1. Overview
This handheld instrument features an elegant design, compact size, portability, and flexible operation. The function key menu interface is clear and intuitive, and the screen utilizes a 3.5-inch IPS full-view color display, with the multimeter showing up to 6000 counts. This product combines the functionalities of an oscilloscope, signal generator, and multimeter into one. With superior performance and powerful features, it can be used in various measurement scenarios to meet a wide range of user measurement needs.
Safety Instructions
To avoid potential electric shock, fire hazards, and personal injury, please read the safety precautions before use. Only use the product for its intended purpose, as misuse may compromise the protection provided by the product.
- Before using the product, please check the casing for any cracks or plastic damage. Carefully inspect the insulator near the input ports.
- Please follow this user manual to use the correct input ports and settings, and perform measurements within the specified range outlined in this manual.
- Do not use this product in the presence of explosive gases and vapors or in humid environments.
- Keep your fingers behind the protective device of the test probes or tips.
- Do not touch unused input ports when the product is connected to the circuit under test.
- Before changing the test setting, disconnect the test probes from the circuit. When the direct current voltage is above 36V or the alternating current voltage is above 30V, it may pose a serious risk of injury to the user; therefore, precautions should be taken to avoid electric shock.
- Please select the correct test setting and range to avoid damage to the instrument or personal injury.
- Do not use this product with the front or back cover open.
- When the battery voltage is too low, it may affect the accuracy of the test results; please recharge it in a timely manner.
- The ground connection between the two channels is the same; during measurement, the ground clips should always be grounded or connected at the same potential.
- The probe ground is at the same potential as the ground. When connecting the USB cable for charging, the probe ground must not be clipped to high voltage; otherwise, it may damage the product or cause injury to the user.
- When using the oscilloscope probe to measure voltages above (AC 30V or DC 36V), please ensure that the USB protection cover of the product is securely closed to avoid human contact with exposed metal parts; otherwise, there is a risk of personal injury.
2. Setup
Product Overview Video
This video provides a visual overview of the BTMETER BT-50M's features and basic operation.
Probe Inspection
Before connecting probes, always perform a visual inspection and compensation check.
Safety
When using the probe, to avoid electric shock, keep your fingers behind the safety ring on the probe body. Do not touch the metal part at the top of the probe when it is connected to a high-voltage power source. The measured voltage must not exceed the probe's specifications (maximum 150V for 1X setting and 300V for 10X setting; otherwise, it may damage the instrument).
Manual Probe Compensation
- When connecting the probe to the oscilloscope for the first time, it is recommended to perform the following compensation check. Probes that are not compensated or have compensation deviations can lead to measurement errors or inaccuracies.
- After powering on, connect the probe to the signal input terminal and input a 3V/1kHz square wave signal.
Once connected, press the AUTO button on the panel to check the waveform display status. The display should show a 'Compensation Normal' waveform. If not, adjust the capacitance on the probe using the provided adjustment tool until the waveform is correct.
Probe Attenuation Settings
The probe attenuation coefficient setting will affect the vertical scale readings of the signal. Ensure that the attenuation switch multiplier on the probe matches the probe attenuation multiplier option in the oscilloscope system settings. If the switch is set to X1, the oscilloscope should also be set to X1; if the switch is set to X10, the oscilloscope should be set to X10.
Note: When the probe is set to X1, a probe rated at 10MHz/X1 will limit the oscilloscope's bandwidth to 10MHz input. To utilize the full bandwidth of the oscilloscope, ensure the switch is set to X10 or use a probe with a higher specification.
Baseline Correction
The instrument is 100% calibrated before leaving the factory, but if there is a baseline zero offset phenomenon found in situations such as large environmental temperature deviation or long-term disuse, baseline calibration can be performed.
- Select the "Baseline Calibration" mode from the System Function Menu Settings.
- After unplugging the probe as prompted on the screen, briefly press the "MENU" button to start calibration.
Important during calibration:
- Do not connect the probe and input signal during calibration, otherwise it may cause calibration deviation or damage to the instrument.
- Do not perform any other operations during the calibration process; wait patiently until the calibration is completed.
Battery Charging
When the battery level symbol appears in the upper right corner of the screen, charge the battery promptly as follows:
- Open the "USB" cover and insert the TYPE-C data cable.
- Connect the TYPE-C cable to a DC5V output adapter or a computer USB port for charging.
- While charging, the screen displays the charging symbol, and the built-in red light of the power button "REL/" will light up.
- When fully charged, the screen displays the full battery symbol, and the built-in red light of the power button "REL/" will turn off.
3. Operating Instructions
Oscilloscope Interface Description

Diagram of the BTMETER BT-50M Oscilloscope Interface.
- Signal Generator Output Port: Used for outputting various waveforms.
- CH1 Input: Channel 1 input for oscilloscope measurements.
- CH2 Input: Channel 2 input for oscilloscope measurements.
- USB Interface: For charging and data transfer.
- Display Screen: 3.5-inch IPS full-view color display.
- Case Area: Protective casing of the device.
- Multimeter Measurement Input Terminal: Input terminals for multimeter functions (10A, mA, COM, VΩHz).
Oscilloscope Display Interface Description

Detailed view of the oscilloscope display with numbered indicators.
- Operating Status Display: Shows RUN (automatic waveform acquisition), WAIT (normal trigger mode, blinking), T·D (captured trigger waveform data), STOP (lock current waveform, acquisition stopped).
- Time Base Scale: Displays the currently set horizontal time base scale value.
- Time Base Window: Displays the current time base position within the storage depth.
- Waveform Shift: Waveform Shift Indicator. The color of the indicator changes according to the currently set channel color. Adjust waveform position using arrow keys.
- Signal Generator Status: Displays corresponding waveform shape symbols (square, pulse, sine, triangle) according to the signal generator's output state.
- Battery Level: Displays current battery level and charging indication.
- Horizontal Trigger: Displays the current triggered horizontal time base position.
- Vertical Trigger for Channel 1 (CH1): Displays the current triggered vertical voltage position of CH1 channel in yellow.
- Vertical Trigger for Channel 2 (CH2): Displays the current triggered vertical voltage position of Channel 2 (CH2) in blue.
- Vertical Trigger for Triggering Channel: Displays the current triggered vertical voltage position of the triggering channel in purple.
- Displays Waveform for Channel 1 (CH1): Displays the waveform of Channel 1 (CH1) in yellow.
- Displays Waveform for Channel 2 (CH2): Displays the waveform of Channel 2 (CH2) in blue.
- Voltage for Channel 1 (CH1): Displays the coupling mode and voltage scale of Channel 1 (CH1) in yellow.
- Voltage for Channel 2 (CH2): Displays the coupling mode and voltage scale of Channel 2 (CH2) in blue.
- Triggering Channel: Displays the current trigger source: yellow for CH1 and blue for CH2. Trigger modes include Auto, Normal, and Single; trigger edge modes include Rising Edge and Falling Edge.
Button Function Description
The device features several function keys for intuitive operation:
- F1 Key: In oscilloscope mode, sets Channel 1 (CH1). In multimeter mode, selects the measurement function.
- F2 Key: In oscilloscope mode, sets Channel 2 (CH2). In multimeter mode, selects the measurement function.
- F3 Key: In oscilloscope mode, sets the AUTO channel. In multimeter mode, selects the measurement function.
- F4 Key: In oscilloscope mode, selects the measurement cursor when enabled. In multimeter mode, selects the measurement function.
- "REL/" Power Key: Long press for two seconds to turn the device on or off. In multimeter mode, a short press enters relative value (REL) measurement.
- AUTO/RANGE Key: In oscilloscope mode, a short press automatically acquires the measurement waveform. In multimeter mode, a short press toggles to select manual range.
- HOLD/SAVE Key: In oscilloscope mode, a short press stops (STOP) the measurement or resumes (RUN) it. A long press saves the measurement waveform data. In multimeter mode, a short press holds or cancels the hold function.
- MODE Key: A short press of this key toggles between oscilloscope mode and multimeter mode.
- Directional Keys (Up, Down, Left, Right): Used for progressively adjusting related setting parameters, moving the cursor position, and selecting items in the menu.
- MENU Key: A short press brings up the system function menu (5 pages). Use left/right directional keys to switch pages, press again to exit.
Oscilloscope Function Operations
Channel Settings
- In the oscilloscope interface, briefly press F1 to switch to CH1 or F2 to switch to CH2, selecting the channel you need to set up.
- After selecting the channel, briefly press F1 or F2, and the screen will display the "+" symbol. Use the arrow keys to move the marker line, and press the key again to exit.
- Long press F1 or F2 to display setting options:
- Switch ON/OFF: Turn the channel on or off.
- Attenuation X1/X10: Select probe attenuation (must match physical probe setting).
- Coupling AC/DC: Select AC (blocks DC component) or DC (allows both AC and DC components) coupling.
Automatic Measurement
When encountering uncertain waveforms or to avoid manual settings, press the AUTO button. The oscilloscope will automatically identify the waveform type, adjust amplitude and time base, and display the matched waveform.
Manual Measurement
Manually set parameters such as waveform voltage, time base, cursor positions, trigger mode, coupling mode, and probe attenuation. Connect the measurement circuit to the oscilloscope probe, and observe the waveform along with the relevant measured values.
Vertical System
The vertical system allows for setting waveform voltage amplitude, scale size, and position.
- Vertical Voltage Scale Setting: After selecting the channel, press the up arrow key to increase the voltage setting value, and the down arrow key to decrease it.
- X1 probe attenuation: 20 mV/div to 10 V/div.
- X10 probe attenuation: 200 mV/div to 100 V/div.
- Vertical Position: After selecting the channel, briefly press F1 or F2; the screen will display the "+" symbol. Press the up arrow key to move the waveform position upward, and the down arrow key to move it downward. Press this key again briefly to exit.
Horizontal System
The horizontal system allows you to set the waveform voltage time base scale size and position.
- Horizontal Scale: Press the left and right arrow keys to change the horizontal scale (time base), with a range of 10 ns/div to 20s/div. The left arrow key decreases the time base, while the right arrow key increases it.
- Horizontal Position: After selecting the channel, briefly press F1 or F2; the screen will display the "+" symbol. Use the left and right arrow keys to move the waveform position horizontally. Press this key again briefly to exit.
- Rolling Mode: When the horizontal time base is set to 200 ms/div, the oscilloscope automatically enters rolling mode. In this mode, the trigger and horizontal position settings are not controlled, and the waveform scrolls from left to right. This mode is suitable for low-speed signals.
Trigger System
The trigger system allows for conditional settings to capture and analyze specific waveform characteristics.
- Trigger Cursor Setting: On the oscilloscope interface, briefly press the F3 key to select the trigger channel. The screen will display a purple "+" symbol. Use the left/right arrow keys to adjust the horizontal trigger cursor position and the up/down arrow keys to adjust the vertical trigger cursor position.
- Trigger Mode Setting: Press and hold F3 to display three setting options:
- AUTO (Automatic) Mode: Oscilloscope automatically triggers, continuously captures and refreshes the waveform display, preventing freezing.
- NORMAL (Normal) Mode: When the signal amplitude reaches the set trigger level, the trigger system locks and holds the waveform. Updates with new triggers.
- SINGLE (Single) Mode: Locks and holds the waveform when amplitude reaches trigger level, indicating acquisition is complete (STOP status). Press HOLD to cancel STOP and re-trigger.
- Edge Trigger Setting: Press the MENU key briefly to select either rising edge ("↑") or falling edge ("↓") triggering.
- Trigger Source Settings: Short press the "MENU" button to select the trigger source to switch between CH1 (yellow) or CH2 (blue).
System Function Menu Settings
Short press the "MENU" button to access the system function menu (5 pages). Use left/right arrow keys to switch pages, then short press MENU to exit.
- Baseline calibration: F1 to turn on/off. Unplug probe, press MENU to start.
- Restore settings: F2 to turn on/off. Press MENU to restore factory settings.
- Data saving: F3 to enter data saving mode.
- Language selection: F4 to select "Chinese" or "English".
- Automatic shutdown: F1 to select time (1, 10, 30, 60, 120 minutes) or 'shut down' (off).
- Backlight brightness: F2 to select "30%", "50%", "80%" or "100%".
- Sampling mode: F3 to select "Peak Detection" or "High Resolution".
- Version number: Displays current system version.
- Display mode: F1 to select "Y-T" or "X-Y".
- Mathematical operation: F2 to select "FFT-CH1", "FFT-CH2" or "Off".
- Sound: F3 to select system prompt sound "Level 1", "Level 2", "Level 3" or "Off".
- Backlight time: F4 to select screen backlight display time "30 seconds", "60 seconds", "120 seconds", or "off" (always on).
- Image browsing: F1 to browse images in system memory.
- Measurement cursor: F2 to select "horizontal", "vertical", "horizontal and vertical", or "off".
- Numerical display: F3 to enter numerical display settings.
- Signal output: F4 to enter signal output interface.
- Afterglow Time: F1 to enter setting interface. Adjust afterglow time (minimum, 500ms, 1S, 10S, infinite).
Value Display
Select "Numerical Display" from the System Function Menu Settings. Press F1 to select measurement channel "CH1" or "CH2", and use directional keys to display corresponding measurement values. There are 8 sets of values (Peak-to-Peak, Max, Min, RMS, Duty Cycle, Frequency, Period, Frequency Counter). CH1 and CH2 can display up to 4 sets each. Select with MENU, press F3 to exit.
X-Y Display Mode
Select "X-Y" display mode from System Function Menu Settings. The screen displays CH1 and CH2 perpendicular to each other, generating Lissajous graphics based on frequency ratio and phase difference.
Cursor Measurement
Select "Measurement cursor" mode from System Function Menu Settings. Short press F2 to select horizontal, vertical, or combined cursors. Numerical values appear in the upper left corner.
- Horizontal cursor measurement: Open horizontal cursor axis, return to main menu, press F4, select up/down cursor axis to move, read voltage value between two cursors.
- Vertical cursor measurement: Open vertical cursor axis, return to main menu, press F4, select left/right cursor axis to move, read time value between two cursors.
- Horizontal and vertical cursor measurement: Simultaneously open both, return to main menu, press F4, select up/down/left/right cursor axes to move, read values between them.
How to Save Measurement Waveforms
Press and hold the HOLD SAVE button for 2 seconds. When the save prompt cursor appears, release the button, and the oscilloscope will automatically save the current waveform data as a picture with a serial number in memory.
How to Browse Stored Waveforms
Select "Image Browse" mode from System Function Menu Settings. The saved waveform image will appear. Press F1 to open the image, use arrow keys to browse. Press F2 to delete (confirm with MENU). Press F3 to format memory (confirm with MENU). Press F4 to exit image browsing.
How to Retrieve Stored Waveforms
Select "Data Save" mode from System Function Menu Settings. Connect the TYPE-C data cable to the oscilloscope and computer. Open the "Image" folder on the USB disk to view or download waveforms. Press F2 to exit "USB" mode.
Sampling Mode Selection
The oscilloscope has two sampling modes:
- Peak Detection Mode: Maximum sampling rate of 200MSa/s, maximum measurement bandwidth of 30MHz. Recommended for input signals less than 30MHz. Low power consumption, energy-efficient.
- High Resolution Mode: Maximum sampling rate of 280MSa/s, maximum measurement bandwidth of 50MHz. Recommended for input signals higher than 30MHz. High power consumption.
Signal Generator Operations

Examples of various waveforms generated by the device.
Select the "Signal Output" mode from the System Function Menu Settings. The output signal setting window will pop up. Use the left and right arrow keys to change the type of output waveform (sine wave, square wave, triangular wave, half wave, full wave, sawtooth wave, and DC signal).
The signal output setting window has 4 sets of setting fields:
- Output Waveform Amplitude: Press F1, use directional keys to adjust amplitude (0.1V-3V). Press to exit.
- Output Waveform Frequency: Press F2, use directional keys to adjust frequency (1Hz-2MHz). Press to exit.
- Output Waveform Duty Cycle (for square wave signals): Press F3, use directional keys to adjust duty cycle (1%-100%). Press to exit.
- Confirm: Press F4 to confirm settings. Briefly press the "MODE" key to exit "signal output" mode.
After setting parameters, connect the oscilloscope probe to the "SIGNAL OUTPUT" port. Press "HOLD/SAVE" to turn signal output ON/OFF (indicated by "ON" or "OFF" in the upper left corner).
Note: When the signal output waveform is sine or other non-square waveform, the maximum oscilloscope measurement time base is 100us. If the time base is restored, set the signal output as a square wave.
Multimeter Operations

A user performing measurements with the multimeter function.
In oscilloscope mode, press the "MODE" button to enter multimeter mode. Press again to return to oscilloscope mode.
Measurement Function
After entering the multimeter function interface, a measurement function indicator will appear at the bottom. Select the measurement function by pressing the left and right directional keys together with the F1~F4 keys. There are 12 measurement functions:
- DCV (DC Voltage)
- Resistance
- Audible continuity
- ACV (AC Voltage)
- DCmV (DC Millivoltage)
- ACmV (AC Millivoltage)
- Diode
- Capacitor
- DCA (DC Amperage)
- ACA (AC Amperage)
- DCmA (DC Milliampere)
- ACmA (AC Milliampere)
Measuring Method
Measurement of DC Voltage and AC Voltage
- Insert the black probe into the COM socket and the red probe into the V Ω Hz socket.
- For voltages less than 600mV, select "DCmV" or "ACmV". For voltages greater than 600mV, select "DCV" or "ACV".
- Contact the probe with the test point reliably.
- Read the measured voltage value on the display screen.
- When measuring DC voltage, the polarity of the test point contacted by the red probe will be displayed.
- When measuring AC voltage, press the up arrow keys to select frequency and duty cycle measurement, with a sensitivity of 0.5V AC effective value.
Caution:
- The measured voltage should not exceed the maximum rated test value to avoid damage or personal injury.
- When measuring high-voltage circuits, avoid touching them.
Measurement of DC and AC Currents
- Insert the black probe into the COM socket. For currents less than 600mA, insert the red probe into the "mA" socket. For currents greater than 600mA, insert the red probe into the "10A" socket.
- For currents less than 600mA, select "DCmA" or "ACmA". For currents greater than 600mA, select "DCA" or "ACA".
- Connect the probe in series to the circuit under test.
- Read the measured current value on the display screen.
- When measuring direct current, the polarity from the red probe to the black probe will be displayed.
- When measuring AC current, press the up arrow keys to select frequency and duty cycle measurement. Sensitivity: 0.3mA in "mA" range, 30mA in "10A" range.
Caution:
- The measured current should not exceed the maximum rated test value to avoid damage or personal injury.
- If the magnitude of the measured current is unknown, select "DCA" or "ACA" first. If less than 600mA, then select "DCmA" or "ACmA".
- It is strictly prohibited to input voltage in this current measurement state.
Resistance Measurement
- Insert the black probe into the COM socket and the red probe into the V Ω Hz socket.
- Select the "Resistance" measurement function.
- Ensure reliable contact with both ends of the measured resistor.
- Read the measured resistance value on the display screen.
Note:
- Before measuring resistance, ensure all power sources in the circuit are off and all capacitors are fully discharged.
- It is strictly forbidden to input voltage in the resistance measurement state.
Capacitance Measurement
- Insert the black test lead into the COM jack and the red test lead into the V Ω Hz jack.
- Select the "Capacitance" measurement function.
- Ensure reliable contact with both ends of the capacitor under test.
- Read the measured capacitance value on the display screen.
Note:
- Discharge the capacitor under test before testing.
- When measuring small capacitors (≤1 μF), briefly press the "REL/" button to zero before testing to eliminate open-circuit residuals.
- Measuring large capacitors may take longer to stabilize (e.g., 600 μF to 60mF may take about 10 seconds).
- It is strictly forbidden to input voltage in the capacitance measurement state.
Audible Continuity Test
- Insert the black test lead into the COM jack and the red test lead into the V Ω Hz jack.
- Select the "Audible continuity" measurement function.
- Ensure reliable contact with both ends of the resistor under test. If resistance is approximately less than 50Ω, the built-in buzzer will sound.
Diode Measurement
- Insert the black test lead into the COM jack and the red test lead into the V Ω Hz jack.
- Select the "Diode" measurement function.
- Connect the red test lead to the anode and the black test lead to the cathode.
- Read the forward voltage value of the diode.
Note: If the diode is connected in reverse or is open, "OL" will be displayed.
Data Hold
In multimeter mode, briefly press the "HOLD/SAVE" button to hold the current display value. The "HOLD" symbol will appear. Press again to cancel.
Relative Value Measurement
In multimeter mode, briefly press the "REL/" button to enter relative measurement. The "REL" symbol will appear. The displayed value is the difference between the measured value and a reference value. This can also be used for zeroing (e.g., for small capacitors). Press "REL/" again to exit.
Manual Range Function
In multimeter mode, measurement is normally in automatic range. To switch to manual range, briefly press the "AUTO/RANGE" button. Press once to cycle through ranges. Long press to exit manual range. Note that some functions do not have manual range.
4. Maintenance
Fuse Replacement
If the fuse is blown or malfunctions, follow these steps:
- Before replacing the fuse, remove the test leads and turn off the power.
- Remove the protective cover.
- Unscrew the four screws securing the back cover and remove it.
- Remove the faulty old fuse and replace it with a new fuse of the same model.
Fuse Specifications:
- 600mA/250V fast-blow fuse (5x20mm)
- 10A/250V fast-blow fuse (5x20mm)
- Reinstall the back cover, tighten the screws, and put on the protective cover.
Battery Replacement
The instrument's lithium battery is rechargeable but consumable. If standby time significantly shortens, replace it with the same model (4000mAh).
- Before replacing, remove test leads and turn off power.
- Unscrew the screws securing the battery cover on the back and remove it.
- Remove the old battery and replace with a new one, ensuring correct polarity.
- Reinstall the battery cover and tighten screws.
Cleaning the Product
Use a damp cloth and mild detergent to clean the casing. Do not use corrosive agents or solvents. Dust or moisture in the test ports may affect accuracy. Remove all input signals before cleaning.
Battery Storage
If not used for over 6 months, charge the battery to 50%-70% and store in a cool, dry environment. If the lithium battery shows signs of rust, leakage, or swelling, remove and dispose of it immediately.
5. Troubleshooting
This section provides general guidance for common issues. For more complex problems, please contact support.
- Device not powering on: Ensure the battery is charged. Connect to a power source via USB-C. If still unresponsive, check for physical damage or try a hard reset (if applicable, consult the full PDF manual).
- Inaccurate readings: Perform probe compensation (see Setup section). Check probe attenuation settings match the device. Ensure proper connection of test leads. Calibrate the device if environmental conditions have changed significantly or after long disuse (see Baseline Correction in Setup).
- No waveform display: Check if the channel is enabled. Verify trigger settings (mode, level, source). Ensure probes are correctly connected to the signal source and the device. Adjust vertical and horizontal scales.
- Signal Generator output issues: Verify the output waveform type, amplitude, frequency, and duty cycle settings. Ensure the output is enabled.
- Multimeter not measuring: Check test lead connections to the correct input terminals (COM, VΩHz, mA, 10A). Ensure the correct measurement function is selected. Verify the circuit under test is powered (for voltage/current) or unpowered (for resistance/capacitance).
- Device freezing or unresponsive: Try restarting the device by long-pressing the power key. If persistent, check for firmware updates (consult the full PDF manual).
6. Specifications
Universal Multimeter Features
| Feature | Description |
|---|---|
| Display (IPS) | 6000 counts |
| Range | Auto/Manual |
| Material | ABS+PVC |
| Sampling Rate | 3 times/second |
| True RMS | Yes |
| Data Hold | Yes |
| Relative Measurement | Yes |
| Low Battery Indicator | Yes |
| Auto power off | Yes |
Mechanical Technical Features
| Feature | Description |
|---|---|
| Dimension | 183×100×44mm |
| Weight | Around 460g |
| Battery Type | 4000mAh Lithium Battery |
Environmental Technical Features
| Feature | Description |
|---|---|
| Operating Temperature | 0~40°C |
| Operating Humidity | <75% |
| Storage Temperature | -20~60°C |
| Storage Humidity | <80% |
Multimeter Technical Specifications
| Function | Range | Resolution | Accuracy |
|---|---|---|---|
| DC Voltage | 60mV | 0.01mV | ±(0.8%+2) |
| 600mV | 0.1mV | ||
| 6V | 1mV | ±(0.5%+2) | |
| 60V | 10mV | ||
| 600V | 100mV | ||
| 1000V | 1V | ±(0.8%+2) | |
| AC Voltage | 60mV | 0.01mV | ±(1.5%+3) |
| 600mV | 0.1mV | ||
| 6V | 1mV | ±(1.0%+3) | |
| 60V | 10mV | ||
| 600V | 100mV | ||
| 750V | 1V | ±(1.5%+3) | |
| AC voltage frequency response: 40Hz~1kHz | |||
| DC Current | 60mA | 0.01mA | ±(1.2%+2) |
| 600mA | 0.1mA | ||
| 6A | 1mA | ±(2.0%+3) | |
| 10A | 10mA | ||
| AC Current | 60mA | 0.01mA | ±(1.5%+3) |
| 600mA | 0.1mA | ||
| 6A | 1mA | ±(2.5%+5) | |
| 10A | 10mA | ||
| AC current frequency response: 40Hz~1kHz | |||
| Resistance | 600Ω | 0.1Ω | ±(1.0%+3) |
| 6kΩ | 1Ω | ||
| 60kΩ | 10Ω | ||
| 600kΩ | 100Ω | ||
| 6MΩ | 1kΩ | ±(1.5%+3) | |
| 60MΩ | 10kΩ | ||
| Capacitance | 6nF | 0.001nF | ±(10.0%+20) |
| 60nF | 0.01nF | ||
| 600nF | 0.1nF | ±(2.5%+5) | |
| 6μF | 0.001μF | ||
| 60μF | 0.01μF | ±(5.0%+10) | |
| 600μF | 0.1μF | ||
| 6mF | 0.001mF | ±(10.0%+20) | |
| 60mF | 0.01mF | ||
| Frequency | 99.99Hz | 0.01Hz | |
| 999.9Hz | 0.1Hz | ±(0.1%+5) | |
| 9.999kHz | 1Hz | ||
| 30kHz | 10Hz | ||
| Duty cycle | 0.1%-99.9% | 0.1% | ±(2.0%+2) |
| Diode test | Yes | ||
| Continuity | Yes | ||
Oscilloscope Specifications
| Feature | Description |
|---|---|
| Bandwidth | 50MHz Dual channel |
| Sampling method | Real-time sampling |
| Real-time sampling rate | 200M/280MSa/s |
| Channels | 2 Dual channel |
| Input coupling | AC, DC |
| Input impedance | 1MΩ, @16pf |
| Probe attenuation | X1, X10 |
| Maximum input voltage | X1 range <150V, X10 range <300V (DC+AC peak) |
| Sampling rate range | 1.5Sa/s - 280MSa/s |
| Waveform interpolation | (sinx) x |
| Sweep range | 10ns/div-20s/div |
| Time base accuracy | 20ppm |
| Record length | Up to 128kbyte |
| Sensitivity | 20mV/div-10V/div |
| Displacement Range | 4 divisions (positive and negative) |
| Analog Bandwidth | 50MHz |
| Low Frequency Response | > 10Hz |
| Rise Time | <10ns |
| DC Gain Accuracy | ±3% |
| Automatic Measurement | Period, Frequency, Peak-to-Peak, Maximum, Minimum, RMS, Duty Cycle, Frequency Counter |
| Trigger Mode | Auto, Normal, Single |
| Trigger Edge | Rising Edge, Falling Edge |
| Signal Generator Output | Sine Wave, Square Wave, Triangle Wave, Half Wave, Full Wave, Sawtooth Wave, and DC Signal |
| Operating Mode | Normal Mode 200MSa/s, High-Speed Mode 280MSa/s |
| Display Mode | Y-T, X-Y |
| Persistence Mode | Minimum, 500ms, 1S, 10S, Infinite |
The above picture and content are for reference only. Please be subject to the actual products if anything different or updated. Please pardon for not informing in advance.
7. Warranty & Support
Except for replacing the battery and fuse, do not attempt to repair this product or alter the circuit unless you are qualified and have the appropriate calibration, performance testing, and repair operation instructions.
Contact Us
For any problem or concern, welcome to email us for prompt response:
To ensure you receive an immediate solution and your requests are processed quickly, please include the following information in your email:
- Order Number
- Platform of Your Purchase
- Full Model Number
- Description of the Problem (Attaching videos or photos can help us troubleshoot the problems even faster)
Additional Resources
A detailed user manual in PDF format is available for download: BTMETER BT-50M User Manual (PDF)
8. User Tips
Here are some general tips to enhance your experience with the BTMETER BT-50M:
- Regular Calibration Checks: Periodically check probe compensation and perform baseline correction, especially if you notice inconsistent readings or after significant temperature changes.
- Battery Management: For optimal battery life, avoid fully discharging the device frequently. Store the device with a partial charge (50-70%) if not used for extended periods.
- Understand Your Modes: Familiarize yourself with the different oscilloscope trigger modes (Auto, Normal, Single) and multimeter functions to select the most appropriate one for your specific measurement task.
- Safety First: Always prioritize safety. Double-check connections, ensure proper insulation, and never exceed the maximum voltage/current ratings.
- Keep it Clean: Regularly clean the device and its input ports to prevent dust and moisture buildup, which can affect performance and accuracy.