Morayelec ME261P

Contents

Moray ME261P Digital Pressure Gauge

Instruction Manual

Model: ME261P | Brand: Morayelec

1. Overview

The Moray ME261P is a high-precision digital pressure gauge designed for real-time monitoring and measurement calibration in various field applications. It utilizes high-precision pressure sensitive elements, a 24-bit AD processing chip, and an industrial MCU for accurate and reliable performance. Suitable for precision measurement of hydraulic, oil, water, and air pressure, it delivers laboratory-grade accuracy for industrial and professional applications.

Moray ME261P Digital Pressure Gauge main view

Figure 1: Moray ME261P Digital Pressure Gauge

Video 1: ME261P Digital Pressure Gauge Demonstration

2. Safety Information

Please read all safety instructions carefully before operating the device to ensure safe and efficient use.

  • Warning! Indicates a potential hazard which, if not avoided, could result in serious loss or injury.
  • Caution! Indicates a potential hazard which, if not avoided, could result in minor injury, property damage, or environmental hazards.
  • Information! Provides useful tips, suggestions, and information for efficient and fault-free operations.

General Safety Precautions:

  • Before using the pressure gauge, the operator must carefully read and understand these operating instructions.
  • The digital pressure gauge must be used within its specified scope of application and comply with all relevant accident prevention regulations and general safety regulations.
  • During disassembly, repair, or disposal, be aware that residual medium in the digital pressure gauge can pose a danger to people, the environment, and equipment. Take adequate precautionary measures.
  • Before storing the digital pressure gauge, ensure all remaining media are removed. Some media can be harmful (e.g., corrosive, toxic, radioactive).

3. Setup

3.1 Battery Installation

The ME261P is powered by two AA batteries. Low power consumption allows for 6-10 months of continuous use without needing to shut down the device.

  1. Locate the battery compartment cover on the rear of the device.
  2. Open the cover by sliding or lifting it.
  3. Insert two AA batteries, ensuring correct polarity (+/-) as indicated inside the compartment.
  4. Close the battery compartment cover securely.
Battery compartment with AA batteries

Figure 2: Battery Installation

3.2 Installation of Pressure Gauge

The device features a standard threaded interface for connection to your system. Common options include M20*1.5, M14*1.5, G1/2, G1/4, and NPT1-4.

  1. Select a proper sealing method (e.g., PTFE tape) before installing the digital pressure gauge to prevent leaks.
  2. Ensure that the measuring medium (air, water, oil) is compatible with the instrument's wetted parts material (stainless steel 17-4PH).
  3. Carefully thread the gauge into the pressure port. Caution: Do not touch the sensitive element diaphragm in any way.
  4. Tighten the gauge using a wrench on the hexagonal fitting. Caution: Do not install the instrument by rotating the instrument housing, as this may damage internal components.
Diagram of pressure gauge parts and functions

Figure 3: Pressure Gauge Components and Interface

4. Operating Instructions

4.1 Power On/Off and Backlight

  • Power On/Off: Long press the ON/OFF button to turn the device on or off.
  • Backlight: Short press the ON/OFF button to turn the LCD backlight on or off. The high-definition backlight ensures clear readings in various lighting conditions.

4.2 Zero Clearing

  • Function: Long press the ZERO button to set the current pressure as the measurement zero point.
  • Range: The zero clearing range is within 5% of the full scale.
  • Recommendation: It is recommended to use the zero clearing function before each critical measurement for optimal accuracy.

4.3 Unit Switching

  • Function: Short press the UNIT button to cycle through available pressure units.
  • Available Units: psi, kPa, MPa, bar, mbar, kgf/cm², Pa, m.
  • Display: After unit switching, the measured value will automatically display the current equivalent pressure in the selected unit.

4.4 Peak View/Clear

  • Peak View: Short press the PEAK button to switch between displaying the maximum and minimum pressure values recorded.
  • Peak Clear: Long press the PEAK button to clear the recorded maximum or minimum pressure values.

5. Maintenance

5.1 Battery Replacement

Refer to section 3.1 for detailed instructions on battery installation. Always turn off the pressure gauge before replacing batteries.

5.2 Cleaning

Clean the exterior of the device with a soft, dry cloth. Do not use abrasive cleaners, solvents, or immerse the device in liquids.

5.3 Storage

Store the device in a dry place with the following conditions:

  • Temperature: -20°C to +60°C
  • Humidity: <85% relative humidity (non-condensing)

Avoid the following conditions during storage:

  • Direct sunlight or proximity to hot objects.
  • Mechanical vibration or impact.
  • Exposure to smoke, steam, dust, or corrosive gases.
  • Potentially explosive or flammable environments.

6. Troubleshooting

The ME261P digital pressure gauge provides fault codes to assist in diagnosing issues:

  • Er-H: If the display shows 'Er-H', it indicates the pressure exceeds the upper measuring range. Check if the measuring range of the pressure gauge is appropriate for your application.
  • Er-S: If the display shows 'Er-S', it indicates the pressure sensitive element has failed. The device requires professional service; return it to the manufacturer for maintenance.
  • Er-D: If the display shows 'Er-D', it indicates data loss. The device requires professional service; return it to the manufacturer or customer for calibration.
  • Er-Z: If Zero clearing fails ('Er-Z'), ensure the pressure is stable and within the 5% full-scale clearing range. Re-attempt the zero clearing procedure.

7. Specifications

Dimensions diagram of Moray ME261P

Figure 4: ME261P Dimensions

FeatureSpecification
Model NumberME261P
Brand NameMorayelec
Measuring Range-0.1 ~ 100 MPa
Pressure Accuracy±0.5% FS
Display TypeDIGITAL, LCD with backlight, 5-digit
Resolution5 bit
Overpressure2XFS
Shell MaterialStainless Steel (304)
Protection LevelIP54
Power Supply2 x AAA batteries
Operating Temperature-10 ℃ ~ +60 ℃
Threaded InterfaceNPT1-2, M20-1.5, G1-4, G1-2, NPT1-4
Medium TypeAir, water, oil
Pressure TypeGauge pressure (standard), absolute pressure or sealed gauge pressure
Unit Switchingpsi, kPa, MPa, bar, mbar, kgf/cm², Pa, m
Dial Diameter80mm (2 - 3.9 Inches)

8. User Tips

  • Always ensure the pressure gauge is properly sealed to prevent leaks, especially when dealing with liquids or high-pressure gases.
  • Regularly check battery life to ensure continuous operation and accurate readings. Replace batteries promptly when the low battery indicator appears.
  • Utilize the zero-clearing function before critical measurements for optimal accuracy, especially if the gauge has been moved or subjected to temperature changes.

9. Warranty and Support

The digital pressure gauge is designed for reliable operation and is generally maintenance-free under normal conditions. For recalibration, repairs, or any technical support, please contact the manufacturer or your authorized dealer.

The manufacturer is not liable for any damage caused by the use of products that do not conform to the intended use, non-compliance with operating instructions, assignment of unqualified technicians, or repair by unauthorized persons.

For further assistance, please refer to the official user manual PDF: View User Manual (PDF)

Documents and downloads

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.