1. Introduction
Thank you for choosing the COXO CP-1 Dental Air Polisher. This handheld dental device is powered by a dental unit and is designed to jet sandblasting powder onto the surface of a patient's teeth.
The COXO CP-1 Dental Air Polisher effectively removes dental plaque, soft deposits, and surface stains from pits, grooves, interproximal spaces, and smooth surfaces of the teeth. It is suitable for both supragingival and subgingival applications.
2. Specifications
2.1. General Specifications (COXO CP-1)
- Input air pressure: 250-450kPa
- Input water pressure: 70-220kPa
- Water flow: >30ml/min
- Nozzle caliber for powder measures: 0.7mm
- Nozzle caliber for water measures: 1.4mm
- Weight: 145g
- Operation: Gun head can be 134° high temperature disinfected. Nozzle can be rotated 360°. Quick-connect design for convenient operation.
- Modes: Two-in-one supragingival and subgingival mode.
- Sandblasting Design: Creative triangle sandblasting design ensures smooth powder flow and prevents blockage. Large central hole for water spray.
- Cleaning: Water spray in three points for thorough cleaning, 360° rotating head.
2.2. Technical Data (AirJet 1 / AirJet 1S)
| Attribute | Value |
|---|---|
| Model | AirJet 1 / AirJet 1S |
| Classification 93/42 EEC | Class IIa |
| Service Pressure to turbine connection: Water | 0.7-0.22MPa, 50-80ml/min |
| Service Pressure to turbine connection: Air | 0.35-0.45MPa (3.5-4.5bar) |
| Net Weight | Approx. 0.16kg |
| Operating Conditions | 10°C-40°C, Relative Humidity: 30%-75%, Air Pressure: 500hPa-1060hPa |
| Storage and transport conditions | -10°C-40°C, Relative Humidity: 10%-95%, Air Pressure: 500hPa-1060hPa |
| Manufacturing Date | See the packing |
| Life time | 7 years |
2.3. Performance Characteristics (CMANTH Dental Air Polisher)
- Suitable for 65μm enamel sandblasting powder
- Working air pressure: 0.2-0.5MPa
- Water pressure: 0.1-0.3MPa
- Air consumption: 10-15L/min
- Water consumption: 50-100mL/min
- Noise: <80dB
- Sandblasting flow: >2g/min
- 120° nozzle for general use
- Tungsten steel nozzle can rotate 360°
- 0.7mm small diameter nozzle for concentrated powder output and remarkable effect
- Precise and reliable sandblasting flow, finger press control
- The whole machine can be autoclaved
- Not easy to block
- Can be used under low pressure conditions (Lowest 45psi)
- Increases adhesion to 400%
- Used to remove tartar and stains on the surface of the teeth or to polish the surface of the teeth
3. Components
The COXO CP-1 Dental Air Polisher consists of the following main components:
- Handpiece
- Powder chamber cap
- Body
- Connector
- Nozzle
- Cap ring
- Cap dome
- Powder chamber seal ring
- Powder chamber
- Air inlet pipe
- Air outlet pipe
- Seal O-ring 1
- Seal O-ring 2
- Long cleaning needle
- Short cleaning needle
- Cleaning spray nozzle
4. Installation and Set Up
⚠️ Before installation, ensure the quick connector or the hose of the dental unit, handpiece, powder chamber, powder chamber cap, air inlet, and outlet pipe are completely dry. A 3-way syringe can be used for drying.
4.1. Connecting to Dental Unit
- Connect the device to the quick connector or dental unit hose.
- Connect the handpiece to the body of the device.
4.2. Filling the Powder Chamber
⚠️ Use only powder provided or recommended by VOLDENT. Ensure the powder chamber is absolutely dry.
⚠️ Do not fill beyond the 'MAX' limit. The opening of the air inlet and outlet pipe must not be covered by powder.
⚠️ Clean the threads of the powder chamber before screwing on the cap. Do not shake the device as this could cause the powder to clog the tubes.
5. Operating Instructions
⚠️ Use only dry, clean, and oil-free air.
5.1. Setting Water and Air Flow
You can modulate the cleaning and polishing effect by adjusting the water and air pressure:
- Increasing the air pressure increases the cleaning effect and reduces the polishing effect.
- Increasing the water flow rate increases the polishing effect and reduces the cleaning effect.
Set the water flow rate to obtain a uniform spray before using the unit for the first time while the powder chamber is empty.
5.2. Patient and Operator Protection
- Patients should wear goggles at all times to prevent powder from entering the eyes.
- Operators should wear goggles and protective masks throughout the process to prevent bacteria, viruses, or sandblasting powder from being inhaled.
5.3. Treatment Procedure
- Place absorbent cotton rolls under the patient's lips to prevent powder from entering saliva and to protect gums.
- Use the high-speed evacuator of your dental unit to evacuate the air and powder mixture deviated by the treated tooth. The evacuator must be handled by the same operator.
- Do not direct the nozzle directly towards the surface of the teeth. Maintain a distance of 3 to 5mm. Vary the angle between the nozzle and tooth from 30 to 60 degrees.
- After treatment, polish the teeth surfaces by setting the water flow rate to the maximum.
⚠️ The air powder jet is powerful. It can cause injury to the gums or an emphysema caused by the introduction of air into soft tissue spaces. Do not direct the nozzle directly at the gum tissue or into the gingival sulcus.
⚠️ After treatment, the keratin and protein layer on the tooth surface are completely removed. The teeth do not have any natural protection with respect to coloring. Advise patients that for 2 to 3 hours following treatment, they should neither smoke nor consume food or drinks which could strongly color the teeth (e.g., tea, coffee).
6. Maintenance
6.1. Daily Cleaning and Disinfecting
- Unscrew the chamber cap and empty the powder chamber.
- Keep operating the unit after shutting off the water supply. Use compressed air to blow out any remaining water and powder in the pipe.
- Blow out any remaining powder in the chamber and connector with dry compressed air.
- Remove the handpiece, connect it to the cleaning spray nozzle, then connect the nozzle to the hose of the dental unit. Keep the air running for at least 10 seconds to thoroughly clean the handpiece tubes.
- Wash the powder chamber cap, cap ring, cap dome, and handpiece with distilled or deionized water.
- Clean the threads of the powder chamber, body surface, and handpiece surface with alcohol (ethanol, isopropanol).
- Enclose the handpiece in a sterilization bag and seal the bag.
- Sterilize the handpiece at up to 135°C (121°C for 20 minutes or 132°C for 15 minutes) in an autoclave.
⚠️ Only the handpiece is autoclavable under high temperature and pressure.
⚠️ Ensure no chemical liquid is attached to the handpiece surface, and the handpiece is completely dry before sterilization.
⚠️ Do not clean or scour this device with high-acidity cleanser or disinfectant.
⚠️ Submerge only the handpiece in a disinfectant bath. Do not submerge the entire unit.
⚠️ The device must be sterilized before the first use and after each patient use.
⚠️ Clean up within 30 minutes after treatment. High temperature and high pressure sterilization treatment should take place within 2 hours after cleaning.
6.2. Periodic Maintenance
- Clean the pipes of the device with long and short cleaning needles at the end of each working day.
- The powder chamber seal ring, Seal O-ring 1, and Seal O-ring 2 need to be replaced every 6 months.
- Ensure the handpiece is cleaned by hot water ultrasonic cleaning once a week to avoid clogging.
7. Troubleshooting
| Type of Problem | Solutions |
|---|---|
| No powder/air jet coming from the unit |
The interior of the handpiece is clogged:
The handpiece is clogged:
|
| Water comes into the powder chamber | Check the connector of the dental unit hose and quick connector. Replace the seal ring if necessary. |
| Water leakage from the connection between handpiece and body | Replace Seal O-ring 1 and O-ring 2. |
| Air or powder leakage from the chamber cap | Check the seal and the cleanliness of the thread on the powder chamber and on the cap. Replace the seal O-ring if necessary. |
| Weak air and powder jet, low cleaning efficiency | Refill or change the powder. Clean the handpiece and all the relevant tubes. |
In case of any fault that cannot be solved by the above methods, please contact your dealer or the nearest service center.
8. Precautions and Contra-indications
⚠️ Please read the following warnings to avoid potential injury to people or damage to the device.
8.1. General Precautions
- This product must be used only by trained and qualified personnel.
- This device can only be used by dentists or professional operators.
- During operation, the operator should wear protective glasses and protective masks throughout the process. Patients should also wear protective glasses.
- It is recommended to use a suction device to remove excess powder during operation.
- If powder accidentally gets into the eyes, rinse immediately with plenty of water and consult an ophthalmologist.
- Do not aim the nozzle directly at fillings, crowns, or dentures, as this may cause damage to these restorations.
8.2. Contra-indications (Do not use this device on the following patients)
- Patients suffering from chronic bronchitis or asthma, as the jet of air and powder could cause respiratory difficulties.
- Patients with respiratory illness.
- Patients with injury in mucous membranes of more than 3mm pocket depth. (Long-term jetting near soft tissues or salivary glands could introduce powder into the injury, increasing the risk of emphysema).
- Patients with grave ulcers or mucosal inflammation in the mouth or digestive organs.
- Patients with pulmonary ventilation disorder or liver dysfunction or heart dysfunction.
- Patients with allergies.
- Patients who wear contact lenses.
- Patients who need to limit sodium intake (e.g., Hypernatremia, Poisoned during pregnancy, Kidney deficiency, Chronic respiratory disease, Chronic diarrhea).
9. User Tips
- For optimal performance, always ensure the powder chamber and all connections are completely dry before use.
- Regularly check and replace O-rings and seals to prevent leakage and maintain efficient operation.
- Utilize the 360° rotatable nozzle to reach all tooth surfaces effectively, including the tongue, lip, and adjacent gaps.
10. Warranty
The warranty for this product is valid for 1 year from the date of purchase. Damages due to non-adherence to the operation instructions or wear and tear of parts are excluded from warranty.
⚠️ The warranty of your product will be canceled if you attempt to open it.
Manufacturer Information:
VOLDENT MEDICAL TECHNOLOGY CO., LTD.Add: No. 8 Information Industrial Park, National High-Tech Zone, Guilin, 541004, P.R.China
Tel: 0773-6751115
E-mail: voident.china@hotmail.com
Manufacturing Date: See the labels on the package
Validity Period: 10 years
11. Packing List
The product package typically includes:
- 1 x AirJet 1 / AirJet 1S main unit
- 1 x Powder chamber seal ring
- 1 x Seal O-ring 1
- 1 x Seal O-ring 2
- 1 x Long cleaning needle
- 1 x Short cleaning needle
- 1 x Cleaning spray nozzle