MYOSLIM MYOslim NEO Professional 6500W

Contents

MYOslim NEO Professional 6500W HI-EMT EMS RF Nova Machine User Manual

Model: MYOslim NEO Professional 6500W | Brand: MYOSLIM

1. General Information

The MYOslim NEO Professional 6500W HI-EMT EMS RF Nova Machine utilizes advanced HI-EMT technology, employing high-energy focused electromagnetic waves (HI-EMT) designed to induce powerful muscle contractions, thereby promoting muscle growth and reshaping the internal structure of muscles. Additionally, it triggers significant fat breakdown through intense muscle contractions, causing fatty acids to be released from triglycerides and metabolized by the body over several weeks.

1.1 Indications for Use

  • Improvement of abdominal tone, strengthening of the abdominal muscles, development of fitter abdomen.
  • Strengthening, Toning and Firming of buttocks, thighs and calves.
  • Improvement of muscle tone and firmness, for strengthening muscles in arms.

1.2 Operating Environment

The operating environment of the EMS Body Sculpt instrument series is crucial to ensuring the effectiveness and safety of the device. Ensure the following conditions:

  • Clean and Hygienic Environment: The operating environment should be clean and hygienic to prevent any external contamination or bacterial infections. Regular cleaning and disinfection of the equipment and its accessories are essential.
  • Appropriate Temperature and Humidity: Ensure that the operating environment's temperature and humidity are within the appropriate range specified by the device manufacturer. Extreme temperatures or excessive dryness/humidity may impact the device's performance.
  • Adequate Lighting: Provide sufficient lighting to ensure that operators can clearly see and operate the equipment correctly. This is especially important for conducting precise beauty procedures.
  • Good Ventilation: The area where the equipment is operated should have adequate ventilation to ensure air circulation and prevent the accumulation of any harmful gases.
  • Safety Measures: The operating environment should have appropriate safety measures, including fire extinguishers and first aid equipment, to address emergencies. Operators should be familiar with the equipment's safety procedures and know how to handle emergency situations.
  • Avoidance of Interference Sources: The operating environment for beauty equipment should be kept away from any electromagnetic fields or other devices that could potentially cause interference, ensuring the normal operation of the equipment.
  • Operator Training: All operators using beauty equipment should undergo proper training to understand the correct usage and safety procedures of the equipment.

1.3 Patient Profile

This machine is suitable for patients who are seeking healthy weight loss, who have no physical diseases, and who do not have metal substances in their bodies. Please note that this machine is only suitable for adults.

1.4 Description of the Device

The EMS Body Sculpt enables a simple and effective treatment process. You can easily perform by adjusting the digital energy selection treatment plan or diagnostic procedure. Any treatment parameter can be manually adjusted by simply touching the screen. You can also create precise and effective treatment plans by configuring the age group and gender of the person being treated within the machine.

The device features a touch-controlled display for easy operation and clear visibility. Real-time monitoring provides information on remaining treatment time, utilized channels, treatment methods, and attachments. The EMS Body Sculpt series also offers flexible upgrade options to meet future treatment requirements.

1.5 Contraindications

Do NOT use this device if any of the following apply:

  • Wounds on the skin (or avoid wounds when using).
  • Patients with heart disease (such as those with arrhythmias), especially those with pacemakers.
  • Cancer patients.
  • Patients with infectious diseases.
  • Pregnant women, lactating women, or during menstrual period.
  • Major surgery patients, diabetic patients.
  • Having any metal in the body.
  • People with cerebrovascular disease (stroke patients).

1.6 Possible Side Effects

These effects may include but are not limited to:

  • Muscle soreness.
  • Reduction in muscle fat.
  • Temporary muscle spasms.
  • Transient joint or muscle tendon pain.
  • Localized redness or skin flushing.

2. Safety Precautions and Warnings

Please observe the following precautions before using the device:

  • Do not use the device when feeling full; wait at least one hour after meals.
  • Before using the instrument, remove any metal jewelry you are wearing.
  • Before starting the operation, touch the probe vertically to the skin, then turn on the switch of the main unit and adjust the intensity of the operation mode.
  • Always start at the lowest intensity level, and then increase the intensity sequentially according to the user's skin tolerance.
  • It is forbidden to eat or drink during the operation. It is recommended to eat after one hour following the operation.

3. Device and Accessories

3.1 Front and Rear Panel of the Main Unit

Front View of the Instrument:

Front view of the MYOslim NEO machine, showing the display screen and two attached EMS handles.
Front view of the MYOslim NEO machine, showing the display screen and two attached EMS handles.
  • Display screen
  • EMS Handle
  • Switch

Rear View of the Instrument:

Rear view of the MYOslim NEO machine, showing handle connections, power plug, and main power switch.
Rear view of the MYOslim NEO machine, showing handle connections, power plug, and main power switch.
  • Handle connection
  • Power plug
  • Fuse
  • Switch

3.2 Handle Types

The device comes with various handle types for different applications:

Standard treatment handles for the MYOslim NEO machine.
Standard treatment handles for the MYOslim NEO machine.
Cushion-style treatment handles, often used for VIP taxes or specific applications.
Cushion-style treatment handles, often used for VIP taxes or specific applications.

3.3 List of Accessories

Serial Name Quantity Picture
1 Host (Main Unit) 1 set Main unit of the MYOslim NEO machine.
2 Handle (Standard) 2 Standard treatment handles.
Handle (Optional Cushion) 1 Cushion-style treatment handle.
3 Belt (8*80CM, 8*90CM, 8*100CM) 3 Treatment belts of various sizes.
4 Power cable 1 Power cable for the device.
5 Electronic manual 1 copy

4. Device Installation

Unboxing and Initial Setup

Watch this video for a visual guide on unboxing and setting up your MYOslim NEO machine:

Video: Unboxing, assembly, and initial power-on demonstration of the MYOslim NEO Professional 6500W machine.

Step-by-Step Installation:

  1. Install the Handle Bracket: Use a screwdriver to install the handle brackets on both sides of the main unit.

    Diagram showing how to install the handle brackets on the MYOslim NEO machine.
    Installation of handle brackets on the main unit.
  2. Apply Handle Film: Stick the black film to the working part of the handle.

    Image demonstrating the application of protective film to the treatment handles.
    Applying protective film to the treatment handles.
  3. Install the Handle: Align the handle with the hole on the bracket and fasten the clip. Lock the key ring to secure it.

    Diagram showing how to attach the treatment handle to the machine bracket.
    Attaching the treatment handle to the machine.
  4. Connect Power: Connect one end of the power cord to the machine and plug the other end into a suitable power socket. Do not connect multiple machines to the same socket.

    Image showing the power cable and connection points on the machine and wall socket.
    Connecting the power cable to the machine and power outlet.
  5. Press the Back Switch: Locate the manual switch on the rear panel and move it to the 'o' position to enable power.

    Close-up of the rear power switch on the MYOslim NEO machine.
    Activating the main power switch on the rear panel.
  6. Press the Front Switch: Press the power button on the front panel. A single click will power on the device, and a long press will power it off.

    Close-up of the front power button and display screen on the MYOslim NEO machine.
    Powering on the device using the front panel button.

5. Basic Device Operation

5.1 Handle Exchange

Important: The handle must be calibrated every time it is installed or replaced. When replacing the handle, you must turn off the power of the machine. The handle cannot be replaced when the machine is powered on.

5.2 Device Start Up/Shut Down

  1. Connect one end of the power cord to the machine and plug the other end into the socket. Plug the power cord directly into the main socket. Do not connect multiple machines to the socket.
  2. Move the manual switch on the rear panel to the "o" position.
  3. Press the button on the front panel.
  4. When the treatment is finished, press and hold the power button to turn off the machine.

5.3 Navigation Controls

The pre-treatment regimen can be modified via on-screen controls. The treatment interface displays the following items and functions:

Screenshot of the MYOslim NEO machine's main operation interface, showing controls for time, intensity, and handle selection.
Main operation interface with navigation controls.
  • List: Access saved treatment protocols.
  • Pattern: Select treatment patterns (e.g., Pattern 1, Pattern 2).
  • Records: View treatment history.
  • Time: Adjust treatment duration.
  • Intensity (A/B): Control the intensity level for each handle.
  • Frequency: Adjust the frequency of the electromagnetic pulses.
  • Start/Pause Button: Initiate or pause the treatment.

5.4 Unit Settings

The device offers various settings accessible through the interface:

  • Mode 1: This mode's operation interface allows adjustment of handle operation time according to customer needs. The frequency for Mode 1 is 5-60Hz.
  • Mode 2: This mode's operation interface allows adjustment of intensity/stage frequency/time. It features 3 different periods of frequency, which are freely adjustable. The frequency for Mode 2 is 30-100Hz.
  • Language Settings: Select your preferred language for the interface.
  • Sound Adjustment: Control the device's audio feedback volume.
  • Brightness Adjustment: Adjust the screen brightness for optimal visibility.
  • Set Date Time: Configure the current date and time.
  • System Adjustment: Access advanced system settings (may require technician password).
  • Technician: Access technician-level settings (requires password).

Note: Adjust the corresponding intensity/stage frequency/time according to different customer needs. Click the start/pause work button to enter the operation. If the customer requests to pause, you can click the start/pause work button to pause the work, and after the work is completed, you can directly click the return key to return.

6. Therapy

6.1 Preparing the Therapy

  1. Communicate with patients to understand their needs and comfort levels.
  2. Ensure the patient is in good health and has no contraindications.
  3. Ensure the patient is adequately hydrated.
  4. Ensure the patient wears appropriate clothing that allows for proper handle placement.
  5. Advise the patient to avoid eating in large quantities before the session.
  6. Familiarize yourself with the treatment process and settings.
  7. Advise the patient to avoid drinking alcohol before treatment.

6.2 Placing the Handle

Proper placement of the handles is crucial for effective treatment. Refer to the diagrams below for guidance:

Diagram showing handle placement for abdominal treatment.
Belly Treatment
Diagram showing handle placement for buttocks treatment.
Buttocks Treatment
Diagram showing handle placement for arm treatment.
Arm Treatment
Diagram showing handle placement for thigh treatment.
Thigh Treatment
Diagram showing handle placement for calf treatment.
Calf Treatment
Diagram showing handle placement for gluteal (arse) treatment.
Gluteal (Arse) Treatment

Please operate according to the controller model you purchased and follow the manufacturer's official documentation to ensure the best experience. If you have any questions, please contact customer service for support.

7. Maintenance and Repair

7.1 Maintenance

7.1.1 Daily Inspection

Regularly inspect instruments and components. Check whether the power socket has a reliable connection to the ground. Please check whether the power cord of the instrument is damaged before use, and stop using the instrument if it is damaged.

7.1.2 Cleaning

Instrument cleaning is performed on a regular basis to ensure sanitation and proper functioning of the equipment. The device housing can be wiped with a damp cloth, soapy water or a neutral detergent while the instrument is powered off.

The parts that come into contact with the skin should be cleaned and disinfected with 75% alcohol before and after use. Do not soak directly in alcohol-based detergent as a whole, wipe with pure cotton material.

7.1.3 Replacing the Fuse

When replacing the fuse, in order to prevent electric shock, please turn off the power and unplug the power plug. The fuse is located below the handle junction on the back of the machine. An additional spare fuse is included.

7.2 Repair

After purchasing the instrument, you need to fill in the warranty card and send it to the manufacturer or agent to have the effect of warranty. From the date of the purchase of the instrument, if a product quality problem occurs, a free warranty is provided for one year. (Warranty requires a warranty card and purchase bill, and the scope of warranty is limited to the instrument host, excluding accessories and other consumables that need to be replaced regularly). After the warranty period has expired, please contact the manufacturer directly to provide paid after-sales service. To repair the machine, you must contact the company's business department or dealer, and do not repair it without authorization. Any unpacking without consent will not be guaranteed by our company.

8. Common Faults and Troubleshooting

Fault Phenomenon Causes Exclusions and Workarounds
Device not working
  • The power plug is not inserted into the socket
  • Power switch not on
  • Power fuse failure
  • Check the power
  • Check the power switch
  • Replace the fuse
Handle not working Check the handle for loose connection Reseat the handle

9. Technical Parameters

Attribute Value
NameE-02
Input Power SupplyAC110V-220V 50/60Hz
Output Maximum Power0-6500W
Frequency5-200HZ
Screen Size8.0 inches
Machine ColorWhite
MaterialABS
ApplicationBODY
Core TechnologyEMT (Electromagnetic Therapy) + EMS (Electrical Muscle Stimulation) Dual-Mode System
Treatment TypeNon-invasive body sculpting, muscle building, fat reduction, and skin tightening
Primary FunctionSupramaximal muscle contractions for toning, lifting, and shaping
Target AreasAbdomen, buttocks, thighs, arms, back customizable treatment zones
Technology UsedHigh-intensity focused electromagnetic field
Frequency Range20-100 Hz adjustable for optimal muscle stimulation
Treatment Heads4 detachable treatment heads (can be used individually or in pairs)
Muscle ResponseInduces supramaximal contractions (up to 200+ contractions per session)
Energy DeliveryAlternating magnetic field generates induced current in muscle tissue
Nerve ActivationDepolarizes nerve membranes and excites motor units for natural muscle contraction
Skin EffectImproves skin elasticity and reduces sagging through combined EMF + thermal effects
DowntimeZero downtime; no pain, no surgery, no recovery period
Session Duration20-30 minutes per treatment
Results TimelineVisible changes after 4-6 sessions; peak results at 6-8 weeks
Touchscreen Interface10-inch HD color touchscreen with intuitive navigation
Multilingual SupportSupports English, Spanish, French, German, Chinese, Russian, Arabic, etc.
CertificationsCE Certified

10. User Tips

  • Always ensure the device is powered off before changing or adjusting handles to prevent injury or damage.
  • Begin each treatment session with the lowest intensity setting and gradually increase it to a comfortable and effective level based on the user's tolerance.
  • Maintain a clean and hygienic environment around the device and ensure all skin-contact accessories are disinfected before and after each use for optimal performance and user safety.
  • For best results, advise patients to stay well-hydrated and avoid heavy meals immediately before a treatment session.
  • Regularly check the power cord and connections for any signs of wear or damage to ensure safe operation.

11. Warranty and Support

Warranty Period

  • The main machine is guaranteed for one year from the date of purchase (excluding fragile parts and other non-warranty scope or expired warranty period).
  • The warranty period for accessories is 3 months (excluding man-made damage).
  • The machine can be replaced with a new one within 7 days from the date of purchase (with the receipt) if it has an intact appearance.

Warranty Exclusions

Free warranty service is not provided for failures caused by personal reasons, including:

  • Faults caused by unauthorized disassembly and modification of this product.
  • Faults caused by accidental fall.
  • Failure caused by lack of reasonable maintenance.
  • Failure to operate according to the correct instructions in this instruction manual.

Manufacturer

Guangzhou Xinwei Electronic Technology Co., Ltd.

For after-sales service or support, please contact the manufacturer or your authorized dealer directly. Do not attempt to repair the machine without authorization, as this will void the warranty.

A detailed warranty card and purchase bill are required for warranty claims. For more information, please refer to the User Manual (PDF).

12. Electromagnetic Compatibility (EMC)

The device has undergone Radiated Electromagnetic Field Immunity Testing to ensure compliance with relevant standards.

Test Details:

  • Basic Standard: IEC 61000-4-3:2020
  • Test Level: 30V/m with 80% AM, 1kHz Modulation.
  • Standard Requirement: A
  • Temperature: 25°C
  • Humidity: 54%

Test Equipment:

Description Manufacturer Model Identifier Cal. Date Cal. Due
Signal GeneratorIFA2023BN/A15/03/202214/03/2023
Power AmplifierAR150W1000N/A15/03/202214/03/2023
Power AntennaAR25S1G4AN/A15/03/202214/03/2023

Block Diagram of Test Setup

Block diagram illustrating the setup for electromagnetic compatibility testing, including the 7x4x3 chamber, EUT, power amplifier, signal generator, and control room.
Block Diagram of EMC Test Setup.