1. Introduction
The OWON HDS200 Series Handheld Digital Oscilloscope is a versatile 3-in-1 instrument combining an oscilloscope, a multimeter, and a waveform generator (available on 'S' models). Designed for outdoor use, rapid on-site measurement, and automobile maintenance, it features a 3.5-inch high-resolution color LCD display and supports PC software connection via a USB Type-C interface. This manual provides essential information for safe and effective use of your device.

2. Product Overview
2.1 Key Features
- 3-in-1 Functionality: Oscilloscope, Multimeter, and Waveform Generator (available on 'S' models).
- Display: 3.5-inch high-resolution, high-contrast color LCD display.
- Power: 18650 lithium battery, providing ≥4 hours working time. Comprehensive power consumption ≤3W.
- Connectivity: USB Type-C interface, supports power bank and PC software connection.
- Usability: Self-calibration function, SCPI supported for secondary development.
- Design: Integrated encapsulation design for durability and comfortable hand feel. Hangable bracket for saving working space.

2.2 Device Layout and Dimensions
The device features a compact design with clearly labeled inputs and controls.

- Dimensions: Approximately 198mm (height) x 96mm (width) x 38mm (depth).
- Weight: Approx. 0.6 kg (main device, without battery).
- Waveform Generator Output: BNC connector for signal output.
- Oscilloscope Input: Input terminals for oscilloscope probes.
- Calibration Signal Output: 1KHz square wave signal output.
- Charging and Communication Interface: USB Type-C port.
- Multimeter Input Connectors: Terminals for multimeter leads.
- Integrated Encapsulation Design: Prevents peeling and ensures long-term use.
3. Setup
3.1 Battery Installation
The device uses a single 18650 lithium battery with a protection board. Ensure the battery size is within 19 mm x 70 mm. The rated voltage and capacity are 3.6 VDC/3.35 Ah (12.06 Wh).
- Locate the battery compartment on the back of the device.
- Open the battery compartment cover.
- Insert the 18650 battery, ensuring correct polarity.
- Close the battery compartment cover securely.

3.2 USB Connection
The USB Type-C interface allows for charging and data transmission.
- Turn on the device and lift the Type-C interface cover.
- Connect the device to a power adapter using a Type-C cable to power on the device or charge the battery.
- To transmit data, connect the device to a PC using the Type-C cable.
3.3 App Connection
The device supports connection to a mobile app for firmware upgrades, browsing/saving data, remote control, and more.
- Scan the QR code below to download and install the app on your smartphone.
- On the device menu, select the option to enable the hotspot.
- Turn on WLAN on your phone and connect to the device's hotspot. The hotspot name is typically 'HIK-IPTS Serial No.'
- To find the hotspot password, navigate to 'Hotspot' in the device menu.
- Open the app on your phone and connect to the device. You can now view the device interface on your phone.

Note: If the password is not displayed, please upgrade the device to the latest version or enter the Serial No. as the password.
4. Operation
4.1 Button Functions
The device features several buttons for navigation and control:
| Button | Action | Function |
|---|---|---|
| Power | Press | Standby mode / Wake up device |
| Hold | Power on / off (Power indicator stays solid red when device is on) | |
| Capture | Press (Non-Menu Mode) | Capture snapshots |
| Hold (Non-Menu Mode) | Start / stop recording videos | |
| Menu | Press (Non-Menu Mode) | Switch palettes |
| Hold (Non-Menu Mode) | Enter menu | |
| Press (Menu Mode) | Confirm / Set parameters | |
| Hold (Menu Mode) | Save and exit menu | |
| Zoom | Press (Non-Menu Mode) | Switch digital zoom |
| Hold (Non-Menu Mode) | Correct non-uniformity of display (FFC) | |
| Up/Down Arrows | Press (Menu Mode) | Move up / down in menu |
4.2 Oscilloscope Functions
The oscilloscope function allows for detailed waveform analysis.
- Two Acquisition Modes: Peak detect is used for the detection of jamming blur and the possibility of reducing the confusion.
- High Refresh, Deep Storage: Max 10,000 wfrms/s waveform refresh rate, 8K record length, easy to capture exceptional and low probability events.
- Automatic Measurement: Period, Frequency, Mean, PK-PK, Max, Min, Amplitude.
- Cursor Measurement: Measures voltage difference (ΔV), time difference (ΔT), and both (ΔT&ΔV) between cursors.
- Save Function: Save 2 settings, 4 reference waveforms, 4 waveform images, 4 CSV waveform files. Reference waveforms can be displayed on the same screen for convenient comparison.

4.3 Multimeter Functions
The built-in multimeter provides true RMS measurements with 20,000 counts.
- Measurements: Voltage, current, resistance, diode, capacitance, and continuity test.
- Input Impedance: 10MΩ.
- Max Input Voltage: AC 750V, DC 1000V.
- Max Input Current: DC 10A, AC 10A.

4.4 Waveform Generator Functions (Only for 'S' models)
The waveform generator supports various output signals and built-in special waveforms.
- Frequency Output: Sine (0.1Hz-25MHz), Square (0.1Hz-5MHz), Ramp (0.1Hz-1MHz), Pulse (0.1Hz-5MHz), Arbitrary Waveform (0.1Hz-5MHz).
- Sample Rate: 125MSa/s.
- Channel: 1-CH.
- Amplitude: 20mVpp-5Vpp (high impedance), 10mVpp-2.5Vpp (50Ω).
- Waveform Length: 8K.
- Vertical Resolution: 14bits.
- Load Impedance: 50Ω.
- Built-in Waveforms: Supports 8 built-in special waveforms, including Sinc, Bessel, BesselJ, StairUp, StairDown, AttAtt, and AmpAtt.

5. Specifications
5.1 Oscilloscope Specifications
| Parameter | HDS242 | HDS272 | HDS2102 | HDS2202 | HDS242S | HDS272S | HDS2102S | HDS2202S |
|---|---|---|---|---|---|---|---|---|
| Bandwidth | 40MHz | 70MHz | 100MHz | 200MHz | 40MHz | 70MHz | 100MHz | 200MHz |
| Channel | 2-CH Oscilloscope | 2-CH Oscilloscope + 1-CH Generator | ||||||
| Sample Rate | 250MSa/s | 500MSa/s | 1GSa/s | 250MSa/s | 500MSa/s | 1GSa/s | ||
| Acquisition Model | Sample, Peak detect | |||||||
| Record Length | 8K | |||||||
| Display | 3.5 inch LCD | |||||||
| Waveform Refresh Rate | 10000 wfrms/s | |||||||
| Input Coupling | DC, AC, Ground | |||||||
| Input Impedance | 1 MΩ±2%, in parallel with 16 pF±10 pF | |||||||
| Probe Attenuation | 1X, 10X, 100X, 1000X, 10000X | |||||||
| Max. Input Voltage | 400V (DC+AC, PK - PK) | |||||||
| Bandwidth Limit (typical) | 20MHz / Full bandwidth | |||||||
| Sampling Rate Range | 5ns/div~500s/div, step by 1-2-5 | 2ns/div~1000s/div, step by 1-2-5 | 2ns/div~1000s/div, step by 1-2-5 | 5ns/div~500s/div, step by 1-2-5 | 2ns/div~1000s/div, step by 1-2-5 | 2ns/div~1000s/div, step by 1-2-5 | ||
| Sensitivity Resolution | 10mV/div~10V/div | |||||||
| Low Frequency (-3db) | ≥10 Hz (at input, AC coupling) | |||||||
| Rise Time (at input, Typical) | ≤8ns | ≤5ns | ≤3.5ns | ≤1.75ns | ≤8ns | ≤5ns | ≤3.5ns | ≤1.75ns |
| Trigger Type | Edge | |||||||
| Trigger Model | Auto, Normal, Single | |||||||
| Trigger Level Range | ±4 div from the screen center | |||||||
| Automatic Measurement | Period, Frequency, Mean, PK-PK, Max, Min, Amplitude | |||||||
| Cursor Measurement | ΔV, ΔT, ΔT&ΔV between cursors, auto cursor | |||||||
| Communication Interface | USB Type-C | |||||||

5.2 Multimeter Specifications
| Parameter | Specification |
|---|---|
| Full scale reading | 4 digitals (max 20000 counts) |
| Testing Modes | Voltage, Current, Resistance, Capacitance, Diode, Continuity |
| Maximum Input Voltage | AC 750V, DC 1000V |
| Maximum Input Current | DC: 10A AC: 10A |
| Diodes Test | 0–2V |
| DC Voltage Measurement Range | 200.00mV, 2.0000V/20.000V/200.00V, 1000.0V |
| DC Voltage Accuracy | ±(0.3%+10 digits) for 200.00mV, ±(0.3%+5 digits) for others |
| AC Voltage Measurement Range | 200.00mV/2.0000V/20.000V/200.00V, 750.00V |
| AC Voltage Accuracy | ±(0.8%+10 digits) for 200.00mV-200.00V, ±(1%+10% off) for 750.00V |
| DC Current Measurement Range | 200.00mA, 10.000A |
| DC Current Accuracy | ±(0.8%+10 digits) for 200.00mA, ±(2.5%+10 digits) for 10.000A |
| AC Current Measurement Range | 200.00mA, 10.000A |
| AC Current Accuracy | ±(1%+10% off) for 200.00mA, ±(2.8%+10 digits) for 10.000A |
| Resistance (Ω) Measurement Range | 200.00Ω/2.0000kΩ/20.000kΩ/200.00kΩ/2.0000MΩ, 20.000MΩ, 100.00MΩ |
| Resistance (Ω) Accuracy | ±(0.8%+10 digits) for 200.00Ω-2.0000MΩ, ±(1%+3% digits) for 20.000MΩ, ±(5%+10% off) for 100.00MΩ |
| Capacitance (F) Measurement Range | 20.000nF/200.00nF/2.0000μF/20.000μF/200.00μF/2.0000mF |
| Capacitance (F) Accuracy | ±(3.0%+10% off) |
| Frequency Response (Hz) | (40 - 1000) Hz |

5.3 Waveform Generator Specifications (Only for 'S' models)
| Parameter | Waveform Type | Frequency Range |
|---|---|---|
| Frequency Output | Sine | 0.1Hz~25MHz |
| Square | 0.1Hz~5MHz | |
| Ramp | 0.1Hz~1MHz | |
| Pulse | 0.1Hz~5MHz | |
| Arbitrary Waveform | 0.1Hz~5MHz | |
| Sample Rate | 125MSa/s | |
| Channel | 1 | |
| Amplitude | 20mVpp-5Vpp (high impedance), 10mVpp-2.5Vpp (50Ω) | |
| Waveform Length | 8K | |
| Vertical Resolution | 14bits | |
| Load Impedance | 50Ω | |

6. Accessories
The following accessories are typically included with your OWON HDS200 Series device:

- Main device
- Bag
- Probe
- Probe Adjust
- USB Cable
- Multimeter Leads
- Quick Guide
- BNC to Alligator Clip
7. Maintenance & Safety
7.1 General Safety Instructions
- Read all safety information carefully before using the device.
- Use of the product must be in strict compliance with local electrical safety regulations.
- Do not drop the product or subject it to physical shock. Keep the device away from magnetic interference.
- Do not expose the device to high electromagnetic radiation or dusty environments.
- Do not aim the lens at the sun or any other bright light.
- Place the device in a dry and well-ventilated environment.
7.2 Battery Safety and Storage
- The external rechargeable battery type is 18650. The rated voltage is 3.6 VDC, and the battery capacity is 3200 mAh.
- CAUTION: Risk of explosion if the battery is replaced by an incorrect type. Replace with the same or equivalent type only.
- Batteries of improper size cannot be installed and may cause abnormal shutdown.
- Dispose of used batteries according to local regulations.
- Remove the battery if you do not use the device for a long time.
- For long-term storage, fully charge the battery every six months to ensure its quality. Otherwise, damage may occur.
- Ensure the battery temperature is between 0°C to 45°C (32°F to 113°F) when charging.
- Do not dispose of the battery into fire or a hot oven, or mechanically crush or cut the battery, which may may result in an explosion.
- Do not leave the battery in an extremely high temperature or low air pressure environment, which may result in an explosion or the leakage of flammable liquid or gas.
- Do not place the device with battery or the battery alone near heating or fire source. Avoid direct sunlight.
- Do not place the battery in the reach of children. Do not swallow the battery to avoid chemical burns.
7.3 Cleaning
- If the product does not work properly, contact your dealer or the nearest service center.
- Wipe the device gently with a clean cloth and a small quantity of ethanol, if necessary.
- Clean the lens with a soft and dry cloth or wiping paper to avoid scratching it.
8. Troubleshooting
8.1 Emergency Procedures
If smoke, odor, or noise arises from the device, immediately turn off the power, unplug the power cable, and contact the service center.
9. User Tips
- Outdoor Use: The HDS200 series is designed for outdoor use. Its compact size and robust design make it suitable for field work.
- Power Management: Utilize the USB Type-C interface for charging with a power bank during extended outdoor sessions.
- Data Analysis: Connect to PC software for more in-depth analysis and management of saved waveforms and data.
- Regular Calibration: Perform self-calibration periodically to ensure measurement accuracy.
10. Support & Compliance
10.1 Manufacturer Information
Hangzhou Microimage Software Co., Ltd.
Room 313, Unit B, Building 2, 399 Danfeng Road, Xixing Subdistrict, Binjiang District, Hangzhou, Zhejiang 310052, China
10.2 Compliance Notices
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment.
FCC Information
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation.
EU Conformity Statement
This product and, if applicable, the supplied accessories too are marked with "CE" and comply therefore with the applicable harmonized European standards listed under the Directive 2014/30/EU (EMCD), Directive 2014/35/EU (LVD), Directive 2011/65/EU (RoHS), Directive 2014/53/EU.
WEEE Directive (2012/19/EU)
Products marked with this symbol cannot be disposed of as unsorted municipal waste in the European Union. For proper recycling, return this product to your local supplier upon the purchase of equivalent new equipment, or dispose of it at designated collection points. For more information see: www.recyclethis.info.
Battery Directive (2006/66/EC and 2013/56/EU)
This product contains a battery that cannot be disposed of as unsorted municipal waste in the European Union. The battery is marked with this symbol, which may include lettering to indicate cadmium (Cd), lead (Pb), or mercury (Hg). For proper recycling, return the battery to your supplier or to a designated collection point. For more information see: www.recyclethis.info.
10.3 User Manual (PDF)
For more detailed operations and comprehensive information, please refer to the full user manual available as a PDF document. You can view the PDF at: OWON HDS200 Series User Manual (PDF).