OWON HDS200 Series

Rate this product if you’ve used it
Contents

OWON HDS200 Series Handheld Digital Oscilloscope User Manual

Models: HDS242, HDS242S, HDS272, HDS272S, HDS2102, HDS2102S, HDS2202, HDS2202S

1. Introduction

The OWON HDS200 Series Handheld Digital Oscilloscope is a versatile 3-in-1 instrument combining an oscilloscope, a multimeter, and a waveform generator (available on 'S' models). Designed for outdoor use, rapid on-site measurement, and automobile maintenance, it features a 3.5-inch high-resolution color LCD display and supports PC software connection via a USB Type-C interface. This manual provides essential information for safe and effective use of your device.

OWON HDS242 Handheld Oscilloscope with kickstand, displaying a waveform.
Figure 1.1: OWON HDS242 Handheld Oscilloscope with kickstand.

2. Product Overview

2.1 Key Features

  • 3-in-1 Functionality: Oscilloscope, Multimeter, and Waveform Generator (available on 'S' models).
  • Display: 3.5-inch high-resolution, high-contrast color LCD display.
  • Power: 18650 lithium battery, providing ≥4 hours working time. Comprehensive power consumption ≤3W.
  • Connectivity: USB Type-C interface, supports power bank and PC software connection.
  • Usability: Self-calibration function, SCPI supported for secondary development.
  • Design: Integrated encapsulation design for durability and comfortable hand feel. Hangable bracket for saving working space.
OWON HDS2102S Handheld Oscilloscope with detailed function breakdown.
Figure 2.1: OWON HDS200 Series functions overview.

2.2 Device Layout and Dimensions

The device features a compact design with clearly labeled inputs and controls.

Diagram showing the physical dimensions and key components of the OWON HDS200 Series.
Figure 2.2: Device dimensions and component layout.
  • Dimensions: Approximately 198mm (height) x 96mm (width) x 38mm (depth).
  • Weight: Approx. 0.6 kg (main device, without battery).
  • Waveform Generator Output: BNC connector for signal output.
  • Oscilloscope Input: Input terminals for oscilloscope probes.
  • Calibration Signal Output: 1KHz square wave signal output.
  • Charging and Communication Interface: USB Type-C port.
  • Multimeter Input Connectors: Terminals for multimeter leads.
  • Integrated Encapsulation Design: Prevents peeling and ensures long-term use.

3. Setup

3.1 Battery Installation

The device uses a single 18650 lithium battery with a protection board. Ensure the battery size is within 19 mm x 70 mm. The rated voltage and capacity are 3.6 VDC/3.35 Ah (12.06 Wh).

  1. Locate the battery compartment on the back of the device.
  2. Open the battery compartment cover.
  3. Insert the 18650 battery, ensuring correct polarity.
  4. Close the battery compartment cover securely.
Image showing battery compartment and Type-C charging.
Figure 3.1: Battery compartment and Type-C charging port.

3.2 USB Connection

The USB Type-C interface allows for charging and data transmission.

  1. Turn on the device and lift the Type-C interface cover.
  2. Connect the device to a power adapter using a Type-C cable to power on the device or charge the battery.
  3. To transmit data, connect the device to a PC using the Type-C cable.

3.3 App Connection

The device supports connection to a mobile app for firmware upgrades, browsing/saving data, remote control, and more.

  1. Scan the QR code below to download and install the app on your smartphone.
  2. On the device menu, select the option to enable the hotspot.
  3. Turn on WLAN on your phone and connect to the device's hotspot. The hotspot name is typically 'HIK-IPTS Serial No.'
  4. To find the hotspot password, navigate to 'Hotspot' in the device menu.
  5. Open the app on your phone and connect to the device. You can now view the device interface on your phone.
QR codes for Android and iOS app downloads.
Figure 3.2: QR codes for app download (Android and iOS).

Note: If the password is not displayed, please upgrade the device to the latest version or enter the Serial No. as the password.

4. Operation

4.1 Button Functions

The device features several buttons for navigation and control:

ButtonActionFunction
PowerPressStandby mode / Wake up device
HoldPower on / off (Power indicator stays solid red when device is on)
CapturePress (Non-Menu Mode)Capture snapshots
Hold (Non-Menu Mode)Start / stop recording videos
MenuPress (Non-Menu Mode)Switch palettes
Hold (Non-Menu Mode)Enter menu
Press (Menu Mode)Confirm / Set parameters
Hold (Menu Mode)Save and exit menu
ZoomPress (Non-Menu Mode)Switch digital zoom
Hold (Non-Menu Mode)Correct non-uniformity of display (FFC)
Up/Down ArrowsPress (Menu Mode)Move up / down in menu

4.2 Oscilloscope Functions

The oscilloscope function allows for detailed waveform analysis.

  • Two Acquisition Modes: Peak detect is used for the detection of jamming blur and the possibility of reducing the confusion.
  • High Refresh, Deep Storage: Max 10,000 wfrms/s waveform refresh rate, 8K record length, easy to capture exceptional and low probability events.
  • Automatic Measurement: Period, Frequency, Mean, PK-PK, Max, Min, Amplitude.
  • Cursor Measurement: Measures voltage difference (ΔV), time difference (ΔT), and both (ΔT&ΔV) between cursors.
  • Save Function: Save 2 settings, 4 reference waveforms, 4 waveform images, 4 CSV waveform files. Reference waveforms can be displayed on the same screen for convenient comparison.
Image showing excellent oscilloscope performance with two acquisition modes, high refresh/deep storage, 7 kinds of auto measurement, cursor measurement function, and save function.
Figure 4.1: Oscilloscope measurement examples.

4.3 Multimeter Functions

The built-in multimeter provides true RMS measurements with 20,000 counts.

  • Measurements: Voltage, current, resistance, diode, capacitance, and continuity test.
  • Input Impedance: 10MΩ.
  • Max Input Voltage: AC 750V, DC 1000V.
  • Max Input Current: DC 10A, AC 10A.
Image showing 20,000 counts multimeter with True RMS, supporting voltage, current, resistance, diode, capacitance, and continuity test.
Figure 4.2: Multimeter voltage reading example.

4.4 Waveform Generator Functions (Only for 'S' models)

The waveform generator supports various output signals and built-in special waveforms.

  • Frequency Output: Sine (0.1Hz-25MHz), Square (0.1Hz-5MHz), Ramp (0.1Hz-1MHz), Pulse (0.1Hz-5MHz), Arbitrary Waveform (0.1Hz-5MHz).
  • Sample Rate: 125MSa/s.
  • Channel: 1-CH.
  • Amplitude: 20mVpp-5Vpp (high impedance), 10mVpp-2.5Vpp (50Ω).
  • Waveform Length: 8K.
  • Vertical Resolution: 14bits.
  • Load Impedance: 50Ω.
  • Built-in Waveforms: Supports 8 built-in special waveforms, including Sinc, Bessel, BesselJ, StairUp, StairDown, AttAtt, and AmpAtt.
Image showing waveform generator functions (Sine, Sinc) and supporting 8 built-in special waveforms.
Figure 4.3: Waveform Generator interface.

5. Specifications

5.1 Oscilloscope Specifications

ParameterHDS242HDS272HDS2102HDS2202HDS242SHDS272SHDS2102SHDS2202S
Bandwidth40MHz70MHz100MHz200MHz40MHz70MHz100MHz200MHz
Channel2-CH Oscilloscope2-CH Oscilloscope + 1-CH Generator
Sample Rate250MSa/s500MSa/s1GSa/s250MSa/s500MSa/s1GSa/s
Acquisition ModelSample, Peak detect
Record Length8K
Display3.5 inch LCD
Waveform Refresh Rate10000 wfrms/s
Input CouplingDC, AC, Ground
Input Impedance1 MΩ±2%, in parallel with 16 pF±10 pF
Probe Attenuation1X, 10X, 100X, 1000X, 10000X
Max. Input Voltage400V (DC+AC, PK - PK)
Bandwidth Limit (typical)20MHz / Full bandwidth
Sampling Rate Range5ns/div~500s/div, step by 1-2-52ns/div~1000s/div, step by 1-2-52ns/div~1000s/div, step by 1-2-55ns/div~500s/div, step by 1-2-52ns/div~1000s/div, step by 1-2-52ns/div~1000s/div, step by 1-2-5
Sensitivity Resolution10mV/div~10V/div
Low Frequency (-3db)≥10 Hz (at input, AC coupling)
Rise Time (at input, Typical)≤8ns≤5ns≤3.5ns≤1.75ns≤8ns≤5ns≤3.5ns≤1.75ns
Trigger TypeEdge
Trigger ModelAuto, Normal, Single
Trigger Level Range±4 div from the screen center
Automatic MeasurementPeriod, Frequency, Mean, PK-PK, Max, Min, Amplitude
Cursor MeasurementΔV, ΔT, ΔT&ΔV between cursors, auto cursor
Communication InterfaceUSB Type-C
Table of Oscilloscope Specifications for HDS200 Series models.
Figure 5.1: Oscilloscope Specifications.

5.2 Multimeter Specifications

ParameterSpecification
Full scale reading4 digitals (max 20000 counts)
Testing ModesVoltage, Current, Resistance, Capacitance, Diode, Continuity
Maximum Input VoltageAC 750V, DC 1000V
Maximum Input CurrentDC: 10A AC: 10A
Diodes Test0–2V
DC Voltage Measurement Range200.00mV, 2.0000V/20.000V/200.00V, 1000.0V
DC Voltage Accuracy±(0.3%+10 digits) for 200.00mV, ±(0.3%+5 digits) for others
AC Voltage Measurement Range200.00mV/2.0000V/20.000V/200.00V, 750.00V
AC Voltage Accuracy±(0.8%+10 digits) for 200.00mV-200.00V, ±(1%+10% off) for 750.00V
DC Current Measurement Range200.00mA, 10.000A
DC Current Accuracy±(0.8%+10 digits) for 200.00mA, ±(2.5%+10 digits) for 10.000A
AC Current Measurement Range200.00mA, 10.000A
AC Current Accuracy±(1%+10% off) for 200.00mA, ±(2.8%+10 digits) for 10.000A
Resistance (Ω) Measurement Range200.00Ω/2.0000kΩ/20.000kΩ/200.00kΩ/2.0000MΩ, 20.000MΩ, 100.00MΩ
Resistance (Ω) Accuracy±(0.8%+10 digits) for 200.00Ω-2.0000MΩ, ±(1%+3% digits) for 20.000MΩ, ±(5%+10% off) for 100.00MΩ
Capacitance (F) Measurement Range20.000nF/200.00nF/2.0000μF/20.000μF/200.00μF/2.0000mF
Capacitance (F) Accuracy±(3.0%+10% off)
Frequency Response (Hz)(40 - 1000) Hz
Table of Multimeter Specifications.
Figure 5.2: Multimeter Specifications.

5.3 Waveform Generator Specifications (Only for 'S' models)

ParameterWaveform TypeFrequency Range
Frequency OutputSine0.1Hz~25MHz
Square0.1Hz~5MHz
Ramp0.1Hz~1MHz
Pulse0.1Hz~5MHz
Arbitrary Waveform0.1Hz~5MHz
Sample Rate125MSa/s
Channel1
Amplitude20mVpp-5Vpp (high impedance), 10mVpp-2.5Vpp (50Ω)
Waveform Length8K
Vertical Resolution14bits
Load Impedance50Ω
Table of Waveform Generator Specifications.
Figure 5.3: Waveform Generator Specifications.

6. Accessories

The following accessories are typically included with your OWON HDS200 Series device:

Product Accessories image, listing Main device, Bag, Probe, Probe Adjust, USB Cable, Multimeter Leads, Quick Guide, BNC to Alligator Clip.
Figure 6.1: Product Accessories.
  1. Main device
  2. Bag
  3. Probe
  4. Probe Adjust
  5. USB Cable
  6. Multimeter Leads
  7. Quick Guide
  8. BNC to Alligator Clip

7. Maintenance & Safety

7.1 General Safety Instructions

  • Read all safety information carefully before using the device.
  • Use of the product must be in strict compliance with local electrical safety regulations.
  • Do not drop the product or subject it to physical shock. Keep the device away from magnetic interference.
  • Do not expose the device to high electromagnetic radiation or dusty environments.
  • Do not aim the lens at the sun or any other bright light.
  • Place the device in a dry and well-ventilated environment.

7.2 Battery Safety and Storage

  • The external rechargeable battery type is 18650. The rated voltage is 3.6 VDC, and the battery capacity is 3200 mAh.
  • CAUTION: Risk of explosion if the battery is replaced by an incorrect type. Replace with the same or equivalent type only.
  • Batteries of improper size cannot be installed and may cause abnormal shutdown.
  • Dispose of used batteries according to local regulations.
  • Remove the battery if you do not use the device for a long time.
  • For long-term storage, fully charge the battery every six months to ensure its quality. Otherwise, damage may occur.
  • Ensure the battery temperature is between 0°C to 45°C (32°F to 113°F) when charging.
  • Do not dispose of the battery into fire or a hot oven, or mechanically crush or cut the battery, which may may result in an explosion.
  • Do not leave the battery in an extremely high temperature or low air pressure environment, which may result in an explosion or the leakage of flammable liquid or gas.
  • Do not place the device with battery or the battery alone near heating or fire source. Avoid direct sunlight.
  • Do not place the battery in the reach of children. Do not swallow the battery to avoid chemical burns.

7.3 Cleaning

  • If the product does not work properly, contact your dealer or the nearest service center.
  • Wipe the device gently with a clean cloth and a small quantity of ethanol, if necessary.
  • Clean the lens with a soft and dry cloth or wiping paper to avoid scratching it.

8. Troubleshooting

8.1 Emergency Procedures

If smoke, odor, or noise arises from the device, immediately turn off the power, unplug the power cable, and contact the service center.

9. User Tips

  • Outdoor Use: The HDS200 series is designed for outdoor use. Its compact size and robust design make it suitable for field work.
  • Power Management: Utilize the USB Type-C interface for charging with a power bank during extended outdoor sessions.
  • Data Analysis: Connect to PC software for more in-depth analysis and management of saved waveforms and data.
  • Regular Calibration: Perform self-calibration periodically to ensure measurement accuracy.

10. Support & Compliance

10.1 Manufacturer Information

Hangzhou Microimage Software Co., Ltd.
Room 313, Unit B, Building 2, 399 Danfeng Road, Xixing Subdistrict, Binjiang District, Hangzhou, Zhejiang 310052, China

10.2 Compliance Notices

This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment.

FCC Information

This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation.

EU Conformity Statement

This product and, if applicable, the supplied accessories too are marked with "CE" and comply therefore with the applicable harmonized European standards listed under the Directive 2014/30/EU (EMCD), Directive 2014/35/EU (LVD), Directive 2011/65/EU (RoHS), Directive 2014/53/EU.

WEEE Directive (2012/19/EU)

Products marked with this symbol cannot be disposed of as unsorted municipal waste in the European Union. For proper recycling, return this product to your local supplier upon the purchase of equivalent new equipment, or dispose of it at designated collection points. For more information see: www.recyclethis.info.

Battery Directive (2006/66/EC and 2013/56/EU)

This product contains a battery that cannot be disposed of as unsorted municipal waste in the European Union. The battery is marked with this symbol, which may include lettering to indicate cadmium (Cd), lead (Pb), or mercury (Hg). For proper recycling, return the battery to your supplier or to a designated collection point. For more information see: www.recyclethis.info.

10.3 User Manual (PDF)

For more detailed operations and comprehensive information, please refer to the full user manual available as a PDF document. You can view the PDF at: OWON HDS200 Series User Manual (PDF).

© 2023 OWON. All rights reserved.


Documents and downloads

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.

Was this page helpful?