COXO CP-1 Dental Air Polisher User Manual

Model: CP-1

1. Introduction

This manual provides detailed instructions for the safe and effective use, maintenance, and troubleshooting of the COXO CP-1 Dental Air Polisher. The CP-1 is a versatile 2-in-1 prophy jet designed for both supragingival and subgingival applications, utilizing air flow for efficient teeth polishing and cleaning.

COXO CP-1 Dental Air Polisher, Supragingival 4-hole model
Figure 1: COXO CP-1 Dental Air Polisher (Supragingival 4-hole model)

2. Safety Precautions

⚠ Warning: Read the following information carefully before using this product.

  • This instrument is intended for use by professional health care providers only, in hospital environments, clinics, or dental offices.
  • In case of failure or obvious aging/abnormal phenomena, stop using the product and contact professionals for maintenance.
  • Regular testing and maintenance are crucial. Do not use if testing and maintenance are not performed.
  • Use only a connector that matches the sandblasting scaler's connection part to avoid failure or bursting due to compressed air pressure.
  • Confirm that the sandblasting scaler is securely connected to the handpiece tube before use.
  • Do not screw the nuts on the connections while the sandblasting scaler is in operation.
  • Ensure adequate cooling water is supplied to cool the nozzle and tooth during cleaning to prevent damage to the tooth surface.
  • Always be careful to prevent scattered powder from entering the patient's eyes and nose.
  • Shake the device before use to eliminate powder agglomeration.
  • Clean the sandblasted tooth powder left in the equipment powder box after each use.
  • Tighten the bottle cap after use to prevent powder deliquescence. Do not store in high temperature and high humidity environments.
  • The subgingival nozzle is provided in a non-sterile state for single use and must be cleaned with alcohol prior to use.
  • Use evacuating suction equipment to absorb air or sand powder mixture ejected during treatment.
  • Do not use the subgingival handpiece with supragingival sand, as this can cause the sandblasting pipe to be blocked.
  • Do not spray the powder directly into the soft tissues of the mouth, gums, or under the tongue.
  • Wear a mask, gloves, and protective glasses during treatment to avoid infection.
  • Rinse the patient frequently to prevent swallowing the powder.
  • To prevent water backflow into the powder box, first vent and then drain the water during operation.

3. Intended Use

The COXO CP-1 Dental Air Polisher is designed for removing tartar, lentigo, or polishing tooth surfaces adhered to the tooth.

4. Contraindications

Do not use this device on patients with:

  • Hemophilia.
  • Allergy to sandblasted tooth cleaning powder.
  • Oral mucosal inflammation or gingival bleeding.
  • Asthma, chronic bronchitis, or other respiratory diseases.

Use with caution for pregnant women and children.

5. Product Structure

The COXO CP-1 features a robust metal construction and an ergonomic design for comfortable operation. It includes interchangeable nozzles for supragingival and subgingival cleaning.

COXO CP-1 Dental Air Polisher, Two-in-one 4-hole model, showing both supragingival and subgingival nozzles
Figure 2: Two-in-one design showing both supragingival (Prophy) and subgingival (Perio) nozzles.
COXO CP-1 Dental Air Polisher, showing 360-degree rotatable head and mode selection
Figure 3: The device features a 360-degree rotatable head for flexible operation and mode selection.
COXO CP-1 Dental Air Polisher, demonstrating ergonomic design in hand
Figure 4: Ergonomic design for comfortable and easy handling.

6. Technical Specifications

Feature Specification
Brand Name YUSENDENT (COXO)
Model Number COXO CP-1
Material Metal
Hand Instrument Yes
Commodity Quality Certification CE
Input Air Pressure 250-450 Kpa
Input Water Pressure 70-220 Kpa
Water Flow >30ml/min
Nozzle Caliber for Power Measuring Instrument 0.7mm
Nozzle Caliber for Water Meters 1.4mm
Weight 0.145kg (145g)
Gas Flow <40NL/min
Noise <68dB

Operating Environment and Storage Conditions

Condition Temperature Humidity Atmospheric Pressure
Operating +5°C to +40°C 20% RH - 80% RH 86kPa - 106kPa
Storage -10°C to +55°C <93% RH 50kPa - 106kPa

7. Setup and Installation

7.1 Attaching the Unit to Handpiece Hose

  1. Align the hose connector of the sandblasting scaler with the handpiece hose interface of the comprehensive treatment table.
  2. Insert and tighten the nut.

⚠ Caution: Fully dry the hose interface of the sandblasting scaler and the handpiece before connecting to prevent nozzle blockage due to damp sand powder.

7.2 Disassembling the Handpiece

  1. Connection: Insert handpiece alignment into the body unit until it clicks.
  2. Disconnecting: Pull it straight out.

⚠ Caution: Thoroughly dry the interior of the handpiece and connection site prior to connection to prevent nozzle blockage due to damp sand powder during use.

7.3 Subgingival Nozzle Disassembly and Assembly

  1. Installing: When installing the subgingival nozzle, the head needs to be oriented to the nozzle D-shaped head.
  2. Disassembling: Use the disassembling wrench to hook up the nozzle and lock it down, then force it outward to pry it out.
COXO CP-1 Subgingival Nozzle
Figure 5: Subgingival Nozzle for specialized cleaning.

8. Operation

8.1 Powder Loading

  1. Unscrew and remove the upper cover of the powder bottle.
  2. Add powder to the bottle. The maximum amount of powder loaded shall not exceed the MAX scale line.
  3. Clean up any powder at the thread position of the powder bottle cover, and then tighten the upper cover.
COXO CP-1 Sandblasting design with powder and water flow
Figure 6: Illustration of the sandblasting design, showing powder and water flow.

8.2 Preparation for Use

  1. Adjust the corresponding pressure regulating valve of the dental unit to the appropriate working pressure: 250kPa - 450kPa.
  2. Adjust the water to a suitable atomization state.

⚠ Caution: Ensure that the dental air polisher is properly installed.

8.3 Sandblasting

8.3.1 Supragingival Sandblasting

  1. Use the supragingival handpiece to spray in an extraoral container for 1-3 seconds in advance to ensure that the air and water are sprayed evenly before treating the patient.
  2. Usually hold the head in a pen holding position.
  3. When performing normal sandblasting and tooth cleaning, maintain a distance of 3-5 mm between the air outlet of the nozzle and the tooth surface. The recommended sandblasting direction is at an included angle of 30-60° with the tooth surface.

8.3.2 Subgingival Blasting

  1. Refer to section 7.3 for nozzle installation.
  2. Use the subgingival handpiece to spray in an extraoral container for 1-3 seconds in advance, ensuring even air and water spray.
  3. Usually hold the head in a pen holding position.
  4. When performing normal subgingival sandblasting, use a nozzle for plaque removal from periodontal pockets (4-9 mm deep) and lift up and down during use.
  5. It is recommended that each periodontal pocket be sandblasted for no more than 5 seconds.
Diagrams showing supragingival and subgingival blasting angles and distances
Figure 7: Diagrams illustrating recommended angles and distances for supragingival (top) and subgingival (bottom) blasting.

9. Maintenance and Care

9.1 Cleaning the Powder

  1. Open the powder cavity cover and clean up any remaining powder.
  2. Use air to blow off any powder remaining in the powder box.
  3. Use an alcohol cotton swab or dry towel to wipe off powder residue on the surface of the sandblasting scaler.

9.2 Cleaning the Nozzle and Handle

  1. Remove the handpiece, unscrew the nozzle, and pull out the nozzle.
  2. Use an air gun to aim at the nozzle and blow off powder in the nozzle.
  3. Use an air gun to blow the front and rear ends of the handpiece to clean residual moisture.
  4. If the nozzle is blocked, use a needle to unclog it.

9.3 Daily Maintenance

Pay attention to daily inspection of the handpiece for cracking, deformation, or falling off. Ensure the sealing ring is complete and clean to maintain normal operation.

  • Keep the sandblasting scaler dry. Rain, moisture, and various liquids can dampen the powder and affect its use.
  • Do not throw, beat, or vibrate the sandblasting scaler, as this will damage its strength and sealing.
  • Do not use paint to smear the instrument, as smearing will leave debris in detachable parts.
  • Do not use a needle to poke the connector end of the tail wire, as this may affect the normal use of the device.

9.4 Sterilization

Sterilizable components (handpiece) should be sterilized using high temperature and high pressure sterilization methods at 134°C for 5 minutes. It is recommended to use sterilization bags that meet national regulations for packaging before sterilization.

⚠ Caution: Except for the handpiece, other components cannot be sterilized under high temperature and high pressure.

10. Troubleshooting

Fault Phenomenon Failure Cause Resolution
Water leakage at O-ring connection Damaged O-ring Replace the O-ring
No air or powder from nozzle Clogged Nozzle In the absence of sandblasted powder in the sand powder bottle, remove the handpiece and clean the nozzle with a needle.
Clogged powder thru hole Clean powder thru hole in sand powder bottle with needle.
Too little powder and air ejection Feed water of sand powder bottle Replace the O-ring on the upper cover of sand powder bottle.
Air path in handpiece is blocked Unloading the handpiece without sand powder bottle and cleaning the air passage of the handpiece with a needle.

11. User Tips

No specific user tips are available at this time.

12. Warranty and Support

The COXO CP-1 Dental Air Polisher comes with a 12-month warranty.

For repairs or any issues with structural components, please contact the after-sales service department of Foshan COXO Medical Instrument Co., Ltd.

Foshan COXO Medical Instrument Co., Ltd
BLDG 4, District A, Guangdong New Light Source Industrial Base, South of Luocun Avenue Nanhai District Foshan 528226 Guangdong China
Web: www.coxotec.com

For further details, you may refer to the official user manual (PDF) available at: User manual (PDF)

13. Recycling and Disposal

Dispose of waste equipment in accordance with national regulations and standards to ensure that all parts are not contaminated during the disposal process.

14. Statement of Compliance

Disclaimer

⚠ This item may be intended for use by professional health care providers only. The seller may ask you to provide evidence that you are a qualified health care provider and are entitled to buy and use certain items in your country.

© 2024 COXO. All rights reserved.