Do not use corrosive chemical cleaning agents.

Warning: Before powering on after cleaning, ensure the instrument is completely dry to prevent electrical short circuits or injury from moisture.

9. Troubleshooting

9.1 Oscilloscope Powered On but No Display

  • Check power connection.
  • Restart the instrument.
  • If the problem persists, contact support.

9.2 Waveform Not Displayed After Acquiring Signal

  • Check if the probe is properly connected to the signal connecting wire and BNC connector.
  • Verify the probe is properly connected to the object being measured.
  • Ensure there is a signal generated from the object.
  • Perform signal acquisition again.

9.3 Measured Voltage Amplitude is Incorrect (10x or 1/10x)

  • Ensure the attenuation coefficient for the input channel and the probe match (refer to Section 3.4).

9.4 Waveform Displayed but Not Stable

  • Check if the Source item in the TRIG MODE menu matches the signal channel used.
  • Verify the trigger Type item (e.g., Edge for common signals, Video for video signals).
  • If Alternate trigger is selected, adjust both CH1 and CH2 trigger levels.

9.5 No Display Response to Run/Stop

  • Check Polarity in the TRIG MODE menu and ensure the trigger level is within the waveform range.
  • Center the trigger level or set the trigger mode to Auto. Use the Autoset button to automatically adjust settings.

9.6 Slow Waveform Display with Increased AVERAGE Value or Longer Persist Duration

  • This is normal behavior as the oscilloscope processes more data points.

10. Technical Specifications

Specifications are for the oscilloscope with probes attenuation set to 10X, unless otherwise specified. The instrument should run for at least 30 minutes continuously under specified operating temperature. A self-calibration should be performed if the operating temperature changes by 5°C or more.

10.1 Performance Characteristics

CharacteristicSpecification
BandwidthSDS220S: 200 MHz
Channels2 channels
Sample Rate (real time)1 GSa/s
Input CouplingDC, AC, Ground
Input Impedance1 MΩ±2%, in parallel with 12 pF±5 pF
Max. Input Voltage400V (DC+AC, PK - PK)
Record Length20M
Vertical Resolution (A/D)8 bits (2 channels simultaneously)
Vertical Sensitivity2 mV/div~10 V/div
Rise Time (BNC, Typical)SDS220S ≤ 1.75 ns
DC Gain Accuracy±3%
Waveform Math+, -, *, /, FFT
Waveform Storage16 waveforms
Communication PortUSB 2.0 (USB storage)
Trigger Level Range±4 div from the screen center

10.2 Waveform Generator (Optional)

CharacteristicSpecification
Standard WaveformsSine, Square, Ramp, Pulse, Arbitrary
Arbitrary WaveformsSinc, exponential rise/decline, Gaussian, >160 kinds
Sine Wave Frequency0.1Hz ~ 25MHz
Square Wave Frequency0.1Hz ~ 5MHz
Ramp Wave Frequency0.1Hz ~ 1MHz
Pulse Wave Frequency0.1Hz ~ 5MHz
Arbitrary Wave Frequency0.1Hz ~ 5MHz
Real-time Sample125MSa/s
Amplitude (50Ω)0.005Vpp ~ 3Vpp
Frequency Resolution0.01%

10.3 General Technical Specifications

CharacteristicSpecification
Display Type7" Colored LCD (Liquid Crystal Display)
Display Resolution800 (Horizontal) × 480 (Vertical) Pixels
Display Colors65536 colors, TFT screen
Mains Voltage100 - 240 VACRMS, 50/60 Hz
Power Consumption< 16 W
Operating Temperature0°C - 40°C
Storage Temperature-20°C - 60°C
Relative Humidity≤ 90%
Operating Height3,000 m
Non-operating Height15,000 m
Cooling MethodNatural cooling
Dimensions (L*H*W)301 mm × 152 mm × 70 mm
WeightAbout 1.1 kg

11. Warranty and Support

11.1 General Warranty

The product is warranted to be free from defects in materials and workmanship for a period of 3 years from the date of purchase by the original purchaser. Accessories such as probes have a 12-month warranty. This warranty applies only to the original purchaser and is not transferable.

If the product is defective during the warranty period, it will be repaired or replaced. Replaced parts may be new or reconditioned. All replaced parts become the property of the company.

To obtain service, notify the company before the warranty period expires. The customer is responsible for packaging and shipping the defective product to a designated service center, along with proof of purchase.

This warranty does not cover defects or damage caused by improper use, inadequate maintenance, unauthorized repair attempts, connection to incompatible equipment, use of non-company supplies, or modifications that increase servicing difficulty.

Except for the after-sales services provided, no other guarantees, express or implied, are offered. The company is not responsible for any indirect, special, or consequential damages.

11.2 Product Support

For product support, visit: www.owon.com.hk/download

Fujian LILLIPUT Optoelectronics Technology Co., Ltd.
No. 19, Heming Road, Lantian Industrial Zone, Zhangzhou 363005 P.R. China
Tel: +86-596-2130430
Fax: +86-596-2109272
E-mail: info@owon.com.cn