1. General Information
The MYOslim NEO Professional 6500W HI-EMT EMS RF Nova Machine utilizes advanced HI-EMT technology, employing high-energy focused electromagnetic waves (HI-EMT) designed to induce powerful muscle contractions, thereby promoting muscle growth and reshaping the internal structure of muscles. Additionally, it triggers significant fat breakdown through intense muscle contractions, causing fatty acids to be released from triglycerides and metabolized by the body over several weeks.
1.1 Indications for Use
- Improvement of abdominal tone, strengthening of the abdominal muscles, development of fitter abdomen.
- Strengthening, Toning and Firming of buttocks, thighs and calves.
- Improvement of muscle tone and firmness, for strengthening muscles in arms.
1.2 Operating Environment
The operating environment of the EMS Body Sculpt instrument series is crucial to ensuring the effectiveness and safety of the device. Ensure the following conditions:
- Clean and Hygienic Environment: The operating environment should be clean and hygienic to prevent any external contamination or bacterial infections. Regular cleaning and disinfection of the equipment and its accessories are essential.
- Appropriate Temperature and Humidity: Ensure that the operating environment's temperature and humidity are within the appropriate range specified by the device manufacturer. Extreme temperatures or excessive dryness/humidity may impact the device's performance.
- Adequate Lighting: Provide sufficient lighting to ensure that operators can clearly see and operate the equipment correctly. This is especially important for conducting precise beauty procedures.
- Good Ventilation: The area where the equipment is operated should have adequate ventilation to ensure air circulation and prevent the accumulation of any harmful gases.
- Safety Measures: The operating environment should have appropriate safety measures, including fire extinguishers and first aid equipment, to address emergencies. Operators should be familiar with the equipment's safety procedures and know how to handle emergency situations.
- Avoidance of Interference Sources: The operating environment for beauty equipment should be kept away from any electromagnetic fields or other devices that could potentially cause interference, ensuring the normal operation of the equipment.
- Operator Training: All operators using beauty equipment should undergo proper training to understand the correct usage and safety procedures of the equipment.
1.3 Patient Profile
This machine is suitable for patients who are seeking healthy weight loss, who have no physical diseases, and who do not have metal substances in their bodies. Please note that this machine is only suitable for adults.
1.4 Description of the Device
The EMS Body Sculpt enables a simple and effective treatment process. You can easily perform by adjusting the digital energy selection treatment plan or diagnostic procedure. Any treatment parameter can be manually adjusted by simply touching the screen. You can also create precise and effective treatment plans by configuring the age group and gender of the person being treated within the machine.
The device features a touch-controlled display for easy operation and clear visibility. Real-time monitoring provides information on remaining treatment time, utilized channels, treatment methods, and attachments. The EMS Body Sculpt series also offers flexible upgrade options to meet future treatment requirements.
1.5 Contraindications
Do NOT use this device if any of the following apply:
- Wounds on the skin (or avoid wounds when using).
- Patients with heart disease (such as those with arrhythmias), especially those with pacemakers.
- Cancer patients.
- Patients with infectious diseases.
- Pregnant women, lactating women, or during menstrual period.
- Major surgery patients, diabetic patients.
- Having any metal in the body.
- People with cerebrovascular disease (stroke patients).
1.6 Possible Side Effects
These effects may include but are not limited to:
- Muscle soreness.
- Reduction in muscle fat.
- Temporary muscle spasms.
- Transient joint or muscle tendon pain.
- Localized redness or skin flushing.
2. Safety Precautions and Warnings
Please observe the following precautions before using the device:
- Do not use the device when feeling full; wait at least one hour after meals.
- Before using the instrument, remove any metal jewelry you are wearing.
- Before starting the operation, touch the probe vertically to the skin, then turn on the switch of the main unit and adjust the intensity of the operation mode.
- Always start at the lowest intensity level, and then increase the intensity sequentially according to the user's skin tolerance.
- It is forbidden to eat or drink during the operation. It is recommended to eat after one hour following the operation.
3. Device and Accessories
3.1 Front and Rear Panel of the Main Unit
Front View of the Instrument:
- Display screen
- EMS Handle
- Switch
Rear View of the Instrument:
- Handle connection
- Power plug
- Fuse
- Switch
3.2 Handle Types
The device comes with various handle types for different applications:
3.3 List of Accessories
| Serial | Name | Quantity | Picture |
|---|---|---|---|
| 1 | Host (Main Unit) | 1 set | ![]() |
| 2 | Handle (Standard) | 2 | ![]() |
| Handle (Optional Cushion) | 1 | ![]() |
|
| 3 | Belt (8*80CM, 8*90CM, 8*100CM) | 3 | ![]() |
| 4 | Power cable | 1 | ![]() |
| 5 | Electronic manual | 1 copy |
4. Device Installation
Unboxing and Initial Setup
Watch this video for a visual guide on unboxing and setting up your MYOslim NEO machine:
Step-by-Step Installation:
-
Install the Handle Bracket: Use a screwdriver to install the handle brackets on both sides of the main unit.
Installation of handle brackets on the main unit. -
Apply Handle Film: Stick the black film to the working part of the handle.
Applying protective film to the treatment handles. -
Install the Handle: Align the handle with the hole on the bracket and fasten the clip. Lock the key ring to secure it.
Attaching the treatment handle to the machine. -
Connect Power: Connect one end of the power cord to the machine and plug the other end into a suitable power socket. Do not connect multiple machines to the same socket.
Connecting the power cable to the machine and power outlet. -
Press the Back Switch: Locate the manual switch on the rear panel and move it to the 'o' position to enable power.
Activating the main power switch on the rear panel. -
Press the Front Switch: Press the power button on the front panel. A single click will power on the device, and a long press will power it off.
Powering on the device using the front panel button.
5. Basic Device Operation
5.1 Handle Exchange
Important: The handle must be calibrated every time it is installed or replaced. When replacing the handle, you must turn off the power of the machine. The handle cannot be replaced when the machine is powered on.
5.2 Device Start Up/Shut Down
- Connect one end of the power cord to the machine and plug the other end into the socket. Plug the power cord directly into the main socket. Do not connect multiple machines to the socket.
- Move the manual switch on the rear panel to the "o" position.
- Press the button on the front panel.
- When the treatment is finished, press and hold the power button to turn off the machine.
5.3 Navigation Controls
The pre-treatment regimen can be modified via on-screen controls. The treatment interface displays the following items and functions:
- List: Access saved treatment protocols.
- Pattern: Select treatment patterns (e.g., Pattern 1, Pattern 2).
- Records: View treatment history.
- Time: Adjust treatment duration.
- Intensity (A/B): Control the intensity level for each handle.
- Frequency: Adjust the frequency of the electromagnetic pulses.
- Start/Pause Button: Initiate or pause the treatment.
5.4 Unit Settings
The device offers various settings accessible through the interface:
- Mode 1: This mode's operation interface allows adjustment of handle operation time according to customer needs. The frequency for Mode 1 is 5-60Hz.
- Mode 2: This mode's operation interface allows adjustment of intensity/stage frequency/time. It features 3 different periods of frequency, which are freely adjustable. The frequency for Mode 2 is 30-100Hz.
- Language Settings: Select your preferred language for the interface.
- Sound Adjustment: Control the device's audio feedback volume.
- Brightness Adjustment: Adjust the screen brightness for optimal visibility.
- Set Date Time: Configure the current date and time.
- System Adjustment: Access advanced system settings (may require technician password).
- Technician: Access technician-level settings (requires password).
Note: Adjust the corresponding intensity/stage frequency/time according to different customer needs. Click the start/pause work button to enter the operation. If the customer requests to pause, you can click the start/pause work button to pause the work, and after the work is completed, you can directly click the return key to return.
6. Therapy
6.1 Preparing the Therapy
- Communicate with patients to understand their needs and comfort levels.
- Ensure the patient is in good health and has no contraindications.
- Ensure the patient is adequately hydrated.
- Ensure the patient wears appropriate clothing that allows for proper handle placement.
- Advise the patient to avoid eating in large quantities before the session.
- Familiarize yourself with the treatment process and settings.
- Advise the patient to avoid drinking alcohol before treatment.
6.2 Placing the Handle
Proper placement of the handles is crucial for effective treatment. Refer to the diagrams below for guidance:
Please operate according to the controller model you purchased and follow the manufacturer's official documentation to ensure the best experience. If you have any questions, please contact customer service for support.
7. Maintenance and Repair
7.1 Maintenance
7.1.1 Daily Inspection
Regularly inspect instruments and components. Check whether the power socket has a reliable connection to the ground. Please check whether the power cord of the instrument is damaged before use, and stop using the instrument if it is damaged.
7.1.2 Cleaning
Instrument cleaning is performed on a regular basis to ensure sanitation and proper functioning of the equipment. The device housing can be wiped with a damp cloth, soapy water or a neutral detergent while the instrument is powered off.
The parts that come into contact with the skin should be cleaned and disinfected with 75% alcohol before and after use. Do not soak directly in alcohol-based detergent as a whole, wipe with pure cotton material.
7.1.3 Replacing the Fuse
When replacing the fuse, in order to prevent electric shock, please turn off the power and unplug the power plug. The fuse is located below the handle junction on the back of the machine. An additional spare fuse is included.
7.2 Repair
After purchasing the instrument, you need to fill in the warranty card and send it to the manufacturer or agent to have the effect of warranty. From the date of the purchase of the instrument, if a product quality problem occurs, a free warranty is provided for one year. (Warranty requires a warranty card and purchase bill, and the scope of warranty is limited to the instrument host, excluding accessories and other consumables that need to be replaced regularly). After the warranty period has expired, please contact the manufacturer directly to provide paid after-sales service. To repair the machine, you must contact the company's business department or dealer, and do not repair it without authorization. Any unpacking without consent will not be guaranteed by our company.
8. Common Faults and Troubleshooting
| Fault Phenomenon | Causes | Exclusions and Workarounds |
|---|---|---|
| Device not working |
|
|
| Handle not working | Check the handle for loose connection | Reseat the handle |
9. Technical Parameters
| Attribute | Value |
|---|---|
| Name | E-02 |
| Input Power Supply | AC110V-220V 50/60Hz |
| Output Maximum Power | 0-6500W |
| Frequency | 5-200HZ |
| Screen Size | 8.0 inches |
| Machine Color | White |
| Material | ABS |
| Application | BODY |
| Core Technology | EMT (Electromagnetic Therapy) + EMS (Electrical Muscle Stimulation) Dual-Mode System |
| Treatment Type | Non-invasive body sculpting, muscle building, fat reduction, and skin tightening |
| Primary Function | Supramaximal muscle contractions for toning, lifting, and shaping |
| Target Areas | Abdomen, buttocks, thighs, arms, back customizable treatment zones |
| Technology Used | High-intensity focused electromagnetic field |
| Frequency Range | 20-100 Hz adjustable for optimal muscle stimulation |
| Treatment Heads | 4 detachable treatment heads (can be used individually or in pairs) |
| Muscle Response | Induces supramaximal contractions (up to 200+ contractions per session) |
| Energy Delivery | Alternating magnetic field generates induced current in muscle tissue |
| Nerve Activation | Depolarizes nerve membranes and excites motor units for natural muscle contraction |
| Skin Effect | Improves skin elasticity and reduces sagging through combined EMF + thermal effects |
| Downtime | Zero downtime; no pain, no surgery, no recovery period |
| Session Duration | 20-30 minutes per treatment |
| Results Timeline | Visible changes after 4-6 sessions; peak results at 6-8 weeks |
| Touchscreen Interface | 10-inch HD color touchscreen with intuitive navigation |
| Multilingual Support | Supports English, Spanish, French, German, Chinese, Russian, Arabic, etc. |
| Certifications | CE Certified |
10. User Tips
- Always ensure the device is powered off before changing or adjusting handles to prevent injury or damage.
- Begin each treatment session with the lowest intensity setting and gradually increase it to a comfortable and effective level based on the user's tolerance.
- Maintain a clean and hygienic environment around the device and ensure all skin-contact accessories are disinfected before and after each use for optimal performance and user safety.
- For best results, advise patients to stay well-hydrated and avoid heavy meals immediately before a treatment session.
- Regularly check the power cord and connections for any signs of wear or damage to ensure safe operation.
11. Warranty and Support
Warranty Period
- The main machine is guaranteed for one year from the date of purchase (excluding fragile parts and other non-warranty scope or expired warranty period).
- The warranty period for accessories is 3 months (excluding man-made damage).
- The machine can be replaced with a new one within 7 days from the date of purchase (with the receipt) if it has an intact appearance.
Warranty Exclusions
Free warranty service is not provided for failures caused by personal reasons, including:
- Faults caused by unauthorized disassembly and modification of this product.
- Faults caused by accidental fall.
- Failure caused by lack of reasonable maintenance.
- Failure to operate according to the correct instructions in this instruction manual.
Manufacturer
Guangzhou Xinwei Electronic Technology Co., Ltd.
For after-sales service or support, please contact the manufacturer or your authorized dealer directly. Do not attempt to repair the machine without authorization, as this will void the warranty.
A detailed warranty card and purchase bill are required for warranty claims. For more information, please refer to the User Manual (PDF).
12. Electromagnetic Compatibility (EMC)
The device has undergone Radiated Electromagnetic Field Immunity Testing to ensure compliance with relevant standards.
Test Details:
- Basic Standard: IEC 61000-4-3:2020
- Test Level: 30V/m with 80% AM, 1kHz Modulation.
- Standard Requirement: A
- Temperature: 25°C
- Humidity: 54%
Test Equipment:
| Description | Manufacturer | Model | Identifier | Cal. Date | Cal. Due |
|---|---|---|---|---|---|
| Signal Generator | IFA | 2023B | N/A | 15/03/2022 | 14/03/2023 |
| Power Amplifier | AR | 150W1000 | N/A | 15/03/2022 | 14/03/2023 |
| Power Antenna | AR | 25S1G4A | N/A | 15/03/2022 | 14/03/2023 |
Block Diagram of Test Setup