1. Introduction
The MYOSLIM EMS Body Sculpting Machine utilizes advanced HI-EMT technology, employing high-energy focused electromagnetic waves (HI-EMT) designed to induce powerful muscle contractions. This process promotes muscle growth and reshapes the internal structure of muscles. Additionally, it helps trigger significant fat breakdown through intense muscle contractions, releasing fatty acids from triglycerides for metabolism over several weeks.
Key Features:
- Advanced HI-EMT Technology: Employs high-energy focused electromagnetic waves for intense muscle contractions, promoting muscle growth and deep internal structural reshaping.
- Touch-Controlled Display: Simplifies device operation with clear visibility and adjustable brightness for various lighting conditions.
- Real-time Monitoring and Guidance: Provides real-time information on treatment time, channels, methods, and attachments, aiding in effective treatment progress.
- Upgradeability: Offers flexible upgrade options to meet evolving therapeutic needs.


2. Indications for Use
The EMS Body Sculpting Machine is intended for:
- Improvement of abdominal tone, strengthening of abdominal muscles, and development of a fitter abdomen.
- Strengthening, toning, and firming of buttocks, thighs, and calves.
- Improvement of muscle tone and firmness, for strengthening muscles in arms.
3. Operating Environment
To ensure the effectiveness and safety of the device, the operating environment should meet the following criteria:
- Clean and Hygienic: The environment must be clean to prevent contamination. Regular cleaning and disinfection of the equipment and accessories are essential.
- Appropriate Temperature and Humidity: Maintain temperature and humidity within the manufacturer's specified range to avoid impacting device performance.
- Adequate Lighting: Sufficient lighting is necessary for clear visibility and correct operation, especially for precise procedures.
- Good Ventilation: Ensure adequate air circulation to prevent accumulation of harmful gases.
- Safety Measures: The environment should have appropriate safety measures, including fire extinguishers and first aid equipment. Operators should be familiar with safety procedures.
- Avoidance of Interference Sources: Keep away from electromagnetic fields or other devices that could cause interference.
- Operator Training: All operators must undergo proper training for correct usage and safety procedures.
4. Patient Profile
This machine is suitable for adults seeking healthy weight loss, who have no physical diseases, and who do not have metal substances in their bodies.
5. Contraindications
Do NOT use the device if any of the following apply:
- Wounds on the skin (avoiding wounds when using).
- Patients with heart disease (e.g., arrhythmias), especially those with pacemakers.
- Cancer patients.
- Patients with infectious diseases.
- Pregnant women, lactating women, or during menstrual period.
- Major surgery patients, diabetic patients.
- Presence of any metal in the body.
- People with cerebrovascular disease (stroke patients).
6. Possible Side Effects
Possible side effects may include, but are not limited to:
- Muscle soreness.
- Reduction in muscle fat.
- Temporary muscle spasms.
- Transient joint or muscle tendon pain.
- Localized redness or skin flushing.
7. Safety Precautions and Warnings
Precautions before use:
- Do not use when feeling full; wait at least one hour after meals.
- Remove all metal jewelry before using the instrument.
- Before starting operation, touch the probe vertically to the skin, then turn on the main unit switch and adjust the intensity.
- Start at the lowest intensity level and gradually increase according to user's skin tolerance.
- It is forbidden to eat or drink during operation. Wait at least one hour after operation before eating.
8. Device and Accessories
8.1 Front and Rear Panel of the Main Unit


8.2 Handles


Note: The handle can only be used by the original machine and cannot be used interchangeably.
8.3 List of Accessories
| Serial | Name | Quantity |
|---|---|---|
| 1 | Host | 1 set |
| 2 | Handle | 2 |
| 3 | 7*100CM belt | 2 |
| 4 | Power cable | 1 |
| 5 | Electronic manual | 1 copy |
9. Setup (Device Installation)
9.1 Handle Film Application
- Apply the black film to the working part of the handle.
9.2 Installing the Handle
- Align the hole on the handle with the machine controller and fasten the clip.
- Lock the key ring to secure the handle.
9.3 Connecting Power
- Connect one end of the power cord to the machine.
- Plug the other end into a suitable power socket. Do not connect multiple machines to a single socket.
9.4 Operating the Switches
- Back Switch: Move the manual switch on the rear panel to the "o" position (power off). To power on, move to the "I" position (light on).
- Front Switch: Single-click the button on the front panel to power on. Long press to power off.
10. Operating (Basic Device Operation)
10.1 Handle Exchange
Important: The handle must be calibrated every time it is installed or replaced. When replacing the handle, you must turn off the power of the machine. The handle cannot be replaced when the machine is powered on.
10.2 Device Start Up/Shut Down
- Connect the power cord as described in the Setup section.
- Move the manual switch on the rear panel to the "I" position.
- Press the button on the front panel to start the machine.
- When treatment is finished, press and hold the power button to turn off the machine.
10.3 Navigation Controls and Setting Therapy
The pre-treatment regimen can be modified via on-screen controls. The treatment interface displays various items and functions:
The main screen typically shows options for 'List', 'Pattern', and 'Records'.
- List: Access predefined treatment protocols.
- Pattern: Customize treatment parameters.
- Records: View past treatment data.
Setting Therapy by Controls (Example: Mode 1)
The operation interface for Mode 1 allows adjustment of parameters. Ensure the handle is removed from its frame and fixed to the treatment area with a belt before operation. The frequency for Mode 1 is typically 5-60Hz.
To adjust settings:
- Enter the desired pattern interface (e.g., Pattern 1 or Pattern 2).
- Adjust the handle operation time according to customer needs.
- Click the start/stop button to begin the treatment.
- Adjust the corresponding intensity according to different customer needs.
Note: If the customer requests to pause, you can click start/pause. The work button suspends the work, and you can directly click the back button to return after the work is completed.
Unit Settings
Access the settings menu (gear icon) to configure various aspects of the machine:
- Language Settings: Select your preferred language (e.g., English, Portuguese, Spanish, Japanese, Korean, French, Russian, Chinese).
- Sound Adjustment: Control the device's audio output.
- Brightness Adjustment: Adjust the screen brightness for optimal visibility.
- Set Date/Time: Configure the current date and time.
- System Adjustment: Access system-level configurations.
- Technician: This section may require a password for advanced settings or diagnostics.

Please operate according to the controller model you purchased and follow the manufacturer's official documentation to ensure the best experience. If you have any questions, please contact customer service for support.
11. Maintenance
11.1 Daily Inspection
Regularly inspect instruments and components. Ensure the power socket has a reliable ground connection. Check the power cord for damage; if damaged, stop using the instrument immediately.
11.2 Cleaning
Instrument cleaning should be performed regularly. The device housing can be wiped with a damp cloth, soapy water, or a neutral detergent while the instrument is powered off. Do not use alcohol-based detergents on painted surfaces.
11.3 Cleaning of Accessories
Parts that come into contact with the skin should be cleaned and disinfected with 75% alcohol before and after use. Wipe with pure cotton material; do not soak directly in alcohol-based detergent.
11.4 Fuse Replacement
The fuse is located below the handle junction on the back of the machine. An additional spare fuse is included. To prevent electric shock, turn off the power and unplug the power plug before replacing the fuse.
11.5 Transport and Storage
- Storage: Store the instrument in its original packaging box in a well-ventilated room. The ambient temperature should be -10°C to 40°C, and relative humidity should not exceed 93%. Keep away from flammable, explosive, and corrosive gases.
- Transportation: The instrument is suitable for road, railway, air, and water transportation in its packaging. Prevent severe vibration and shock, and protect from moisture. Do not mix with flammable or corrosive substances.
12. Troubleshooting
| Fault Phenomenon | Causes | Exclusions and Workarounds |
|---|---|---|
| Device not working | Power plug not inserted into socket Power switch not on Power fuse failure | Check the power Check the power switch Replace the fuse |
| Handle not working | Handle is loose | Reseat the handle |
13. Specifications (Technical Parameters)
| Parameter | Value |
|---|---|
| Name | E-11 |
| Input Power Supply | AC110V-220V 50/60Hz |
| Output Maximum Power | 0-6000W |
| Frequency | 5-200HZ |
| Voltage | 50-1000V AC |
| Material | ABS+Steel |
| Screen Size | 7.0 inches |
| Machine Color | White |
| Item Type | Massage & Relaxation |
| Style | PORTABLE / Desktop |
| High Quality Assurance | CE, ROHS certificates |
| Application | For Commercial, Beauty Center & Click |
13.1 Electromagnetic Compatibility (EMC)
The device has undergone Radiated Electromagnetic Field Immunity Testing according to IEC 61000-4-3:2020 standards. The test level was 30V/m with 80% AM, 1kHz Modulation.
For detailed EMC information, please refer to the official user manual PDF: View User Manual (PDF)
14. Warranty and Support
14.1 Warranty Information
- The main machine has a warranty period of 1 year from the date of purchase (excluding man-made damage, fragile parts, and other non-warranty scope or expired periods).
- Accessories have a warranty period of 3 months (excluding man-made damage).
- Free warranty service is not provided for failures caused by: unauthorized disassembly/modification, accidental fall, lack of reasonable maintenance, or failure to operate according to instructions.
- If the instrument needs repair within six months of purchase, the buyer bears return shipping costs, and the seller bears shipping-in costs. After six months, the buyer bears all shipping costs.
14.2 Manufacturer
Guangzhou Xinwei Electronic Technology Co., Ltd.
14.3 Service
For after-sales service, please contact the company's business department or dealer. Do not attempt repairs without authorization, as this will void the warranty.
- Our Service: 24 hours online service, instructions operation video training.
- After-sales Service Provided: Free spare parts, video technical support, online support.
15. User Tips
- Always start with the lowest intensity and gradually increase to find your comfortable and effective level.
- Ensure proper handle placement and secure attachment with belts for optimal results and safety.
- Regularly clean the handles and machine surface to maintain hygiene and prolong device lifespan.
- Refer to the digital user manual (PDF) for detailed diagrams and advanced settings.
16. Frequently Asked Questions
- Q: Can I use the machine if I have metal implants?
- A: No, the device is contraindicated for individuals with any metal in their body, including implants.
- Q: How often should I clean the device?
- A: The device housing should be wiped regularly with a damp cloth. Accessories that contact the skin should be cleaned and disinfected with 75% alcohol before and after each use.
- Q: What should I do if the machine is not turning on?
- A: First, check if the power plug is securely inserted and if the power switch on the rear panel is in the 'on' position. If it still doesn't work, check and replace the fuse if necessary.